Alarms
Icon Name Alarm
priority Message Therapy
interrupted Check
Watertemperature
deviation
Medium,5
minute delay
Watertemperatureis±0.8°C(1.4°F)
outsideoftargettemperature
No Temporaryconditionuponstartup,additionofthermaltransferdevice,
oraddition of water
No water Medium,60
second delay
No water Yes Check for leaks
Addaminimumof2litersofsteriledistilledwaterordistilledwaterthat
hasbeenpassedthroughalterlessthanorequalto0.22microns
Nowaterow Medium,20
second delay
Noowdetected Yes Checkforleaksandobstructionsatconnections,hoses,andthermal
transfer devices
Checkwaterow Medium,10
minute delay
Reducedowdetected No TapConrm,ifthewaterportwasremovedintentionallyCheckforleaks
andobstructionsatconnections,hoses,andthermaltransferdevices
Checkpatientprobe
(AorB)
Medium Abnormalchangeinpatient
temperature
Yes Checkprobecondition,location,andconnections
Probeoradapter
malfunction(AorB)
Medium,30
second delay
Notemperaturesignaldetected Yes Checkprobeoradaptercablecondition,location,andconnections
Adaptercable
disconnected(AorB)
Medium,30
second delay
Adaptercablenotdetected Yes Reinserttheadaptercable.Ifdamaged,replacetheadaptercable
Patienttemperature
deviation
Medium Patienttemperatureis>±1.0°C
(0.9°F)outsideoftarget
temperature
No Checkpatientcondition,placementofthermaltransferdevices,andall
connections
Patienttemperature
deviation
Low Patienttemperatureis>±0.5°C
(0.9°F)outsideoftarget
temperature
No Checkpatientcondition,placementofthermaltransferdevices,andall
connections
Normothermia
deviation
Low Patienttemperatureisoutsideof
36.0°C(96.8°F)to38.0°C(100.4°F)
No Checkpatientcondition,placementofthermaltransferdevices,andall
connections
Therapypause Medium Therapyiscurrentlypaused Yes Toresume,pressandholdtheplay/pausefor2seconds
Batterylow Low Batteryislow No Maintenanceisrecommended.Ifbatteryisnotreplaced,theproductmay
notfunctiononthenextstartup.
Patienttemperature
output(AorB)
Low Patienttemperatureoutputis
inaccurateontheexternaldevice,
oroutsidesupportedrange
No Checkoutputadaptercableconnection.TapConrmtoreactivatethe
outputport.
Removefromuse
(RFU)
Medium Thesystemhaspoweredoffdueto
a malfunction
Yes Removetheproductfromuseimmediately.Notifytheappropriate
personnel.
Power Loss Medium Notapplicable Yes Checkpowercordconnection
Ifanyofthealarmconditionspersist,callmaintenance. Ifthepageindicatorsappearonthealarmscreen,therearemultipleactivealarms.Onlythehighestalarmisdisplayed.