  • Hello! I am an AI chatbot trained to assist you with the Atmos ENT Operating Instructions Manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Operating Instructions
ATMOS®ENTexiblebre
endoscope
English
GA1GB.150252.0
2017-07Index02
2
Table of contents
1 Introduction ........................................................................................ 4
1.1 Notes on operating instructions ...............................................................4
1.2 Explanationofpicturesandsymbols .......................................................5
1.3 Intended use ...............................................................................................6
1.4 Function .......................................................................................................7
1.5 Intended users ............................................................................................7
1.6 Scopeofdelivery ........................................................................................8
1.7 Transport and Storage ...............................................................................8
2 Hints for your safety .......................................................................... 9
2.1 Generalsafetyinformation .......................................................................9
2.2 Dangerforusers,patientsandthirdparties ..........................................9
2.3 Avoiddamagetothedevice ................................................................... 10
3 Setting up and starting up .............................................................. 11
3.1 Deviceoverview ....................................................................................... 11
3.2 Preparingthedevice ............................................................................... 11
3.3 Usingwithotherdevices ........................................................................ 12
4 Operation .......................................................................................... 14
4.1 Prior to use ............................................................................................... 14
4.2 Potentialmovement................................................................................ 14
4.2.1 Deectingthetip .................................................................................. 14
4.3 Focussing .................................................................................................. 14
4.4 Afteruse ................................................................................................... 15
5 Cleaning and disinfection ............................................................... 16
5.1 Prepareforcleaning ................................................................................ 17
5.2 Manualcleaninganddisinfection ......................................................... 17
5.2.1 Stage I .................................................................................................... 18
5.2.2 Stage II ................................................................................................... 18
5.2.3 Stage III .................................................................................................. 19
5.2.4 Stage IV .................................................................................................. 19
5.2.5 PhaseV .................................................................................................. 19
5.2.6 Removestubborndepositsfromglasssurfaces.............................. 19
5.2.7 Concludecleaninganddisinfection ................................................... 19
5.3 Machinecleaninganddisinfection ........................................................ 20
5.4 Recommendeddisinfectants ................................................................. 21
5.4.1 Recommendedendoscopedisinfectants ......................................... 21
5.5 Sterilization .............................................................................................. 22
3
5.5.1 Preparingforsterilization ................................................................... 22
5.5.2 Gassterilization .................................................................................... 22
6 Maintenance and service ................................................................ 23
6.1 Functioncheck ......................................................................................... 23
6.2 Furthertests ............................................................................................. 23
6.2.1 Leakagetest .......................................................................................... 23
6.2.2 Checkingthebreoptics .................................................................... 24
6.2.3 Checkingtheglasssurfacesandthesurface ................................... 24
6.3 Sendinginthedevice .............................................................................. 25
7 Troubleshooting ............................................................................... 26
8 Disposal ............................................................................................. 28
9 Technical data .................................................................................. 29
10 For your notes .................................................................................. 30
4 Introduction
1 Introduction
1.1 Notes on operating instructions
Theseoperatinginstructionscontainimportantinstructionson
howtooperateyourproductsafely,correctlyandeectively.
Theseoperatinginstructionsservenotonlyfornewoperating
personneltobeinstructedinitsuse,butalsoforuseasa
referencemanual.Anyreprint-eveninextracts-onlyafter
writtenpermissionfromATMOS.
These operating instructions must always be kept available
near the device.
Care,periodtests,regularcleaningandproperapplicationare
indispensable.Thatincreases,amongotherthings,thereliability
andservice-lifeoftheproduct.
Maintenance,repairsandperiodtestsmayonlybecarriedout
bypersonswhohavetheappropriatetechnicalknowledgeand
arefamiliarwiththeproduct.Tocarryoutthesemeasuresthe
personmusthavethenecessarytestdevicesandoriginalspare
parts.
