SAFETY CHECKLIST
Survey the test station. Make sure it is safe & orderly.
Always keep unqualied/unauthorized personnel away from the test area.
Familiarize yourself with safety protocols in the event of a problem.
Exercise caution and never touch products or connections during a test.
Train operators. Never touch clips directly and always connect the return lead rst.
You should always know when a test is being performed.
WARNING: THIS GUIDE WAS CREATED FOR OPERATORS HAVING SOME FAMILIARITY WITH ELECTRICAL SAFETY
TESTING. AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL
OR FATAL ELECTRIC SHOCK. TO PREVENT ACCIDENTAL INJURY OR DEATH, THESE SAFETY PROCEDURES MUST
BE STRICTLY OBSERVED WHEN HANDLING AND USING A TEST INSTRUMENT. CONTACT US AT
INFO@ARISAFETY.COM FOR MORE INFO ON HOW TO GET TRAINED ON ELECTRICAL SAFETY TESTING.
Quick Start Guide
INSTRUMENT SETUP
Connect the female end of the standard NEMA style power cord into the
input power receptacle on the rear panel of the instrument. Plug the male
end of the cord into a grounded power source.
Connect the interlock disable key to the signal input connector on the
rear panel of the instrument. This is required to run a test.
Turn the POWER switch to ON.
After a test has been set, you may edit the test parameters at any time.
ADD NEW TESTS
EDIT TEST PARAMETERS
WARNING: LOCATE A SUITABLE TESTING AREA WITH A CLASS I GROUNDED OUTLET.
BE SURE THAT YOUR CLASS I OUTLET HAS BEEN TESTED FOR PROPER WIRING. READ
THE SAFETY CHECKLIST OF THIS GUIDE BEFORE STARTING TO TEST.
SENSE+
H.V.
SENSE
-
CURRENT
RETURN
SIGNAL INPUT
SIGNAL OUTPUT
REMOTE OUTPUT
277V~MAX. 16A MAX.
EXTERNAL
CASE CND PROBE HI PROBE LO
L
L
N
N
SCANNER 1 SCANNER 2
RS232
T10A 250V
GND
CAT. II
115V~ /230V~
10A MAX., 50Hz/60HZ
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV
16A AX. MAX.
OUTPUT
REMOTE INTERFACE
INPUT
DUT OUTPUTSDUT POWER I/P MD CIRCUITS
SENSE+
H.V.
SENSE
-
CURRENT
RETURN
SIGNAL INPUT
SIGNAL OUTPUT
REMOTE OUTPUT
277V~MAX. 16A MAX.
EXTERNAL
CASE CND PROBE HI PROBE LO
L
L
N
N
SCANNER 1 SCANNER 2
RS232
T10A 250V
GND
CAT. II
115V~ /230V~
10A MAX., 50Hz/60HZ
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV
16A AX. MAX.
OUTPUT
REMOTE INTERFACE
INPUT
DUT OUTPUTSDUT POWER I/P MD CIRCUITS
TESTRESET
POWER
ELECTRICAL SAFETY COMPLIANCE ANALYZER
OMNIA
®
SERIES
My
Menu
1 2
ABC DEF
GHI JKL MNO
PQRS TUV WXYZ
3
4 5 6
7
EXIT ENTER
SCANNER STATUS
H.V.
GND
HI
LO
CH
1 2 3 4 5 6 7 8
ON
8 9
0 <
- -
CAUTION
HIGH VOLTAGE
5KV MAX.
H.V.
SENSE+ SENSE
-
CURRENT RETURN
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
8206/8256 8207/8257
8206/8256 8207/8257
TEST SETUP
To Grounded
Power Source
2 3
START
Select the Add soft key.From the Main Menu select
the Setup Tests soft key.
Select the a Test Type.
