MRU Instruments OPTIMA 7 Biogas User manual

Type
User manual

This manual is also suitable for

AMPRO 2000 / OPTIMA 7
Taking good care of your analyzer will pay off on the long run!
MRU Instruments, Inc. / 18838 S Memorial Drive Suite 103 / Humble, Texas 77338 / Tel.: 832-230-0155
Taking good care of your analyzer
AMPRO 2000 / OPTIMA 7
Taking good care of your analyzer will pay off on the long run!
MRU Instruments, Inc. / 18838 S Memorial Drive Suite 103 / Humble, Texas 77338 / Tel.: 832-230-0155
Let your analyzer pump run at ambient air
for at least 3 minutes after each
measurement!
This will ensure that the gas channels and
sensors are free of gas.
Clean and empty the condensate separator
after each measurement!
Condensate is slightly acidic, you dont want
this condensate to stay in the analyzer over
time.
This can damage parts of the analyzer!
AMPRO 2000 / OPTIMA 7
Taking good care of your analyzer will pay off on the long run!
MRU Instruments, Inc. / 18838 S Memorial Drive Suite 103 / Humble, Texas 77338 / Tel.: 832-230-0155
Replace the pleated filter (10155) regularly!
These filters are washable (washing machine)
and reusable once they are dry.
If replaced too late, dirt will eventually make
its way into the analyzer!
IMPORTANT!!
Disconnect the hoses from the analyzer.
Clean and empty the sampling line with
compressed air!
Remove the probe tube from the probe
handle then use compressed air to clean the
hoses.
IMPORTANT!!
Disconnect the hoses from the analyzer.
Condensate is slightly acidic, you dont want
this condensate to stay in the analyzer over
time.
This can damage parts of the analyzer!
AMPRO 2000 / OPTIMA 7
Taking good care of your analyzer will pay off on the long run!
MRU Instruments, Inc. / 18838 S Memorial Drive Suite 103 / Humble, Texas 77338 / Tel.: 832-230-0155
Clean the probe tube both inside and
outside!
Clean the probe tube with compressed air (A)
Use sanding paper to clean the probe tune
(C)
Make sure that you dont damage the
thermo-couple!
Apply grease to the two O-Rings (B)
NEVER operate the analyzer inside the case!
The gas exits are blocked, and the sensors
will get damaged and have a very short life
span.
Also, the condensate separator has its best
effect when operated upright!
AMPRO 2000 / OPTIMA 7
Taking good care of your analyzer will pay off on the long run!
MRU Instruments, Inc. / 18838 S Memorial Drive Suite 103 / Humble, Texas 77338 / Tel.: 832-230-0155
Use the CO sensor protection as much as
possible!
You might damage the CO sensor by
changing the CO threshold to higher limits.
The standard CO threshold setting is at
500ppm!