Perusechapter„2Hintsforyoursafety“onpage9,before
usingtheproductforthersttime.Thishelpsyouavoid
potentiallydangeroussituations.
TheproductbearstheCEmarkingCEaccordingtotheEC
Directiveofthecouncilformedicalproducts93/42/EECand
meetsthebasicrequirementsofAppendixIofthedirective.
Theproductcomplieswithallapplicablerequirementsofthe
Directive2011/65/ECrestrictingtheuseofcertainhazardous
substancesinelectricalandelectronicequipment(“RoHS”).
Thedeclarationofconformityandourgeneralstandard
termsandconditionscanbeobtainedonourwebsiteat
www.atmosmed.com.
ThequalitymanagementsystematATMOShasbeencertied
accordingtointernationalstandardsENISO13485.
Theseoperatinginstructionsarevalidforthefollowingdevices:
ATMOS®nasopharyngoscope3.2mm 950.0243.0
5Introduction
1.2 Explanation of pictures and symbols
In the operating instructions
DANGER
Warningofadangerwhichcausesimmediatedeathorseriousinjury.
Observethenecessarymeasures.
WARNING
Bewareofadangerwhichcancausedeathorseriousinjury.Observethe
necessarymeasures.
CAUTION
Bewareofadangerwhichcaneasilyhurtyou.Observethenecessary
measures.
ATTENTION
Indicationofadangerwheretheproductorotheritemscanbedamaged.
Observethenecessarymeasures.
Warningofadangerwhichcancausedeathorseriousinjury.
Informationregardingpossiblematerialdamagewhichcanbe
caused.
&Usefulinformationonthehandlingofthedevice.
1. Action.Gostepbystep.
»Resultofanaction.
On device and type plate
Refertotheoperating
instructions
Followoperating
instructions(blue)
Thisproductcomplieswith
therelevantrequirements
oftheEUDirective
TheUSFederalLaw
restrictsthisproductto
salebyorontheorderofa
physician.
Manufacturer Manufacturingdate
6 Introduction
SN Serialnumber REF Ordernumber
ApplicationparttypeBF
MR
Donotuseinmagnetic
resonance areas
NON
STERILE
Notsterile Professionaldisposal
1.3 Intended use
Name: ATMOS®ENTexiblebreendoscopes
Main function: TheapplicationoftheATMOS®ENTexible
breendoscopesisindicatedforendoscopic
procedures,fortheshorttermapplicationin
theoralcavityuptothepharynx,intheauditory
canaltotheeardrumandinthenasalcavityfor
thevisualizationoforicesandvisceralcavities.
Medical indication /
application:
Forapplicationonhumans
Specication of the
main function:
Highresolutionexiblebreendoscopewitha
pliableexibletip,whichenablesanextended
lineofvisionfrom0°to140°intwoopposite
directions.AllstandardC-mountadapterswith
cameraheadscanbeattachedtotheeyepiece,
thisenablesavideotechnicalimagetransmission
aslive-imageorarecordingasapictureand/or
lmfordocumentation.
User prole: Doctorsandmedicallytrainedpersonnel
Patient groups (e.g.
age, weight, health,
state):
Patientsofallageswithoutrestrictions,where
aendoscopicstandardexaminationistobe
performed.
Application organ: Oralcavity,nasalcavity,auditorycanal
Application time: Applicationupto60minutes
Application site: InENTpractices,outpatientdepartments,clinics,
operatingroomsandphoniatrists.
7Introduction
Contraindications: TheapplicationoftheATMOS®HNOexible
breendoscopeiscontraindicatedifendoscopic
procedures are contraindicated.
The product is: Notactive
Sterility: Notsterile
Single-use product/
reprocessing:
Nosingle-useproduct.Seecleaningand
disinfectinginstructionsintheoperating
instructions.
1.4 Function
ATMOS®endoscopesareindicatedforthevisualizationoforicesandbody
cavities.Dependingonthecross-section,lengthandlockingpossibilities,
eachoftheendoscopesisdesignedforaspecicpurposeinaparticular
eldofmedicine.
