Traceable 4367 Owner's manual

  • Hello! I am an AI chatbot trained to assist you with the Traceable 4367 Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
SPECIFICATIONS
Salinity—
Range: 0.00 to 10.00 ppt (NaCl)
Resolution: 0.01
Range: 10.1 to 70.0 ppt (NaCl)
Resolution: 0.01
Automatic Temperature Compensation—
Range: 32.0 to 122.0 °F / 0.0 to 50.0 °C
Normalization: 25°C(xed)
Temperature Measurement—
Range: 32.0 to 122.0 °F / 0.0 to 50.0 °C
Resolution: 0.1°
OPERATION
Thissectioncontainsbasicoperatinginstructions,to
insureaccurateresultsandformoreadvancedfunctions,
refertotheadditionalsectionsthatfollow.
1.Removetheprotectivecoverfromtheprobehead.
2.Turnthemeteronbypressingthe
ON/OFF
SET
button.
3.Placetheprobeheadintothesolutiontobemeasured.
4.ToinsureAutomaticTemperatureCompensation,
immersetheprobeheadandtemperaturesensor
completely.ForaccurateAutomaticTemperature
Compensation,thetemperaturesensormustreach
equilibriumwiththeliquidbeingmeasured,thiscould
takeseveralminutes.
5.Insurethatthereisadequatemovementoftheliquid
aroundthetipoftheprobehead(0.2-0.3metersper
secondminimum).Thiscanbeachievedbystirringthe
meterintheliquid.
Themeterwillautomaticallyselecttherangethatis
themostappropriateforthevaluebeingmeasured.To
manuallyselecttherange,pressandholdthe
Ç
button
forapproximatelythree(3)seconds(seethe“Range
ModeSelection”section).
6.Takethedesiredmeasurements.
7.Pressthe
ON/OFF
SET
buttontoturnthemeteroff.
Topreservebatterylife,alwaysturnthemeteroffwhen
notinuse.
8.Rinsetheprobeheadwithdistilled/deionizedwaterand
replacetheprotectivecover.Alwaysreplacethe
protectivecovertopreventdamagetotheprobehead.
Ifnobuttonispressedfortwenty(20)minutes,themeter
willturnoffautomaticallytopreservebatterylife(seethe
“AutomaticShutoff”section).
POWER ON DISPLAY SEQUENCE
Whenthemeteristurnedon(bypressingthe
ON/OFF
SET
button),thedisplaywillcyclethroughshowingthe
followingitems:
AllSegments-Displaycheck
tnr Normalizationtemperaturevalue(seethe
“SelectingTemperatureNormalizationValue”
section)
ran Rangedisplaymode(seethe“RangeMode
Selection”section).
SALINITY CALIBRATION
Onecalibrationpointmaybeenteredforeachof
themeasurementranges(0.00to10.00pptand
10.1 to 70.0 ppt).
Toinsureaccuracy:
–Selectaknownstandardwithavalueascloseas
possibletothevalueoftheunknownsample.
–Ifpossible,theknownstandardandunknown
sampleshouldbeatthesametemperature.Theideal
temperatureis25°C.
Previouscalibrationpointsareclearedwhenthemeteris
calibratedtoanewvalue.
1.Removetheprotectivecoverfromtheprobehead.
2.Turnthemeteronbypressingthe
ON/OFF
SET
button.
3.Placetheprobeheadintotheknownstandard.
4.ToinsureAutomaticTemperatureCompensation,
immersetheprobeheadandtemperaturesensor
completely.ForaccurateAutomaticTemperature
Compensation,thetemperaturesensormustreach
equilibriumwiththeliquidbeingmeasured,thiscould
takeseveralminutes.
5.Insurethatthereisadequatemovementoftheliquid
aroundthetipoftheprobehead(0.2-0.3metersper
secondminimum).Thiscanbeachievedbystirringthe
meterintheliquid.
6.Pressandholdthe
È
HLD/CAL
buttonforapproximately
three(3)seconds,CALwillappearatthetopofthe
displayandtheconductivitymeasurementwillash.
7.Pressthe
Ç
and/or
È
HLD/CAL
buttonstoadjustthe
displaytotheknowvalueoftheknownstandard.
The meter may only be adjusted by ±30%
of the detected value.
8.Withthetheknowvalueoftheknownstandard
appearingandCALnolongerashing,pressthe
ON/OFF
SET
button,themeterwillreturntothe
measurementdisplay.
VIEWING CALIBRATION DATA
1.Turnthemeteronbypressingthe
ON/OFF
SET
button.
2.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,P1.0willappearatthebottomof
thedisplay.
3.Pressthe
Ç
buttontwo(2)times,P4.0will
appearatthebottomofthedisplay.
4.Pressthe
ON/OFF
SET
button,P4.1willappearatthe
bottomofthedisplay.Thetopofthedisplaywillshow
thevaluethatthemeterwascalibratedtoforthe
rangeof0.00to10.00ppt.(Ifnousercalibrationhas
beenperformed,“- - - ”willappear.)
5.Pressthe
Ç
button,P4.2willappearatthe
bottomofthedisplay.Thetopofthedisplaywill
showthevaluethatthemeterwascalibratedtoforthe
rangeof10.1to70.0ppt.(Ifnousercalibrationhas
beenperformed,“- - - ”willappear.)
6.Pressthe
ON/OFF
SET
button,P4.0 willappearatthe
bottomofthedisplay.
7.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,themeterwillreturntothe
measurementdisplay.
SELECTING °F OR °C
1.Turnthemeteronbypressingthe
ON/OFF
SET
button.
2.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,P1.0willappearatthebottomof
thedisplay.
3.Pressthe
ON/OFF
SET
button,thetemperatureunitof
measurewillashonthedisplay.
4.Pressthe
Ç
button,thetemperaturereadingwill
switchtotheoppositeunitofmeasure.Ifthemeter
hadbeendisplaying°F,itwillnowdisplay°C,and
vice-versa.Eachpressofthe
Ç
buttonwillswitch
thedisplaytotheoppositeunitofmeasure.
5.Withthedesiredunitofmeasureappearing,pressthe
ON/OFF
SET
buttonthree(3)times,P1.0 willappearatthe
bottomofthedisplay.
6.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,themeterwillreturntothe
measurementdisplay.
Theunitofmeasureselectedwillbethedefaultunitof
measureuntilchanged.
RANGE MODE SELECTION
Themeterwillautomaticallyselecttherangethatis
themostappropriateforthevaluebeingmeasured. To
manuallyselectthedisplayrange,pressandholdthe
Ç
buttonforapproximatelythree(3)seconds.
Eachtimethe
Ç
buttonforpressedandheld,therange
modewillcyclethroughtheavailablemodes:
ATO Automaticrangeselection
1 0.00 to 10.00 ppt
2 10.1 to 70.0 ppt
HOLD FUNCTION
Theholdfunctionallowsthereadingsonthedisplaytobe
“frozen”sothattheymayberecorded.
1.Pressandreleasethe
È
HLD/CAL
buttonto“freeze”
thedisplayatthecurrentreading.“HOLD”will
appearatthetopofthedisplay.
2.Oncethereadinghasbeenrecorded,pressand
releasethe
È
HLD/CAL
buttontoreturnthedisplayto
thecurrentreading.“HOLD”willnolonger
appearonthedisplay.
AUTOMATIC SHUTOFF (Disable)
Ifnobuttonhasbeenpressedfortwenty(20)minutes,
theAutomaticShutofffeaturewillturnthemeteroffto
preservebatterylife.
TodisabletheAutomaticShutofffeature:
1. Withthemeterturnedoff,pressandholdthe
È
HLD/CAL
button,thenpressandholdthe
ON/OFF
SET
buttonuntil
“n ”appearsonthedisplay.
2. Releasethe
È
HLD/CAL
and
ON/OFF
SET
buttons.Themeter
willpoweronandtheAutomaticShutofffeaturewill
bedisabled.
TheAutomaticShutofffeatureisautomaticallyenabled
whenthemeterispoweredoffandturnedbackon.
RESET TO FACTORY DEFAULTS
Resettingthemeterwillclearallcalibrationdataand
resetthetemperaturedisplayto°C.
Thefollowingprocedureisusedtoresetthemetertothe
factorydefaultvalues.
1.Turnthemeteronbypressingthe
ON/OFF
SET
button.
2.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,P1.0willappearatthebottomof
thedisplay.
3.Pressthe
Ç
buttonone(1)time,P3.0will
appearatthebottomofthedisplay.
4.Pressthe
ON/OFF
SET
button,the
n
willashonthedisplay.
5.Pressthe
Ç
button,the
y
willashonthedisplay.
6. With
y
appearing,pressthe
ON/OFF
SET
button,P3.0
willappearatthebottomofthedisplayandthevalues
willhavebeenresettothefactorydefaults.
7.Pressandholdthe
ON/OFF
SET
buttonforapproximately
three(3)seconds,themeterwillreturntothe
measurementdisplay.
TRACEABLE
SALINITY METER
INSTRUCTIONS
®
DISPLAY MESSAGES
Upper Display Messages:
--- Indicatesthatthemeterisinthemanual
rangeselectionmodeandthevaluebeing
measuredishigherthan10.00ppt.
E03 Indicatesthatthevalueofthesolutionbeing
measuredexceedsthemeasurementrangeof
theunit.
E04 Indicatesthatthevalueofthesolutionbeing
measuredexceedsthemeasurementrangeof
theunitbasedonatemperaturemeasurement
error(seethe“LowerDisplayMessages”section).
Lower Display Messages:
E0
1
Indicatesthatthetemperaturesensorisdamaged.
E0
2
Indicatesthatthetemperatureofthesolution
beingmeasuredisbelowthemeasurement
rangeoftheunit.
E0
3
Indicatesthatthetemperatureofthesolution
beingmeasuredisabovethemeasurement
rangeoftheunit.
ALL OPERATION DIFFICULTIES
Ifthismeterdoesnotfunctionproperlyforanyreason,
replacethebatterieswithanew,highqualitybatteries
(seethe“BatteryReplacement”section).Lowbattery
powercanoccasionallycauseannumberof“apparent”
operationaldifculties.Replacingthebatterieswithnew
freshbatterieswillsolvemostdifculties.
BATTERY REPLACEMENT
Anerraticdisplay,faintdisplay,nodisplay,orabattery
symbolappearingonthedisplayareallindicatorsthat
thebatteriesneedreplacement.Thebatterycoveris
locatedatthetopoftheunit.Unscrewthebatterycover
byturningitcounter-clockwise.Removetheexhausted
batteriesandreplacethemwithfour(4)new#357/LR44
sizesilveroxidebatteries.Makecertaintoinsertthe
newbatterieswiththeproperpolarityasindicatedbythe
“+”and“–”symbolsinthebatterycompartment.The
“+”sideofthesilveroxidebatteryistheatsidewith
theengraving.Replacethebatterycoverandtighten
securely.
WARRANTY, SERVICE, OR RECALIBRATION
Forwarranty,service,orrecalibration,contact:
CONTROL COMPANY
12554 Old Galveston Rd. Suite B230
Webster, Texas 77598 USA
Ph. 281 482-1714 • Fax 281 482-9448
www.traceable.com
Control Company is ISO 9001:2008 Quality-Certied
by DNV and ISO/IEC 17025:2005 accredited as a
Calibration Laboratory by A2LA.
Control Company Cat. No. 4367