My
Menu
0001
Add
Edit
Delete
Prompt
File
Fail Stop
ON
SETUP TESTS
My
Menu
File Name: 0001 TEST
0001
Dielectric
Withstand
Insulation
Resistance
AC Ground Bond
DC Continuity
Run Test
Line Leakage
SETUP TESTS
My
Menu
AC - WITHSTAND
Voltage
HI-Limit
Total
LO-Limit
Total
Ramp Up
Dwell Time
Ramp Down
Range :
Voltage 0V-5000V
Frequency
Arc Detect
1240 V
10.00mA
1.0s
5
0.00mA
10.00mA
0.00mA
0.00mA
0.1s
0.0s
60Hz
OFF
Arc Sense
HI-Limit
Real
LO-Limit
Real
Offset
Step Name
Defaults
More
Use the arrow keys to choose a
parameter to edit.
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
My
Menu
AC - WITHSTAND
Voltage
HI-Limit
Total
LO-Limit
Total
Ramp Up
Dwell Time
Ramp Down
Frequency
Arc Detect
1240 V
10.00mA
1.0s
5
0.00mA
10.00mA
0.00mA
0.00mA
0.1s
0.0s
60Hz
OFF
Arc Sense
HI-Limit
Real
LO-Limit
Real
Offset
Step Name
Defaults
More
Enter the value for each parameter by
using the alpha/numeric key pad. Select
ENTER to save and go to the next parameter.
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
My
Menu
AC - WITHSTAND
Voltage
HI-Limit
Total
LO-Limit
Total
Ramp Up
Dwell Time
Ramp Down
Frequency
Arc Detect
1240 V
10.00mA
1.0s
5
0.00mA
10.00mA
0.00mA
0.00mA
0.1s
0.0s
60Hz
OFF
Arc Sense
HI-Limit
Real
LO-Limit
Real
Offset
Step Name
Defaults
More
Select EXIT key followed by the ENTER
key to save your selection and return to
the Setup Tests screen.
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
My
Menu
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
START
Use the arrow keys to choose a test
to edit within the Setup Tests screen.
My
Menu
0002 IR Settings
500V 0.00M
0.5s 0.05M
File Name: 0001
Add
Edit
Delete
Prompt
File
Fail Stop
ON
SETUP TESTS
Use the arrow keys to choose a
parameter to edit.
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
Select EXIT key. You will return to the Setup Tests screen.
0001 ACW Settings
1240V 10.00mA
1.0s 0.000mA
Select the Edit soft key.
My
Menu
0002 IR Settings
500V 0.00M
0.5s 0.05M
File Name: 0001
Add
Edit
Delete
Prompt
File
Fail Stop
ON
SETUP TESTS
0001 ACW Settings
1240V 10.00mA
1.0s 0.000mA
My
Menu
0002 IR Settings
500V 0.00M
0.5s 0.05M
File Name: 0001
Add
Edit
Delete
Prompt
File
Fail Stop
ON
SETUP TESTS
0001 ACW Settings
1240V 10.00mA
1.0s 0.000mA
My
Menu
INSULATION RESISTANCE
Voltage
HI-Limit
LO-Limit
Ramp Up
Dwell Time
Delay Time
Range :
Voltage 0V-6000V
500 V
0.00M
0.5s
0.0s
0.05M
0.000uA
0.1s
1.0s
Ramp Down
Charge-Lo
OFF
DUT Output
Step Name
Defaults
Enter the value for each parameter by
using the alpha/numeric key pad. Select
ENTER to save and go to the next parameter.
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
My
Menu
INSULATION RESISTANCE
Voltage
HI-Limit
LO-Limit
Ramp Up
Dwell Time
Delay Time
Range :
Hi-Limit 0.05M - 50000M 0=OFF
500 V
0.00M
0.5s
0.0s
0.05M
0.000uA
0.1s
1.0s
Ramp Down
Charge-Lo
Step Name
Defaults
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
My
Menu
INSULATION RESISTANCE
Voltage
HI-Limit
LO-Limit
Ramp Up
Dwell Time
Delay Time
Range :
Hi-Limit 0.05M - 50000M 0=OFF
500 V
0.00M
0.5s
0.0s
0.05M
0.000uA
0.1s
1.0s
Ramp Down
Charge-Lo
Step Name
Defaults
OFF
DUT Output
OFF
DUT Output
Adding multiple tests to the same screen automatically creates a multiple step test
sequence.
Note: The test will run in a sequential order unless Single Step control is set to ON.
Test programs can be saved and loaded as les for easy recall of test conditions.
4 5
TEST SETUP
SET TEST SEQUENCES
CREATE A NEW TEST FILE
SET TEST SEQUENCES
DELETE A TEST STEP
START
Select the Single Step soft key
to toggle ON or OFF.
To perform a Single Step test,
select the Perform Tests soft key.
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
My
Menu
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
My
Menu
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
My
Menu
0002 IR Settings
500V 0.00M
0.5s 0.05M
File Name: 0001
PERFORM TESTS
0001 ACW Settings
1240V 10.00mA
1.0s 0.000mA
Single Step
Fail Stop
Load Files
Results
OFF
ON
Select the EXIT key.
You will return to the Main Menu.
My
Menu
0002 IR Settings
500V 0.00M
0.5s 0.05M
File Name: 0001
PERFORM TESTS
0001 ACW Settings
1240V 10.00mA
1.0s 0.000mA
Single Step
Fail Stop
Load Files
Results
ON
ON
START
From the Main Menu select
the Setup Tests soft key.
Select the File soft key.
Select the New File soft key.
My
Menu
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
My
Menu
My
Menu
0001
Add
Edit
Delete
Prompt
File
Fail Stop
ON
SETUP TESTS
This returns you to the Setup Tests
menu with a blank sequence. The
file name is displayed at the bottom.
You may begin setting up a full test
sequence by selecting the Add soft key.
My
Menu
File Name: 0002 TEST
Add
Edit
Delete
Prompt
File
Fail Stop
ON
SETUP TESTS
0002
File Name:
New File
Save
Save As
Delete
Load
FILE SETUP
Use the alpha/numeric key pad to give
the file a name. Select ENTER to save.
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
My
Menu
CREATE FILE
Caps Lock
Symbol Key
ON
File Name:
My
Menu
File Name: 0002 TEST
Add
Edit
Delete
Prompt
File
Fail Stop
ON
SETUP TESTS
0002 TEST
0001
0002
0002
0001
START
From the Main Menu select
the Setup Tests soft key.
If a test file is not already loaded,
load a file by selecting the File
soft key.
Use the arrow keys to choose the file.
Select the Load soft key to load the file.
My
Menu
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
My
Menu
My
Menu
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
File Name: 0001 TEST
0001 ACW Settings
1240 V 10.00mA
1.0s 0.000mA
Add
Edit
Delete
Prompt
File
Fail Stop
ON
SETUP TESTS
0001
0002 TEST
0003 SEAN
0004
File Name:
New File
Save
Save As
Delete
Load
FILE SETUP
The test file is loaded.
My
Menu
File Name: 0002 TEST
Single Step
Fail Stop
Load Files
Results
OFF
ON
PERFORM TESTS
0002 LLT Settings
10000uA
0.5s 0.0uA
0001 ACW Settings
1240 V 10.00mA
1.0s 0.000mA
START
From the Main Menu select
the Setup Tests soft key.
If a test file is not already loaded,
load a file by selecting the File
soft key.
Select the Load soft key to load the file. Use the arrows to select the test
you would like to delete. Select
the delete soft key to delete the
Test Step.
My
Menu
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
My
Menu
My
Menu
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
File Name: 0001 TEST
0001 ACW Settings
1240 V 10.00mA
1.0s 0.000mA
Add
Edit
Delete
Prompt
File
Fail Stop
ON
SETUP TESTS
0001
0002 TEST
0003 SEAN
0004
File Name:
New File
Save
Save As
Delete
Load
FILE SETUP
My
Menu
0001
0002 TEST
0003 SEAN
0004
File Name:
New File
Save
Save As
Delete
Load
FILE SETUP
TEST LEAD CONNECTIONS
If there is no ground circuit present on the DUT, you can use test leads to connect
the OMNIA II to the DUT.
Plug the black return lead (P/N 38490) into the Return terminal. Plug black
sense lead into the Sense (-) terminal.
Attach the clip-terminated end of the return lead to the dead metal on
the chassis of the DUT.
Connect the high voltage lead (P/N 4040A-08) into the High Voltage
terminal on the front panel of the instrument.
Attach the high voltage lead to the current-carrying conductors of
the DUT.
Did you know? When performing a Ground Bond test on a Class I product
connect the red high current lead (P/N 38489) into the current terminal and red
sense lead into the sense (+) terminal. The clip at the other end will connect to the
ground pin of the DUT plug.
Did you know? If necessary, a 10 x 20 cm piece of foil can be attached to the
enclosure to simulate a full hand contact to the DUT. This foil acts as the return
point. The return and sense (-) terminal are attached to the foil. Always check to
make sure you made a good connection between the DUT and the return clip.
Class I Product
Class II Product
TESTRESET
POWER
ELECTRICAL SAFETY COMPLIANCE ANALYZER
OMNIA
®
SERIES
My
Menu
1 2
ABC DEF
GHI JKL MNO
PQRS TUV WXYZ
3
4 5 6
7
EXIT ENTER
SCANNER STATUS
H.V.
GND
HI
LO
CH
1 2 3 4 5 6 7 8
ON
8 9
0 <
- -
CAUTION
HIGH VOLTAGE
5KV MAX.
H.V.
SENSE+ SENSE
-
CURRENT RETURN
TESTRESET
POWER
ELECTRICAL SAFETY COMPLIANCE ANALYZER
OMNIA
®
SERIES
My
Menu
1 2
ABC DEF
GHI JKL MNO
PQRS TUV WXYZ
3
4 5 6
7
EXIT ENTER
SCANNER STATUS
H.V.
GND
HI
LO
CH
1 2 3 4 5 6 7 8
ON
8 9
0 <
- -
CAUTION
HIGH VOLTAGE
5KV MAX.
H.V.
SENSE+ SENSE
-
CURRENT RETURN
FOIL
6 7
TEST CONNECTIONS
WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.
ADAPTER BOX CONNECTIONS (8204/8254)
The adapter box (P/N 38482) provides an easy way to connect the OMNIA II to a
device under test (DUT).
Note: If you are testing a Class I product, safety agencies may require you to perform
a Ground Bond test on the ground circuit of the DUT along with the Hipot test.
Using the output ports on the rear panel, plug the black high voltage lead in
the High Voltage terminal.
Plug the sense lead into the Sense (+) terminal. Plug the high current test
lead into the Current terminal. These two red leads act as the ground
connection.
Connect the ground return clip (P/N 38490) to the dead metal on the chassis
of the DUT. Plug the sense lead into the Sense (-) terminal. Plug the return
lead into the Return terminal. Always check to make sure a good connection
is made between the DUT and the return clip.
Plug the power cord of the DUT into the adapter box receptacle.
SENSE+
H.V.
SENSE
-
CURRENT
RETURN
SIGNAL INPUT
SIGNAL OUTPUTSCANNER 1 SCANNER 2
RS232
T10A 250V
GND
CAT. II
115V~ /230V~
10A MAX., 50Hz/60HZ
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV MAX.
GROUND
TEST
OUTPUT
CH
H.V.
TEST
OUTPUT
OUTPUT
REMOTE INTERFACE
INPUT
G-COM. CH1 CH2 CH3 CH4 CH5 CH6 CH7 CH8
CAUTION
HIGH VOLTAGE
5KVAC MAX.
6KVDC MAX.
NL
CAUTION
HIGH VOLTAGE
5KVAC MAX.
6KVDC MAX.
ADAPTER BOX CONNECTIONS (8206/8256 & 8207/8257)
Note: Use the rear output terminal for connecting the adapter box (P/N 38578) on the
8206/8256 and 8207/8257 models.
Using the DUT output ports on the rear panel, plug the line lead from the
adapter box into the L terminal and the neutral lead into the N terminal.
Attach the red ground lead from the adapter box to the GND terminal.
Attach the black case return (P/N CBLHC40-10TL) lead into the case terminal.
Connect the ground return clip to dead metal on the chassis of the DUT.
Plug the line power cord from the DUT into the adapter box receptacle.
REMOTE OUTPUT
277V~MAX. 16A MAX.
EXTERNAL
CASE CND PROBE HI PROBE LO
L
L
N
N
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV
16A AX. MAX.
DUT OUTPUTSDUT POWER I/P MD CIRCUITS
SENSE+
H.V.
SENSE
-
CURRENT
RETURN
SIGNAL INPUT
SIGNAL OUTPUT
REMOTE OUTPUT
277V~MAX. 16A MAX.
EXTERNAL
CASE CND PROBE HI PROBE LO
L
L
N
N
SCANNER 1 SCANNER 2
RS232
T10A 250V
GND
CAT. II
115V~ /230V~
10A MAX., 50Hz/60HZ
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV MAX.
CAUTION
HIGH VOLTAGE
5KV
16A AX. MAX.
OUTPUT
REMOTE INTERFACE
INPUT
DUT OUTPUTSDUT POWER I/P MD CIRCUITS
CAUTION
HIGH VOLTAGE
5KVAC MAX.
6KVDC MAX.
NL
CAUTION
HIGH VOLTAGE
5KVAC MAX.
6KVDC MAX.
TEST CONNECTIONS
View the results from performed tests. Results include test type, reason for failure, and
test parameters.
RESULTS
START
From the Perform Tests screen, use
the arrow keys to choose the file to
see the test results.
Select the Results soft key.
Results include test type, reason
for failure and test parameters.
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
Select the EXIT key to return to the
Perform Tests screen
My
Menu
File Name: 0002 TEST
Single Step
Fail Stop
Load Files
Results
OFF
ON
0001 ACW Settings
1240 V 10.00mA
1.0s 0.000mA
My
Menu
File Name: 0002 TEST
Single Step
Fail Stop
Load Files
Results
OFF
ON
PERFORM TESTS
0002 LLT Settings
10000uA
0.5s 0.0uA
0001 ACW Settings
1240 V 10.00mA
1.0s 0.000mA
My
Menu
File Name: 0002 TEST
RESULTS
0002 LLT Settings
10000uA
0.5s 0.0uA
0001 ACW Settings
1240 V 10.00mA
1.0s 0.000mA
0002 LLT Settings
10000uA
0.5s 0.0uA
PERFORM TESTS
1 2
ABC
3
DEF
4 5
JKLGHI
6
MNO
7 8
TUVPQRS
9
EXIT ENTER
WXYZ
0
<
- -
My
Menu
File Name: 0002 TEST
Single Step
Fail Stop
Load Files
Results
OFF
ON
0001 ACW Settings
1240 V 10.00mA
1.0s 0.000mA
0002 LLT Settings
10000uA
0.5s 0.0uA
PERFORM TESTS
LLT
PASS
264.0
179.9
Volta
g
e
Leaka
g
e RMS
V
uA
179.9
I-Maximum
uA
179.6
MC Volt RMS
Neutral
Reverse
Ground
Probe Ph-Pl
1.0
Time
s
mV
For additional information about these and other key features of the OMNIA
®
II, please consult the
full Operation and Service Manual or call us toll-free 1-800-858-TEST (8378) or +1-847-367-4077
©2017 Associated Research arisafety.com
01/2017
Safety Is Our Only Focus
®
SCAN FOR
QUICK
START
VIDEO
FOLLOW US!