  • Hello! I'm your chat assistant. I've reviewed the ELK Quick Start Guide, which provides instructions for using the ELK patient lifting aid and the Airflo 24. The guide covers important steps like hose connections, patient positioning, device operation, and cleaning. I'm here to help you understand these instructions and answer any questions you might have.
  • How do I connect the hoses?
    How do I position the patient on the ELK?
    How do I switch the device ON?
    What is the lifting sequence for the ELK?
    How do I clean and disinfect the ELK?
MI0443 Issue 2 (UK)
Connect hoses:
1 = 1
2 = 2
3 = 3
4 = 4
1
Switch ‘ON’
ELK Quick Start Guide
Make the Connecons
Posion Paent
Operate Hand Control Buons.
(See ‘Liing Sequence’ page 2)
1
Buon 11
Buon 22
Buon 44
Buon 33
4
Posion paent centrally on ELK with waist band
level with edge of ELK
2
Momentarily press the
power buon to switch ‘On’.
3
See ‘BATTERY NOTE’ page 2 before proceeding. Please read the full product user
instrucons for the Airo 24 and ELK.
Mangar Health
Tel: +44(0)1544 267674
Email: sales@mangarhealth.com
Web: www.mangarhealth.com
Liing Sequence
Cleaning and Disinfecng
The ELK should be fully inated and pre-cleaned by washing with detergent and water, thoroughly
rinsed and then disinfected using a 1:10 soluon of household liquid bleach and water (1 part
bleach to 10 parts cold water). Leave the soluon on the ELK for approximately 10 minutes
before thoroughly rinsing o with clean cold water. Allow to air dry before storage. Do not dry
using hot air.
BATTERY NOTE: Charge for 12 hours before rst use. Before each use, check the baery charge
state by pressing the ‘Baery State Indicator’ buon (please read full Airo 24 User Instrucons).
2
5Bring the paent to
a sing posion and
provide support.
Whilst providing support:
1. Fully inate secon 1
2. Fully inate secon 2
3. Fully inate secon 3
4. Fully inate secon 4
6
NOTE: The compressor will automacally
stop as each secon reaches full pressure.
WARNING: While in use, never allow a secon to become parally deated while any secons
above it are inated. Secons must always be fully inated from the boom up, in sequence, in
order to maintain stability.
Press Baery
State Indicator
Buon RED
Recharge
Immediately
2 GREEN
Recharge
soon
3 GREEN
OK to Use
4 GREEN
OK to Use
1 GREEN
Recharge
before next use
/