  • Hello! I am an AI chatbot trained to assist you with the Benchmark Scientific B1000-RA Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Replacement of the Heating Element
1. Turn off BactiZapper unit and allow it to cool completely. Unplug the unit from the wall.
2. Remove the two screws holding the heating element to the base of the unit (Screws A):
3. Pull the unit off of the base gently, pulling up the electrical wires with it.
4. Unplug the head units two electrical wires from the base.
5. Plug the replacement elements two wires to those of the base and push the connected wires
into the body of the BactiZapper until the replacement heating element is properly aligned with
the body of the unit.
6. Replace the two screws removed in Step 2.
/