NOTE For time critical situations, you can switch the system o faster by pressing the System
o button for at least 3 seconds (shutdown time in this case is 3 seconds). This may
cause data to be lost and may increase the system startup time (up to 5 minutes) for the
next start.
2Wait 5 seconds.
3Press the System on button on the mobile view station.
The startup time for radioscopy imaging could be higher in case of abrupt shut down.
The system starts with default settings and a new patient.
4Select the previously used examination type and acquisition mode.
When the procedure is nished, you can modify the patient name and other administration items for
the current acquisition patient. You cannot continue the previous examination.
2.3 Reporting a Serious Incident
If a serious incident occurs in relation to the device, it should be reported to the manufacturer and the
competent authority of the country where you are located.
For contact details, see Contacting the Manufacturer (page 15)
A serious incident means any incident that directly or indirectly led to, might have led to, or in case of
recurrence, could lead to, any of the following:
• The serious deterioration of the state of health (temporary or permanent) of a patient, user or other
person.
• A serious public health threat.
2.4 Electrical Safety
System covers or cables should only be removed by the qualied and authorized service personnel.
In this context, qualied means ‘those legally permitted to work on this type of medical electrical
equipment in the jurisdiction(s) where the equipment is being used’, and authorized means ‘those
authorized by the responsible organization’.
WARNING
Do not remove system covers or cables from this equipment, unless expressly instructed to do so in
this manual. High electrical voltages are present within this equipment. Removing system covers or
cables could lead to serious or fatal personal injury.
WARNING
Do not touch the pins of the mobile view station C-arm cable or the central pin of the video/USB
connectors when touching the patient. Connector contact pins may carry low voltages that are safe
to touch, but that may be harmful to the patient.
WARNING
Do not touch the pins of the mobile view station C-arm cable plug directly after the system is
powered o, some residual voltage may remain.
WARNING
In case of changing the X-ray on indicator light on the mobile view station, do not touch the
electrical contacts and the patient simultaneously. Connector contact pins may carry low voltages
that are safe to touch, but that may be harmful to the patient.
Safety Reporting a Serious Incident
Zenition 30 Release 1.0 Instructions for Use 18 Philips 3000 095 56112