  • Hello! I am an AI chatbot trained to assist you with the SENTRON SI400 7400-0 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
SI line pH METER MANUAL
For products:
SI400
7400-0xx
SI600
7600-0xx
2
Warning! The only user replaceable parts in this meter are the batteries. These can be accessed by
opening the battery compartment. Do not open the meter casing. Opening the meter may
cause damage and will void the guarantee. Also, do not remove the SD memory card from
the meter unless instructed to do so by a Sentron service engineer.
All information provided in this manual is applicable at the moment of publication. Sentron holds the
right to change components, functions and procedures without notice.
Manual SI line pH meter
Ref. number E7500330.02
In the event your Sentron pH meter has this symbol on its label, you are requested to
return this meter to Sentron after use. Sentron will ensure the meter is disposed of in
an environmentally friendly way.
3
Contents
1 General .......................................................................................................................................4
1.1 Declaration of compliance ...................................................................................................4
1.2 Guarantee ...........................................................................................................................4
1.3 Specifications ......................................................................................................................5
2 Installation ..................................................................................................................................6
2.1 Batteries ..............................................................................................................................6
2.2 Connection of probe to meter .............................................................................................7
2.3 SD memory card ..................................................................................................................8
3 Operation....................................................................................................................................8
3.1 Keypad ................................................................................................................................8
3.2 Home screen .......................................................................................................................9
4 Configuration ............................................................................................................................ 12
4.1 Power save ........................................................................................................................ 12
4.2 Time .................................................................................................................................. 14
4.3 Button sound ..................................................................................................................... 14
4.4 Temperature ..................................................................................................................... 15
4.5 Language ........................................................................................................................... 15
5 Calibration ................................................................................................................................ 16
5.1 Determining the calibration type ....................................................................................... 17
5.2 Before calibration .............................................................................................................. 18
5.3 Performing a calibration .................................................................................................... 19
5.4 Calibration reports............................................................................................................. 20
5.5 Calibration reminder ......................................................................................................... 23
5.6 Buffer selection (SI600 only) .............................................................................................. 24
6 Options ..................................................................................................................................... 27
6.1 Alarms (SI600 only) ............................................................................................................ 27
6.2 Storing measurements....................................................................................................... 30
6.3 Bias setting ........................................................................................................................ 38
7 Maintenance ............................................................................................................................. 39
8 Troubleshooting guide .............................................................................................................. 40
9 Contact information .................................................................................................................. 42
4
1 General
The Sentron SI line meter allows you to perform accurate pH measurements together with Sentron SI
line probes. We advise you to read the ‘Quick Startguide before use and keep this guide for future
reference. We also advise you to read this manual to learn about the meter’s many built in features.
The Sentron pH meters and probes are designed for pH measurements. Do not use the meters or
probes for other applications as this may cause damage.
1.1 Declaration of compliance
Sentron Europe B.V. located in Leek, the Netherlands, declares that the Sentron SI line pH meters are
designed and produced in accordance with EMC-standards EN 61326-1: 2006.
As a result of this and adherence to other standards this meter is permitted to carry the CE-mark.
1.2 Guarantee
All Sentron meters are produced, packaged and transported with the utmost care and Sentron
provides a 12 month guarantee for material or manufacturing defects.
All Sentron probes have a guarantee of 6 months. You can find information on the guarantee provided
on probes in the probe manual.
Sentron will, according to its own judgment, either replace or repair a defect meter within 12 months
of shipment, provided the cause is a defect in material of manufacturing and the defect appeared
during correct use of the meter. The guarantee explicitly excludes normal wear and tear and misuse of
the meter.
The decision whether or not the guarantee is applicable is subject to Sentron’s assessment of the
defect cause. Sentron holds the right, at wholly its own discretion, to refuse guarantee claims in the
event it suspects the meter has been used incorrectly. In cases of incorrect use Sentron may, but is not
obliged to, offer to repair the meter at regular repair fees.
Any representations and guarantees made by any person, including distributors, representatives and
employees of Sentron that are contradictory to the abovementioned guarantee conditions are void
unless these are made in writing and signed by an authorized person.
Sentron holds the right to request proof of purchase in the event guarantee claims are made.
Sentron is not liable to users for any damages, either direct or indirect, relating to the use or intended
use of Sentron meters or accessories.
5
1.3 Specifications
6
2 Installation
To start using your Sentron SI pH meter for the first time you will need to connect a probe to the meter
and remove the battery isolator. In the sections below you will find instructions on how to do this. In
the Batteries section you will also find instructions for charging and replacing batteries.
