PSM1735 user manual
2-4
2.3 Basic operation
Once the unit has powered on and is displaying the default
RMS voltmeter screen, the required function can be
selected by pressing the function key, FRA, PAV, LCR,
RMS, POWER, or HARM. Pressing the same function key
again brings up the menu options specific to that function.
For example, after the unit has powered on in RMS mode,
press PAV to select phase angle voltmeter mode; press
PAV again to gain access to the PAV specific menu options.
Alternatively, press MODE to bring up the function specific
menus.
There are a number of menus available to configure the
instrument. All of these menus are directly accessible with
one press of the appropriate menu key. The cursor keys
allow navigation around the menus and also control the
generator when not accessing a menu.
For example, to turn on the output, press OUT to bring up
the generator control menu, press UP to select the output
on/off option, press RIGHT to turn on the output. Press
ENTER, or HOME twice, to exit the menu. Press FRA to
select frequency response analyser mode and the
instrument will now display the gain and phase of the
transfer function of the circuit under test at the spot
frequency specified by the output control menu.
Press LEFT or RIGHT to adjust the frequency, Press UP or
DOWN to adjust the amplitude (use the OUT menu to
change the size of the steps).
Press START and the PSM1735 will start a frequency
sweep over the specified range (set in the SWEEP menu);
press TABLE to see the table of results, GRAPH to see a
graph of the results and REAL TIME to return to the real
time data.