Rosemount BINOS 100 Series Analyzers including OXYNOS 100, HYDROS 100 and CAT 100-2nd Ed. Owner's manual

Type
Owner's manual
Instruction Manual
HAS1xE-IM-HS
January 2006
www.EmersonProcess.com
BINOS
®
100 Series Gas Analyzers
BINOS
®
100, BINOS
®
100 M
BINOS
®
100 2M, BINOS
®
100 F
OXYNOS
®
100, HYDROS
®
100, CAT 100
Emerson Process Management
Manufacturing GmbH & Co. OHG
Industriestrasse 1
D-63594 Hasselroth
Germany
T +49 (0) 6055 884-0
F +49 (0) 6055 884-209
Internet: www.EmersonProcess.com
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Emerson Process Management designs, manufactures and tests its products to meet
many national and international standards. Because these instruments are sophisticated
technical products, you MUST properly install, use, and maintain them to ensure
they continue to operate within their normal specifications. The following instructions
MUST be adhered to and integrated into your safety program when installing, using and
maintaining Emerson Process Management products. Failure to follow the proper
instructions may cause any one of the following situations to occur: Loss of life; personal
injury; property damage; damage to this instrument; and warranty invalidation.
·
Read all instructions prior to installing, operating, and servicing the product.
· If you do not understand any of the instructions, contact your Emerson Process
Management representative for clarification.
· Follow all warnings, cautions, and instructions marked on and supplied with the
product.
· Inform and educate your personnel in the proper installation, operation, and
maintenance of the product.
· Install your equipment as specified in the Installation Instructions of the
appropriate Instruction Manual and per applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
· To ensure proper performance, use qualified personnel to install, operate, update,
program, and maintain the product.
· When replacement parts are required, ensure that qualified people use replacement
parts specified by Emerson Process Management. Unauthorized parts and procedures
can affect the product’s performance, place the safe operation of your process at risk,
and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical
hazards, or improper operation.
· Ensure that all equipment doors are closed and protective covers are in place,
except when maintenance is being performed by qualified persons, to prevent
electrical shock and personal injury.
The information contained in this document is subject to change without notice.
1
st
edition 3/2005
2
nd
edition 1/2006
Emerson Process Management Manufacturing GmbH & Co. OHG S-I
Instruction Manual
HAS1xE-IM-HS
03/2005
BINOS
®
100 Series
PREAMBLE
This instruction manual provides information about BINOS
®
100 series gas
analyzers concerning subassemblies, functions, procedures, installation,
operation and maintenance.
This instruction manual covers several BINOS
®
100 series analyzer variations
and therefore may describe configurations and/or options not part of your
specific analyzer.
Installation and operation of instruments intended to be installed and operated
in HAZARDOUS AREAS is NOT COVERED by this instruction manual, but part
of the specific instruction manual shipped together with such analyzers
because of the special requirements for working in hazardous environments!
DEFINITIONS
The following definitions apply to WARNINGS, CAUTIONS and NOTES found throughout
this publication.
Highlights an operation or
maintenance procedure,
practice, condition, state-
ment, etc.
If not strictly observed, could
result in injury, death, or long-
term health hazards of
personnel.
Highlights an operation or
maintenance procedure,
practice, condition, state-
ment, etc.
If not strictly observed, could
result in damage to or
destruction of equipment, or
loss of effectiveness.
NOTE
Highlights an essential operating
procedure, condition or statement.
Emerson Process Management Manufacturing GmbH & Co. OHGS-II
BINOS
®
100 Series
Instruction Manual
HAS1xE-IM-HS
03/2005
All graphical symbols used in this product are from one or more of the following standards:
EN61010-1, IEC417, and ISO3864.
Explosion Hazard!
Harmful (to Health)!
Elektrostatic discharge (ESD)
Toxic!
Heavy Instrument!
Disconnect from Mains!
IMPORTANT
SAFETY INSTRUCTIONS
WIRING AND INSTALLATION OF THIS APPARATUS
The following safety instructions apply specifically to all EU member states. They should be
strictly adhered to in order to assure compliance with the Low Voltage Directive. Non-EU states
should also comply with the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points, internal and external,
where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced.
The integrity of all earth terminals must be maintained at all times.
3. To ensure safe operation of this equipment, connection to the mains supply should only be
made through a circuit breaker which will disconnect
all circuits carrying conductors during
a fault situation. The circuit breaker may also include a mechanically operated isolating
switch. Circuit breakers or switches must comply with a recognized standard such as
IEC947. All wiring must conform with any local standards.
