IQS-1500
SPECIFICATIONS
1
GENERAL SPECIFICATIONS
Size (H × W × D) 12.5 cm × 3.6 cm × 28.2 cm (4
15
/16 in × 1
7
/16 in × 11
1
/8 in)
Weight 0.63 kg (1.4 lb)
Temperature operating 23 °C ± 5 °C (73 °F ± 9 °F)
storage —40 °C to 70 °C (—40 °F to 158 °F)
Relative humidity 0 % to 80 % non-condensing
Model IQS-1502-Q0 IQS-1502-Q1
Detector type Ge Ge
Detector size (mm) 5 5
Power range
1
(dBm) +5 to —60 +5 to —60
Wavelength range (nm) 750 to 1800 750 to 1800
Linearity
1,2
(dB) ± 0.01 ± 0.01
(+5 dBm to —50 dBm) (+5 dBm to —50 dBm)
Uncertainty
6
±2 %
3
≤ ±0.9 %
4
at 3 λ
(at 1310.0 ± 0.1 nm and 1550.0 ± 0.1 nm) (—10 dBm CW)
(0 dBm to —10 dBm CW)
±3 %
3
±3 %
4
(750 nm to 1700 nm) (750 nm to 1700 nm)
(—30 dBm to —40 dBm CW) (—30 dBm to —40 dBm CW)
Power resolution (dB) 0.001 0.001
Wavelength resolution (nm) 0.1 0.1
Applicable fiber type (µm) 9/125 (B); 50/125 (C); 62.5/125 (D) 9/125 (B); 50/125 (C); 62.5/125 (D)
Connector type
5
B: 89 B: 58; 88; 89; 90; 91
C/D: 50; 54; 74
EXFO America 4275 Kellway Circle, Suite 122 Addison, TX 75001 USA Tel.: 1 800 663-3936 Fax: 1 972 836-0164
EXFO Europe Le Dynasteur, 10/12 rue Andras Beck 92366 Meudon la Forêt Cedex FRANCE Tel.: +33.1.40.83.85.85 Fax: +33.1.40.83.04.42
EXFO Asia-Pacific 151 Chin Swee Road, #03-29 Manhattan House SINGAPORE 169876 Tel.: +65 6333 8241 Fax: +65 6333 8242
EXFO China Beijing New Century Hotel Office Tower, Room 1754-1755 Beijing 100044 P. R. CHINA Tel.: +86 (10) 6849 2738 Fax: +86 (10) 6849 2662
No. 6 Southern Capital Gym Road
Corporate Headquarters > 400 Godin Avenue, Vanier (Quebec) G1M 2K2 CANADA
Toll-free: 1 800 663-3936 (USA and Canada) www.exfo.com
SPIQS1500.2AN © 2004 EXFO Electro-Optical Engineering Inc. All rights reserved. Printed in Canada 04/11
EXFO is certified ISO 9001 and attests to the quality of these products. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. EXFO
has made every effort to ensure that the information contained in this specification sheet is accurate. However, we accept no responsibility for any errors or omissions, and we
reserve the right to modify design, characteristics and products at any time without obligation. Units of measurement in this document conform to SIstandards and practices.
Contact EXFO for prices and availability or to obtain the phone number of your local EXFO distributor.
For the most recent version of this spec sheet, please go to the EXFO website at http://www.exfo.com/specs
In case of discrepancy, the Web version takes precedence over any printed literature.
IQS-150X-XX-X-XX
ORDERING INFORMATION
Model
IQS-1502-Q0-B = 5 mm TEC-Ge detector, 2 % uncertainty at (1310 ± 0.1) nm and (1550 ± 0.1) nm, Singlemode 9/125 µm
7
IQS-1502-Q1-B = 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Singlemode 9/125 µm
IQS-1502-Q1-C = 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Multimode 50/125 µm
IQS-1502-Q1-D = 5 mm TEC-Ge detector, 0.9 % uncertainty calibrated at NIST, Multimode 62.5/125 µm
Test jumper connector
50 = FC/PC
9
54 = SC/PC
9
58 = FC/APC narrow key
8
Example: IQS-1502-Q1-B-89
74 = ST/PC
9
88 = SC/APC
8
89 = FC/UPC narrow key
8
90 = ST/UPC
8
91 = SC/UPC
8
STANDARD ACCESSORIES
User guide, fiber-optic adapter
(FOA), one reference test jumper
and Certificate of Calibration
Notes
1. At 1310 nm and 1550 nm.
2. At a ± 1 °C constant temperature.
3. Q0 option, uncertainty at EXFO
reference conditions:
• ± 2 % uncertainty at (1310.0 ± 0.1) nm
and (1550.0 ± 0.1) nm, with 9/125 µm
(B) fiber, and source spectral width
(FWHM) < 10 nm.
• ± 3 % uncertainty from 750 nm to
1700 nm, with 62.5/125 µm (D) fiber,
and source spectral width (FWHM)
≤ 12 nm.
• FC/UPC connector (ceramic ferrule)
with FOA-322.
4. Q1 option, calibration at NIST
a) • ≤ ± 0.9 % uncertainty at three
user-specified wavelengths within
the following ranges at NIST
reference conditions:
786 nm ± 0.5 nm
(840 to 860) nm ± 0.25 nm
(1280 to 1330) nm ± 0.25 nm
(1520 to 1625) nm ± 0.25 nm
• 9/125 to 62.5/125 µm (B to D) fiber.
• FC, ST or SC connector.
• Fiber-optic adapter (FOA) used
at NIST.
• Source spectral width (FWHM)
< 10 nm.
b) ± 3 % uncertainty from 750 nm to
1700 nm, 62.5/125 µm (D) fiber,
source spectral width (FWHM)
< 12 nm, and FC/UPC connector
(ceramic ferrule) with FOA-322.
5. See Ordering Information for details.
6. All uncertainties are valid on the
day of calibration after a warm-up
time of 20 minutes, and specified
with a 95 % confidence level.
7. Test jumper FC/UPC only
8. Singlemode only
9. Multimode only