Beurer Oxymètre de pouls PO40 Owner's manual

Category
Heart rate monitors
Type
Owner's manual

This manual is also suitable for

Storage
-20°C
55°C
RH ≤93%
Permissible storage temperature and humidity
IP22
Device protected against foreign objects ≥ 12.5 mm and against
falling drops of water
Operating
10°C
40°C
RH 15-93%
Permissible operating temperature and humidity
Disposal in accordance with the Waste Electrical and Electronic
Equipment EC Directive – WEEE
SN
Serial number
5. Warnings and safety notes
Non-observance of the following information may result in personal injury or material damage. Store these instructions for use and make them accessible to other users.
Make sure you include these instructions for use when handing over the device to third parties.
WARNING
Check to ensure that the package contains all the parts that should be included in the delivery.
Check the pulse oximeter regularly before use to ensure that there is no visible damage to the device and the batteries are still suciently charged. In case of doubt, do
not use the device and contact Beurer customer services or an authorised retailer.
Do not use any additional parts that are not recommended by the manufacturer or oered as equipment.
Under no circumstances should you open or repair the device yourself, as faultless functionality could no longer be guaranteed thereafter. Failure to comply will result in
voiding of the warranty. For repairs, please contact Beurer customer services or an authorised retailer.
D
o NOT use the pulse oximeter
if you are allergic to rubber products.
if the device or the finger you are using is damp.
on small children or babies.
during an MRI or CT scan.
while transporting a patient other than within a medical establishment.
whilst taking a blood pressure measurement on the same arm using a cu.
on fingers that have nail varnish on, are dirty or have a plaster or other dressing on them.
on large fingers that do not fit into the device easily (fingertip: width approx. > 20 mm, thickness approx. >15 mm).
on fingers with anatomical changes, oedemas, scars or burns.
on fingers that are too small, as with small children for example (width approx. < 10 mm, thickness < 5 mm).
on patients who are not steady at the site of application (e.g. trembling).
near flammable or explosive gas mixtures.
Using the device for long periods may cause pain for people with circulatory disorders. Therefore do not use the pulse oximeter for longer than 30 minutes on one
finger.
This is essential to ensure correct sensor orientation and to safeguard the integrity of the skin.
The pulse oximeter displays an instantaneous measurement but cannot be used for continuous monitoring.
The pulse oximeter does not have an alarm function and is therefore not suitable for evaluating medical results.
Do not self-diagnose or self-medicate on the basis of the measurements without consulting your doctor. In particular, do not start taking any new medication or change
the type and/or dosage of any existing medication without prior approval.
Do not look directly inside the housing during the measurement. The red light and the invisible infra-red light in the pulse oximeter are harmful to your eyes.
This device is not intended for use by people (including children) with restricted physical, sensory or mental skills or a lack of experience and/or a lack of knowledge,
unless they are supervised by a person who has responsibility for their safety or they receive instructions from this person on how to use the device. Children should be
supervised around the device to ensure they do not play with it.
The displays for the pulse wave and pulse bar allows the strength of the pulse or circulation to be evaluated at the measurement site. Rather, they are exclusively
used to display the current visual signal variation at the measurement site and do not enable reliable diagnostics for the pulse.
N
on-observance of the following instructions can lead to incorrect or failed measurements:
There must not be any nail varnish, artificial nails or other cosmetics on the finger to be measured.
Ensure that the finger nail on the finger to be measured is short enough that the fingertip covers the sensor element in the housing.
If the person moves while the measurement is being taken. Keep your hand, finger and body steady during the measurement.
For people with cardiac arrhythmia, the oxygen saturation level (SpO₂) readings and the heart rate (PRbpm) may be incorrect or the measurement may not be possible at all.
If an electronic surgical device or defibrillator is used, the functioning of the pulse oximeter may be impaired.
In cases of carbon monoxide poisoning, the pulse oximeter displays a measurement value that is too high.
To avoid falsifying the measuring result, there should not be any strong light sources (e.g. fluorescent lamps or direct sunlight) in the immediate vicinity of the pulse oximeter.
People with low blood pressure, who suer from jaundice or take medication for vascular contraction may experience incorrect or falsified measurements.
Incorrect measurements are likely for patients who have been administered medical dye in the past or for those who have abnormal haemoglobin levels. This applies in
particular for cases of carbon monoxide poisoning and methaemoglobin poisoning, which can occur for example from the administration of local anaesthetics or from an
existing methaemoglobin reductase deficiency.
The measurement may be falsified in patients with an arterial catheter, hypotension, severe vascular constriction, anaemia or hypothermia.
Protect the pulse oximeter from dust, shocks, moisture, extreme temperatures and explosive materials.
6. Unit description
Device
Finger opening
Lanyard holder
Function button
Display
1. Oxygen saturation (value in
percent)
2. Pulse rate (value in beats per
minute)
3. Pulse bar
4
. Pulse modulation index (value
in percent)
5. Pulse wave (plethysmographic
wave)
Display formats (7 dierent formats)
7.
Initial use
7.1 Inserting the batteries
1. Slide the battery
compartment lid
open.
2. Insert the two sup-
plied batteries into
the pulse oximeter
battery compart
-
ment following the
correct polarity as
shown.
3. Close the battery
compartment
cover again.
7.2 Attaching the lanyard
To transport the pulse oximeter more easily (e.g. whilst on the move) you can attach a lanyard to the device.
1. Insert the narrow end of the
lanyard through the holder as
shown.
2. Draw the other end of the lanyard through the
loop at the narrow end and tighten.
8. Operation
1. Insert one finger
into the finger
opening of the
pulse oximeter
as shown and
hold it steady.
98
65
%SpO
2
PRbpm
2. Press the function
button. The pulse
oximeter begins
its measurement.
Do not move
during the meas
-
urement.
98
65
%SpO
2
PRbpm
3. Your measurement
values will appear on
the screen after a few
seconds.
Note
• If the ?, symbol appears on the display this indicates that the measurement signal is unstable, and the readings shown are invalid.
• When you remove your finger from the pulse oximeter, the device will automatically switch o after approx. 8 seconds.
• To select your desired display format, hold down the function button briefly during operation.
• To select your desired display brightness, hold down the function button for slightly longer during operation.
9. Evaluating measurement results
WARNING
The following table for evaluating your measurements does NOT apply to
people with certain pre-existing conditions (e.g. asthma, heart failure, res
-
piratory diseases) or whilst staying at altitudes above 1500 metres. If you
have a pre-existing condition, always consult your doctor to evaluate your
measurements.
Decline in oxygen saturation depending on altitude
Note
The following table informs you of the eects of various altitudes on oxygen saturation
value and its impact on the human body. The following table does NOT apply to peo
-
ple with certain pre-existing conditions (e.g. asthma, heart failure, respiratory diseases
etc.). People with pre-existing conditions can show signs of illness (e.g. hypoxia) at
lower altitudes.
SpO₂ (oxygen saturation)
measurement in %
Classification/measures to be taken Altitude Expected SpO₂ value (oxy-
gen saturation) in %
Impact on human body
99-94 Normal range 1500-2500 m > 90
No altitude sickness (normally)
94-90 Decreased range:
Visit to the doctor recommended
2500-3500 m ~90 Altitude sickness, acclimatisa
-
tion recommended
< 90 Critical range:
Seek medical attention urgently
3500-5800 m <90 Very frequent altitude sickness,
acclimatisation absolutely
essential
Evaluating pulse modulation index
The pulse modulation index (PMI) may lie between 0.3% and 20%, and
varies depending on the patient, measurement location and state of health.
A very low PMI value can impair the measurement.
5800-7500 m <80
Severe hypoxia, only limited
length of stay possible
7500-8850 m <70
Immediate, acute danger to life
Source: Hackett PH, Roach RC: High-Altitude Medicine. In: Auerbach PS
(ed): Wilderness Medicine, 3rd edition; Mosby, St.Louis, MO 1995; 1-37.
10. Maintenance/cleaning
IMPORTANT:
Do not use high pressure or ethylene oxide sterilisation on the pulse oximeter! The device is not suitable for sterilisation.
Under no circumstances should you hold the pulse oximeter under water, as this can cause liquid to enter and damage the pulse oximeter.
Clean the housing and the interior rubber surface with a soft cloth dampened with medical alcohol after each use.
If a low battery status appears on the display of the pulse oximeter, change the batteries.
If you are not going to use the pulse oximeter for more than one month, remove both batteries from the device to avoid possible leaking.
11. Storage
IMPORTANT:
Store the pulse oximeter in a dry place (relative humidity ≤93 %). If the humidity is too high it may shorten the service life of the pulse oximeter or damage it. Store the pulse
oximeter in a place where the ambient temperature is between -20°C and 55°C.
12. Disposal
Please dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and Electronic Equipment). If you have any questions, please contact the local
authorities responsible for waste disposal.
The empty, completely flat batteries should be disposed of through specially designated collection boxes,
recycling points or electronics retailers. You are legally required to dispose of the batteries.
Note: The codes below are printed on batteries containing harmful substances: Pb = Battery contains lead, Cd = Battery contains cadmium,
Hg = Battery contains mercury.
13. What if there are problems?
Problem Possible cause Solution
"
Finger out
" appears on the display The finger on which the measurement is being taken has not
been inserted properly in the pulse oximeter
Insert the finger in the pulse oximeter again
Measurement values are not correctly displayed The measured SpO₂ is too low (<70%) Do the measurement again. If the problem occurs
repeatedly and the device is functioning properly,
seek medical advice as a matter of urgency
There is a strong light source (e.g. fluorescent lamp or direct
sunlight) in the vicinity
Remove pulse oximeter from the vicinity of these
light sources
The pulse oximeter is displaying measurement
interruptions or high measurement value jumps
Insucient circulation in the measurement finger Observe the warnings and safety notes in section 5
Measurement finger is too large or too small Fingertip must have the following measurements:
Width between 10 and 20 mm
Thickness between 5 and 15 mm
Finger, hand or body is moving
Keep your finger, hand and body still during the
measurement.
Cardiac arrhythmia Seek medical attention
2
1
3
5
4
Problem Possible cause Solution
Pulse oximeter will not switch on. Batteries are flat Replace the batteries
The batteries have not been inserted correctly Reinsert the batteries
The pulse oximeter is faulty. Contact the retailer or Customer Services
Indicator light goes out suddenly The pulse oximeter switches o automatically after 8 seconds
if it is not receiving a signal
Switch the pulse oximeter on again using the ON/
OFF button.
Batteries are flat Replace the batteries
"
Error 3
" appears on the display The red light receiving LED is faulty Contact the retailer or Customer Services
"
Error 4
" appears on the display The infrared light receiving LED is faulty Contact the retailer or Customer Services
"
Error 6
" appears on the display The display is faulty. Contact the retailer or Customer Services
"
Error 7
" appears on the display The receiving LEDs are faulty Contact the retailer or Customer Services
14. Technical data
Model no. PO 40
Measurement method Non-invasive measurement of arterial oxygen saturation of haemoglobin, pulse rate and pulse modulation index in finger.
Measurement range SpO₂ (oxygen saturation): 70 - 100%,
pulse: 30 - 250 beats/minute
PMI: 0.3 - 20 %
Accuracy SpO₂ (oxygen saturation): 70 - 100%, ± 2%,
pulse: 30 - 250 bpm, ± 2 beats/minute
PMI: 0.3% - 1%; ±0.2 digits; >1.1% ± 20%
Dimensions L 58.4 mm x W 33.5 mm x H 37 mm
Weight Approx. 57 g (including batteries)
Sensor to measure SpO₂ Red light (wave length 660 nm); infra-red (wave length 905 nm); silicon receiver diode
Permissible operating conditions +5 °C to +40 °C, ≤15 - 93 % relative humidity,
86 - 106 kPa
ambient pressure
Permissible storage conditions
-20 °C to +55 °C, ≤93 % relative humidity,
86 - 106 kPa
ambient pressure
Power supply
2 x 1.5V
AAA batteries
Battery life 2 AAA alkaline batteries last for approx. 2 years of operation at 1 measurements per day (each of 60 seconds).
Classification IP22, application part, type BF
Technical information is subject to change without notification to allow for updates.
This device complies with European Standard EN60601-1-2 and is subject to particular precautions with regard to electromagnetic compatibility. Please note that porta-
ble and mobile HF communication systems may interfere with this unit. More details can be requested from the stated Customer Services address or found at the end of
the instructions for use.
This device complies with EU Directive 93/42/EC concerning medical devices, the Medizinproduktegesetz (German Medical Devices Act) and the DIN EN ISO 80601-2-61
standard (Medical electrical equipment – Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use).
751.815 - 0515 Irrtum und Änderungen vorbehalten
Guidance and manufacture’s declaration-electromagnetic emission
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration –electromagnetic emission
The PO40 Pulse Oximeter is tended for use in the electromagnetic environment specified below. The customer of the user of the PO40 Pulse Oximeter should assure
that it issued in such an environment.
Emission test compliance Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1 The PO40 Pulse Oximeter uses RF energy only for their internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B The PO40 Pulse Oximeter is suitable for use in all establishments, including domestic estab
-
lishments and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emission
IEC 61000-3-3
Not applicable
Guidance and manufacture’s declaration-electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration-electromagnetic immunity
The PO40 Pulse Oximeter is intended for use in the electromagnetic environment specified specified below. The user of PO40 Pulse Oximeter should assure that it is
used in such an environment.
Immunity test IEC60601 test level Compliance level Electromagnetic environment-guidance
Electrostatic discharge (ESD) IEC 61000-4-2 ±6KV contact
±8KV air
±6KV contact
±8KV air
Floors should be wood, concrete or ceramic
tile. If floor are covered with synthetic
material, the relative humidity should be at
least 30%.
Power frequency (50Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment
Guidance and manufacture’s declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration-electromagnetic immunity
The PO40 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of PO40 Pulse Oximeter should assure that
it is used in such an environment.
Immunity test IEC60601 test level Compliance
level
Electromagnetic environment -guidance
Radiated RF
ICE 61000-4-3
3V/m
80MHz to 2.5GHz 3V/m
Portable and mobile RF communication equipment should be used no closer to any part of the
PO40 Pulse Oximeter, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
recommended separation distance
Where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
a
should be less than the compliance level in each frequency range
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures,
objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broad-
cast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which The PO40 Pulse Oximeter is used exceeds the applicable RF compliance level
above, the PO40 Pulse Oximeter should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the PO40 Pulse Oximeter.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM
for EQUIPMENT or SYSTEM that not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the PO40 Pulse Oximeter
The PO40 Pulse Oximeter is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the
PO40 Pulse Oximeter can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the PO40 Pulse Oximeter as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150KHz to 80MHz 80MHz to 800MHz 800MHz to 2.5GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can be estimated using the equation appli
-
cable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.