2.1 Batteries
2.1.1 First time use
The SI meter is delivered with three pre-installed pre-charged AA rechargeable batteries. To activate
the meter for the first time, pull the plastic strip (battery isolator) out of the battery compartment and
press the power button.
2.1.2 Recharging batteries
When fully charged, the three AA rechargeable batteries with which the Sentron SI meter was shipped
provide 7 hours of operating under normal use.
To recharge the batteries, connect the meter to the power adapter using the supplied USB cable and
place the power adaptor in a power outlet. The meter will recharge in approximately 7 hours.
In case the meter batteries are empty and will not be recharged soon, remove the batteries.
2.1.3 Replacing batteries
The three AA batteries required for operation of the meter are replaceable. The rechargeable batteries
may lose capacity over their lifetime and replacement may be necessary when operating time on a
battery charge becomes too short. When replacing the rechargeable batteries, only batteries supplied
by Sentron should be installed. To replace the batteries, open the battery compartment cover by
turning the half turn screw counter clockwise and pulling the bottom of the battery cover outwards.
When inserting the batteries ensure the orientation of the batteries have the correct + and - alignment
as indicated in the figure below.
7
It is possible to temporarily install non-rechargeable batteries when the rechargeable batteries have
run out and there are no other fully charged rechargeable batteries available (i.e. when using the meter
in a remote location). This should be done with care as the nonrechargeable batteries should be
removed immediately after use, before the meter is connected to a power adaptor or USB port.
Warning! Do not connect the meter to the power adaptor or to a USB when non-rechargeable
batteries are installed. This may damage the batteries and/or meter.
2.1.4 Batteries and storage
In the event the meter is not used for a longer time period the batteries should be fully recharged and
removed from the meter.
2.2 Connection of probe to meter
The probe is designed to work in combination with a Sentron SI line pH meter. Push the probe’s
connector into the socket on the upper side of the meter. The (white) notch on the probe connector
needs to be aligned with the meter, having the notch facing towards the front side of the meter. The
connector is pushed down until it clicks into the meter’s socket.
The probe can be disconnected from the meter by simply pulling the probe connector.
Note: Do not turn the probe connector once inserted into the meter’s socket. Also, make sure
both meter and probe connector are clean and free from debris before connection.
Do not pull the probe’s cable, always pull the connector.
8
2.3 SD memory card
The battery compartment contains a SD card slot holder with a pre-installed SD card. This SD card is
required for the meter’s operation and should not be removed unless instructed by a Sentron service
engineer.
3 Operation
Operation of the SI meter is done via the 6 keys below the screen.
3.1 Keypad
The SI meter contains a keypad with six keys
Navigation
Display on/off
(save power)
Move back one menu step
Power meter
on/off
Confirm or select menu
steps
Return to the main (Home)
screen
9
3.2 Home screen
The home screen displays all the key information required when making pH measurements. It contains
five sections:
Date, time and battery status
Measurement values
Notifications
Status bar
Main menu
The SI meter’s home screen
Data, time and
battery status
Measurement values
Notifications
Status bar
Main menu
3.2.1
3.2.2 Date, time and battery status
The top of the screen contains the actual date (DD-MM-YYY) and the time (user settable as 12 or 24
hour notation).
In the right hand corner the status of the battery is displayed:
Indicator
Battery status
Charging
100% capacity
75% capacity
50% capacity
25% capacity
Almost empty
3.2.3 Measurement values
The measurement values section on the home screen contains the pH and temperature recorded by
the probe. When a probe is connected the pH value should be between 1 and 14. When no probe is
connected the pH value will be 16.00(0) and the temperature will read -20°C (or -4°F).
10
3.2.4 Notifications
Only when the main screen is displayed, some additional notifications regarding certain settings are
can being displayed on top of the status bar.
Alarm (SI600 only):
An alarm can be set when pH and / or temperature values go beyond user definable upper and lower
limits. See chapters 6.1.1 and 6.1.2 for setting or changing the pH and temperature alarm settings.
Notification
of a set Alarm
Notification of
an activated Alarm
Calibration reminder:
A calibration reminder can be set to perform calibrations within a specific time interval. The time
interval is user definable. See chapter 5.5 for setting or changing the calibration reminder function.
Notification of a set
calibration reminder
Notification of an
activated calibration
reminder
Stability check:
The stability check can be set to provide a warning when the rate of change of pH exceeds a certain
threshold. See chapter 6.1.3 for setting or changing the stability check.
Notification of a set
stability check
Notification of an
activated stability check
Bias:
A user definable bias value can be set to adjust the displayed pH measurement value. See chapter 6.3
for setting or changing the bias.
Notification of a set
bias
Range:
A range notification is a default alarm (not user definable) and provides a warning when pH and / or
temperature conditions exceed the specification limits of the equipment.
Notification of a
range alarm
11
3.2.5 Status bar
The status bar displays additional information about the meter status. What is displayed depends on
whether the meter is on the home screen, in a sub menu or a function is being selected.
On the home screen, the status bar will display information regarding the probe, e.g. Probe
disconnected or Calibration OK. When sub menu’s are selected, it will show the menu name. When a
function is being selected the status bar will provide useful instructions or status information such as
Place probe in next buffer, Take sample, Send data etc.
3.2.6 Main menu
The meter’s main menu has three icons: Config, Calibrate and Options. Selecting one of these main
menu icons will display the subsequent menu.
The Config submenu contains the general configuration options of the meter, such as display,
language and time settings.
The Calibrate submenu contains all the functions and options related to the calibration of the
probe. For instance, performing a calibration, setting buffer type and recalling calibration
report(s).
The Options submenu allows the user to set parameters and perform actions related to
measurements and measurement values. In this submenu functions such as alarm thresholds,
data acquisition and bias settings can be accessed.
12
4 Configuration
In the Config submenu a wide range of features such as energy saving, time and display language
settings can be selected. All features in this menu relate to the operation of the meter itself, options
and settings related to the measurements can be found in the Options menu (Page 27).
4.1 Power save
The Sentron SI meter has two options to save power and extend battery operation time: Display-off
and Meter-off. Set either one, or both, of the options to maximize use of the battery capacity:
Display-off: switch off the display after a user definable inactive time.
When the display-off option is activated, the screen will switch off after the selected time has expired
since the last key has been pressed. Pressing any key will turn the screen back on.
While the screen is turned off, the meter will still fully function and perform such actions as logging
and measuring the signals from the connected probe. Pressing a key during the display-off mode will
not evoke any action other than reactivating the display.
To set or change the Display-off setting:
Go to the Config menu,
and choose Power save.
Now select Display-off.
Toggle the on/off button and select the
display-off time between 1 and 60
minutes.
Confirm and apply with the Set button, or exit
with the Back button to discard any changes.
Meter-off: switch off the meter after a user definable inactive time.
With the meter-off function activated, the meter will power down entirely after the selected time has
expired since the last key action. The meter has to be turned back on in the same way as a “normal”
power-up; by pressing the ON button on the keypad.
When the meter is switched off using this function, the meter is entirely powered down. This means
the ISFET pH sensor circuit will be powered down as well. Since a new calibration is recommended
after a startup, it is therefore recommended to perform a new calibration after an automatic Meter-
off as well.
The meter will not power-down when the AC-adapter is connected.
13
To set or change the Meter-off setting:
Go to the Config menu,
and choose Power save.
Now select Meter-off.
Toggle the on/off button and select the
display-off time between 1 and 60
minutes.
Confirm and apply with the Set button, or exit
with the Back button to discard any changes.
Note: For meters with logging capabilities (SI600), the meter-off function will be automatically
disabled for the period of active logging. Hence, setting this function will not affect
measurements during continuous logging.
14
4.2 Time
The Sentron SI meter is equipped with a highly accurate internal clock. The date and time are factory
set to Central European Time (CET), which is 1 hour ahead of Coordinated Universal Time (UTC +01:00).
The meter’s time setting can be adjusted to the local time and either 12 or 24 hour display can be
selected.
To set the meter to your local time the time can be adjusted 24 hours forwards or backwards with
respect to the factory preset time. The date will adjust accordingly when the 0:00 time is crossed. Time
zones west of UTC should adjust time backwards and time zones east of UTC should adjust time
forwards.
Daylight savings time is not set automatically and, if required, should be set manually.
To change the time and/or set the 12- or 24-hour clock:
Go to the Config menu,
and select Time.
Toggle the 12/24 button to set the
display into either a 12- or 24-hour clock
notation.
Toggle the Hours and Minutes buttons to
change the meter’s current setting.
Confirm and apply with the Set button, or exit
with the Back button to discard any changes.
4.3 Button sound
When operating the Sentron SI meter, audible sounds can be generated to provide feedback on button
operation. Sound can be set to on or off, the volume of the sound cannot be altered.
The button sound setting has no influence on the audible notifications for set alarms.
To set or disable the button sound:
Go to the Config menu,
now select and enter the Button sound
option.
Toggle and choose the button sound to
on or off.
Confirm and apply with the Enter key, or exit
with the Back key to discard any changes.
15
4.4 Temperature
The Sentron SI meter can display temperature in either Celsius or Fahrenheit. The resolution is not
influenced by the chosen unit setting.
To change the temperature unit:
Go to the Config menu,
now select and enter the Temperature
option.
Toggle and choose the desired unit
setting.
Confirm and apply with the Enter key, or exit
with the Back key to discard any changes.
4.5 Language
The language in which the Sentron SI meter displays menus and options can be changed into one of
the five pre-installed languages; Dutch, English, French, German, and Spanish.
To change the language:
Go to the Config menu,
now select and enter the Language
option.
Toggle and choose one of the available
languages.
Confirm and apply with the Enter key, or exit
with the Back key to discard any changes.
16
5 Calibration
In order to perform accurate pH measurement with a Sentron SI meter, it must be calibrated before
use. Generally, a calibration should be performed each time the meter is switched on or when a
different probe is attached. In practice this means calibration should be performed at least daily.
The calibration process consists of placing a probe attached to the Sentron pH meter in one, or a series
of, buffer solutions with a predetermined pH. The type of calibration, and the number of calibration
points depend on several factors. In general, the more buffers that are used, the more accurate the
calibration will be. A detailed explanation of the calibration process, and the considerations for
choosing the number of calibration points, can be found in section 5.1.
The Sentron SI meter displays a calibration results at the end of each calibration and stores the
calibration results in a report. This report is available for review at a later time. Please review section
5.4 for detailed information on accessing and interpreting calibration reports.
The Sentron SI meter can provide a warning as a reminder to perform a calibration. The warning will
be activated after a user definable time period since the last calibration. It is recommended to set this
calibration reminder time at an appropriate setting for the intended application. This function is
explained in section 5.5
The Sentron SI 400 meter can only be calibrated using Sentron buffer sets. The SI 600 meter can be
calibrated using Sentron, DIN, NIST and JIS buffer sets. The process for changing the buffer set
configuration is explained in section 5.6 below.
17
5.1 Determining the calibration type
When starting a new pH measurement, the first step is to determine what calibration process is most
appropriate for this application. Choosing the wrong calibration process may result in inaccurate
measurements and possibly require the measurements to be repeated.
Calibrations can be either single or multiple point. The SI400 meter allows 1 and 2 point calibration,
whilst the SI600 allows 1, 2, 3 and 5 point calibration. In general, increasing the number of points used
for calibration will increase the measurement accuracy. However, the calibration range should be as
close as possible to the range over which measurements are expected to take place and extend to
above and below the minimum and maximum expected measurement values.
1 point calibration, an option for quick pH measurements spanning 1 2 pH.
Example:
A single point calibration would be an option for measurements where the objective is to quickly
detect a difference from one single point e.g. pH 7, and the variation is always lesser than 1 pH in
either direction. For a 1 point calibration a buffer should be available with a pH close to the point from
which the deviations are to be measured
2 point calibration, recommended for accurate pH measurements spanning < 3 pH and
quick measurements spanning 3 6 pH.
Examples:
If, for instance, measurement values are expected to lie between pH 7 and pH 10, a 2 point calibration
would be appropriate. A 2 point calibration would be performed with buffers with pH 7 and pH 10.
If, on the other hand, measurement values are expected lie between pH 6 and pH 10, a 2 point
calibration performed with a pH 4 and pH 10 buffer would be acceptable. For most accurate
measurements, however, a 3 point calibration with pH 4, pH 7 and pH 10 buffers should be
considered.
3 point calibration, recommended for accurate pH measurements spanning 3 6 pH and
quick measurements spanning >6 pH.
Examples:
A series of measurements that are expected, for example, to take place between 4 and 10 pH would
most accurately be performed when preceded by a 3 point calibration with pH 4 and pH 7 and pH 10
buffers.
Measurements expected to take place over a wider range of 7 pH, for example, between pH 2 and pH
11 could also be performed using a 3 point calibration with pH 2, pH 7 and pH 12 buffers, but would
more accurately be measured using a 5 point calibration.
18
5 point calibration, recommended for accurate pH measurements spanning > 6 pH.
Examples:
Measurements expected to take place between pH 3 and pH 11 would require a full 5 point with pH
2, pH 4, pH 7, pH 10 and pH 12 buffers to achieve the most accurate measurements.
A number of experiments conducted sequentially that each have a relatively narrow range of
expected values, but whose values overall are spread over a wide range, for which pH will be taken
without interim calibrations between the experiments will require a 5 point calibration
Experiment 1 values are expected between pH 2 and pH 3
Experiment 2 values are expected between pH 7 and pH 9
Experiment 3 values are expected between pH 9 and pH 12
5.2 Before calibration
To ensure the best possible calibration results, always observe the following guidelines:
Only use the buffer set with which your meter is configured
Only use fresh buffers for a calibration
Make sure your buffer sets are at the same temperature range as your sample to be measured
Before, during and after calibration, always clean and flush your probe. Please see the probe
manual for instructions on how to clean the probe.
Make use of the calibration reminder function to ensure calibrations are made at appropriate
time intervals (see chapter 5.5)
Make use of Sentron twin-neck calibration fluid bottles as they minimize contamination of the
buffer and allow convenient storage and calibration using a single bottle.
When a new probe is attached to a meter, or when a probe has not been used for a while, the probe
may react slow or has low slope readings after calibration. In this case the probe will need to be
“soaked” (revitalized). This will reactivate an inactive probe and ensures it will produces stable
readings. For more information on revitalization and treatment of the probe, please review the probe
manual.
19
5.3 Performing a calibration
The calibration sequence is set up in such a way that it is easy to perform and minimal input is required.
To rule out erroneous multi-point calibrations, the calibration can only take place for an increasing or
decreasing pH sequence. Hence, a multi-point calibration must be started with either the highest or
lowest pH value in the intended range.
To perform a calibration, first chose the number of calibration points for which the calibration is to be
performed. Next, select the pH value of first buffer with which the calibration with be started.
In most cases, the first selected buffer type combined with the chosen number of calibration points, is
sufficient to determine the calibration path. In some multiple point calibrations cases, however, a
second buffer solution has to be selected in order for the meter to understand the desired calibration
path.
The screen will continuously provide information about the status of the calibration and which step is
to be taken next. After the calibration is completed, a report is shown with details of the performed
calibration such as time, date, temperature, slope and buffer type used.
To perform a calibration:
Go to the Calibrate menu,
and choose Calibrate.
Now toggle and select number of
calibration points.
Toggle and select the first calibration
value to start with. (some cases require a
second calibration value to be entered)
Follow the instructions provided by the meter and when done, a calibration
report is displayed. Exit the calibration report with the Close button.
Successful calibrations result in the display of the message calibration OK on the home screen status
bar. If, for any reason the calibration process has not been successful a message will be displayed on
the screen. Also, a calibration process may be aborted. In both cases the meter will automatically
revert back to the last successful calibration results. Be aware that this last calibration may have
exceeded the recommended or user defined calibration interval.
5.3.1 Aborting a calibration
Aborting a calibration by pressing the “back” or homebutton during the calibration process will
evoke the message “probe disconnected” for a maximum of 120 seconds at the status bar on the main
screen. As well, the measurement values will be out of range in this time period. Hereafter, the probe
will return to the same calibration state as it was before.
Recommendation: disconnect the probe from the meter to prevent the above described waiting time.
The calibration procedure has to be repeated completely.
20
5.4 Calibration reports
After each calibration a report is generated and stored in the meter’s memory. The date of calibration,
the buffers used, measured temperature and, in the case of a multi-point calibration, the calculated
slopes are summarized in this report. Calibration reports are shown after each calibration and are
stored by the meter.
5.4.1 Review calibration reports
A calibration report can be viewed to review how the last calibration process was performed. To see
the report and results again at a later time, it can be opened form the calibration reports menu.
The SI400 stores only the latest calibration report and only this last report can be reviewed:
To review a calibration report (SI400):
Go to the Calibrate menu,
and choose Calibration report.
Exit the report with the Close button, or remove
the report with the Delete button.
The SI600 has the possibility to save, review and send the last seven calibration reports to a PC:
To review a calibration report (SI600):
Go to the Calibrate menu,
and choose Calibration reports.
Now select Browse reports.
Choose and select one of the calibration
reports in the list.
Exit the report with the Close button, or remove
the report with the Delete button.
5.4.2 Interpreting the displayed slope percentage
The slope percentage displayed in the calibration report is an indication of the calibration’s success
and the state of the probe.
Normal slopes should be between 105% 95%.
Slopes outside these values, indicate polluted or aging probes. Although calibrations and
measurements can be performed the measured values may be less accurate. In this case, be sure to
replace the probe.
/