Bacharach PCA 55 User manual

Category
Oxygen Equipment
Type
User manual

This manual is also suitable for

Portable Combustion
Analyzer (PCA)
Instruction 0024-9219
Operation & Maintenance
Rev. 10 – May 2010
Product Leadership • Training • Service • Reliability
WARRANTY
Bacharach, Inc. warrants to Buyer that at the time of delivery this Product
will be free from defects in material and manufacture and will conform
substantially to Bacharach Inc.’s applicable specifications. Bacharach’s
liability and Buyer’s remedy under this warranty are limited to the repair or
replacement, at Bacharach’s option, of this Product or parts thereof returned
to Seller at the factory of manufacture and shown to Bacharach Inc.’s
reasonable satisfaction to have been defective; provided that written notice
of the defect shall have been given by Buyer to Bacharach Inc. within two (2)
year after the date of delivery of this Product by Bacharach, Inc. The Oxygen
(O
) Sensor has a limited warranty of one (1) year. Routine service and/or
2
calibration are excluded from this warranty.
Bacharach, Inc. warrants to Buyer that it will convey good title to this
Product. Bacharach’s liability and Buyer’s remedy under this warranty of
title are limited to the removal of any title defects or, at the election of
Bacharach, to the replacement of this Product or parts thereof that are
defective in title.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN
AND ACCEPTED IN LIEU OF (I) ANY AND ALL OTHER WARRANTIES,
EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR
PURPOSE:
AND
(II)
ANY
OBLIGATION,
LIABILITY,
RIGHT, CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER OR
NOT
ARISING
FROM
BACHARACH’S
NEGLIGENCE,
ACTUAL
OR
IMPLIED. The remedies of the Buyer shall be limited to those provided
herein to the exclusion of any and all other remedies including, without
limitation incidental or consequential damages. No agreement varying or
extending the foregoing warranties, remedies or this limitation will be
binding upon Bacharach, Inc. unless in writing, signed by a duly authorized
officer of Bacharach.
To Register Your Warranty, Visit
www.mybacharach.com
Notice:
Product improvements and enhancements are continuous, therefore the specifications and information
contained in this document may change without notice.
Bacharach, Inc. shall not be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this material.
No part of this document may be photocopied, reproduced, or translated to another language without the prior
written consent of Bacharach, Inc.
BACHARACH
®
is a registered trademark of Bacharach, Inc.
All other trademarks are the property of their respective owners.
Copyright © 2010, Bacharach, Inc., all rights reserved.
A
Instruction 0024-9219
PCA
Contents
Contents
1.0
INTRODUCTION
............................................................................
1
-
1
1.1 The Portable Combustion A
nalyzer
........................................... 1-1
1.2 Displayed
Data
........................................................................... 1-2
1.3 PCA Model
Configurations
......................................................... 1-3
2.0 TECHNICAL
CHARACTERISTICS
.............................................
2
-
1
3.0 SETTING UP
THE PCA
..................................................................
3
-
1
3.1 Scope
............................................................................................ 3-1
3.2 PCA
Power
.................................................................................. 3-1
3.2.1 Checking and Replacing
the Batteries
........................... 3-1
3.2.2 Using the Optional
Power Supply
.................................. 3-2
3.3 Connecting
the Probe
................................................................. 3-2
3.4 Configuring the
PCA
.................................................................. 3-4
4.0 OPERATION
.....................................................................................
4
-
1
4.1 Key Pad
Functions
...................................................................... 4-1
4.2 Sampling Hole
Location
............................................................. 4-2
4.3 Performing a
Combustion Test
.................................................. 4-3
4.3.1 Analyzer Turn On
and Warm Up
................................... 4-3
4.3.2 Installing Probe in the Stack
.......................................... 4-4
4.3.3 Starting a
Combustion Test
............................................ 4-5
4.3.4 Ending a
Combustion Test
.............................................. 4-5
4.3.5 Turning Off the Analyzer and Purging the CO Sensor
. 4-6
4.4 Differential Pressure Measurement
.......................................... 4-7
4.5 Warm-up
Screen
......................................................................... 4-8
4.6 Sensor Status
Screen
.................................................................. 4-9
4.7 Combustion Test
Screen
........................................................... 4-10
4.8 Fuel Sel
ection Screen ............................................................... 4-11
4.9 Draft
Screens
............................................................................ 4-12
4.10 Memory Directory
Screen
...................................................... 4-13
4.11 Memory to
PC Screen
............................................................. 4-14
4.12 ID Setup
Screens
.................................................................... 4-17
4.13 Temperature Setup
Screen
.................................................... 4-19
4.14 Draft Unit
Setup S
creen
........................................................ 4-20
4.15 O
2
Reference Setup Screen
..................................................... 4-21
4.16 Language S
etup S
creen
.......................................................... 4-22
4.17 Display M
ode Setup Screen
................................................... 4-23
4.18 Time / Dat
e Setup Scr
een
....................................................... 4-24
4.19 Printer /
Setup Screen ............................................................ 4-25
4.20 Calibration / Maintenance
Password Screen ........................ 4-26
4.21 Maint
enance Screen
............................................................... 4-27
Instruction 0024-9219
i
Contents
PCA
4.22 User Na
me Screens
................................................................ 4-28
4.23 Saving Test
Data
.................................................................... 4-29
4.24 Printing Test
Data
.................................................................. 4-30
4.25 Clear
Memory Screen
............................................................. 4-32
4.26 Resetting the
Microprocessor
................................................. 4-32
5.0 CALIBRATION
..................................................................................
5
-
1
5.1 Sensor
Check
.............................................................................. 5-1
5.2 Calibration
Fixtures
................................................................... 5-2
5.3 Calibrate Menu
Screen
............................................................... 5-3
5.4 Calibrate
TS-Zero ....................................................................... 5-4
5.5 Calibrate
TS-Span
...................................................................... 5-5
5.6 Calibrate
TA-Zero
....................................................................... 5-6
5.7 Calib
rate TA-Span
...................................................................... 5-7
5.8 Calibrate
NX
............................................................................... 5-8
5.9 Calibrate
CO ............................................................................... 5-9
5.10 Calibrate
Draft
........................................................................ 5-11
6.0 MAINTENANCE
................................................................................
6
-
1
6.1 Routine
Maintenance
................................................................. 6-1
6.2 Disassembly
................................................................................ 6-2
6.3 Cleaning the
Probe
..................................................................... 6-4
6.4 Water Trap/Fil
ter Assembly Mai
ntenance
................................ 6-5
6.5 Replacing the
Particulate Filter
................................................ 6-6
6.6 Replacing the
Oxygen Sensor .................................................... 6-7
6.7 Replacing the Nitric
Oxide Sensor
............................................ 6-8
6.7.1 Replacing the Nitric
Oxide Sensor Filter
....................... 6-8
6.7.2 Replacing the Nitric Oxi
de Sensor Bias Battery
........... 6-9
6.8 Replacing the Carbon
Monoxide Sensor
.................................. 6-10
6.8.1 Replacing the Carbon
Monoxide Sensor Filter
............ 6-11
6.9 Replacing the
Pump Assembly
................................................ 6-12
7.0 TROUBLESHOOTING
.....................................................................
7
-
1
7.1 Analyzer
Repair
.......................................................................... 7-1
7.2 Error
Codes
................................................................................. 7-2
7.3 Troubleshooting Guide
............................................................... 7-3
8.0 PARTS & SERVICE
..........................................................................
8
-
1
8.1 Replacement
Parts
...................................................................... 8-1
8.2 Accessories
.................................................................................. 8-2
8.3 Service
Centers
........................................................................... 8-6
APPENDIX A – Display Screen Translations
.....................................
A
-
1
APPENDIX B – Printout Translations
.................................................
B
-
1
ii
Instruction 0024-9219
PCA
1.0 INTRODUCTION
Introduction
1.1 The Portable Combustion Analyzer
The Portable Combustion Analyzer (PCA) (Figure 1-1) is a commercial
grade, hand held, combustion efficiency analyzer that is designed for
continuous
(on demand) sampling of light industrial and residential
furnaces, appliances, and boilers. The basic instrument is supplied with a
probe, instruction manual, batteries, and carrying case.
Figure 1-1. PCA
Instruction 0024-9219
1-1
Introduction
1.2 Displayed Data
PCA
The PCA directly measures, displays, and stores the following data:
Room Temperature in °F or °C (Primary Air/Ambient Temperature)
Flue Gas Oxygen Content in %
Flue Gas Temperature in
°F or °C
Flue Gas Carbon Monoxide Content (H
2
Compensated) in ppm
(For analyzers having a Carbon Monoxide sensor)
Flue Gas Nitric Oxide content in ppm
(For analyzers having a Nitric Oxide sensor)
Pressure/Draft in Millibars, Pascals, or Inches of Water Column
(For analyzers having a draft sensor)
Differential Pressure in Millibars, Pascals, or Inches of Water Column
(For analyzers having a draft sensor)
The PCA will compute, display, and store the following data for any of the
seven standard fuels:
Combustion Efficiency in %
Excess Air in %
Flue Gas Carbon Dioxide Content in %
Flue Gas Carbon Monoxide Content referenced to Oxygen in ppm
(For analyzers having a Carbon Monoxide sensor)
Flue Gas Nitric Oxide Content referenced to Oxygen in ppm
(For analyzers having a Nitric Oxide sensor)
The seven standard types of fuels that can be selected are:
Natural Gas
Oil #2
Oil #4
Oil #6
Kerosene
Propane
Coal
The PCA continuously monitors flue gas exhaust conditions and updates
the above displayed values during a combustion test. If the analyzer is
equipped with an optional pressure sensor, then draft measurements can
be made simultaneously with the combustion test, or made separately.
The analyzer has the capability of storing data that was collected during
a combustion test or draft measurement. The stored data can then at a
later date be either viewed on the PCA’s display, printed using an op-
tional printer, or downloaded to a computer.
1-2
Instruction 0024-9219
PCA
1.3 PCA Model Configurations
Introduction
TABLE 1-1. PCA SENSOR CONFIGURATIONS
PCA Models
Standard Advanced
Sensors Installed
PCA Part No. PCA Part No. Stack Temp., CO NX Draft
Model 24- Model 24- Air Temp. & O
2
(
P)
10 8040 40 8140 X
15 8041 45 8141 X X
20 8042 50 8142 X X
25 8043 55 8143 X X X
30 8044 60 8144 X X X
35 8045 65 8145 X X X X
PCA 10 & 40
These basic instruments have the capability of measuring, displaying,
and storing combustion tests. They will also display flue gas Oxygen
content, flue gas Carbon Dioxide (CO
2
) content, air temperature, flue gas
temperature,
combustion efficiency, excess air
, and the current fuel
selected. The ‘standard’ PCA 10 stores up to 10 combustion tests, while
the ‘advanced’ PCA
40 can store up to 100 tests.
PCA 15 & 45 with Draft
In addition to the features of the basic PCAs described above, these instru-
ments have the added capability of measuring, displaying, and saving draft
or differential pressure in either Millibars, Pascals, or Inches-of-Water
Column.
PCA 20 & 50 with Carbon Monoxide Measurement
In addition to the features of the basic PCAs described above, these
instruments have the added capability of measuring, displaying, and
saving Carbon Monoxide (CO) content, as well as calculating a CO level
that is referenced to Oxygen. The ‘standard’ PCA 20 calculates CO refer-
enced to 0% Oxygen, while the ‘advanced’ PCA
50 calculates CO refer-
enced to a user selected Oxygen level of between 0–15%.
PCA 25 & 55 with Draft and CO Measurement
These instruments combine the features of all the PCAs listed above.
Instruction 0024-9219
1-3
Introduction
PCA
PCA 30 & 60 with CO and Nitric Oxide Measurement
In addition to the features of the basic PCAs with CO measurement, these
instruments have the added capability of measuring, displaying, and
saving Nitric Oxide (NX) content, as well as calculating a NX level that is
referenced to Oxygen. The ‘standard’ PCA 30 calculates NX referenced to
0% Oxygen, while the ‘advanced’ PCA
60 calculates NX referenced to a
user selected Oxygen level of between 0–15%.
PCA 35 & 65 with Draft, CO & NX
These instruments are capable of measuring, displaying, and saving all
measurements as previously described.
Printout Capability
All PCAs have the ability to print the latest test data, or any of the saved
tests, to an optional printer via an infrared link.
CO Sensor Purge
On all PCAs that measure CO, if
the
ON/OFF
k
e
y
is pressed while the CO
reading is 100 ppm or higher, the analyzer will attempt to purge itself of
CO before turning off.
Advanced PCA Model Features
‘Advanced’ models of the PCA contain the following features that are in
addition to the features of their corresponding ‘standard’ units:
100 memory locations
RS232 output for transferring saved data to a personal computer
Ability to enter three lines of user-identification information which is
printed at the top of each printout
Ability to enter three lines of customer-identification information
which is printed with each test record
Ability to set the Oxygen reference level from 0 to 15% for CO and NX
measurements on analyzers equipped with a CO or NX sensor
1-4
Instruction 0024-9219
PCA
Technical Characteristics
2.0 TECHNICAL CHARACTERISTICS
The PCA Directly Measures and Displays:
Oxygen content in flue gas in the range of 0.1 to 20.9 % O
2
Flue gas temperature in the range of 0 to 2192 ºF (–18 to 1200 ºC)
Primary-air / ambient temperature is in the range of 0 to 999 ºF
(–18 to 999 ºC)
Optional . . .
Differential Pressure and Draft in the range of ±28" H
2
O (±70.0 mb)
Carbon Monoxide* content in flue gas in the range of 0 to 4000 ppm
Nitric Oxide* content in flue gas in the range of 0 to 1000 ppm
The PCA Computes and Displays:
When the measured oxygen level is not above 16.0%, and the Stack (Flue
Gas) temperature is not above 1832 °F (1000 ºC)
Efficiency in the range of 0.1 to 99.9%
Excess Air in the range of 1 to 250%
Carbon Dioxide content in flue gas from 0.1 to a fuel dependent
maximum value in percent
Carbon Monoxide* content referenced to a percentage of Oxygen in
the range of 0
to 9,999 ppm on analyzers equipped with a CO sensor.
Nitrix Oxide* content referenced to a percenage of Oxygen in the
range of 0 to 9,999
ppm on analyzers equipped with a NX sensor.
Standard Fuels** Available for
Combustion Calculations:
Natural Gas
Oil #2
Oil #4
Oil #6
Kerosene
Propane
Coal
*
For the PCA 30, 35, 60, & 65, the display can be set up to show
either measured values of Carbon Monoxide and Nitric Oxide (CO
& NX), or show the calculated values of these gases (CF & NF)
referenced to Oxygen. In either case, all values are listed on the
printout of analzyers equipped with a printer.
**
Custom fuels available upon request. Contact factory for details.
Instruction 0024-9219
2-1
Technical Characteristics
PCA
Normal Operating Conditions:
Temperature:
Analyzer
...................
32 to
104 ºF (0 to 40 ºC)
Probe Tip
..................
1472 ºF
(800 ºC) Max.
Humidity:
Analyzer
...................
15 to
90% Relative Humidity, Non-Condensing
Air Pressure:
Analyzer
...................
Atmospheric
Probe .........................
10"
H
2
O (25 mb) draft max at probe
tip
Performance:
Accuracy:
Oxygen*
.................... ±0.3% O
2
Carbon
Monoxide
.....
±5% of reading or ±10
ppm, whichever is greater
between 0 – 2000 ppm, and ±10% of reading
between 2001 – 4000 ppm.
Nitric Oxide
..............
±5% of reading or ±5 ppm, whichever is greater
Flue Gas Temp.
........
±4 ºF between 32 and 255 ºF
(±2 ºC between 0 and 124 ºC)
±6 ºF between 256 and 480 ºF
(±3 ºC between 125 and 249 ºC)
±8 ºF between 481 and 752 ºF
(±4 °C between 250 and 400 °C)
Ambient
Temp.
........
±2 ºF between 32 and 212 ºF
(±1 °C between 0 and 100 °C)
Pressure
...................
±2% of reading or ±0.02 inches of Water Column
(±0.05 mb), whichever is greater
System Flow Rate:
With
probe ................ 200
cc/min minimum
Front Panel
Controls:
Seven embossed pushbutton switches with tactile feedback (refer to
Section 4.3)
Display:
20 character by 4 line alphanumeric LCD panel with a green backlight.
*
Accuracy referenced in practical flue gas concentrations
(mixtures of O
2,
CO
2
and N
2
)
2-2
Instruction 0024-9219
PCA
Technical Characteristics
Power Requirements:
Four AA alkaline batteries. Battery backup for the real-time clock, RAM,
and bias voltage for the Nitric Oxide sensor is provided by internal
lithium
batteries. Optional AC Power Supplies (110 VAC & 230
VAC) are also
available.
Operating Time:
A fresh set of batteries will provide at least 8
hours of continuous opera-
tion with the pump running and the backlight turned on.
Warm Up Time:
60 seconds.
Printer Interface:
Infrared Communications (refer to Section 4.24).
Materials:
High impact ABS plastic case
Polycarbonate window over the display
Nickel plated, brass quick-connect hose fitting
Stainless steel probe
Dimensions:
Height:
.................. 8.5
in. (215 mm)
Width:
...................
3.8 in.
(96 mm) at display (3.0 in. [75 mm] at controls)
Depth:
................... 2
in. (50 mm)
Weight:
With
Batteries:
.....
Approximately 1.5 lbs (0.7 Kg)
Instruction 0024-9219
2-3
Technical Characteristics
NOTES:
PCA
2-4
Instruction 0024-9219
PCA
3.0 SETTING UP THE PCA
3.1 Scope
Before using the PCA, you MUST:
Setup
Check the batteries or plug in an Optional Power Supply (Section 3.2)
Connect the probe to the analyzer (Section
3.3)
Check the analyzer’s configuration (Section 3.4)
3.2 PCA Power
3.2.1 Checking and Replacing the Batteries
A fresh set of batteries is supplied with the PCA. Install the batteries as
described below. If a
LOW BATTERY
message is displayed when the
analyzer is turned on, replace the batteries.
1.
Remove the battery cover from the back of the PCA (Figure 3-1).
2.
Remove (and properly dispose of) any old batteries.
3.
Insert a fresh set of four AA alkaline batteries, making sure to install
them per the “+” and “–” markings shown in the battery compart-
ment.
4.
Replace the battery cover.
Figure 3-1. Battery Replacement
Instruction 0024-9219
3-1
Setup
3.2.2 Using the Optional Power Supply
If an Optional Power Supply is to be used:
1.
Connect the output plug of the Optional Power Supply to the
analyzer’s power supply jack (Figure 3-2).
PCA
2.
Plug the Optional Power Supply into an appropriate AC wall outlet.
The analyzer will now operate and function normally.
3.3 Connecting the Probe
To attach the probe to the analyzer (Figure 3-2):
1.
Push the yellow-banded, quick-connect Flue Gas Hose (giving a slight
twist) onto the analyzer’s GAS sample-inlet fitting.
2.
Push the blue-banded, quick-connect Draft Hose (giving a slight twist)
onto the analyzer’s DRAFT sample-inlet fitting.
3.
Push the Flue Gas Thermocouple into the T-STACK jack (connector
fits in only one way).
NOTE:
The PCA has a built in room-air thermocouple.
Perform Step 4 only if the Optional Room Air/Primary Air
Thermocouple is used.
4.
Push the Optional Room Air/Primary Air Thermocouple into the
T-AIR jack (connector fits in only one way).
NOTE:
In order for the PCA to correctly calculate combus-
tion efficiency when the burner’s primary-air temperature is
not
the same as room temperature, the primary-air tempera-
ture should be measured using the optional Primary Air
Thermocouple.
Inspect all the hoses for cracks. If any hose is defective, replace the entire
probe assembly. Check that the water trap is empty, and the filter is not
dirty or saturated with water.
3-2
Instruction 0024-9219
PCA
Setup
Reset Button
Room Air /
Primary Air
Thermocouple
(Optional)
Power Supply
110V/60Hz
230V/50Hz
(Optional)
Flue Gas
Thermocouple
Draft Hose
*
Pressure Reference Port
*
(Used in the Measurement
of Differential Pressure)
Flue Gas
Hose
*
Present only on analyzers
equipped with a pressure
sensor.
Probe Tube
Water
Trap / Filter
Assembly
Adjustable
Probe Stop
Flue
Gas
Hose
Probe
Handle
Figure 3-2. Connecting the Probe to the Analyzer
Instruction 0024-9219
3-3
Setup
3.4 Configuring the PCA
PCA
The PCA is configured at the factory for the parameters shown below.
These parameters, however, can be changed by following the instructions
in their associated sections.
Function
Parameters
To Change,
Refer to . . .
Fuel
Natural Gas
Section
4.8
Temperature
º
C
Section 4.13
Draft
W
C
Section 4.14
Oxygen Reference
0
%
Section
4.15
Language
English
Section 4.16
Display
Mode
CO & NX
Section
4.17
Time
HR:MIN:SEC AM/PM
Section
4.18
Date
MM/DD/YY
Section 4.18
3-4
Instruction 0024-9219
PCA
4.0 OPERATION
4.1 Key Pad Functions
Operation
Descriptions of the key pad functions are given below. Note that most of
the front panel key pad buttons perform multiple functions as determined
by what screen is being displayed at the time.
Turns the analyzer ON and OFF. Note that there is a 5
second
OFF-delay that allows an operator to turn the instrument back
ON by pressing the
RUN
key. This feature
prevents the accidental
loss of test data if the analyzer is turned OFF by mistake.
Moves the cursor [
z
] in front of a menu item up through the
displayed items. This key also increases alphanumeric values in
screens requiring a value change.
Moves the cursor [
z
] in front of a menu item down through the
displayed items. This key also decreases alphanumeric values in
screens requiring a value change.
Chooses the highlighted item (the item with the cursor [
z
] in front
of it) in all menus and screens. This key also causes the cursor to
enter the number field in the Password Screen, and causes the
cursor to advance to the next field position in screens requiring
multiple alphanumeric entries.
Starts and stops a combustion test when the Combustion Test
Screen is displayed. Pressing this key in any other screen almost
always returns the instrument to the Combustion Test Screen.
However, there are four situations where this key behaves as an
enter key: 1)
After entering a correct password in the Password
Screen, press the RUN
k
e
y
to display the first calibration screen.
2)
After entering an offset or span value in any of the Calibration
Edit Screens, press the
RUN
k
e
y
to store the new values. 3)
After
entering a time or date value in the Time/Date Setup Screen,
press the
RUN
k
e
y
to store the new values and return the cursor to
the left side of the display. 4)
After entering text in either the ID
Setup or User Name Screens, press the
RUN
k
e
y
to store the text.
Advances the display to the next menu screen.
Turns the screen’s backlight ON and OFF.
Instruction 0024-9219
4-1
Operation
4.2 Sampling Hole Location
PCA
The analyzer requires that a ½" diameter sampling hole be made in the
furnace stack to accommodate the probe stop on the Probe and Hose
Assembly.
Locate the sampling hole downstream from the last heat exchanger,
and upstream from any source of dilution, such as a draft diverter
(Figure 4-1)
.
Important!
As the distance between the last heat exchanger
and sampling point increases, stack loss will falsely decrease
due to heat loss by convection from the flue or stack.
For residential and light-commercial combustion-equipment applications,
the following recommendations are applicable:
Oil Gun Burners
– Locate sampling hole at least 12 in. (30 cm) down-
stream from the furnace breaching, and at least 6 in. (15 cm) upstream
from the furnace side of the draft regulator.
Gas Burners
– Locate sampling hole
at least 6 in. (15 cm) upstream from
the furnace side of the draft diverter
on gas-converted units. For gas-
designed equipment, the probe may
be inserted down into the flue
through the draft diverter or hood.
FLUE GAS FLOW
(Downstream)
DRAFT
DIVERTER/
REGULATOR
SAMPLE
POINT
(Upstream)
FURNACE
BREECHING
Figure 4-1. Sampling Hole Location
4-2
Instruction 0024-9219
PCA
4.3 Performing a Combustion Test
Operation
Important!
Large rapid changes in the temperature of the
analyzer can affect its accuracy. This is important to know if
the analyzer is stored in a cold place (such as an unheated
vehicle in the winter) and then taken into a warm furnace
area. For the most accurate test results, allow the analyzer to
warm up to room temperature before use (about 10 minutes).
4.3.1 Analyzer Turn On and Warm Up
Important!
Be sure the probe is at room temperature before
performing the following steps.
1.
Make sure that the analyzer is properly set up per Section
3.0.
2.
Place probe in an area of fresh, ambient air; then press the analyzer’s
ON/OFF
key.
3.
Wait for the analyzer to countdown through its 60 second warm-up
period; then perform one of the following:
If no errors were detected during warm-up,
the Combustion
Test Screen will be displayed. Skip Step 4, and go to Section
4.3.2.
If an error was detected during warm-up,
proceed with Step
4.
4.
If one or more errors were detected by the analyzer’s microprocessor
during warm-up, these errors will be displayed at the bottom of the
Sensor Status Screen (Section 4.6). Address any problems now per
Section 7.2; then repeat this procedure starting with Step 1.
NOTE:
If the error detected is not critical to your test, the
instrument can still perform any test not using the function
disabled by the error.
Instruction 0024-9219
4-3
Operation
4.3.2 Installing Probe in the Stack
PCA
1.
After making a sampling hole in the stack (Section 4.2), and turning
on the analyzer (Section 4.3.1), screw the probe stop supplied with the
Probe and Hose Assembly into the sampling hole (Figure
4-2).
2.
Insert the probe through the hole in the probe stop, then position the
probe tip inside the stack, near its center. Tighten the thumbscrew on
the probe stop to secure the probe.
PROBE STOP
THUMBSCREW
Figure 4-2. Installing the Probe
4-4
Instruction 0024-9219
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91