4. Where equipment or covers are marked with the symbol to the right,
hazardous voltages are likely to be present beneath. These covers should
only be removed when power is removed from the equipment — and then
by trained service personnel only.
5. Where equipment or covers are marked with the symbol to the right, there
is a danger from hot surfaces beneath. These covers should only be removed
by trained service personnel when power is removed from the equipment.
Certain surfaces may remain hot to the touch.
6. Where equipment or covers are marked with the symbol to the right, refer
to the Instruction Manual for instructions.
7. Further graphical symbols used in this product:
Emerson Process Management Manufacturing GmbH & Co. OHG S-III
Instruction Manual
HAS1xE-IM-HS
03/2005
BINOS
®
100 Series
This instrument has left the factory in
compliance with all applicable safety
regulations.
To maintain this operating condition, the user
must strictly follow the instructions and consider
the warnings in this manual or provided on the
instrument.
Before switching on the instrument, verify that
the electrical supply voltage matches the
instrument´s operating voltage as set in the
factory.
Any interruption in the instrument´s ground line,
whether inside or outside the instrument, or
removal or interruption of its ground line
connection, could result in hazardous operating
conditions. Intentionally interrupting the
instrument´s protective ground is strictly prohi-
bited.
Opening cover panels could expose voltage-
carrying components. Connectors may also be
under voltage. The instrument must be
disconnected from all electrical supplies before
attempting any calibrations, maintenance
operations, repairs or component
replacements requiring opening of the
instrument. Any calibrations, maintenance
operations, or repairs that need the instrument
to be opened while connected to electrical
supplies should be subject to qualified
technicians familiar with the hazards involved
only!
Use only fuses of the correct type and current
ratings as replacements. Using repaired fuses
and short circuiting of fuse holders is prohibited.
Observe all applicable regulations when
operating the instrument from an auto-
transformer or variac.
Substances hazardous to health may emerge
from the instrument‘s exhaust.
Please pay attention to the safety of your
operation personnel. Protective measures
must be taken, if required.
OPERATING AND MAINTAINING THIS APPARATUS
SAFETY INSTRUCTIONS
Emerson Process Management Manufacturing GmbH & Co. OHGS-IV
BINOS
®
100 Series
Instruction Manual
HAS1xE-IM-HS
03/2005
SAFETY INSTRUCTIONS
TOXIC GASES !
This unit’s exhaust may contain toxic gases such as sulfur dioxide.
These gases can cause serious injuries.
Aviod inhalation of the exhaust gases at the exhaust fitting.
Connect exhaust outlet to a safe vent. Check vent line and connections for leakage.
Keep all fittings tight to avoid leaks. See section 9-2, page 9-3 for leak test information.
ELECTRICAL SHOCK HAZARD !
Do not operate without covers secure. Do not open while energized. Installation requires
access to live parts which can cause death or serious injury.
For safety and proper performace this instrument must be connected to a properly
grounded three-wire source of power.
SAFETY SUMMARY
If this equipment is used in a manner not specified in these instructions, protective systems
may be impaired.
AUTHORIZED PERSONNEL
To avoid loss of life, personal injury and damage to this equipment and on-site property, do
not operate or service this instrument before reading and understanding this
instruction manual and receiving appropriate training. Save these instructions.
BINOS
®
100 series gas analyzers are intended to be used as analyzers for industrial
purposes. They must not be used in medical, diagnostic or life support applications,
and no independent agency certifications or approvals are to be implied as covering
such applications!
INTENDED USE STATEMENT
Emerson Process Management Manufacturing GmbH & Co. OHG S-V
Instruction Manual
HAS1xE-IM-HS
03/2005
BINOS
®
100 Series
SAFETY INSTRUCTIONS
BINOS
®
100 F and CAT 100 -- HEAVY INSTRUMENTS !
The analyzer variations BINOS
®
100 F and CAT 100 intended to be wall
mounted and/or outdoor installed weigh up to 30 kg resp. 70 kg, depending
on included options!
Use two person and/or suitable tools for transportation and lifting these
instruments!
Take care to use anchors and bolts specified to be used for the weight of
the units!
Take care the wall or stand the unit is intended to be installed at is solid and
stable to hold the units!
HIGH TEMPERATURES !
While working at photometers and/or thermostated components inside the
analyzers hot components may be accessible!
EXPLOSION HAZARD !
Do not operate nor install these instruments in hazardous areas without
additional measures!
Emerson Process Management Manufacturing GmbH & Co. OHGS-VI
BINOS
®
100 Series
Instruction Manual
HAS1xE-IM-HS
03/2005
SAFETY INSTRUCTIONS
Take care of the safety instructions applicable for the gases (sample
gases and test gases) and for the gas bottles containing these gases!
GASES AND GAS CONDITIONING
Before opening gas paths they must be purged with ambient air or
neutral gas (N2) to avoid hazards caused by toxic, flammable,
explosive or harmful to health sample gas components!
Supplying explosive gases is not permitted !
EXPLOSION HAZARD !
Supplying flammable gases of concentrations above 25 % of lower
explosion limit (LEL) we recommend to utilize one or more of the following
measures:
• Purging the housing with inert gas
• Internal tubing with stainless steel
• Flame arrestors at gas input and output fittings
• Intrinsically safe paramagnetical or thermal conductivity sensors
Emerson Process Management Manufacturing GmbH & Co. OHG S-VII
Instruction Manual
HAS1xE-IM-HS
03/2005
BINOS
®
100 Series
SAFETY INSTRUCTIONS
POWER SUPPLY
The mains socket has to be nearby the power supply unit and easily
accessible! Disconnecting from mains requires unplugging the power plug!
To comply with the CE mark requirements use only power supply units of
type VSE 2000, UPS 01 T, DP 157, SL5, SL10 (DP 157 and SL for rack
installation only) or equivalent units. Equivalent units must provide SELV
output voltages!
Verify proper polarity when connecting DC 24 V operated analyzers !
These instruments provide a protective earth terminal. To prevent electrical
shock hazards the instrument must be connected to a protective earth.
Therefore the instrument has to be connected to mains by using a three
wire mains cable with earth conductor!
Any interruption of the earth connector inside or outside the instrument or
disconnecting the earth terminal may cause potential electrical shock
hazzard! Intended interruption of protective earth connections is not
permitted!
ELECTRICAL SHOCK HAZARD !
Verify the mains voltage at site of installation corresponds to the analyzer´s
rated voltage as given on the nameplate label!
Verify the safety instruction given by power supply unit manufacturer !
Instruments with External Power Supply Unit
Instruments with Internal Power Supply Unit
Emerson Process Management Manufacturing GmbH & Co. OHGS-VIII
BINOS
®
100 Series
Instruction Manual
HAS1xE-IM-HS
03/2005
Installation and connecting mains and signal cables are subject to qualified
personnel only taking into account all applicable standards and legislative
requirements!
Failure to follow may cause warranty invalidation, property damage and/or
personal injury or death! Mains and signal cables need to be connected to
internal srew terminals requiring to work at open housing near life parts!
Installation of this instrument is subject to qualified personnel only, familiar
with the resulting potential risks!
These gas analyzer do not provide a mains power switch and are operable
when connected to power.
These gas analyzers do not provide a mains switch! A mains switch or circuit
breaker has to be provided in the building installation. This switch has to be
installed near by analyzer, must be easily operator accessible and has to be
assigned as disconnector for the analyzer.
ADDITIONAL HINTS FOR MODELS BINOS
®
100 F AND CAT 100
Cables for external data processing must be double insulated for mains
voltage when used inside the instrument!
If double insulation is not available signal cables inside the analyzer must
be installed in a way that a distance of at least 5 mm is ensured permanently
(e.g. by utilizing cable ties).
ADDITIONAL HINTS FOR MODEL BINOS
®
100 F
SAFETY INSTRUCTIONS
Emerson Process Management Manufacturing GmbH & Co. OHG S-IX
Instruction Manual
HAS1xE-IM-HS
03/2005
BINOS
®
100 Series
Supplying external components utilizing the DC 24 V power output provided
by the BINOS
®
100 2M internal power supply unit requires a fuse to be
connected in series to the load limiting the power to 2 A max!
Verify proper DC 24 V polarity when connecting any external load!
SAFETY INSTRUCTIONS
ADDITIONAL HINTS FOR MODEL BINOS
®
100 2M
Emerson Process Management Manufacturing GmbH & Co. OHGS-X
BINOS
®
100 Series
Instruction Manual
HAS1xE-IM-HS
03/2005
General Operating Instructions
• Installation area has to be clean, free from moisture, excessive vibration and frost-protected.
Take care to meet the permissible ambient temperatures as given in the technical data section!
Instruments must not be exposed to direct sunlight, fluorescent lamps nor sources of heat.For
outdoor installation it is recommended to mount the instruments into a cabinet. At least sheltering
against rain is required. Do not cover venting openings and take care to mount the instrument in
a distance to walls not affecting venting.
• Do not interchange gas inlet and outlet! All gases must be conditioned before supplying! When
supplying corrosive gases ensure that gas path components are not affected!
• Max. permissible gas pressure: 1,500 hPa, except with integrated valve blocks (see hint at
page 3-4) and/or paramagnetic Oxygen sensor (see page 2-15)!
• Exhaust lines must be installed in a descending way, need to be pressureless, frost-protected
and in compliance with applicable legislative requirements!
• When it is necessary to open gas paths seal the analyzer‘s gas fittings by using PVC caps to
avoid pollution of the internal gas path by moisture, dust, etc.
• To stay in compliance with regulations regarding electromagnetic compatibility it is recommended
to use only shielded cables, as optionally available from Emerson Process Management or
equivalent. Customer has to take care that the shield is connected in proper way (see section
3-5). Shield and signal connector enclosure need to be conductively connected, submin-d plugs
and sockets must be screwed to the analyzer.
• Using external submin-d-to-terminal adaptor elements (option) affects electromagnetic
compatibility. In this case the customer has to take measures to stay in compliance and has to
declare conformity, when required by legislation (e.g. European EMC Directive).
General operating instructions
DANGER TO LIFE ! EXPLOSION HAZARD !
Verify all gas lines are connected as described within this manual and tight!
Improper gas connections may cause explosion, serious injury or death!
Exhaust may contain hydrocarbons and other toxic gases, e.g. carbon
monoxide. Carbon monoxide is toxic!
Emerson Process Management Manufacturing GmbH & Co. OHG S-XI
Instruction Manual
HAS1xE-IM-HS
03/2005
BINOS
®
100 Series
Magnetically Operated Front Panel
Magnetically Operated Front Panel
DANGER TO LIFE !
Persons with cardiac pacemakers should absolutely avoid magnetic
fields !
Negative effects on persons beyond those described above caused by
magnetic fields are not known. It is presumed that persons showing allergic
reaction on contact with ceramic or metallic material show the same
behavior on contact with magnetic material.
Permanent magnets are surrounded by magnetic fields. These magnetic
fields can disturb and even destroy sensitive electronic measuring devices,
but also mechanical watches, credit cards, etc.
Usually a distance of 0.5 m is enough to avoid damages. All sintered
permanent magnets are hard and brittle. Hitting of sintered permanent
magnets by the magnetic attraction causes splitting into fragments with many
sharp edges. This especially occurs with high energy magnets, and can
also cause skin bruises by high attraction.
High energy magnets made of rare-earth materials have to be stored dry,
otherwise the surfaces would oxidise. Unprotected operation in a humid
environment may cause corrosion. Avoid damaging the protective galvanic
coating.
A storage in a hydrogen atmosphere destroys these magnets. A
demagnetisation is caused when permanent magnet materials have been
exposed in a radioactive radiation for a long time.
For air transportation of magnetic material the IATA instructions have to be
observed:
Magnetic fields are not allowed to penetrate the package, if necessary the
magnets have to be shorted using a metal plate.
Emerson Process Management Manufacturing GmbH & Co. OHGS-XII
BINOS
®
100 Series
Instruction Manual
HAS1xE-IM-HS
03/2005
Emerson Process Management Manufacturing GmbH & Co. OHG TOC-1
Instruction Manual
HAS1xE-IM-HS
03/2005
BINOS
®
100 Series
Table of Contents
PREAMBLE ................................................................................................................... I
DEFINITIONS ................................................................................................................... I
SAFETY INSTRUCTIONS ..................................................................................................II
CHAPTER 1 TECHNICAL DESCRIPTION................................................................... 1-1
1-1 General Description................................................................................................................ 1-1
1-2 BINOS
®
100, BINOS
®
100 M, OXYNOS
®
100, HYDROS
®
100.............................................. 1-5
1-3 BINOS
®
100 2M ...................................................................................................................... 1-9
1-4 BINOS
®
100 F....................................................................................................................... 1-15
1-5 CAT 100 ................................................................................................................................ 1-22
1-6 Internal Gas Paths ................................................................................................................ 1-30
1-6-1 Gas Paths Materials........................................................................................................... 1-30
1-6-2 Safety Filter......................................................................................................................... 1-30
1-6-3 Fittings.................................................................................................................................1-30
1-6-4 Piping .................................................................................................................................. 1-30
1-6-5 Gas Path Variations............................................................................................................ 1-31
1-7 Technical Data ...................................................................................................................... 1-34
1-7-1 Common Technical Data.................................................................................................... 1-34
1-7-2 Model Specific Data............................................................................................................1-36
1-7-2-1 BINOS
®
100, BINOS
®
100 M, OXYNOS
®
100, HYDROS
®
100 ....................................... 1-36
1-7-2-2 BINOS
®
100 2M................................................................................................................ 1-41
1-7-2-3 BINOS
®
100 F ..................................................................................................................1-48
1-7-2-4 CAT 100 ........................................................................................................................... 1-55
1-7-2-5 External Power Supply Unit for Tabletop and Rack Mount Analyzers.............................. 1-63
1-7-2-6 5 A Power Supply Unit for DIN Rail Installation ...............................................................1-65
1-7-2-7 External 10A Power Supply Unit for DIN Rail Installation ................................................. 1-67
1-7-2-8 10 A Tabletop Power Supply Unit ....................................................................................1-69
1-7-3 Specifications at the Nameplate Label ............................................................................... 1-70
CHAPTER 2 MEASURING PRINCIPLES..................................................................... 2-1
2-1 Infrared Measurement (IR)......................................................................................................2-1
2-1-1 Interference Filter Correlation (IFC)......................................................................................2-1
2-1-2 Opto Pneumatic Measuring Principle...................................................................................2-3
2-1-3 Technical Implementation..................................................................................................... 2-4
2-2 Oxygen Measurement ............................................................................................................ 2-9
2-2-1 Paramagnetic Measurement................................................................................................ 2-9
2-2-2 Electrochemical Measurement .......................................................................................... 2-11
2-2-3 Special Hints on Oxygen Sensors ..................................................................................... 2-13
2-3 Thermal Conductivity Measurement..................................................................................... 2-14
2-3-1 Principle of Operation ......................................................................................................... 2-14
2-3-2 Technical Implementation................................................................................................... 2-15
2-4 Measurement Specification .................................................................................................. 2-16
Emerson Process Management Manufacturing GmbH & Co. OHGTOC-2
BINOS
®
100 Series
Instruction Manual
HAS1xE-IM-HS
03/2005
Table of Contents
CHAPTER 3 INSTALLATION ....................................................................................... 3-1
3-1 Abstract................................................................................................................................... 3-1
3-2 Gas Conditioning .................................................................................................................... 3-2
3-3 Gas Connections.................................................................................................................... 3-3
3-4 Analyzer Specific Instructions.................................................................................................3-5
3-4-1 1/4 19" Instruments...............................................................................................................3-5
3-4-2 1/2 19" Instruments...............................................................................................................3-9
3-4-3 BINOS
®
100 F.....................................................................................................................3-14
3-4-4 CAT 100 .............................................................................................................................. 3-21
3-5 Wiring Signal Terminals........................................................................................................ 3-27
3-5-1 Electrical Connections in General ...................................................................................... 3-27
3-5-2 Wiring Inductive Loads ....................................................................................................... 3-29
3-5-3 Driving Multiple Loads ........................................................................................................ 3-29
3-5-4 Driving High Current Loads ................................................................................................ 3-30
CHAPTER 4 OPERATION ........................................................................................... 4-1
4-1 Abstract................................................................................................................................... 4-1
4-2 Button "FUNCTION"................................................................................................................ 4-4
4-3 Button "ENTER"...................................................................................................................... 4-6
4-4 Buttons "INPUT - CONTROL" ................................................................................................ 4-6
4-5 Button "PUMP" ........................................................................................................................ 4-7
4-6 Software Menu Structure ........................................................................................................ 4-7
4-7 Magnetically Operated Front Panel ........................................................................................ 4-9
4-7-1 Front Panel Elements ........................................................................................................... 4-9
4-7-2 Accessory............................................................................................................................. 4-9
4-7-3 Operating the Front Panel...................................................................................................4-10
4-7-4 Storing the Actuator Tool..................................................................................................... 4-11
CHAPTER 5 STARTUP ............................................................................................... 5-1
5-1 Abstract................................................................................................................................... 5-1
5-2 Software Menu Structure ........................................................................................................ 5-2
5-3 Checking System Parameter Settings................................................................................... 5-3
5-3-1 Menu "System Parameter"..................................................................................................5-3
5-3-2 Menu "Druckkorrektur"........................................................................................................5-4
5-3-3 Menu "Internal Cross Compensation´"................................................................................5-4
5-3-4 Menu "Calibrate Cross Compensation" ..............................................................................5-5
5-3-5 Menu "Hold" ......................................................................................................................... 5-6
5-3-6 Menu "Automatic Calibration".............................................................................................. 5-6
5-3-7 Menu "Tolerance Check"..................................................................................................... 5-7
5-3-8 Menu "Display Off" .............................................................................................................. 5-8
5-3-9 Menu "Gas Path Purge Time" ............................................................................................. 5-9
5-3-10 Menu "Access Code" ..........................................................................................................5-9
Emerson Process Management Manufacturing GmbH & Co. OHG TOC-3
Instruction Manual
HAS1xE-IM-HS
03/2005
BINOS
®
100 Series
Table of Contents
5-3-11 Menus "Response Time t90"...........................................................................................5-10
5-3-12 Menu "Reset"..................................................................................................................... 5-11
5-3-13 Menu "Software Revision" ................................................................................................. 5-12
5-3-14 Menus "Analyzer Serial Number"...................................................................................... 5-12
5-3-15 Menu "Pump Status" *)......................................................................................................5-13
5-3-16 Menu "Pump Control" *) ....................................................................................................5-13
5-4 Analog Outputs ....................................................................................................................5-14
5-4-1 Menu "Life-Zero for Analog Outputs"................................................................................. 5-15
5-4-2 Menu "Adjusting Analog Outputs"......................................................................................5-16
5-4-3 Menu "Analog Output Offset" ............................................................................................5-18
5-4-4 Menu "Analog Output Upper Limit"....................................................................................5-19
5-5 Digital Outputs ...................................................................................................................... 5-20
5-5-1 Concentration Threshold Operation Modes ....................................................................... 5-20
5-5-2 Setting Concentration Thresholds ...................................................................................... 5-24
5-5-3 Valve Control.......................................................................................................................5-26
5-5-4 Wiring Digital Outputs......................................................................................................... 5-26
5-6 Calibration Procedures......................................................................................................... 5-27
5-6-1 Manual Calibration .............................................................................................................. 5-28
5-6-1-1 Zero Calibration ................................................................................................................ 5-28
5-6-1-2 Span Calibration ............................................................................................................... 5-30
5-6-2 Time Controlled Calibration ................................................................................................ 5-32
5-6-2-1 Zero Calibration ................................................................................................................ 5-32
5-6-2-2 Combined Zero and Span Calibration.............................................................................. 5-34
5-7 Remote Calibration (Option)................................................................................................. 5-35
5-8 Measuring ............................................................................................................................ 5-36
5-9 Taking Out of Operation ........................................................................................................ 5-37
CHAPTER 6 DESCRIPTION OF OPTIONS ................................................................. 6-1
6-1 Relay Status Signals............................................................................................................... 6-2
6-2 Serial Interface........................................................................................................................ 6-3
6-2-1 Enabling / Disabling Serial Interface..................................................................................... 6-5
6-2-2 Configuring the Interface.......................................................................................................6-5
6-2-3 Communication..................................................................................................................... 6-7
6-2-3-1 Telegram Syntax ................................................................................................................6-8
6-2-3-2 Status Telegram .................................................................................................................6-9
6-2-3-3 Number Formats.............................................................................................................. 6-10
6-2-3-4 Longitudinal Parity Test .................................................................................................... 6-10
6-2-3-5 Command Syntax ............................................................................................................ 6-10
6-2-3-6 List of Available Commands ............................................................................................ 6-11
6-2-3-7 Reply Telegrams .............................................................................................................. 6-12
6-2-3-7-1 Status Messages (Command 030) ................................................................................ 6-12
6-2-5-7-2 Pump Status (Command 008) *).................................................................................... 6-13
Emerson Process Management Manufacturing GmbH & Co. OHGTOC-4
BINOS
®
100 Series
Instruction Manual
HAS1xE-IM-HS
03/2005
Table of Contents
6-3 Option Digital Inputs / Foundationâ„¢ Fieldbus ...................................................................... 6-14
6-3-1 Digital Inputs ....................................................................................................................... 6-14
6-3-2 Starting a Calibration Procedure......................................................................................... 6-15
6-3-3 Valve Control....................................................................................................................... 6-15
6-3-4 Pump Control...................................................................................................................... 6-15
6-3-5 Foundationâ„¢ Fieldbus....................................................................................................... 6-16
6-4 Cross Compensation (TCD) ................................................................................................ 6-17
6-5 Fine Dust Filter / Needle Valve ........................................................................................... 6-18
6-5-1 Fine Dust Filter.................................................................................................................... 6-18
6-5-2 Needle Valve....................................................................................................................... 6-18
6-6 Flowmeter............................................................................................................................. 6-19
CHAPTER 7 BUG FIXING ........................................................................................... 7-1
7-1 Error Codes ............................................................................................................................7-1
7-2 Other Problems ...................................................................................................................... 7-6
CHAPTER 8 TROUBLESHOOTING ON ELECTRONICS............................................ 8-1
8-1 Getting Access to Internal Analyzer Components .................................................................. 8-2
8-1-1 BINOS
®
100 (M), OXYNOS
®
100, HYDROS
®
100 ............................................................. 8-2
8-1-2 BINOS
®
100 2M.................................................................................................................... 8-8
8-1-3 BINOS
®
100 F..................................................................................................................... 8-14
8-1-4 CAT 100 .............................................................................................................................. 8-16
8-2 Measuring Points .................................................................................................................. 8-23
8-2-1 Measuring Points at BKS ...................................................................................................8-23
8-2-1-1 Supply voltage +6 V ......................................................................................................... 8-23
8-2-1-2 Positive reference voltage................................................................................................ 8-23
8-2-1-4 Temperature sensor......................................................................................................... 8-25
8-2-1-3 Negative reference voltage .............................................................................................. 8-25
8-2-1-5 Light barrier signal............................................................................................................ 8-25
8-2-1-6 Analog Preamplifier .......................................................................................................... 8-26
8-2-2 Measuring Points at OXS (Electrochemical Oxygen Measurement) ............................... 8-27
8-2-2-1 Sensor Signal................................................................................................................... 8-27
8-3 Position and Assignment of Connectors on Electronics Boards .......................................8-28
8-3-1 BKS..................................................................................................................................... 8-28
8-3-1-1 IR Measurement w/o Oxygen Sensor.............................................................................. 8-29
8-3-1-2 Oxygen Measurement w/o IR .......................................................................................... 8-29
8-3-1-3 Combination of IR and Oxygen Measurement................................................................ 8-29
8-3-1-4 TCD w/o IR channel......................................................................................................... 8-29
8-3-1-5 Combined IR and TCD Measurement ............................................................................8-30
8-3-1-6 Combined Oxygen and TCD Measurement .................................................................... 8-30
8-3-2 OXS (Electrochemical Oxygen Measurement)................................................................. 8-31
8-3-3 WAP 100 (TC Measurement)............................................................................................. 8-32
Emerson Process Management Manufacturing GmbH & Co. OHG TOC-5
Instruction Manual
HAS1xE-IM-HS
03/2005
BINOS
®
100 Series
8-4 BKS 20 Jumper Configuration.............................................................................................. 8-33
CHAPTER 9 MAINTENANCE ...................................................................................... 9-1
9-1 Replacing the Fine Dust Filter ................................................................................................9-2
9-2 Leak Testing............................................................................................................................ 9-3
9-3 Cleaning the Analyzer´s Outside ............................................................................................ 9-4
9-4 Replacing the Electrochemical Oxygen Sensor .................................................................... 9-5
9-4-1 Analyzer with Oxygen Measurement only, no IR Channel................................................... 9-6
9-4-2 Analyzer with Combined IR and Oxygen measurement...................................................... 9-7
9-4-3 Replacing the Sensor ........................................................................................................... 9-7
9-4-4 Reassembling the Sensor Unit............................................................................................. 9-8
9-4-5 Sensor Amplifier Adjustment ................................................................................................9-8
9-5 Magnetically Operated Front Panel ........................................................................................ 9-9
APPENDIX A-1
A-1 Block Diagrams ......................................................................................................................A-2
A-1-1 BINOS
®
100 (M), OXYNOS
®
100, HYDROS
®
100 ............................................................ A-2
A-1-2 BINOS
®
100 2M................................................................................................................... A-3
A-1-3 BINOS
®
100 F...................................................................................................................... A-4
A-1-4 CAT 100 ............................................................................................................................... A-5
A-2 Declarations of Conformity .....................................................................................................A-9
A-2-1 BINOS
®
100 (M), OXYNOS
®
100, HYDROS
®
100 ............................................................ A-9
A-2-2 BINOS
®
100 2M................................................................................................................. A-10
A-2-3 BINOS
®
100 F.....................................................................................................................A-11
A-2-4 CAT 100 ............................................................................................................................. A-12
Table of Contents
Emerson Process Management Manufacturing GmbH & Co. OHGTOC-6
BINOS
®
100 Series
Instruction Manual
HAS1xE-IM-HS
03/2005
Fig. 4-1: BINOS
®
100 (M), OXYNOS
®
100, HYDROS
Fig. 4-2: BINOS
®
100 2M (standard version) front sid
Fig. 4-3: Magnetically operated front panel ...............
