Invacare Perfecto2 IRC5PO2VAW User manual

Category
Oxygen Equipment
Type
User manual

This manual is also suitable for

Invacare®Perfecto
Oxygen
Concentrator
IRC5PO2AW,IRC5PO2VAW
enHomeFill®OxygenSystemCompatible
ServiceManual
DEALER:Keepthismanual.
TheproceduresinthismanualMUSTbeperformedbyaqualified
technician.
©2016Invacare®Corporation
Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithoutpriorwritten
permissionfromInvacare.Trademarksareidentifiedby™and®.AlltrademarksareownedbyorlicensedtoInvacare
Corporationoritssubsidiariesunlessotherwisenoted.
MakingLife’sExperiencesPossibleisaregisteredtrademarkintheU.S.A.
TeflonisaregisteredtrademarkofDuPontCorp.
TorxisaregisteredtrademarkofTextron,Inc.
RexRothisaregisteredtrademarkofBOSCHREXROTHAG,GERMANY
PhillipsisaregisteredtrademarkofPhillipsScrewCompany.
WD-40and3–IN-ONEareregisteredtrademarksofWD-40Company.
Contents
1General.......................................4
1.1Symbols.....................................4
1.2LimitedWarranty—EuropeandAustralia.............5
1.3ComponentIdentification.........................5
2Safety.........................................6
2.1LabelLocations................................6
2.2GeneralGuidelines.............................6
2.3RadioFrequencyInterference......................8
3Installation.....................................9
3.1VerificationofBatteryFreePowerLossAlarm..........9
3.2SequenceofOperation..........................9
3.3SensO
2
OxygenSensorTechnology-CeramicZirconia
Sensor.....................................9
3.3.1TechnicalDescription.........................9
3.3.2OperatingSequence..........................9
3.4SensO
2
AlarmIndicators..........................10
4PneumaticDiagram.............................11
4.1PneumaticDiagram.............................11
5Maintenance...................................12
5.1CleaningtheCabinetFilter........................12
5.2CheckingtheOutletHEPAFilter...................12
5.3ReplacingtheOutletHEPAFilter...................13
5.4ReplacingtheCompressorInletHEPAFilter...........13
5.5ReplacingtheMufflerAssembly.....................14
5.6CleaningtheHeatExchanger......................15
5.7PreventiveMaintenanceChecklist...................16
6Service........................................17
6.1RemovingCabinet..............................17
6.2ReplacingCompressorAssembly...................18
6.2.1ReplacingCompressorAssemblyBeforeSeptember
2015....................................18
6.2.2ReplacingCompressorAssemblyAfterSeptember
2015....................................20
6.3ReplacingCapacitor.............................21
6.3.1ReplacingCapacitorBeforeSeptember2015........21
6.3.2ReplacingCapacitorAfterSeptember2015.........22
6.4ReplacingP.E.Valve.............................22
6.5ReplacingSieveBeds............................24
6.5.1ReplacingSieveBeds(IRC5PO2AW)..............24
6.5.2ReplacingSieveBeds(IRC5PO2VAW).............25
6.6ReplacingCheckValves..........................26
6.7ReplacingRegulator.............................27
6.8AdjustingRegulator.............................27
6.9ReplacingHeatExchangerAssembly.................28
6.10Removingand/orReplacingControlPanel............29
6.10.1RemovingControlPanel......................29
6.10.2ReplacingtheControlPanel....................30
6.11ReplacingCoolingFan..........................31
6.12ReplacingP.C.Board...........................32
6.13ReplacingtheTransformer.......................33
6.14ReplacingOn/OffSwitch.........................34
6.15ReplacingFlowmeter...........................35
6.16ReplacingHourMeter..........................36
6.17Replacing4-WayValveand/orManifoldAssembly......37
6.18ReplacingPilotAssembly........................38
7Autotuning....................................40
7.1Autotuning...................................40
8Testing........................................41
8.1LeakTest....................................41
8.2PowerLossAlarm..............................42
8.3LowPressureTest..............................43
8.4Time-OutTest................................43
8.5HighPressureTest.............................44
8.6OxygenSensor................................44
8.7P.E.ValveCoilTest.............................45
8.8CheckingSieveBedPressureIRC5PO2AW............46
8.9CheckingSieveBedPressure(ModelIRC5PO2VAW
Only)......................................47
9Troubleshooting................................48
9.1Troubleshooting...............................48
10Technicaldata.................................54
10.1Specifications.................................54
10.2O2Indicators................................55
11CustomerService..............................56
11.1CustomerServiceandWarrantyInformation..........56
Invacare®PerfectoOxygenConcentrator
1General
1.1Symbols
Signalwordsareusedinthismanualandapplytohazardsorunsafe
practiceswhichcouldresultinpersonalinjuryorpropertydamage.
Seetheinformationbelowfordefinitionsofthesignalwords.
DANGER!
Dangerindicatesanimminentlyhazardoussituation
which,ifnotavoided,willresultindeathorserious
injury.
WARNING!
Warningindicatesapotentiallyhazardoussituation
which,ifnotavoided,couldresultindeathorserious
injury.
CAUTION!
Cautionindicatesapotentiallyhazardoussituation
which,ifnotavoided,mayresultinpropertydamage
orminorinjuryorboth.
Refertothefollowingtableforthemeaning
ofsymbolsmarkedontheequipmentand/or
packaging:
Givesusefultips,recommendationsandinformationfor
efficient,trouble-freeuse.
SymbolsinDocumentation
GeneralWarningSign
Thebackgroundcolorinsidethe
triangleisyellowonproductlabels.
Readmanual
Thecolorofthesymbolbackgroundis
blueonproductlabels.
NoSmoking
Thecolorofthecirclewithdiagonal
barisredonproductlabels.
NoOpenFlame
Thecolorofthecirclewithdiagonal
barisredonproductlabels.
ClassII,DoubleInsulated
Protectedagainstsolidforeignobjectsof12.5
mmdiameterandgreater.
Protectedagainstverticallyfallingwaterdrops.
IndoorUseONLY
KeepDry
65C
MAX
-29C
Min
TransportandStorageTemperature
15%
MIN
95%
MAX
TransportandStorageHumidity
AlternatingCurrent
TypeBFequipment
Recycle
DONOTdisposeofinhouseholdwaste
ElectricalHazard
Manufacturer
ECRepresentative
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.
Thelaunchdateofthisproductisstatedinthe
CEdeclarationofconformity.
FormoredetailsaboutCEmarks,referto10.1Specifications,page54.
SymbolsonProduct
Unitrunning
Unitnotrunning
4
1154245-F-02
General
O
2
Indicators
SymbolO
2
Purity
IndicatorLights
(LED)
O
2
SYSTEMOKAY
O
2
over85%
GREENIndicator
Light
O
2
between73%
to85%
YELLOWIndicator
Light
A.YELLOWSolid
B.YELLOWFlashing
Sensor
Failure
Callaqualified
technician
SYSTEMFAILURE
O
2
Below73%
REDIndicatorLight
ContinuousAudible
AlarmSieve-
GARD™
Compressor
Shutdown
1.2LimitedWarranty—Europeand
Australia
Termsandconditionsofthewarrantyarepartofthegeneralterms
andconditionsparticulartotheindividualcountriesinwhichthis
productissold.
ContactinformationforyourlocalInvacareofficeislocatedinside
thebackcoverofthismanual.
1.3ComponentIdentification
Thisoutletfittingistobeusedonlyforfillingoxygencylinders
withtheHomeFillhomeoxygencompressor.Theoutlet
fittingdoesnotaffectconcentratorperformance.Refer
totheHomeFillowner'smanual,partnumber1145804,
(Northern,1145805,Southernor1145806,Scandinavianfor
connectionandoperatinginstructions.Whennotinuse,the
plugprovidedwiththeconcentratorshouldbeinsertedinto
theoutletfitting.FormoreinformationabouttheHomeFill,
contactyourInvacaredealer.
A
OxygenOutlet
B
Flowmeter
C
ElapsedTimeMeter
D
OxygenPurityIndicator
Lights/FaultandPowerIndicator
Lights
E
CircuitBreaker
F
PowerSwitch
G
PowerCord
H
OutletFitting
I
CabinetFilter
NotshownPreciseRX™Pediatric
Humidifier/Flowmeter
AccessoryIRCPF16AW
NotshownHomeFillhomeoxygen
compressorIOH200AW
1154245-F-025
Invacare®PerfectoOxygenConcentrator
2Safety
2.1LabelLocations
A
Serialnumberlabelislocatedontheresonatorintake
assembly.
B
SpecificationLabelislocatedonthebackofthe
concentratoratthebase.
C
EURepresentativelabelislocatedonthebase.
D
Phthalatelabelislocatedonthebase.
E
2.2GeneralGuidelines
DANGER!
RiskOfDeath,Injury,OrDamage
Improperuseoftheproductmaycausedeath,injuryor
damage.Thissectioncontainsimportantinformationfor
thesafeoperationanduseofthisproduct.
DONOTusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand
understandingtheseinstructionsandanyadditional
instructionalmaterialsuchasusermanuals,service
manualsorinstructionsheetssuppliedwiththis
productoroptionalequipment.
Ifyouareunabletounderstandthewarnings,cautions
orinstructions,contactahealthcareprofessional,
dealerortechnicalpersonnelbeforeattemptingtouse
thisequipment.
CheckALLexternalcomponentsandcartonfor
damage.Incaseofdamage,oriftheproductisnot
workingcorrectly,contactatechnicianorInvacare
forrepair.
THEINFORMATIONINTHISDOCUMENTIS
SUBJECTTOCHANGEWITHOUTNOTICE.
DANGER!
RiskOfDeath,Injury,OrDamageFromFire
Textiles,oilorpetroleumsubstances,grease,greasy
substancesandothercombustiblesareeasilyignitedand
burnwithgreatintensityinoxygenenrichedairand
whenincontactwithoxygenunderpressure.Toavoid
fire,death,injuryordamage:
DONOTSMOKEwhileusingthisdevice.
DONOTusenearOPENFLAMEorIGNITION
SOURCES.
DONOTuseanylubricantsonconcentratorunless
recommendedbyInvacare.
NOSMOKINGsignsshouldbeprominentlydisplayed.
Avoidcreationofanysparknearoxygenequipment.
Thisincludessparksfromstaticelectricitycreatedby
anytypeoffriction.
Keepallmatches,lightedcigarettes,electronic
cigarettesorothersourcesofignitionoutoftheroom
inwhichthisconcentratorislocatedandawayfrom
whereoxygenisbeingdelivered.
Keeptheoxygentubing,cord,andconcentratorout
fromundersuchitemsasblankets,bedcoverings,
chaircushions,clothing,andawayfromheatedorhot
surfacesincludingspaceheaters,stoves,andsimilar
electricalappliances.
DANGER!
RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,deathor
injurytopersons:
DONOTdisassemble.Referservicingtoqualified
servicepersonnel.Therearenouserserviceableparts.
Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the
concentratormustbelocatedinanotherroomatleast
7ft(2.1m)fromthebath.
DONOTcomeincontactwiththeconcentrator
whilewet.
DONOTplaceorstoreconcentratorwhereitcan
dropintowaterorotherliquid.
DONOTreachforconcentratorthathasfalleninto
water.UnplugIMMEDIATELY.
DONOTusefrayedordamagedACpowercords.
61154245-F-02
Safety
DANGER!
RiskofDeath,Injury,fromElectricShock
Toreducetheriskofburns,electrocution,deathor
injurytopersons:
DONOTdisassemble.Referservicingtoqualified
servicepersonnel.Therearenouserserviceableparts.
Avoidusingwhilebathing.Ifcontinuoususage
isrequiredbythephysician'sprescription,the
concentratormustbelocatedinanotherroomatleast
7ft(2.1m)fromthebath.
DONOTcomeincontactwiththeconcentrator
whilewet.
DONOTplaceorstoreconcentratorwhereitcan
dropintowaterorotherliquid.
DONOTreachforconcentratorthathasfalleninto
water.UnplugIMMEDIATELY.
DONOTusefrayedordamagedACpowercords.
WARNING!
RiskOfInjuryOrDamage
Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare
accessories.Accessoriesdesignedbyother
manufacturershavenotbeentestedbyInvacareand
arenotrecommendedforusewithInvacareproducts.
Therearemanydifferenttypesofhumidifiers,oxygen
tubing,cannulasandmasksthatcanbeusedwiththis
device.Youshouldcontactyourlocalhomecare
supplierforrecommendationsonwhichofthese
deviceswillbebestforyou.Theyshouldalsogiveyou
adviceontheproperusage,maintenance,andcleaning.
DANGER!
RiskofInjuryorDeath
Toavoidchokingoringestionofchemicalsfromairway
contamination:
DONOTusetheconcentratorinthepresence
ofpollutants,smoke,fumes,flammableanesthetics,
cleaningagents,orchemicalvapors.
WARNING!
RiskofInjuryorDeath
Topreventinjuryordeathfromproductmisuse:
Closelysupervisewhenthisconcentratorisusedbyor
nearchildrenorimpairedindividuals.
Monitorpatientsusingthisdevicewhoareunableto
hearorseealarmsorcommunicatediscomfort.
WARNING!
RiskofInjuryorDeath
Toavoidchokingand/orstrangulationfromtubing
entanglement:
Keepchildrenandpetsawayfromnasalcannulaand
tubing.
Closesupervisionisnecessarywhenthenasalcannula
isusedbyornearchildrenand/orimpairedpersons.
WARNING!
RiskofInjuryorDeath
Toreducetheriskofinjuryordeathfromillness:
Replacethenasalcannulaonaregularbasis.Check
withyourequipmentsupplierorphysicianto
determinehowoftenthecannulashouldbereplaced.
DONOTsharecannulasbetweenpatients.
WARNING!
RiskofInjury
Achangeinaltitudemayaffecttotaloxygenavailableto
you.Topreventoxygendeprivation:
Consultyourphysicianbeforetravelingtohigheror
loweraltitudestodetermineifyourflowsettings
shouldbechanged.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamagefromcordmisuse:
DONOTmoveorrelocateconcentratorbypulling
onthecord.
DONOTuseextensioncordswithACpowercords
provided.
Properlystoreandpositionelectricalcordsand/or
tubingtopreventatrippinghazard.
WARNING!
RiskofInjuryorDamage
Topreventinjuryordamagefrommisuse:
NEVERleaveconcentratorunattendedwhenplugged
in.
Makesureconcentratorisoffwhennotinuse.
WARNING!
RiskofInjuryorDamage
Invacareoxygenconcentratorsarespecificallydesigned
tominimizeroutinepreventivemaintenance.Toprevent
injuryordamage:
Onlyprofessionalsofthehealthcarefieldor
personsfullyconversantwiththisprocesssuchas
factorytrainedpersonnelshouldperformpreventive
maintenanceorperformanceadjustmentsonthe
oxygenconcentrator,exceptfortasksdescribedin
thismanual.
UsersshouldcontactyourdealerorInvacarefor
service.
CAUTION!
RiskofDamage
Topreventdamagefromliquidingress:
Iftheconcentratorisnotworkingproperly,ifit
hasbeendroppedordamaged,ordroppedinto
water,callequipmentsupplier/qualifiedtechnicianfor
examinationandrepair.
NEVERdroporinsertanyobjectorliquidintoany
opening.
ForindooruseONLY.
WARNING!
Perfecto
2
PartsCompatibility
ThePlatinumconcentratorandthePerfecto
2
concentratorsharemanysimilarcomponents.
However,therearesomecomponentsthatareNOT
crosscompatibleandtheyarespecifictoeachmodel.
PartsthatcanbeusedforPerfecto
2
repairsarelisted
inthePerfecto
2
partscatalog,formnumber93_038.
Pleasereferencethiscatalogbeforereplacingany
parts.Useofincorrectpartscouldresultininjuryor
propertydamage.
1154245-F-027
Invacare®PerfectoOxygenConcentrator
WARNING!
RiskofInjury
Itisveryimportanttoselecttheprescribedlevelof
oxygenflow.Tovoidinjuryfromoxygendeprivation:
DONOTincreaseordecreasetheflowunlessa
changehasbeenprescribedbyyourphysicianor
therapist.
ALWAYSconfirmprescribeddosebefore
administeringtopatientandmonitoronafrequent
basis.
CAUTION!
Riskofdamage
Shorterperiodsofoperationmayreducemaximum
productlife.Foroptimumperformance:
Theconcentratorshouldbeonandrunningfora
minimumof30minutesatatime.
2.3RadioFrequencyInterference
WARNING!
RiskofInjuryorDamage
Toreducetheriskofinjuryorproductdamagefrominterferencewithwirelessequipment:
Keepconcentratoratleast9.8ft(3.0m)awayfromwirelesscommunicationequipmentsuchaswirelesshomenetworkdevices,
mobilephones,cordlessphonesandbasestations,walkie-talkies,etc.
ThisequipmenthasbeentestedandfoundtocomplywithEMClimitsspecifiedbyIEC/EN60601-1-2.Theselimitsaredesignedtoprovidea
reasonableprotectionagainstelectromagneticinterferenceinatypicalmedicalinstallation.
Otherdevicesmayexperienceinterferencefromeventhelowlevelsofelectromagneticemissionspermittedbytheabovestandards.Todetermine
iftheemissionsfromtheconcentratorarecausingtheinterference,turntheconcentratorOff.Iftheinterferencewiththeotherdevice(s)stops,
thentheconcentratoriscausingtheinterference.Insuchrarecases,interferencemaybereducedorcorrectedbyoneofthefollowingmeasures:
Reposition,relocate,orincreasetheseparationbetweentheequipment.
Connecttheequipmentintoanoutletonacircuitdifferentfromthattowhichtheotherdevice(s)isconnected.
81154245-F-02
Installation
3Installation
3.1VerificationofBatteryFreePowerLossAlarm
Checktheconcentratorforproperoperatingconditions.
1.Iftheunithasbeeninbelow-freezingtemperatures,allowittowarmuptoroomtemperaturebeforeoperating.
2.TheconcentratormayneedtobeturnedonforfourtofivesecondstochargetheBatteryFreePowerLossAlarm.Connectpowercord
tooutletandturntheconcentratoron.TurnflowcontrolknobcounterclockwiseandflowwillbeginIMMEDIATELY.Setflowrate
tofiveL/min.Lettheunitrunfor30minutes,thenturnunitoff.
3.UnplugthepowercordandpressOn/Off(I/O)switchtotheOn(I)position.Anintermittentaudiblealarmwillsound.Thisconfirms
properoperationoftheBatteryFreePowerLossAlarm.TurnOn/OffswitchOff(O).
4.Connectpowercordtooutletandturnonconcentrator.Unitwillbeeponstart-up.
5.Checktheoxygenconcentrationperspecificationsafter30-40minutesrunningtime.
3.2SequenceofOperation
TurningOn(I)thepowerswitchapplies230VACtothecompressormotor,hourmeter,transformer,coolingfanandtheprintedcircuit
(P.C.)board.
Roomairentersthecompressorviathecabinetfilterandthecompressorinletfilter.TheairiscompressedbytheWob-Lpistonsinthe
compressortoapressureof144.79kPa(21p.s.i.).
Asincreasedpressurecreatesincreasedtemperature,aheatexchangerisusedtolowerthetemperaturebeforetheairentersthe4-way
valve.Itisthenchanneledtoasievebedcontainingtheadsorptionmaterial.Restrictiondownstreamofthesievebedcausespressureto
buildupinsidethesievebedwhichisnecessaryfortheadsorptionprocess.Asmallamountofrelativelypureoxygenentersthetopofthe
secondbedthrougharestrictiveleakinthepressureequalization(P.E.)valvewiththebalanceenteringastoragetank.Thenitrogenremoved
isexhaustedbackfromthebedthroughthe4-wayvalveintoroomair.Amufflerislocatedattheexhaustendofthevalvetomufflethe
soundoftheexhaustasitexitstheconcentrator.
Theoxygennotbeingusedtoexhaustischanneledintothestoragetank.Thepressurizedoxygenisregulateddownto5p.s.i.(34.4kPa),enters
anaccurateflow-measuringdevice,flowsthroughtheoutletHEPAfilterandcheckvalve,thenouttothepatient.
Theelectricalactivationofthe4-wayValveisaccomplishedevery8to15secondsbythepressuresensorandP.C.Boardelectronicswhenthe
pressurereachesasetpointof144.79kPa(21p.s.i.)outputflows4L/minandaboveor16p.s.i.(110.32kPa)outputflows3L/minandbelow.The
timebetweencyclesisdependentonaltitude,flowrateandinternalenvironmentalfactors.
AP.E.valveopensjustpriortotheshiftofthe4-wayvalve.Thisallowshighlyconcentratedoxygentoenterthejustexhaustedbedfromthetop.
Thisadditionalpressureallowsthebedtostartitscycleatahigherpressure.TheP.E.valvewillclosejustaftertheshiftofthe4-wayvalve.
Ifmainpowerislost,theBatteryFreePowerLossAlarmwillsoundashort"BEEP",withalongpauseafter.Allunitsareequippedwithadiagnostic
alarmsystemthatsignalsifthepneumaticpressureorelectricalsystemsmalfunction.Thetroubleshootingguideinthismanualexplainsthe
alarmsystemsignalsandreasons,indetail.RefertoTroubleshooting.
3.3SensO
2
OxygenSensorTechnology-CeramicZirconiaSensor
3.3.1TechnicalDescription
Theoxygenbeingproducedbytheconcentratorflowsoutoftheproducttankandintotheflowmeter.Asmallflowofoxygenproducedbythe
unitissentthroughaprecisionorificetotheoxygensensormountedontheprintedcircuitboard.
Astheoxygenentersthesensor,itpassesthroughascreenandcontactsthesensingdisk.
Electriccurrentflowingthroughametalfilmresistorheatsthediskinexcessof300°C.
Oxygenmoleculescontacttheelectrodeofthediskandpick-upextraelectronstobecomeoxygenions.Theseoxygenionsareattractedto
theelectrodeonthebottomofthezirconiasensingdisk.Becauseofthecrystalstructureofthezirconia,onlyoxygenionscanpassthrough.
Whentheoxygenionsreachthebottomelectrode,theextraelectronsarereleasedfromtheoxygenionsandoxygenmoleculesreturntothe
air.Thenumberofelectronsisdirectlyrelatedtotheoxygenconcentration.TheelectronstraveltotheP.C.boardwheretheyarecounted
andtheoxygenconcentrationreadingiscalculated.
AmicroprocessorontheP.C.boardcontainssoftwarethatinterpretsthesignalbeingreceivedfromthesensor.Itcomparesthesignalto
clinicallyacceptablelimits.Signalsoutsideoftheclinicallyacceptablelimitsgenerateresponsesintheformoflights,audibleindicators,and/or
systemshut-down.
3.3.2OperatingSequence
OncethepowerswitchhasbeenturnedOn(I),theSensO
2
circuitwillwaitfiveminutesfortheconcentratortobeginproducingclinically
acceptableoxygenandtheoxygensensortostabilize.TheGREENlightwillilluminate(indicatingnormalsystemoperation)whiletheoxygen
sensoriswarmingup.
Afterfiveminutes,iftheoxygenpurityexceeds85%±2%,theGREENlightwillcontinuetoilluminate.
1154245-F-029
Invacare®PerfectoOxygenConcentrator
Iftheoxygenlevelisnotabove85%±2%afterthefirstfiveminutes,thesystemwillcontinuetomonitortheO
2
andwaitforamaximumof
30minutesfromstart-uptoreach85%±2%beforeactivatinganalarm.Environmentalfactorssuchaslowvoltage,highaltitude,orage
ofthemachinewillaffectthetimerequiredtoreach85%±2%.
Iftheoxygenlevelisnotabove85%±2%withinthefirst30minutes,theoxygenconcentrationalarmsequencewillactivateandtheunit
willshutdown.
Whenoxygenconcentrationisabove85%±2%,thesensormeasuresoxygenpurityevery10minutes.Ifareadingfallsbelow85%±2%,a
YELLOWlightwillilluminate.Iftheoxygenpurityfallsbelow73%±3%theREDlight/Alarm/Shut-Downmodewillactivate.
3.4SensO
2
AlarmIndicators
DANGER!
RiskofDeath,InjuryorDamage
Topreventelectricalshock
ALWAYSdisconnectconcentratorfromelectricaloutletbeforeservicing.
YELLOWlightB
CallsupplierIMMEDIATELY.Youmay
continuetousetheconcentratorunless
instructedotherwisebyyoursupplier.
Becertainthatbackupoxygenisnearby.
GREENlightA
(O
2
)-NormalOperation.
REDlightC
Totalunitshutdown.Switch
IMMEDIATELYtobackupoxygensupply.
CallsupplierIMMEDIATELY.
GREENlightA
withYELLOWlight
Bflashing
CallsupplierIMMEDIATELY.Oxygen
sensormalfunctioning;youmaycontinue
tousetheconcentrator.Indicators-If
yourunitdoesnotfeaturetheO
2
Sensor.
Operation
Theconcentratorisequippedwithanoxygenpurityindicator.Thisfeaturemonitorsthepurityleveloftheoxygengeneratedbytheconcentrator.
Ifpurityfallsbelowalarmthresholds,indicatorlightsonthecontrolpanelwillilluminate.
Concentratormaybeusedduringtheinitialstartwarm-uptime(approximately30minutes)whilewaitingfortheO
2
puritytoreach
maximum.
Whentheunitisturnedon,theGREENlightwillcomeon(SYSTEMOK/O
2
greaterthan85%).Afterfiveminutes,theoxygensensorwillbe
operatingnormallyandwillcontroltheindicatorlightsdependingonoxygenconcentrationvalues.Theexplanationoftheindicatorlight
functionsareasfollows:
IndicatorLights
Indicatorlightsarevisibleonlywhenlit.
101154245-F-02
PneumaticDiagram
4PneumaticDiagram
4.1PneumaticDiagram
1154245-F-02
11
Invacare®PerfectoOxygenConcentrator
5Maintenance
5.1CleaningtheCabinetFilter
DANGER!
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutletbeforeservicing.
Ataminimum,preventivemaintenanceMUSTbeperformedaccordingtothemaintenancerecordguidelines.Inplaceswithhighdustor
sootlevels,maintenancemayneedtobeperformedmoreoften.
Refertorecyclinginformationfollowinglocalgoverningordinancesandrecyclingplansregardingdisposaloftheconcentratoror
components.
CAUTION!
RiskofDamage
Toavoiddamagetotheinternalcomponentsoftheunit:
DONOToperatetheconcentratorwithoutthefilterinstalledorwithadirtyfilter.
Thereisonecabinetfilterlocatedonthebackofthecabinet.
1.RemovetheairscoopAfrombackofunit.
2.RemovethefilteraccesspanelBbypressingdownoncentertabCandpullingpanelout.
3.RemovecabinetfilterDfromfilteraccesspanelandcleanasneeded.
Environmentalconditionsthatmayrequiremorefrequentinspectionandcleaningofthefiltersincludebutarenotlimitedto:
highdust,airpollutants,etc.
4.Cleanthecabinetfilterwithavacuumcleanerorwashwithamildliquiddishdetergentandwater,rinsethoroughly.
5.Thoroughlydrythefilterandinspectfromfraying,crumbling,tearsandholes.Replacefilterifanydamageisfound.
6.Reinstallcabinetfilterontofilteraccesspanel
7.Reinstallfilteraccesspanelintounit.
8.Reinstallairscoopontobackofunit.
5.2CheckingtheOutletHEPAFilter
TheoutletHEPAfiltercanbecheckedduringpreventivemaintenanceorbetweenpatientsbyperformingthefollowingprocedure.
1.TurntheconcentratorOn(I)andadjusttheflowmetertothemaximumflowrateoftheunit.
2.Observetheflowmeter'sflowindicatorwhileconnectinga15.2m(50ft)cannulatubetotheoutletbarboftheconcentrator(notshown).
3.Iftheflowindicatorfluctuates,theoutletHEPAfiltermayneedreplacement.Referto9.1Troubleshooting,page48
12
1154245-F-02
Maintenance
5.3ReplacingtheOutletHEPAFilter
PerformthisprocedureduringpreventivemaintenanceorbetweenpatientsdependingontheoutletHEPAfiltercheckresults.
1.TurnpowerOff(O)andunplugtheunit.
2.Removethecabinet.Referto6.1RemovingCabinet,page17.
3.ForeasieraccesstotheoutletHEPAfilterA,removethecompressorinletHEPAfilterbygraspingthecompressorinletfilter,pulloutward
andupuntilfilterisdislodgedfromrubberbase.
4.Withaflatheadscrewdriver,removethetubingBfrombothsidesoftheexistingoutletHEPAfilterbypryingthetubingawayfrom
theoutletHEPAfilter.
5.DiscardexistingoutletHEPAfilter.
6.ConnecttubingtobothsidesofthenewoutletHEPAfilterasshownbelow.
7.Rununitandinspectforleaks.Referto8.1LeakTest,page41.
8.Ifnoleaksdetected,turnpowerOff(O)andunplugtheunit.
9.Reinstallthecabinet.Referto6.1RemovingCabinet,page17andreverseallsteps.
5.4ReplacingtheCompressorInletHEPAFilter
Performthisprocedureduringpreventivemaintenanceorbetweenpatientsdependingupontheenvironmenttheconcentratorisusedin.
1.TurnpowerOff(O)andunplugtheunit.
2.RemovetheairscoopA.
3.RemovethefilteraccesspanelBbypressingdownoncentertabCandpullingpanelout.
4.GraspthecompressorinletHEPAfilterD,pulloutwardandupuntilfilterisdislodgedfromrubberbaseE.
5.DiscardtheexistingcompressorinletHEPAfilter.
6.Installnewcompressorinletfilterbyinsertingthefilterintotherubberbase.
7.Pushthefilterdownuntiltherubberbasetouchestheedgeofthefilter.
8.Reinstallthefilteraccesspanel.
9.Reinstalltheairscoop.
1154245-F-0213
Invacare®PerfectoOxygenConcentrator
5.5ReplacingtheMufflerAssembly
Performthisprocedureduringpreventivemaintenanceorbetweenpatientsdependingupontheenvironmenttheconcentratorisusedin.
IRC5PO2AW
Filter,capacitor,andPEvalvenotshown. PEvalvenotshown.
IRC5PO2VAW
Filter,capacitor,andPEvalvenotshown. PEvalvenotshown.
1.TurnpowerOff(O)andunplugtheunit.
2.Removecabinet.Referto6.1RemovingCabinet,page17.
3.CuttiewrapAthatsecuresthebottomofthemanifoldexhausttubeBtothe“F”tubeC.
4.Removethebottomofthemanifoldexhausttubefromthe“F”tube.
5.DisconnectandremovethetopofthemanifoldexhausttubeDfromthegrommetE.
6.PushthemufflerassemblydownthroughtheopeninginthesoundboxF.
7.CutthetwotiewrapsGaroundthemufflerassembly.
8.Separatethemufflerassembly.
14
1154245-F-02
Maintenance
9.Inspectthecomponentstoensuretheyareclear.
10.Dooneofthefollowing:
ReplaceexhaustcanisterHandthrottlingmufflerIifpluggedorrestricted.
Installonlynewthrottlingmufflerifexhaustcanisterisingoodcondition.
11.Installnew/existingmufflerassemblybyreversing3-8.
12.Rununitandinspectforleaks.Referto8.1LeakTest,page41.
13.Ifnoleaksdetected,turnpowerOff(O)andunplugtheunit.
14.Reinstallcabinet.Referto6.1RemovingCabinet,page17andreverseallsteps.
5.6CleaningtheHeatExchanger
DANGER!
RiskofDeath,Injury,orDamage
Topreventelectricalshock
ALWAYSdisconnectconcentratorfromelectricaloutletbeforeservicing.
CAUTION!
RiskofDamage
UsecarenottodeformheatexchangerAwheninstalling,removingorcleaning.
IRC5PO2AWIRC5PO2VAW
PEvalveisnotshown. PEvalveisnotshown.
