LINET Eleganza 1 Operating instructions

Type
Operating instructions
ELEGANZA 1
Hospital bed for standard care
D9U001GT0-0101
Version: 10
Publication Date: 2023-01
Instructions for Use
and Technical Description
D9U001GT0-0101_102
Manufacturer:
L I N E T spol. s r.o.
Želevčice 5
274 01 Slaný
Tel.: +420 312 576 111
Fax: +420 312 522 668
http://www.linet.com
Service department: [email protected]
Authorized Representative in Great Britain:
LINET UK Ltd
11 Brunel Way
Segensworth East
Fareham
Hampshire
PO15 5TX
United Kingdom
Authorized Representative in Switzerland:
Bigla Care AG
Bernstrasse 3
CH-3421 Lyssach
Switzerland
Authorized Representative in Malaysia:
Emergo Malaysia Sdn. Bhd.
Level 16, 1 Sentral, Jalan Stesen Sentral 5
KL Sentral
50470 Kuala Lumpur
Malaysia
Authorized Representative in Peru:
Drogería Emergo Peru S.R.L.
Calle Las Orquídeas Nro. 585, Int. 1301
San Isidro, Lima
Director Técnico Renato Delgado Rivera
RUC: 205 52 75 65 35
Peru
Authorized Representative in Philippines:
Biomedica Healthcare Inc.
Unit 2103 City land 10, Tower 1
Ayala Corner, Dela Costa Street
Makati City, Metro Manila
Philippines
D9U001GT0-0101_10
3
Eleganza 1
Hospital bed for standard care
Author: L I N E T, s.r.o.
Related links: www.linet.cz
D9U001GT0-0101
Version: 10
Publication Date: 2023-01
Copyright © L I N E T, s.r.o., 2023
Translation © L I N E T, 2023
All rights reserved. All trademarks and brands are the property of the appropriate owners. The manufacturer reserves the right to
changes in the contents of the instructions for use that relate to the product´s technical regulations.
It is for this reason that the contents of the instructions for use may indicate dierences from the current manufacture of the pro-
duct. Reproduction, also excerpts, only with prior permission of the publisher. Subject to changes due to technical developments.
All technical data are rated data and are subject to construction and manufacturing tolerances.
D9U001GT0-0101_104
1 Symbols and Denitions .............................................5
1.1 Warning Notices .......................................................... 5
1.1.1 Types of Warning Notices ........................................5
1.1.2 Structure of Warning Notices ...................................5
1.2 Instructions ..................................................................5
1.3 Lists .............................................................................5
1.4 Symbols on the Package ............................................6
1.5 Symbols and Labels on the Bed .................................7
1.6 Acoustic signalisation (Eleganza 1 with PB43 control unit)
..........................................................................................10
1.7 Acoustic signalisation (Eleganza 1 with PB11 control unit)
..........................................................................................10
1.8 Visual signalisation......................................................11
1.8.1 Mains Power LED (Attendant Control Panel) ...........11
1.8.2 Accumulator Indicator (Attendant Control Panel, MiniACP)
..........................................................................................11
1.8.3 Lock LED (Attendant Control Panel, MiniACP) ........ 11
1.9 Denitions ...................................................................12
1.10 Abbreviations ............................................................13
2 Safety Instructions .......................................................14
2.1 Additional safety instructions for Extra Safe Bed ....... 17
3 Intended use ................................................................. 18
3.1 User population ........................................................... 18
3.2 Contraindications ........................................................18
3.3 Operator ...................................................................... 18
4 Incorrect Use ................................................................18
5 Product Description ..................................................... 19
5.1 Eleganza 1 (1GTP) with Single Collapsible Siderails .19
5.2 Eleganza 1 (1GTP) with Split Plastic Siderails (SR22)
..........................................................................................20
5.3 Eleganza 1 (1GTP) with Split Plastic Siderails (SR46)
..........................................................................................21
5.4 Eleganza 1 (1GTP) – Single collapsible siderails controlled
