System Setup Configuration
Parameter Description Range
Slow Averaging
Sets the number of mV RMS readings that are
averaged when the Display Averaging parameter
is set to Slow. A higher number will cause the
display to update slower, but will give a more
stable reading
Default = 150
1-200 Readings
Medium Averaging
Sets the number of mV RMS readings that are
averaged when the Display Averaging parameter
is set to Medium. A higher number will cause the
display to update slower, but will give a more
stable reading
Default = 15
1-200 Readings
Fast Averaging
Sets the number of mV RMS readings that are
averaged when the Display Averaging parameter
is set to Fast. A higher number will cause the
display to update slower, but will give a more
stable reading
Default = 4
1-200 Readings
Averaging Window
Sets the range of input readings that will be
averaged. If a new mV reading deviates from the
average by less than this amount, it will be
averaged with the rest of the readings in the
display averaging buffer. Otherwise, the input is
considered a step change and the display
averaging buffer is flushed.
0.0 to 100.0 mV
Load Set 1
Assigns a resistance value for Set 1 from a
combination of the available loads as determined
by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more
information.)
Default = None Selected
0-6,500.0 ohms
Load Set 2
Assigns a resistance value for Set 2 from a
combination of the available loads as determined
by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more
information.)
Default = None Selected
0-6,500.0 ohms
Load Set 3
Assigns a resistance value for Set 3 from a
combination of the available loads as determined
by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more
information.)
Default = None Selected
0-6,500.0 ohms
Load Set 4
Assigns a resistance value for Set 4 from a
combination of the available loads as determined
by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more
information.)
Default = None Selected
0-6,500.0 ohms