1.TurnpowerOff(O)andunplugtheunit.
2.Removecabinet.Referto6.1RemovingCabinet,page17.
3.RemoveexcessdirtfromtheheatexchangerAusingcompressedairorvacuum.
4.Reinstallcabinet.Referto6.1RemovingCabinet,page17andreverseallsteps.
1154245-F-0215
Invacare®PerfectoOxygenConcentrator
5.7PreventiveMaintenanceChecklist
ModelNo:
SerialNo:
ONEACHINSPECTION
RecordDateofService
RecordElapsedHourson
HourMeter
CleanCabinetFilter(s)
CheckPrescribedL/min.
FlowRate
DURINGPREVENTIVEMAINTENANCESCHEDULE,ORBETWEENPATIENTS
UNITSWITHSensO
2
EVERY26,280HOURS
CheckOxygen
Concentration*
Clean/ReplaceCabinetFilter
Check/ReplaceOutletHEPA
Filter*
Check/ReplaceCompressor
InletFilter*
CheckPowerLossAlarm
26,820hoursareequivalenttousage24hoursperday,7daysperweekfor3years.
*Tobeconductedbyproviderorqualifiedservicetechnician.Referto8.6OxygenSensor,page44.
161154245-F-02
Service
6Service
6.1RemovingCabinet
DANGER!
RiskofDeath,Injury,orDamage
Topreventelectricalshock
ALWAYSdisconnectconcentratorfromelectricaloutletbeforeservicing.
1.TurnpowerOff(O)andunplugtheunit.
2.RemovethefourmountingscrewsAthatsecurecabinetassemblyBtothebaseassemblyC.
3.Liftthecabinetstraightup.
Whenrequired,vacuuminsideofthecabinetandexposedfoaminsulation.
4.Tore-installcabinet,reverseSTEPS2-3.
1154245-F-0217
Invacare®PerfectoOxygenConcentrator
6.2ReplacingCompressorAssembly
6.2.1ReplacingCompressorAssemblyBeforeSeptember2015
DANGER!
RiskofDeath,Injury,orDamage
Topreventelectricalshock
ALWAYSdisconnectconcentratorfromelectricaloutletbeforeservicing.
Refertorecyclinginformationfollowinglocalgoverningordinancesandrecyclingplansregardingdisposaloftheconcentratoror
components.
IRC5PO2AW
PEvalvenotshown. PEvalvenotshown.
IRC5PO2VAW
PEvalvenotshown. FilterandPEvalvenotshown.
1.TurnpowerOff(O)andunplugtheunit.
2.Removethecabinet.Referto6.1RemovingCabinet,page17.
3.CutthetiewrapAthatsecurestheintakehoseBtothecompressorassemblyC.
4.Disconnectintakehosefromcompressorassembly.
5.Removecompressorwiresfromwireclampsortie-wraps(notshown).
6.Disconnectcompressorconnectorfromthemainharnessconnector(notshown).
7.DisconnectcapacitorwiresDfromthetopofthecapacitorE.
181154245-F-02
Service
8.DisconnectbrassnutFfromcompressor.
IRC5PO2AWIRC5PO2VAW
9.RemovethefourmountingscrewsGandgrommetsHthatsecurethecompressortothebaseassembly.Leavespringsonpostinthebase.
10.Tiltcompressorassemblyforwardandliftout.
11.ReverseSTEPS3-10toinstallnewcompressorassembly.Torquescrewsto30±5in-lbswhenassemblingthefourmountingscrewsinSTEP9.
12.Rununitandinspectforleaks.Referto8.1LeakTest,page41.
13.Ifnoleaksdetected,turnpowerOff(O)andunplugtheunit.
14.Reinstallthecabinet.Referto6.1RemovingCabinet,page17andreverseallsteps.
1154245-F-0219
Invacare®PerfectoOxygenConcentrator
6.2.2ReplacingCompressorAssemblyAfterSeptember2015
DANGER!
RiskofDeath,Injury,orDamage
Topreventelectricalshock
ALWAYSdisconnectconcentratorfromelectricaloutletbeforeservicing.
IRC5PO2AW
PEvalvenotshown. PEvalvenotshown.
IRC5PO2VAW
PEvalvenotshown. PEvalvenotshown.
1.TurnpowerOff(O)andunplugtheunit.
2.Removethecabinet.Referto6.1RemovingCabinet,page17.
3.CutthetiewrapAthatsecurestheintakehoseBtothecompressorassemblyC.
4.Disconnectintakehosefromcompressorassembly.
5.Cuttie-wrapsthatsecurecompressorwirestogusset.
6.Disconnectthecompressorwires(notshown)fromthejumperwires(notshown).
201154245-F-02
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60