from top ............................................................................22
5.5 Eleganza 1 (1GTP) – Single collapsible siderails controlled
from bottom ......................................................................23
6 Technical Specication ...............................................24
6.1 Type B Applied Parts ...................................................24
6.2 Mechanical Specication (standard version) .............24
6.3 Mechanical Specication (shortened version) ...........25
6.4 Safe Working Load (SWL) ..........................................26
6.5 Maximum Patient Weight ............................................26
6.6 Mass of Mobile Hospital Bed ......................................26
6.7 Environment Conditions .............................................. 26
6.8 Electrical Specications ..............................................27
6.9 Electromagnetic compatibility .....................................27
6.9.1 Manufacturer instructions - electromagnetic emissions
..........................................................................................28
6.9.2 Manufacturer instructions - electromagnetic susceptibility
..........................................................................................28
7 Use and Storage Condition ......................................... 29
8 Scope of Delivery and Bed Variants ........................... 30
8.1 Scope of Delivery ........................................................ 30
8.2 Bed Variants ................................................................30
8.2.1 Eleganza 1 Model 1GTP Hospital Bed.....................30
8.2.2 Eleganza 1 Model 1GTP Extra Safe Bed .................30
9 Putting into Service .....................................................32
9.1 Accumulator Activation ................................................ 33
9.1.1 Placement of Control Section ..................................33
9.1.2 Removing the Isolating Foil ......................................33
9.1.3 Isolating Foil ............................................................. 33
9.2 Head Board and Foot Board .......................................34
9.3 Mattress support platform ...........................................36
9.3.1 Shortened Bed .........................................................37
9.4 Potential Equalisation .................................................38
9.5 Before Use ..................................................................39
9.6 Transport ..................................................................... 39
9.7 Firmware .....................................................................39
10 Power Cable ...............................................................40
10.1 Lockable box for mains cable (Extra Safe Version) .. 40
11 Accumulator ...............................................................41
11.1 Status Faulty Accumulator .........................................42
11.2 Status Discharged Accumulator ................................ 42
11.3 Removing the Bed from Service...............................43
11.4 Deactivating the Accumulator ................................... 43
12 Manipulation ............................................................... 44
12.1 MiniACP (Supervisor board) .....................................44
12.2 ACP – Attendant Control Panel ................................. 45
12.2.1 Central STOP Button .............................................45
12.2.2 Activating GO Button ..............................................46
12.2.3 Function Buttons ....................................................46
12.2.4 Lock Buttons ..........................................................46
12.2.5 Mains Power LED ..................................................46
12.2.6 Position Buttons .....................................................46
12.3 Handset .....................................................................47
12.4 Satellite Control Panel ..............................................48
12.5 Control Element Integrated in Siderail ......................49
12.6 Foot Control ..............................................................50
12.7 CPR backrest release ...............................................51
12.8 Siderails ....................................................................52
12.8.1 Collapsible siderail (3/4 siderail) ............................52
12.8.2 Single collapsible siderails – controlled from the bottom