Thenaturaloricesareilluminatedviaextremelythinbreglassbres
whicharethinnerthanhumanhair.Withtheaidofacameraanimagecan
betransferredfromtheoricetothescreenordisplayviaasubstantially
largeramountofsimilarbreglassbres.
1.5 Intended users
Endoscopicproceduresmayonlybecarriedoutbyqualiedprofessionals
(e.g.doctors)withcorrespondingtrainingandfamiliaritywithendoscopic
procedures.Inthatregarditistheresponsibilityoftheuserto
continuouslyinformhimselffromthemanufacturerregardingindication,
contraindication,possiblecomplicationsandrisksaswellasfurther
developmentsintheendoscopicprocedures.
Athoroughunderstandingoftheprinciplesandmethodsusedinlaser
endoscopyandelectro-surgicalproceduresisnecessarytoavoidshockand
riskofburnstopatientsandusersaswellasdamagetootherequipment
andinstruments.
8 Introduction
1.6 Scope of delivery
Flexiblebreendoscope
Aluminiumcarryingcase
Leakagetester
Valvecap
Operating Instructions
1.7 Transport and Storage
Onlytransportthedeviceinashippingcontainer,whichispaddedand
oerssucientprotection.
Ifdamageoccursduringtransport:
1. Documentandreportthetransportdamage.
2. SendtheproducttoATMOS,seeChapter„6.3Sendinginthedevice“on
page 25.
Environmental conditions for transport and storage:
Temperature: -20...+70°C
Relativeairhumidity: 5...95%withoutcondensation
Airpressure: 70...106kPa
9Hintsforyoursafety
2 Hints for your safety
Thesafetyoftheproductcomplieswithalltherecognizedrulesof
technologyandtheDirectivesoftheMedicalDevicesAct.
Readandfollowthesafetyinformationcarefullybeforeusingtheproduct.
2.1 General safety information
Neveroperatethedevice,ifitshowsanyobvioussafetydefects.Checkthe
unitatregularintervalsforsafetyandfunction.
2.2 Danger for users, patients and third parties
Reduce the risk of infection for you and your patients!
Deadlydiseasescanbetransmitted.
Cleananddisinfecttheproductaftereveryuse.
Cleananddisinfectthedeviceaccordingtotheoperatinginstructions.
Avoid misapplication.
Yourpatientmaybeseriouslyinjured.
Priortoeachuseafunctioncheckhastobeperformed.
Alwayshaveareplacementlightsourceathandoruselightsources
whichhaveasubstitutebulb.
ThesimultaneoususeofNMR(NuclearMagneticResonance)and
endoscopescanleadtodangersandartefacts.Observethenecessary
safetyinstructions.
Heat development on the light source.
Burnsarepossible.
Attentionmustbepaidthattheendofthelightconductordoesnot
touchthetissue.
Ifapplicable,rinsethesurgicaleldsuciently.
Onlyafullyfunctionalproductmeetsthesafetyrequirementsofusers,
patientsandthirdparties.Therefore,pleasealsoobservethefollowing
guidelinesconcerningyourproduct.
10 Hintsforyoursafety
2.3 Avoid damage to the device
Incorrect or careless handling of the product.
Theproductmaybedamaged.
Avoidbendingtheexiblebreendoscopetightly,wrapittightly,or
subjectittostrongimpact.
Donotrotate,pullorputpressureontheexibleparts.
Alwaysallowtheproducttocooldowntoroomtemperaturewithout
anyadditionalcoolingmeasures.
Improper cleaning, disinfection and sterilization.
Theproductmaybedamaged.
Observetheinstructionsforcleaning,disinfectionandsterilizationin
chapter„5Cleaninganddisinfection“onpage16.
