  • Hello! I am an AI chatbot trained to assist you with the Taylor Technologies K-1770 Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
InstructionNumber:5092
Component Description
> DROPTEST
> CALCIUMHARDNESS(1drop=10ppm)
COMPONENTS:
1x5092 Instruction
1x9198 SampleTube,Graduated,25mL,plastic
w/cap
1xR‐0010 CalciumBuffer,DB
1xR‐0011LCalciumIndicatorLiquid,DB
1xR‐0012 HardnessReagent,DB
TOORDERREPLACEMENTPARTSANDREAGENTSCALLTOLL‐FREE
800‐TESTKIT(800‐837‐8548).
PROCEDURE:
CAREFULLYREADANDFOLLOWPRECAUTIONSONREAGENTLABELS.
KEEPREAGENTSAWAYFROMCHILDREN.
1. Rinseandfill25mLsampletube(#9198)to25mL
markwithwatertobetested.*
2. Add20dropsR‐0010CalciumBuffer.Swirltomix.
3. Add5dropsR‐0011LCalciumIndicatorLiquid.
Swirltomix.Samplewillturnredifcalcium
hardnessispresent.
4. AddR‐0012HardnessReagentdropwise,swirlingand
countingaftereachdrop,untilcolorchangesfrom
redtoblue.Alwaysholdbottleinvertical
position.
5. MultiplydropsofR‐0012HardnessReagentby10.
Recordaspartspermillion(ppm)calciumhardness
ascalciumcarbonate.
*Whenhighcalciumisanticipated,thisprocedure
maybeused:Use10mLsample,10dropsR‐0010,
3dropsR‐0011L,andmultiplydropsinStep4
by25.
5/08