Omega LVM-50 Series Owner's manual

Type
Owner's manual
1
2
3
INTRODUCTION/TABLEOFCONTENTS StepOne
About Omega Engineering’s SingleSwitch Fitting Assembly:Omega Engineering’s SingleSwitch Fitting
Assemblyisanindustrialmountingsystemforinstalling a liquid levelsensorverticallywithinatank.Omega
Engineering’s LVCN11 or LVCN12 electrical housing, and LVCN100, LVCN110 or LVCN20 compact relay
controllers are designed to be mounted directly to the 3/4" NPT fitting at the top of the assembly.Single
SwitchFittingAssemblymountsverticallythroughastandard2"NPTtankadapteroronaside mountbracket
(suchastheVM30).
SingleSwitchFittingAssembly:TheSingleSwitchFittingAssemblyconsistsofthefollowing:
Pipe:TheSingleSwitchFittingAssemblyfeaturesastandard¾”pipeastheextensionforthesensor.
When assembling a Single Switch Fitting Assembly, use a thermal weld for the LVM51 series (PP
version) and LVM52 series (PVDF version) and a solvent weld for the LVM50 series (PVC version).
WeldsmustbemadefortheFitting[2”]andfortheCoupling(3/4”).
Fitting[2”]:Thepipeslidesintotheslipconnectionofalargefittingwhichhasa¾”threadsonthetop
and a 2” standard threads on the bottom.The slip connection is located on the same side asthe 2”
threads.The¾”threadisavailableforattachingajunctionboxorconduit.
Coupling (3/4”):The coupling has a slip end and a ¾” threaded end.The other end of the pipe
attachestothe slipendofthecoupling.The3/4"threadontheotherendattachesthelevelsensorto
theSingleSwitchFittingAssembly.
TableofContents
Specifications:.........................................................................................................................................................4
Dimensions:.............................................................................................................................................................4
SafetyPrecautions:.................................................................................................................................................5
MakeaFailSafeSystem:............................................................................................................................5
AssemblyofaSingleSwitchAssembly:.................................................................................................................6
Components:..............................................................................................................................................6
Installation..............................................................................................................................................................7
MountingtoaSensor:...............................................................................................................................8
SelectingaLocation:.................................................................................................................................8

4
SPECIFICATIONS/DIMENSIONS StepTwo
SingleSwitchFittingAssemblyFitting:
Fittinglengths: 6”to6’(150mmto1.82m)
Fittingtolerance: 1/8”(3mm)typical
Temperaturerange.: F:‐40to176°
C:‐40to80°
Pressurerange: 150psi(10bar)@25°C.,derated@
1.667psi(0.113bar)per°Cabove25°C.
Trackmaterial: Polypropylene(PP),
PolyvinylideneFluoride(PVDF)or
Polyvinylchloride(PVC)
Mountingthread: 2”NPT
JunctionBox:
Enclosurerating: NEMA4X(IP65)
Enclosurematerial: Polypropylene(U.L.94VO)
Enclosurerotation: 300°swivelbase
Conduitconnection: ½”NPT
Termination: 6polesocketterminalstrip
Temperaturerange.: F:‐40to158°
C:‐40to70°
SideMountBracket:
BracketMaterial: Polypropylene
Mountingthread: 2”NPT
Tankinstallation: Boltorplasticweld
SwitchAddersforOmegaEngineeringLevelSensors:
NominalSensorAdders
Technology Lengthadder
LVU150series *2.0”(51mm)
LVC152series 4.0”(102mm)
LVF212Series *2.0”(51mm)
LVV110Series 4.3”(109mm)
*Basedupontheshortswitchlength
UserSuppliedPipeSizes:
PartNumber Material PipeSize
LVM50*
LVM51*
LVM52*
PVC
PP
PVDF
¾”(schedule40)
25mm(metric)
25mm(metric)
SingleSwitchFittingAssemblySideView

5
SAFETYPRECAUTIONS StepThree
AboutThisManual:PLEASEREADTHEENTIREMANUALPRIORTOINSTALLINGORUSINGTHISPRODUCT.
This manual includes information on the SingleSwitch Fitting Assembly installation fitting from Omega
Engineering: LVM50/51/52 series.Please refer to the part number located on the label to verify the
exactmodelwhichyouhavepurchased.Manyaspectsofinstallationandusearesimilarbetweenmodels.
User’s Responsibility for Safety:OMEGA ENGINEERING manufactures several models of liquid level
sensors, controller and mounting systems.It is the user’s responsibility to select components that are
appropriatefor the application, install itproperly, perform tests ofthe installed system, and maintainall
components.Thefailuretodosocouldresultinpropertydamageorseriousinjury.
Proper Installation and Handling:Use a proper sealant with all installations.Never over tighten the
components.Alwayscheckforleakspriortosystemstartup.
MaterialCompatibility:Thepipe,fittingandcouplingforalltheLVM50/51/52seriesaremadeofoneof
threematerials,Polypropylene(PP),PolyvinylideneFluoride(PVDF)orPolyvinylchloride(PVC).Makesure
that the application liquids are compatible with the materials that will be wetted.To determine the
chemical compatibility between the components and its application liquids, refer to the Compass
CorrosionGuide.
TemperatureandPressure:Single SwitchFittingAssemblyisdesignedforuseinapplicationtemperatures
upto80°C(176°F).Itisdesignedforuseinapplicationpressuresupto150psi(10bar).
Wiring and Electrical:Electrical wiring of any liquid level control system should be performed in
accordance with all applicable national, state, and local codes.Take care not to cut or break the outer
insulation jacket of wiring that may be immersed while routing cables in the Smart Trak system.Such
breaksoftheliquidsealofthesensorsystemmayleadtocomponentfailure.
Flammable or Explosive Applications: SingleSwitch Fitting Assembly may be used within flammable or
explosiveapplicationsonlyiftheassociatedcomponentsareratedintrinsicallysafeforsuchuse.Referto
the National electric Code (NEC) for all applicable installation requirements in hazardous locations.In
hazardous applications, use redundant measurement and
control points, each having a different sensing
technology.
MakeaFailSafeSystem:Designafailsafesystemthataccommodatesthepossibilityofsensororpower
failure.Incriticalapplications,OmegaEngineeringrecommendstheuseofredundantbackupsystemsand
alarmsinadditiontotheprimarysystem.

6
ASSEMBLYOFSINGLESWITCHFITTINGASSEMBLY StepFour
SingleSwitchFittingUnassembledDrawing
(sideview)
SingleSwitchFittingAssemblyDrawing
(sideview)
Components:One SingleSwitch Fitting Assembly
(LVM50*, LVM51*, LVM52* series) includes
thefollowingparts:
1Fitting(oneofthefollowing)
o LVM50*
o LVM51*
o LVM
52*
o *=Fittingsonly,nopipesupplied
1Coupling
Owner’sManual
Pipeisusersupplied.
PartNumber Material PipeSize
LVM50*
LVM51*
LVM52*
PVC
PP
PVDF
¾”(schedule40)
25mm(metric)
25mm(metric)
Components:One SingleSwitch Fitting Assembly
(LVM511.0, LVM511.5, LVM512.0, LVM512.5,
LVM513.0, LVM513.5, LVM514.0, LVM514.5,
LVM515.0, LVM515.5 & LVM516.0 series)
includesthefollowingparts:
1Fitting(oneofthefollowing)
o LVM511.0’
o LVM511.5’
o LVM512.0’
o LVM512.5’
o LVM513.0’
o LVM513.5’
o LVM514.0’
o LVM514.5’
o LVM515.0’
o LVM515.5’
o LVM516.0’
1Coupling
1CustomerSpecifiedPipe
Owner’sManual
7
INSTALLATION StepSix
SingleSwitchFitting Assemblycanbeshippedinoneoftwomethods,AssembledandUnassembled.Withthe
assembled version, the LVM50/51/52 series has been shipped a one complete fitting and the first step
belowcanbeshipped.Withtheunassembledversion,SingleSwitchFittingAssemblyisshippedwithjustthe
fittingandcouplingandtheusersuppliedpipewillneedtobeadded.WhenassemblingaSingleSwitchFitting
Assembly, use a solven t weld for the LVM50* (PVC version) and a thermal weld for the LVM51* (PP
version)andLVM52*(PVDFversion).
1. AssemblinganunassembledSingleSwitchFittingAssembly:
a. Makesurethepipebeingusedisthesamematerialofconstructionasthefittingandcoupling
received.
i. LVM50*ismadeof¾”(schedule40)Polyvinylchloride(PVC)
ii. LVM51*ismadeof25mm(metric)Polypropylene(PP)
iii.
LVM52*ismadeof25mm(metric)PolyvinylideneFluoride(PVDF)
b. Cutthepipetothecorrectlengthtakingintoaccountthetotalheightofthefittingandcoupling
andthedepththepipewillextendintothepipesockets.
c. Prepare the pipe as well as the pipe
sockets on the fitting and coupling for assembly making
sure they are clean for either the thermal weld (PP or PVDF versions) or solvent weld (PVC
version).
d. Weld the pipe to the fitting and to the coupling using a thermal weld for the PP and PVDF
fittingsandasolventweldforthePVCfittings.
Allowtimefortheweldstocurebeforecontinuingontothenextstep.
2. Attachsensortothecoupling:
a. Threadthesensor’swirethroughthethreadedpartofthe couplingandallthewaythroughthe
LVM50/51/52seriesuntilitextendsouttheotherend.
i. Makesuretheendofthewireissufficientforconnectionsontheotherend.Ifnot,you
may be required to extend the wire within the LVM50/51/52 series to achieve the
correctwirelength.
b. Applyasealant,suchasTeflontapetothethreadsofthesensor.
c. Threadthesensorintothecouplingmakingsureitishandtight.
3. Attach the termination hardware:At this point, a junction box housing (LVCN11/12 series) or
Compact Relay Controller (LVCN20/100/110 series) may be screwed onto the top of the assembly,
and wiring attachments made following the instructions in its manual.Assuming that each sensor is
alreadymountedinposition,cutthesensorwirestothelengthapproximately11/2”abovethetopof
the housing.However, be sure to make allowances when cutting the sensor leads for future
adjustmentstothesensor
position.
Note:Alwaystightenthecontrollerfrom
the wrench flat located on the swivel
base.Nevertightenfromthebodyofthe
controller.
8
INSTALLATION(continued) StepSix
SelectingaLocation:TheSingle SwitchFittingAssemblysystemshouldbemountedverticallyatapointinthe
tankwhereitwillnotbeexposedtoexcessivestress.Whenmountinginatankwithamixer,mounttheLVM
50/51/52 series close to the tank wall were velocities are lowest.Choose a mounting location where the
sensors will function correctly (away from inlet pipes that may spray on the sensors causing false readings)
andwherethesensortechnologiesfunctionbest.
It is the user’s responsibility to identify the proper placement and method of installation for the specific
application.
TankTopMounting:
1. Installastandard2”NPTtankadapterthroughthetopofthetankatthe
desiredlocation.Ifthetanktopisnotflat,usea2”NPTselfaligningtank
adapter.Makesurethethreadsoftheadapterarenotdamagedorworn.
Ifdamaged,useanewadapter.
2. Insert the assembled SingleSwitch Fitting Assembly system through the
tankadapter.Ifseveralsensorsaretooclosetogethertofitthroughthe
tankadapter,slidethemtoadifferent locationssothatonlyonesensoris
goingthroughtheadapteratatime.
3. To ensure a proper seal, apply an appropriate sealant to the threads of
the SingleSwitch Fitting Assembly system.Screw the LVM50/51/52
seriesassemblyintothetankadapter.
4. Makeadjustmentsasrequiredtothesensorposition.
5. Connect the external wiring and conduit to the terminal strip or
controller,followingtheinstructionsinthatmanual.
SideMountBracketMounting:
1. Install the side mount bracket (LVM30 series) at the desired location.
Make sure the threads of the bracket are not damaged or worn.If
damaged,useanewbracket.
2. Insert the assembled SingleSwitch Fitting Assembly system into the
bracket
3. ScrewtheLVM50/51/52seriesassemblyintothebracket.
4. Makeadjustmentsasrequiredtothesensorposition,andtothebracket
tiltingarm.
5. Connect the external wiring and conduit to the terminal strip or
controller,followingtheinstructionsinthatmanual.

9

10

11

12
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12

Omega LVM-50 Series Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI