Thermo Fisher Scientific Pierce Biotin Coated Clear 8-Well Strip Plates User guide

Type
User guide
INSTRUCTIONS
Pierce Biotechnology
PO Box 117
(815) 968-0747
www.thermoscientific.com/pierce
3747 N. Meridian Road
Rockford, lL 61105 USA
(815) 968-7316 fax
Number Description
15151 Pierce Biotin Coated Clear 8-Well Strip Plates, 5 plates
Coat volume: 200µL
Blocking volume: 300µL
Storage: Upon receipt store desiccated at 4°C. Product is shipped at ambient temperature.
Introduction
Biotin binds with high affinity to avidin, streptavidin and Thermo Scientific NeutrAvidin Protein. The bond formation is
rapid and, once formed, is unaffected by most pH extremes, organic solvents and other denaturing agents. The strength and
specificity of the avidin-biotin interaction makes it useful for developing various assays. Avidin-biotin chemistry is an
effective way to increase sensitivity in several types of immunoassays including ELISA, blotting and immunohistochemical
applications. Most of these assays use conjugates of avidin, streptavidin or NeutrAvidin® Protein with enzymes (such as
horseradish peroxidase or alkaline phosphatase) or fluorophores (FITC). Also, biotinylated secondary antibodies provide
versatility and increased sensitivity via signal amplification.
Procedure for Testing Enzyme Conjugates
A. Additional Materials Required
Phosphate-buffered saline (PBS; 25mM phosphate, 150mM sodium chloride; pH 7.2; Product No. 28372)
Horseradish peroxidase (HRP) conjugated to avidin (Product No. 21123), streptavidin (Product No. 21126) or
NeutrAvidin Protein (Product No. 31001)
Thermo Scientific 1-Step Ultra TMB-ELISA (Product No. 34028)
Stop solution: 2M H2SO4
B. Method
1. Rinse each well three times with 200µL of PBS.
2. Add 100µL of the HRP conjugate at 0.2-1µg/mL.
3. Incubate the plate at room temperature for 30 minutes.
4. Wash each well three times with 200µL of PBS.
5. Add 100µL the TMB substrate and incubate at room temperature for 10 to 30 minutes.
6. Add 100µL of stop solution. Measure the absorbance at 450nm.
Related Thermo Scientific Products
37070 SuperSignal® ELISA Pico Chemiluminescent Substrate, 100mL
37515 SuperBlock® (PBS) Blocking Buffer, 1L
15169 QuantaBlu Fluorogenic Peroxidase Substrate Kit
15159 QuantaRed Enhanced Chemifluorescent HRP Substrate
0710.2
15151
Pierce® Biotin Coated Clear 8-Well
Strip Plates
Pierce Biotechnology
PO Box 117
(815) 968-0747
www.thermoscientific.com/pierce
3747 N. Meridian Road
Rockford, lL 61105 USA
(815) 968-7316 fax
37621 1-Step PNPP colorimetric AP substrate, 100mL
15075 Reagent Reservoirs, 200/pkg
15082 Microtube Racked System, 960 tubes
15036 Sealing Tape for 96-Well Plates, 100/pkg
34028 1-Step Ultra TMB-ELISA, 250mL
21123 Avidin, Horseradish Peroxidase Conjugated, 2mg
21126 Streptavidin, Horseradish Peroxidase Conjugated, 2mg
31001 NeutrAvidin, Horseradish Peroxidase Conjugated, 2mg
This product (“Product”) is warranted to operate or perform substantially in conformance with published Product specifications in effect at the time of sale,
as set forth in the Product documentation, specifications and/or accompanying package inserts (“Documentation”) and to be free from defects in material and
workmanship. Unless otherwise expressly authorized in writing, Products are supplied for research use only. No claim of suitability for use in applications
regulated by FDA is made. The warranty provided herein is valid only when used by properly trained individuals. Unless otherwise stated in the
Documentation, this warranty is limited to one year from date of shipment when the Product is subjected to normal, proper and intended usage. This
warranty does not extend to anyone other than the original purchaser of the Product (“Buyer”).
No other warranties, express or implied, are granted, including without limitation, implied warranties of merchantability, fitness for any particular
purpose, or non infringement. Buyer’s exclusive remedy for non-conforming Products during the warranty period is limited to replacement of or
refund for the non-conforming Product(s).
There is no obligation to replace Products as the result of (i) accident, disaster or event of force majeure, (ii) misuse, fault or negligence of or by Buyer, (iii)
use of the Products in a manner for which they were not designed, or (iv) improper storage and handling of the Products.
Current product instructions are available at www.thermoscientific.com/pierce. For a faxed copy, call 800-874-3723 or contact your local distributor.
© 2012 Thermo Fisher Scientific Inc. All rights reserved. Unless otherwise indicated, all trademarks are property of Thermo Fisher Scientific Inc. and its
subsidiaries. Printed in the USA.
  • Page 1 1
  • Page 2 2

Thermo Fisher Scientific Pierce Biotin Coated Clear 8-Well Strip Plates User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI