SPECIFICATIONS
Salinity—
Range: 0.00 to 10.00 ppt (NaCl)
Resolution: 0.01
Range: 10.1 to 70.0 ppt (NaCl)
Resolution: 0.01
Automatic Temperature Compensation—
Range: 32.0 to 122.0 °F / 0.0 to 50.0 °C
Normalization: 25°C(xed)
Temperature Measurement—
Range: 32.0 to 122.0 °F / 0.0 to 50.0 °C
Resolution: 0.1°
OPERATION
Thissectioncontainsbasicoperatinginstructions,to
insureaccurateresultsandformoreadvancedfunctions,
refertotheadditionalsectionsthatfollow.
1.Removetheprotectivecoverfromtheprobehead.
2.Turnthemeteronbypressingthe
ON/OFF
SET
button.
3.Placetheprobeheadintothesolutiontobemeasured.
4.ToinsureAutomaticTemperatureCompensation,
immersetheprobeheadandtemperaturesensor
completely.ForaccurateAutomaticTemperature
Compensation,thetemperaturesensormustreach
equilibriumwiththeliquidbeingmeasured,thiscould
takeseveralminutes.
5.Insurethatthereisadequatemovementoftheliquid
aroundthetipoftheprobehead(0.2-0.3metersper
secondminimum).Thiscanbeachievedbystirringthe
meterintheliquid.
Themeterwillautomaticallyselecttherangethatis
themostappropriateforthevaluebeingmeasured.To
manuallyselecttherange,pressandholdthe
Ç
button
forapproximatelythree(3)seconds(seethe“Range
ModeSelection”section).
6.Takethedesiredmeasurements.
7.Pressthe
ON/OFF
SET
buttontoturnthemeteroff.
Topreservebatterylife,alwaysturnthemeteroffwhen
notinuse.
8.Rinsetheprobeheadwithdistilled/deionizedwaterand
replacetheprotectivecover.Alwaysreplacethe
protectivecovertopreventdamagetotheprobehead.
Ifnobuttonispressedfortwenty(20)minutes,themeter
willturnoffautomaticallytopreservebatterylife(seethe
“AutomaticShutoff”section).
POWER ON DISPLAY SEQUENCE
Whenthemeteristurnedon(bypressingthe
ON/OFF
SET
button),thedisplaywillcyclethroughshowingthe
followingitems:
AllSegments-Displaycheck
tnr Normalizationtemperaturevalue(seethe
“SelectingTemperatureNormalizationValue”
section)
ran Rangedisplaymode(seethe“RangeMode
Selection”section).
SALINITY CALIBRATION
Onecalibrationpointmaybeenteredforeachof
themeasurementranges(0.00to10.00pptand
10.1 to 70.0 ppt).
Toinsureaccuracy:
–Selectaknownstandardwithavalueascloseas
possibletothevalueoftheunknownsample.
–Ifpossible,theknownstandardandunknown
sampleshouldbeatthesametemperature.Theideal
temperatureis25°C.
Previouscalibrationpointsareclearedwhenthemeteris
calibratedtoanewvalue.
1.Removetheprotectivecoverfromtheprobehead.
2.Turnthemeteronbypressingthe
ON/OFF
SET
button.
3.Placetheprobeheadintotheknownstandard.
4.ToinsureAutomaticTemperatureCompensation,
immersetheprobeheadandtemperaturesensor
completely.ForaccurateAutomaticTemperature
Compensation,thetemperaturesensormustreach
equilibriumwiththeliquidbeingmeasured,thiscould
takeseveralminutes.
5.Insurethatthereisadequatemovementoftheliquid
aroundthetipoftheprobehead(0.2-0.3metersper
secondminimum).Thiscanbeachievedbystirringthe
meterintheliquid.
6.Pressandholdthe
È
HLD/CAL
buttonforapproximately
three(3)seconds,CALwillappearatthetopofthe
displayandtheconductivitymeasurementwillash.
7.Pressthe
Ç
and/or
È
HLD/CAL
buttonstoadjustthe
displaytotheknowvalueoftheknownstandard.
The meter may only be adjusted by ±30%
of the detected value.
8.Withthetheknowvalueoftheknownstandard
appearingandCALnolongerashing,pressthe
ON/OFF
SET
button,themeterwillreturntothe
measurementdisplay.
VIEWING CALIBRATION DATA
1.Turnthemeteronbypressingthe
ON/OFF
SET
button.
2.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,P1.0willappearatthebottomof
thedisplay.
3.Pressthe
Ç
buttontwo(2)times,P4.0will
appearatthebottomofthedisplay.
4.Pressthe
ON/OFF
SET
button,P4.1willappearatthe
bottomofthedisplay.Thetopofthedisplaywillshow
thevaluethatthemeterwascalibratedtoforthe
rangeof0.00to10.00ppt.(Ifnousercalibrationhas
beenperformed,“- - - ”willappear.)
5.Pressthe
Ç
button,P4.2willappearatthe
bottomofthedisplay.Thetopofthedisplaywill
showthevaluethatthemeterwascalibratedtoforthe
rangeof10.1to70.0ppt.(Ifnousercalibrationhas
beenperformed,“- - - ”willappear.)
6.Pressthe
ON/OFF
SET
button,P4.0 willappearatthe
bottomofthedisplay.
7.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,themeterwillreturntothe
measurementdisplay.
SELECTING °F OR °C
1.Turnthemeteronbypressingthe
ON/OFF
SET
button.
2.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,P1.0willappearatthebottomof
thedisplay.
3.Pressthe
ON/OFF
SET
button,thetemperatureunitof
measurewillashonthedisplay.
4.Pressthe
Ç
button,thetemperaturereadingwill
switchtotheoppositeunitofmeasure.Ifthemeter
hadbeendisplaying°F,itwillnowdisplay°C,and
vice-versa.Eachpressofthe
Ç
buttonwillswitch
thedisplaytotheoppositeunitofmeasure.
5.Withthedesiredunitofmeasureappearing,pressthe
ON/OFF
SET
buttonthree(3)times,P1.0 willappearatthe
bottomofthedisplay.
6.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,themeterwillreturntothe
measurementdisplay.
Theunitofmeasureselectedwillbethedefaultunitof
measureuntilchanged.
RANGE MODE SELECTION
Themeterwillautomaticallyselecttherangethatis
themostappropriateforthevaluebeingmeasured. To
manuallyselectthedisplayrange,pressandholdthe
Ç
buttonforapproximatelythree(3)seconds.
Eachtimethe
Ç
buttonforpressedandheld,therange
modewillcyclethroughtheavailablemodes:
ATO Automaticrangeselection
1 0.00 to 10.00 ppt
2 10.1 to 70.0 ppt
HOLD FUNCTION
Theholdfunctionallowsthereadingsonthedisplaytobe
“frozen”sothattheymayberecorded.
1.Pressandreleasethe
È
HLD/CAL
buttonto“freeze”
thedisplayatthecurrentreading.“HOLD”will
appearatthetopofthedisplay.
2.Oncethereadinghasbeenrecorded,pressand
releasethe
È
HLD/CAL
buttontoreturnthedisplayto
thecurrentreading.“HOLD”willnolonger
appearonthedisplay.
AUTOMATIC SHUTOFF (Disable)
Ifnobuttonhasbeenpressedfortwenty(20)minutes,
theAutomaticShutofffeaturewillturnthemeteroffto
preservebatterylife.
TodisabletheAutomaticShutofffeature:
1. Withthemeterturnedoff,pressandholdthe
È
HLD/CAL
button,thenpressandholdthe
ON/OFF
SET
buttonuntil
“n ”appearsonthedisplay.
2. Releasethe
È
HLD/CAL
and
ON/OFF
SET
buttons.Themeter
willpoweronandtheAutomaticShutofffeaturewill
bedisabled.
TheAutomaticShutofffeatureisautomaticallyenabled
whenthemeterispoweredoffandturnedbackon.
RESET TO FACTORY DEFAULTS
Resettingthemeterwillclearallcalibrationdataand
resetthetemperaturedisplayto°C.
Thefollowingprocedureisusedtoresetthemetertothe
factorydefaultvalues.
1.Turnthemeteronbypressingthe
ON/OFF
SET
button.
2.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,P1.0willappearatthebottomof
thedisplay.
3.Pressthe
Ç
buttonone(1)time,P3.0will
appearatthebottomofthedisplay.
4.Pressthe
ON/OFF
SET
button,the
n
willashonthedisplay.
5.Pressthe
Ç
button,the
y
willashonthedisplay.
6. With
y
appearing,pressthe
ON/OFF
SET
button,P3.0
willappearatthebottomofthedisplayandthevalues
willhavebeenresettothefactorydefaults.
7.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,themeterwillreturntothe
measurementdisplay.