Vantage Startup Guide 003-10353-00 Rev. AA1 Page 1 of 2
System Function Checklist April 2020
System Function Checklist
CAUTION
Do not use the device if it cannot perform the functions in this checklist.
Task Area Description
P
Floor and Wall
Ensure that the wall support is adequate and that the system is securely attached to the wall
and oor. For freestanding units, the structure must be rmly attached to the oor.
Mechanical Safety
Inspect the structure for mechanical integrity of the telescoping column. Inspect the main
drive lead screw, bearings, overhead attachment and pivot points, C-arm attachment and
pivot points, and the patient table attachment.
Electrical Safety Verify the integrity of the power line, its connections, and the connection to earth ground.
Consumable Items
Verify that these are present and in working order:
For panoramic systems: chin rest, bite piece, TMJ positioner, sheaths, wands
For cephalometric systems: naison post, ear post, ear speculum
If replacements are needed, see the Consumable Items for Panoramic and Cephalometric
Systems section of the Vantage user guide.
Diagnostic Source
Assembly
Under the tubehead cover, evaluate the condition of these items: the collimator attachment,
the collimator blade mechanism, the electrical connections, and the mechanical attachment
of the tubehead and collimator to the structure. Check for oil leaks.
Power Switch
Verify that the switch is working and that the operator panel illuminates when the power
switch is in the on position.
Operator Panel / Touch
Control Panel
A few seconds after power up, the splash screen should appear. The screen should display
the status when the Vantage Client software is started. The panel must switch to the main
screen when the VantageTouch software and all related devices are operational. The panel
functions must respond to interrogation.
Movement Control
Switches
Verify wand positioning and column vertical drive.
Laser Operation and
alignment
The mid-sagittal, Frankfort plane, and cuspid lasers operate on demand by way of the activa-
tion switches on the patient table. Verify alignment through the use of the laser test xtures.
Exposure Switch
Verify that the exposure switch is functioning properly. Make an exposure, follow the proce-
dures outlined in the user guide.
checklist continued...