Issue Date: 2007-12-21 Page 7 of 10 Report Reference # E179824-A38-CB-1
Amendment 3 2009-06-30
TRF No.: IEC60601_1C UL International Demko A/S
The original report was modified on 2009-06-30 to include the following changes/additions:
Amendment to cover addition of 2 optional fans, Sanyo-Denki, type 109P0812M7xx. No testing conducted.
This report is only to be used and read in conjunction with the following reports:
E179824-A38-CB-1, dated 2007-12-21 with its amendments 1 and 2, dated 2008-04-07 and 2008-09-09.
Product Description
The product covered by this report is a landscape type LCD display system for use with medical electrical
equipment.
The product covered is intended for use in a hospital or related health care facility. The display does not
employ any applied parts.
Model Differences
N/A
Additional Information
- Maximum load conditions during testing is defined as units displaying test screen pattern, max. contrast and
brightness.
- The equipment includes an external power supply which has not been investigated as part of the
evaluation, but has been separately investigated for reinforced insulated output. The corresponding
certificates are included in the enclosure "Licenses".
The monitor shall be supplied from an external power supply of the following type(s):
- Jerome Industries Corp, type WSX824M; Skynet, type SNP-129-M or XP Power, type AMM120PS24
series.
- A full test report of the power supply may be required when submitting this CB Test Report to a National
Certification Body (NCB) for obtaining certification at national level.
- Model MDSC-2124 is considered Class I.
- Max. ambient temperature of model MDSC-2124 is 40°C.
- This product contains visible indicator LEDs (with diffusing lens) only. No IEC 60825-1 evaluation was
deemed necessary. Additional testing and evaluation may be required based on auditing agency's discretion.
Technical Considerations
The product was investigated to the following additional standards: CAN/CSA C22.2 No. 601.1-M90 (R1997),
EN 60601-1: 1990 + A1:1993 + A2:1995 + A13:1996, UL 60601-1, 1st Edition, 2006-04-26 (includes National
Differences for USA)
The product was not investigated to the following standards or clauses: Clause 36, Electromagnetic
Compatibility (IEC 601-1-2), Clause 48, Biocompatibility (ISO 10993-1), Clause 52.1, Programmable
Electronic Systems (IEC 601-1-4)
The product is Classified only to the following hazards: Casualty, Shock, Fire
The degree of protection against harmful ingress of water is: Ordinary