..........................................................................................53
12.8.3 Split Plastic Siderails (SR22) .................................54
12.8.4 Split Plastic Siderails (SR46) .................................55
12.9 Castor Control and Bed Transport ............................ 56
12.9.1 Central braking control ........................................... 56
12.9.2 Individual braking control .......................................57
13 Equipment ..................................................................58
13.1 Accessory Rails .........................................................58
13.2 Green light (optional) ................................................58
14 Mattress ......................................................................58
15 Accessories ................................................................59
15.1 Lifting Pole ................................................................60
15.2 Infusion Stands .........................................................60
15.3 Name holder .............................................................61
15.4 Urinary bottle holder ..................................................61
15.5 Urinary bag holder (optional) ...................................61
15.6 Eleganza Protector® ................................................. 61
15.7 Siderail extension – „Extender®“ for siderails controlled
from top (SR55) ................................................................ 62
15.8 Siderail extension – „Extender®“ for siderails controlled
from bottom (SR57) .......................................................... 63
16 Cleaning/Disinfection ................................................65
16.1 Safety Instructions for Cleaning and Disinfection of the Bed
..........................................................................................66
16.2 General Instructions for Cleaning and Disinfection ... 66
16.2.1 Daily Cleaning ........................................................ 66
16.2.2 Cleaning before Changing Patients .......................66
16.2.3 Complete Cleaning / Cleaning before First Use .....66
16.2.4 Cleaning of Spilled Fluids ......................................66
16.2.5 Damaged Foam Mattress ......................................66
16.3 Modes of Cleaning and Disinfection .........................67
17 Troubleshooting ......................................................... 68
18 Maintenance ...............................................................69
18.1 Regular maintenance ................................................ 69
18.2 Spare Parts ...............................................................69
18.3 Safety Technical Checks ........................................... 69
19 Disposal ......................................................................70
19.1 Environment Protection .............................................70
19.2 Disposal ....................................................................70
19.2.1 Within Europe ........................................................70
19.2.2 Outside Europe ......................................................70
20 Warranty ......................................................................71
21 Standards and Regulations ......................................71
Table of Contents
D9U001GT0-0101_10
5
1 Symbols and Denitions
1.1 Warning Notices
1.1.1 Types of Warning Notices
Warning notices are dierentiated by the type of danger using the following key words:
CAUTION warns about the risk of material damage.
WARNING warns about the risk of physical injury.
DANGER warns about the risk of fatal injury.
1.1.2 Structure of Warning Notices
SIGNAL WORDS!
Type and source of danger!
► Measures to avoid the danger.
1.2 Instructions
Structure of instructions:
► Perform this step.
Results, if necessary.
1.3 Lists
Structure of bulleted lists:
■   List level 1
List level 2
●   List level 3
D9U001GT0-0101_106
1.4 Symbols on the Package
FRAGILE, HANDLE WITH CARE
THIS WAY UP
KEEP DRY (PROTECT FROM HUMIDITY)
PAPER RECYCLING SYMBOL
DO NOT USE HAND TRUCK HERE
DO NOT STACK DURING STORAGE
D9U001GT0-0101_10
7
1.5 Symbols and Labels on the Bed
READ INSTRUCTIONS FOR USE
WARNING
THERMAL PROTECTION OF TRANSFORMER
ONLY SUITABLE FOR INDOOR USE
PROTECTION AGAINST ACCIDENTS DUE TO ELECTRICAL CURRENT – TYPE B
APPLIED PARTS
SAFETY ISOLATING TRANSFORMER, GENERAL
CE MARK OF CONFORMITY WITH EU REGULATION
SAFETY MARK – BED CERTIFIED IN TÜV SÜD
JACK FOR ATTACHMENT OF CONDUCTOR FOR POTENTIAL EQUALISATION
SAFE WORKING LOAD
WARNING AGAINST CRUSHING OR TRAPPING
MAXIMUM WEIGHT OF PATIENT
D9U001GT0-0101_108
USE MATTRESS RECOMMENDED BY MANUFACTURER
WEIGHT OF BED
DO NOT PUT ANY OBJECTS HERE
DESIGNATION OF HOSPITAL BED FOR ADULTS
ALTERNATING CURRENT
DIRECT CURRENT
EAC MARKING
WEEE SYMBOL (RECYCLE AS ELECTRONIC WASTE,
DO NOT PUT INTO THE HOUSEHOLD WASTE)
MEDICAL DEVICE (compatible with Medical Device Regulation)
RECYCLING SYMBOL
DO NOT POLLUTE THE ENVIRONMENT
D9U001GT0-0101_10
9
HEAD BOARD UNLOCKED OR FOOT BOARD UNLOCKED
HEAD BOARD LOCKED OR FOOT BOARD LOCKED
MASS OF MOBILE HOSPITAL BED (MASS OF EMPTY BED + SAFE WORKING LOAD)
UK CONFORMITY ASSESSED (UKCA) MARKING (PRODUCT NORMATIVELY
HARMONIZED FOR GREAT BRITAIN ECONOMIC AREA)
AUTHORIZED REPRESENTATIVE IN GREAT BRITAIN
AUTHORIZED REPRESENTATIVE IN SWITZERLAND
REPCH
D9U001GT0-0101_1010
1.6 Acoustic signalisation (Eleganza 1 with PB43 control unit)
SOUND MEANING
CONTINUOUS SOUND overheating
accumulator overcurrent
actuator overload
REPEATED BEEP: 0,6s sound / 2,6s silence STOP error (all STOP buttons are disabled)
REPEATED BEEP: 0,1s sound / 3s silence switching of the transformer winding error (Brazil)
BEEP lasting 0,3s conrmation
stopping or locked function
optionally: transition from tilt (Trendelenburg, Antitrendelenburg)
to horizontal position
BEEP lasting 0,5s start of service mode or end of service mode
keyboard error (positioning blocked)
BEEP lasting 3s system error
REPEATED BEEP during 3 minutes: 1,1s sound / 1,1s silen-
ce
Brake Signal (only version with Brake Signal)
1.7 Acoustic signalisation (Eleganza 1 with PB11 control unit)
There are no acoustic signals in the case of Eleganza 1 bed with PB11 control unit.
D9U001GT0-0101_10
11
1.8.3 Lock LED (Attendant Control Panel, MiniACP)
VISUAL
SIGNALISATION
LOCK LED
lit
ashing: 
0,6s lit / 0,6s not lit not lit
Thighrest Lock LED locked lock error keyboard error motion blocked unlocked
Backrest Lock LED locked lock error keyboard error motion blocked unlocked
Bed Height, Trendelenburg and Antitrendelenburg
Tilt Lock LED
locked lock error keyboard error motion blocked unlocked
Foot Control Lock 
LED
locked lock error keyboard error motion blocked unlocked
ACCUMULATOR INDICATOR MEANING
lit accumulator disconnected or faulty
ashing: 1,6s lit / 0,2s not lit accumulator deeply discharged
ashing: 0,1s lit / 0,1s not lit accumulator discharged
ashing: 0,2s lit / 1,6s not lit accumulator is charging
not lit accumulator charged
1.8 Visual signalisation
1.8.1 Mains Power LED (Attendant Control Panel)
1.8.2 Accumulator Indicator (Attendant Control Panel, MiniACP)
MAINS POWER LED MEANING
lit connected to the mains
ashing: 0,6s lit / 0,6s not lit keyboard error (ashing inverted to Lock LED)
error (rst fault)
ashing: 0,1s lit / 0,1s not lit service mode
not lit disconnected from the mains power
transformer switching error
D9U001GT0-0101_1012
1.9 Denitions
Basic Bed Conguration the pricelist model conguration, not including a mattress
Bed Weight The value depends on the product conguration, accessories or customer
adjustments.
Clearance of Undercarriage the height from the oor to the lowest point of the undercarriage between
the castors, for the manipulation of accessories under a braked bed in the
standard position
Duty Cycle cycle of operation of the motor: time of activity/time of rest
Ergoframe Ergoframe is the kinematic system of Mattress support platform Adjustment
whose eect is the elimination of pressure on the patient´s abdomen and
pelvic area and frictional forces on the patient‘s back and legs.
Maximum Patient Weight Maximum Patient Weight depends on the application environment according
to IEC 60601-2-52. For application environment 1 (intensive/critical care) and
2 (acute care) reduce Safe working Load by 65 kg. For application environ-
ment 3 (long-term care) and 5 (ambulatory care) reduce Safe working Load
by 35 kg.
Safe Working Load  the highest allowable load on the bed (patient, mattress, accessories and the
load supported by those accessories)
Siderail Height the height of the upper crossbar or the edges of the siderails (not the highest
point of the siderail controls) from the patient surface
Standard Bed Position ■ The height of the patient surface with regard to the oor is 400 mm
■ The mattress support platform, including the individual parts, has to be in a
horizontal (level - 0°) position.
■ The siderails are always locked in the upper position.
■ The basic position of the integrated extension.
Adult Patient having a physical size equal to or more than 146 cm, a mass equal
to or more than 40 kg and a body mass index (BMI) equal to or more than 17
(according to IEC 60601-2-52).
Mass of mobile hospital bed Sum of empty bed mass and Safe Working Load.
D9U001GT0-0101_10
13
AC ( ~ ) Alternating Current
ACP Attendant Control Panel
CE European Conformity
CPR Cardiopulmonary Resuscitation
dBA Sound Intensity Unit
DC ( ) Direct Current
CUC Conguration number
EMC Electromagnetic Compatibility
FET Field-eect transistor
HF High Frequency
HPL High Pressure Laminate
HW Hardware
ICU Intensive Care Unit
INT. Duty Cycle
IP Ingress Protection
IV Intravenous
LED Light Emitting Diodes
ME Medical Electrical (Equipment)
ON Activation
OFF Deactivation
ppm parts per million, millionth (1000 ppm = 0,1%)
REF Reference Number (product type depending on conguration)
SCU System Control Unit (active mattress)
SN Serial Number
SW Software
SWL Safe Working Load
UDI Unique Device Identication (for medical devices)
USB Universal Serial Bus
WEEE Waste Electrical and Electronic Equipment
1.10 Abbreviations
D9U001GT0-0101_1014
2 Safety Instructions
WARNING!
Eleganza 1 bed should be left in its lowest position when the patient is unattended in order to reduce risk of 
injury due to falls!
WARNING!
Siderails of Eleganza 1 should be located in the „up“ position to reduce the risk of the patient accidentally 
slipping or rolling o the mattress!
WARNING!
Incompatible siderails and mattresses can cause an entrapment hazard!
WARNING!
Inappropriate handling of the power supply cord, e. g. by kinking, shearing or other mechanical damages is 
hazardous!
WARNING!
When routing cables from other equipment in the Eleganza 1 bed avoid squeezing those between parts of the
Eleganza 1 bed!
WARNING!
Eleganza 1 bed should not be used with bed hoists and bed lifts!
WARNING!
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
WARNING!
During specic investigations or treatments the signicant risks of reciprocal interference posed by medical 
electrical equipment may occur.
WARNING!
No modication of this equipment is allowed.
WARNING!
Do not modify this equipment without authorization of the manufacturer.
WARNING!
If this equipment is modied, appropriate inspection and testing must be conducted to ensure continued safe 
use of the equipment.
WARNING!
An additional multiple socket-outlet or extension cord shall not be connected to the medical electrical system.
D9U001GT0-0101_10
15
WARNING!
Incompatible mattresses can create hazards.
WARNING!
The bed is intended for adults.
►  Follow the chapter Intended use.
WARNING!
Length adjustment of the bed must be proportional to the height of patient!
WARNING!
Risk of trapping or squeezing because of patient´s body constitution disproportionate to the size of mattress 
support platform!
WARNING!
Risk of injury or death due using of incorrect equipment!
►Always conduct the risk assessments required for the selection of suitable equipment.
WARNING!
Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the
competent authority of the Member State in which the user and/or patient is established!
WARNING!
Only authorised and trained person using the tool is allowed to change fuses and power supplies!
WARNING!
This medical device is not intended for oxygen enriched environment!
WARNING!
This medical device is not intended for use with ammable substances!
WARNING!
This medical device is not portable medical electrical equipment!
WARNING!
Make sure the duty cycle (2 min ON/18 min OFF) is not exceeded during bed positioning!
WARNING!
Patient is allowed to use selected control elements only if hospital personnel had assessed that the patient´s 
physical and psychological state is in accordance with use of them and only if the hospital personnel had
trained the patient in accordance with the instructions for use!
D9U001GT0-0101_1016
► Follow the instructions for use carefully.
Any non-observance of these instructions for use may lead to injuries or damage of property.
► Exclusively use the bed if it is in perfect working order.
► If necessary, check the bed functions daily or at each shift change.
► Use the bed exclusively in its original condition.
► Use the bed exclusively with the correct mains supply.
► Ensure the bed is operated by qualied personnel trained in using the bed in accordance with the instructions for use by
manufacturer or by person approved by the manufacturer, eventually by authorized representative or by person approved
by the authorized representative.
► Ensure that the patient (health permitting) has been informed about the operation of the bed and all applicable safety
instructions.
► Move the bed exclusively on even, hard-surfaced oors.
► Replace any damaged parts immediately with original spare parts.
► Ensure that maintenance and installation are performed exclusively by qualied personnel who have been trained by the
manufacturer.
► Do not apply excess weight or loads to the bed according to SWL (Safe Working Load).
► During peak loads or unavoidable excess loads (CPR), place the mattress platform in the lowest position.
► Ensure that only one adult patient uses the bed at any time.
► Take care to avoid squeezing when operating moving parts.
► When using lifting poles or infusion stands, ensure that nothing will be damaged when you move or adjust the bed.
► Ensure that the castors are braked when the bed is occupied or not being moved.
► Ensure that the castors are braked when the bed is not being moved, regardless of whether it is occupied or empty.
► Keep the mattress platform in the lowest position at any time when the healthcare personnel are not treating the patient in
order to prevent the patient from falling or injuries.
► Ensure that siderails are operated exclusively by healthcare personnel.
► Ensure no injuries or damages will occur when folding the siderails.
► Never use the bed in areas where there is a hazard of explosion.
► Enable or disable functions on the handset using the Attendant Control Panel as appropriate for the patient’s physical and
mental state. Verify that the function is actually disabled.
► Never handle the mains plug with wet hands.
► Remove the mains cable exclusively by pulling on the plug.
► Position the mains cable so that there are no loops or bends in the cable; protect the cable from mechanical wear
and tear.
► Improper handling of mains cable can cause an electric shock hazard, other serious injuries or damage the bed.
► Ensure that the stipulated duty cycle (on-time) is not exceeded (see INT. on product label).
► Ensure that the moving parts of the bed are not blocked.
► To prevent failures, use exclusively the manufacturer’s original accessories and mattresses.
► Ensure that the stipulated safe working load is not exceeded.
► If the patient‘s condition could lead to an entrapment, leave the mattress support platform in the at position whilst
unattended.
Adjust bed height to approx. 20 cm below maximum height when transporting the bed in order to facilitate overcoming
possible obstacles.
► Do not exceed maximum load of 150 kg for mattress platform extension.
► Ensure that the bed and its components are exclusively modied with the manufacturer´s approval.
► Ensure there is no danger of crushing or otherwise injuring the patient’s limbs (e.g. between siderails and mattress
platform, between movable parts etc.) before positioning the bed or collapsing the siderails.
► Close linen shelf before using the Reverse Trendelenburg position.
► Do not put any objects (e.g. accessories, infusions, cables) between or on siderails and movable parts.
Always set mattress platform to its lowest position and single parts of mattress platform to horizontal position in case the
patient is left on the bed without supervision of personnel and if his health and mental status may indicate increase risk of
falling out of the bed or entrapment.
► Personnel must consider overall adjustment of the bed and locking all of the positioning functions in accordance to health
and mental status of patient, especially if the patient is left without supervision (even for short period of time) of the
personnel.
► Manual positioning of parts of the bed which are designed for electronic positioning (e.g. backrest) is forbidden. Otherwise
there is a risk of damaging and dysfunction of the backrest actuator or unprompted fall of the backrest.
D9U001GT0-0101_10
17
2.1 Additional safety instructions for Extra Safe Bed
► Use of removable accessories (lifting pole, infusion stand, Protector, accessories on the accessory rail etc.) is not
recommended. Using this type of accessories must be considered by responsible personnel after evaluation of patient’s
health and mental condition.
► Hospital personnel must consider leaving handset at the bed and leaving bed in the mains while the patient is without
supervision. It is recommended for mentally unstable patients to remove handset and to lock mains cable to the box under
head board.
► Use of bed extension is not recommended for mentally unstable patients.
► Periodically check if the bed placed on the psychiatric department is not damaged. In case of any damage of the bed,
replace the damaged parts.
Always check grounding of the bed in case the bed is equipped with grounding chain.
► Fixation of the bed to the oor must be done only by qualied personnel.
Always x the bed to the ground using 8 screws.
D9U001GT0-0101_1018
3 Intended use
The intended use is the hospitalization of the patient in the acute and long-term care units, which includes above all the
following aspects:
Adjustment of the specic positions needed for the preventive reasons, routine nursing, treatments, mobilization,
physiotherapy, examinations, sleeping, and relaxation. These positions are further specied and described in the clinical
evaluation of this device, together with their potential clinical outcomes and benets.
Providing the safe environment for the patient during all relevant procedures. The particular requirements on patient safety
are the subject of the clinical evaluation, including evaluation of the risk/benet ratio. The relevant safety issues are the
part of the risk management le.
Patient in-bed indoor transport out of the patient room.
Providing the suitable working conditions for the caregivers to perform the routine and specic tasks during the patient
hospitalization.
3.1 User population
Adult patients (weight >= 40 kg, height >= 146 cm, BMI >= 17) in the acute and long-term care units (Application
Environment 2 and 3, as in IEC 60601-2-52)
Caregivers (nurses, doctors, technical personnel, transport personnel, cleaning personnel)
3.2 Contraindications
The medical device is not intended for the pediatric patients use.
Certain positions are not suitable for specic diagnoses/medical conditions (e.g. spinal cord injuries vs. Fowler position,
higher ICP patients vs. Trendelenburg). Sta expert assessment / nursing consideration is needed in all individual case of
contraindication.
3.3 Operator
Caregiver
Patient (based on individual patient status assessment by caregiver the patient can utilize dedicated device functions)
4 Incorrect Use
Eleganza 1 is not suitable for:
■ Patients
Not fullling conditions stated in chapter “Intended use”
■  Use
Home care
Other than described in the instructions for use
LINET®’s eorts in research, design and manufacture ensure LINET® products are of the highest quality and t for their intended
purpose. However, LINET® can take no responsibility for any damage to the products or any harm to patients, sta or other indivi-
duals resulting from:
■    Not following the instructions in the instructions for use, including warning notices.
■    Using the product for a purpose other than the intended purpose stated in the relevant documentation provided by
LINET® (see Intended use).
D9U001GT0-0101_10
19
5 Product Description
5.1 Eleganza 1 (1GTP) with Single Collapsible Siderails
1. Foot Board
2. Removable plastic parts
3. Siderail release mechanism
4. Single collapsible siderails
5. Head Board
6. Handset
7. Castor
8. Fifth Castor
9. CPR lever (Backrest release)
10. Central castor control lever
11. Calfrest Handle
12. Corner Bumper
13. Linen shelf
14. MiniACP (Supervisor board)
NOTE For safe, easy handling, LINET® recommends having two technicians assemble the bed.
Fig. Overview Eleganza 1 – Single collapsible siderails
D9U001GT0-0101_1020
5.2 Eleganza 1 (1GTP) with Split Plastic Siderails (SR22)
1. Foot Board
2. Split collapsible siderails
3. Control element, integrated in siderail
4. Head Board
5. Castors
6. Siderail release mechanism (based on tipping down)
7. Central castor system lever
8. Linen shelf
9. Supervisor board
Fig. Overview Eleganza 1 – Split plastic siderails
NOTE For safe, easy handling, LINET® recommends having two technicians put the bed into service.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71