11Setting up and starting up
3 Setting up and starting up
3.1 Device overview
12
3456
7
8
910 11
12
13
1Flexiblebreendoscopewithventvalve
2Ventvalve
3Diopteradjustment(focussing)
4DeectionleverUP/DOWN
5LightcableconnectionforACM1/BritishStandard(ontheopposite
sideofthehousing)
6Lens
7Transfersleeveforlightcable
8LightcablesleeveforAMCIconnection
9Lightcable
10 Lightcablesleeveforlightsources
11 Valvecap
12 Leakagetester
13 Siliconehose
3.2 Preparing the device
Theproductissuppliednon-sterile.
12 Setting up and starting up
1. Cleananddisinfectthedevicepriortorstuse.
2. Sterilizethedevicewhereitisnecessaryforthemedicalindication.
3.3 Using with other devices
Attentionmustbepaidthatdeviceswhichareusedwiththeendoscope
meetallsafetyrequirements.
Attentionmustbepaidwhenusingtheendoscopewithanelectromedical
device,thatthetermsandconditionsfortheapplicationpartarecomplied
withtypeBF(insulated,oatingapplicationpart).
Combinedwithlaser,HF-surgery,pneumaticorelectrohydraulic
lithotripters,varioustherapeuticapplicationsarepossible.
1. Observetheoperatingandsafetyinstructionsoftheapplieddevices
and accessories.
Using with high-frequency surgical devices
Explosion and  re hazard!
Burnsandinjuriesarepossible.
Takethenecessarymeasurestoremoveoravoidformationof
combustiblegases.
Predominant depression in the tissue can be caused due to low and
unsuitable performance settings or incorrect application.
Seriousinjurytothepatientispossible.
Thepatientmustbesuitablyprepared.
Settheperformancesettingbasedonclinicalreferencesandonlywith
thenecessarytrainingandexperience.
Onlyswitchonthehigh-frequencycurrentwhentheapplicationpart
(electrode)canbeseenthroughtheendoscope.
WARNING
13Setting up and starting up
Using with laser devices
Dangerous laser radiation.
Seriousdamagetotheeyesispossible.
Wearsuitableprotectionglasses.
Burns and undesirable depressions in the tissue
Thelaserperformanceshouldonlybeactivatedwhenthetipofthe
laserbrecanbeseenthroughtheendoscope.
WARNING
14 Operation
4 Operation
Danger of infection due to secretion on the endoscope.
Deadlydiseasescanbetransmitted.
Alwayskeepausedendoscopeseparatefromotherreprocessed
instruments.
Cleananddisinfecttheendoscopeaftereachusewithin6hours.
4.1 Prior to use
1. Connecttheaccessoriesandcomponents.
2. Performafunctioncheck.
&Checkthefunctionalityofaccessoriesandcomponents.
4.2 Potential movement
Theexiblebreendoscopecanbemovedinthreeways:
Itcanbemovedinalldirectionsalongthelongitudinalaxis.
Itcanberotated.
Thedistalendcanbebent.
4.2.1 Defl ecting the tip
1. Pressthedeectionlever 1.
1
4.3 Focussing
1. Turnthediopteradjustment 2.
2
WARNING
15Operation
4.4 After use
1. Removeallusedadaptersfromthelightcableconnection.
2. Removethesealcapfromtheluer-lockconnection.
3. Visibleresidueshouldberemovedimmediatelyaftersurgeryusinga
lint-freedisposablecloth.
4. Transporttheproductdryinacloseddisposalcontainerforcleaning
anddisinfecting.
5. Cleananddisinfecttheendoscopeaftereachusewithin6hours.
16 Cleaninganddisinfection
5 Cleaning and disinfection
Werecommendyoutodocumentanymaintenanceworkandalsoany
exchangeofparts.
Itistheresponsibilityoftheoperatortoensurethattherequiredresultsfor
cleaninganddisinfectionareadheredto.Generallyvalidationandroutine
monitoringoftheprocedureisnecessary.
Theamountofprocessingcyclesdependsondierentfactorsandcannot
begeneralized.
Risk of infection due to secretion on the product and accessories.
Deadlydiseasescanbetransmitted.
Alwaysweardisposableglovesduringanycleaning.
Cleananddisinfecttheproductaftereveryuse.
Cleananddisinfectthedeviceaccordingtotheoperatinginstructions.
Observerelevantnationallegalregulationsandstandardsforcleaning,
disinfectionandsterilization.Thisappliesespeciallytopatientswith
suspectedCreutzfeldt-Jacob-Disease(CJD),orpossiblevariants.
Dried up residue after surgery due to wrong cleaning.
Corrosiondamageanddicultcleaningarepossible.
Cleananddisinfecttheendoscopeaftereachusewithin6hours.
Donotusepre-cleaningtemperatures>45°C.
Donotusexingdetergentwiththesubstancealdehydeoralcohol.
WARNING
ATTENTION
17Cleaninganddisinfection
Incorrect cleaning and disinfectants.
Damagedsurface,fadedlasermarkingorchemicallydamagedproduct.
Observethecorrectdosageforneutralizingagentsandbasiccleaners.
Donotuseabrasives.
Onlyusecleaninganddisinfectantagentswhicharerecommendedby
themanufacturerofendoscopes.
Donotusecleaninganddisinfectantagentswhichcouldcausestress
cracksorcorrosioninplastic.
ObservethatthepHvalueoftheapplicationsolutionis<8.
Incorrect cleaning and disinfection:
Damagedproduct.
Nevercleantheproductinanultrasonicbath.
Onlyusecleaninganddisinfectiondeviceswhicharesuitablefor
cleaningexiblebreendoscopes.
Donotautoclavetheproduct.
Onlyusetherecommendeddisinfectantagentsandsterilization
methods.
Avoidexcessivepressureortensiontotheexibleprobearea.
Ensurethatthebendingradiusoftheexibleprobeareaisatleast
40mm.
5.1 Prepare for cleaning
1. Priortoeverycleaninganddisinfectionaleakagetestmustbe
conducted,seechapter„6.2.1Leakagetest“onpage23.
5.2 Manual cleaning and disinfection
Onlycleanthedevicewhenitisleakproof.Seechapter„6.2.1Leakagetest“
on page 23.
Followingdisinfectionwithnon-virucidalagents,sterilizationmustbe
carried out.
ATTENTION
18 Cleaninganddisinfection
Stage Step Tempera-
ture
Time Concen-
tration
Water
quality
Chemistry
ICleaning 35-45°C/
95-113°F
5min 0.8% Drinking
water
Cidezyme/
Enzol1
II Intermedi-
ate rinsing
Roomtem-
perature
(cold)
3 X
1min
--- Drinking
water
---
III Disinfec-
tion
20-25°C/
68-77°F
12min --- Drinking
water
CidexOPA2
IV Final
rinsing
Roomtem-
perature
(cold)
3 X
2min
--- VE-W
sterile3
---
VDrying Roomtem-
perature
--- --- Drinking
water
---
1Enzymaticcleaner
2Ready-to-usesolution,0,55%orthophtalaldehyd
3Demineralizedwater(demineralised,lowgerms,maximum10germs/ml
andendotoxin-poor,maximum0.25endotoxinunits/ml)
5.2.1 Stage I
&Theconcentrationofthecleaningagentdependsonthecontamination
oftheproduct.Observethemanufacturer'sinstructionsforuseofthe
cleaningsolution.
1. Theproductmustbecompletelyimmersedinthecleaningsolutionso
thatallaccessiblesurfacesarecovered.
Damagedproductduetocleaningwithametalbrush.
2. Placetheproductinthesolutionandcleanwithasoftclothora
suitablesoftcleaningbrush.
» Thereisnovisibleresidueonthesurface.
3. Operatethemovablecomponentslikefocussinganddeectionleverin
thesolution.
4. Thoroughlyrinsethemovablecomponentswiththecleaningsolution,
atleast5times,usingadisposablesyringe(20ml)
5.2.2 Stage II
Usefreshwaterforeachrinsingcycle.
19Cleaninganddisinfection
1. Thoroughlyrinseallaccessiblesurfacesoftheproductthreetimesfor
atleast1minute.
2. Operatethemovablecomponentslikefocussinganddeectionleverin
thesolution.
3. Allowtherestofthewatertodraino.
5.2.3 Stage III
1. Theproductmustbecompletelyimmersedinthecleaningsolutionso
thatallaccessiblesurfacesarecovered.
2. Operatethemovablecomponentslikefocussinganddeectionleverin
thesolution.
5.2.4 Stage IV
Corrosiondamageduetochlorine-andchlorideresidues.
Usefresh,demineralisedwaterforeachrinseandrinsethoroughly.
1. Thoroughlyrinsetheproduct3timescompletely,allaccessible
surfaces,foratleast2minutes.
2. Operatethemovablecomponentslikefocussinganddeectionleverin
thesolution.
3. Allowtheremainingwatertodrainosuciently.
&Iftheproductisnotsucientlyrinsed,stubbornresiduecoulddevelop
ontheglasssurface.
5.2.5 Phase V
1. Drytheproductwithalint-freecloth.
2. Areaswhichcannotbeaccessedwithaclothmustbedriedwith
medical-qualitycompressedair(pmax.=0.5bar).
5.2.6 Remove stubborn deposits from glass surfaces
&Stubborndepositsdevelopwhenthedeviceisnotrinsedsuciently.
1. Cleantheglasssurfaceswithsuitablecleaningpasteorwithacotton
budsoakedinalcohol.
5.2.7 Conclude cleaning and disinfection
1. Checkwhetherallvisiblesurfacesarewithoutresidue.Ifresidueis
found,repeatthecleaninganddisinfection.
20 Cleaninganddisinfection
2. Ifyouhavedisinfectedwithnon-virucidaldisinfectants,youmust
sterilizetheproduct.
&Ifstubbornresiduecannotberemovedbycleaning,theproductmust
besenttoATMOSoranauthorizeddealer.
5.3 Machine cleaning and disinfection
Checkwhetheryourcleaninganddisinfectiondeviceforendoscopes
issuitablefortheATMOS®ENTexiblebreendoscope.Ifnecessary,
checkwiththemanufacturerofyourcleaninganddisinfectiondevicefor
endoscopes.
Onlycleanthedevicewhenitisleakproof.Seechapter„6.2.1Leakagetest“
on page 23.
Followingdisinfectionwithnon-virucidalagents,sterilizationmustbe
carried out.
1. Iftheendoscopeishighlycontaminated,thencleanitmanually
beforehand.
2. Removethevalvecapfromtheventvalve.
Damagedendoscopeduetoconnectionofthewronghose.
3. Connectthepressureequalizationhoseofthecleaninganddisinfection
deviceforendoscopestothevalvecap.Ifnecessaryuseanadapter.
Damagedendoscopefromtemperatures>65°C.
4. Selectasuitablecleaninganddisinfectionprogramwithatemperature
<65°C.
5. Cleananddisinfecttheendoscope.
6. Checkwhetherallvisiblesurfacesarewithoutresidue.Ifresidueis
found,repeatthecleaninganddisinfection.
7. Soakacottonbudinlenscleanerorwith70%-Isopropylalkohol.
8. Cleanthefollowingpartswithacottonbud,wipingwithcircular
movements:
Objective(distalend)
Eyepiece
Lightguideconnection
9. Ifyouhavedisinfectedwithnon-virucidaldisinfectants,youmust
sterilizetheproduct.
&Ifstubbornresiduecannotberemovedbycleaning,theproductmust