Table of Illustrations
CHAPTER 1 TECHNICAL DESCRIPTION
Fig. 1-1: BINOS
®
100 (M), OXYNOS
®
100, HYDROS
®
100, front side view ...................................... 1-6
Fig. 1-2: BINOS
®
100 (M), OXYNOS
®
100 (EO2), HYDROS
®
100, rear side view............................ 1-7
Fig. 1-3: OXYNOS
®
100 (PO2), rear side view................................................................................... 1-8
Fig. 1-4: BINOS
®
100 2M (standard version) front side view............................................................ 1-11
Fig. 1-5: BINOS
®
100 2M, variation A (internal power supply) Rear side view with all options........ 1-12
Fig. 1-6: BINOS
®
100 2M, variation B (external power supply) Rear side view with all options ...... 1-13
Fig. 1-7: BINOS
®
100 2M (special variation), Rear side view with all options .................................1-14
Fig. 1-8: BINOS
®
100 F Standard version, front side view................................................................ 1-18
Fig. 1-9: BINOS
®
100 F Standard front panel incl. sample handling options.................................... 1-19
Fig. 1-10: BINOS
®
100 FMagnetically operated front panel (option)................................................... 1-20
Fig. 1-11: BINOS
®
100 F Interior view (without components)............................................................ 1-21
Fig. 1-12: CAT 100 Exterior view......................................................................................................... 1-25
Fig. 1-13: CAT 100 (Dome and extender housing removed).............................................................. 1-26
Fig. 1-14: CAT 100 Front panel ........................................................................................................... 1-27
Fig. 1-15: CAT 100 Junction box, interior view................................................................................... 1-28
Fig. 1-16: CAT 100 Gas fittings and assignment ................................................................................ 1-29
Fig. 1-17: Single channel or serial tubing............................................................................................ 1-31
Fig. 1-18: Dual channel parallel tubing................................................................................................ 1-31
Fig. 1-19: Single channel resp. serial tubing, with options.................................................................. 1-32
Fig. 1-20: Parallel tubing, with options ................................................................................................ 1-33
Fig. 1-21: BINOS
®
100 (M) / OXYNOS
®
100 / HYDROS
®
100 dimensions........................................ 1-36
Fig. 1-22: Pin assignment 24 V DC input plug.................................................................................... 1-37
Fig. 1-23: Socket X2 pin assignment (analog signal outputs; standard) ............................................1-37
Fig. 1-26: Plug Analog IN pin assignment (cross compensation; option).......................................... 1-38
Fig. 1-25: Plug X3 pin assignment (digital outputs; standard) ............................................................ 1-38
Fig. 1-24: Connecting digital outputs................................................................................................... 1-38
Fig. 1-27: Socket X4 pin assignment (serial interface; option) ......................................................... 1-39
Fig. 1-28: Relay status signals, block diagram................................................................................... 1-40
Fig. 1-29: Plug X1 pin assignment (relay status signals; option)........................................................ 1-40
Fig. 1-30: BINOS
®
100 2M dimensions............................................................................................... 1-41
Fig. 1-31a: Pin assignment 24 V DC input plug.................................................................................... 1-42
Fig. 1-31b: Pin assignment 24 V DC output socket.............................................................................. 1-42
Fig. 1-31c: IEC mains input plug........................................................................................................... 1-42
Fig. 1-32: Socket X2 pin assignment (analog signal outputs; standard) ............................................1-43
Fig. 1-34: Plug Analog IN pin assignment (cross compensation; option).......................................... 1-44
Fig. 1-34: Plug X3 pin assignment (digital outputs; standard) ............................................................ 1-44
Fig. 1-33: Connecting digital outputs................................................................................................... 1-44
Fig. 1-36: Socket X4 pin assignment (serial interface; option) ......................................................... 1-45
Fig. 1-38: Plug X1 pin assignment (relay status signals; option)........................................................ 1-46
Fig. 1-37: Relay status signals, block diagram................................................................................... 1-46
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278

Rosemount BINOS 100 Series Analyzers including OXYNOS 100, HYDROS 100 and CAT 100-2nd Ed. Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI