We reserve the right to technical modifications without prior notice.
GE Inspection Technologies
Xray
©2004 General Electric Company. All rights reserved.
GEIT-30131GB (5/04)
GEInspectionTechnologies.com
The cooling output depends on the ambi-
ent temperature, on the temperature of
the return-flow cooling water, and also on
the flow rate.
In practice, the cooling capacity of the
WL 3001 at up to 25°C ambient tempera-
ture is sufficient for continuous operation
(100% duty cycle) of X-ray tubehousings
(up to 3000 W).
The water cooling pump WL 3001 con-
sists of the following components:
1 storage tank
1 radial ventilators
1 thermostat
1 protective housing
1 radiators
1 connecting cable, 10 m (32 ft) long
with plug-in connection
1 electro motor and pump unit
Description
Technical Data
Dimensions and Weights
Length 479 mm (18.85 inch)
Width 398 mm (15.67 inch)
Height 481 mm (18.94 inch)
Weight Approximately 38.2 kg (62.05 lbs) without water
Tank capacity Approximately 3.7 l
Output Data
Cooling output see power diagram
Circulation rate 6.0 l/min at 6.0 bar
Mains voltage 230 V +/- 10%, 50/60 Hz
Power input 2.5 A
Noise level 55 dB (A) at 50 Hz, measured at a 1 m distance
56 dB (A) at 60 Hz, measured at a 1 m distance
Environmental Conditions
Operating temperature range +5°C … +40°C
Storage temperature range -25°C … +70°C
(storage without water)
Relative air humidity range 20% … 80%
Settings
Maximum pressure 8.0 bar
Water temperature 35°C maximum
Difference between water
temperature and ambient
temberature [°C]
Cooling power [W]