Issue Date: 2009-05-07 Page 5 of 50 Report Reference # E179824-A40-UL-2
TRF No.: IEC60601_1C Underwriters Laboratories Inc.
reasons in model designation only.
CD1.0
Additional Information
CD1.1 This report is re-issued to cover:
- Addition of of power supplies Jerome Industries, WSX824M; Skynet Electronic, SNP-A129-M;
Sinpro Electronics, MPU130-108
- Removal of power supplies Skynet, SNP-A127-M and Sinpro Electronics, MPU100-105
- Addition of alternative backlight inverter Sumida, type IV75270/T
- Descriptive changes in critical component list
- Addition of evaluation according to EN 60601-1: 1990 + A1:1993 + A2:1995.
- This report replaces report E179824-A40-CB-1, dated 2008-06-19.
- The equipment includes a power supply which has been separately investigated for reinforced
insulated outputs. The corresponding licenses are included in the enclosure "Licenses". A full test
report of the power supply may be required when submitting this CB Test Report to a National
Certification Body (NCB) for obtaining certification at national level. The Recognizing National
Certification Body (NCB), may challenge the CB Test Certificate when it is more than three years
old.
- This report does not include the investigation of the external power supplies:
- Skynet, SNP-A129-M
- Sinpro Electronics, MPU130-108
- Jerome Industries, WSX824M
- Supply frequency and power input are marked on the power supply.
- The fans are considered optional and not safety critical . They are intended for improved lamp
cooling only.
- Max. load during testing: max. contrast, max. brightness, 2 A total load on USB ports and other
I/O ports.
- This product contains visible indicator LEDs (with diffusing lens) only. No IEC 60825-1
evaluation was deemed necessary. Additional testing and evaluation may be required based on
auditing agency's discretion.
CE1.0
Technical Considerations
CE1.1 The product was investigated to the following
additional standards:
EN 60601-1: 1990 + A1:1993 + A2:1995,
(except EMC limitations, EN 60601-1-2,
Biocompatibility, EN 10993-1, Programmable
Electronic Systems, IEC 60601-1-4), CAN/CSA-
C22.2 No. 601.1-M90 (R2005) (includes
National Differences for Canada)
CE1.2 The product was not investigated to the
following standards or clauses:
Clause 36, Electromagnetic Compatibility (IEC
601-1-2), Clause 48, Biocompatibility (ISO
10993-1), Clause 52.1, Programmable
Electronic Systems (IEC 601-1-4)
CE1.3
The product is Classified only to the following
Shock, Fire, Casualty