Liing Sequence
Cleaning and Disinfecng
The ELK should be fully inated and pre-cleaned by washing with detergent and water, thoroughly
rinsed and then disinfected using a 1:10 soluon of household liquid bleach and water (1 part
bleach to 10 parts cold water). Leave the soluon on the ELK for approximately 10 minutes
before thoroughly rinsing o with clean cold water. Allow to air dry before storage. Do not dry
using hot air.
BATTERY NOTE: Charge for 12 hours before rst use. Before each use, check the baery charge
state by pressing the ‘Baery State Indicator’ buon (please read full Airo 24 User Instrucons).
2
5Bring the paent to
a sing posion and
provide support.
Whilst providing support:
1. Fully inate secon 1
2. Fully inate secon 2
3. Fully inate secon 3
4. Fully inate secon 4
6
NOTE: The compressor will automacally
stop as each secon reaches full pressure.
WARNING: While in use, never allow a secon to become parally deated while any secons
above it are inated. Secons must always be fully inated from the boom up, in sequence, in
order to maintain stability.
Press Baery
State Indicator
Buon RED
Recharge
Immediately
2 GREEN
Recharge
soon
3 GREEN
OK to Use
4 GREEN
OK to Use
1 GREEN
Recharge
before next use