Simplex GD2 Digital Gauge - 55652_b User manual

Type
User manual
55652 Rev. B 12/2019
Instruction Sheet
GD2 10,000 PSI
Digital Gauge
The model GD2 digital pressure gauge has several features not normally found as standard in digital pressure
gauges. These include the analog scale, tare, and peak holds. The analog scale is a 40 segment display
that appears around the edge of the liquid crystal display. This scale represents the pressure reading as a
percent age of full scale, and each individual segment equals 2.5%. The tare feature provides a method of
zeroing out an existing oset such as the weight of the platform on a hydraulic scale. The hold feature allows
the user to select Hi or Lo. When the hold is enabled the reading displayed is the highest or lowest pressure
experienced in the system since hold was enabled, and this information will continue to be updated at the
rate of six times a second until hold is disabled.
A) ZERO OR “TARE OUT” AND EXISTING OFFSET:
1. Press and hold the “MODE” key until the reading starts flashing. Release “MODE” key.
2. Press “ENTR” key until the reading becomes 0000. NOTE: Tare range is 100% full scale.
3. To return to the gross reading press and hold “ENTR” then with “ENTR” still depressed press “MODE”
until the reading returns to the Gross, then release both keys.
B) TO SELECT PEAK HI OR PEAK LO HOLD FUNCTIONS:
1. Press and hold the “MODE” key until the reading starts flashing. Release “MODE” key, then press
the “MODE” key once more, the display will read “HOLD”. Press “ENTR” key once, the display will
read”NO H”. Using “” or “” keys select “NO H” (no hold), “H LO” (peak low hold), or “H HI” (peak
high hold). Press “ENTR” key to lock your selection.
NOTE: When the hold function is enabled “PEAK” is displayed, this will be the same for “Hi” or “Lo” Hold.
YOU MUST REMEMBER WHICH YOU HAVE ENGAGED.
2. To reset after capturing a Hi Peak or Lo Peak, press and hold “MODE” key until the reading starts
flashing. Release the “MODE” key, then press “ENTR”, hold momentarily, then release.
3. To disengage Hold, press and hold the “MODE” key until the reading starts flashing. Release the
“MODE” key, then press the “MODE” key once more, the display will read “HOLD”. Press the “ENTR”
key once more and the display will return to reading the system pressure.
C) ZERO TRIMMING:
Should it be necessary to readjust the Zero setting, an adjustment has been provided. Prior to ad-
justing the zero, verify that you are displaying the “gross” zero reading by performing step A.3. In the
upper left of the rear of the case, there are two cover screws labeled “Z” and “S”. Remove the screw
covering the ‘7”, and utilizing a jewelers screwdriver turn the small screw to re-zero the unit. The ideal
zero is indicated by a reading of 0000 with an intermittently flashing “-” sign.
NOTE: DO NOT READJUST THE “S” AS THIS WILL INVALIDATE THE FACTORY CALIBRATION.
Description
The GD2 digital gauge combines the best features
of analog and digital instruments. The 0.75 in. high
full 4 digit display is easily read at a distance and
the “analog” bar graph allows instant estimation of
changing pressures and gives visual indication of alarm
set points. The all stainless steel sensor is vibration
and shock resistant, and unlike mechanical gauges is
immune to the eects of pulsating pressures. Available
options include RS232 output, alarm set points,
0-2V analog output, and 4-20 mA transmitter output.
Battery Powered Digital Gauge
Oers Digital and Analog Display
with Standard High or Low Peak Hold
Features
Self Powered
Extra Large Full 4-Digit Display With .75 in.
Characters and 270 Degree Bar Graph “Analog”
Readout
Direct Reading to 10,000 PSI
Splash Proof Phenolic Housing,
NEMA 4 Sealing Optional
Automatic Light-Actuated
Battery-Saver Circuit
Microprocessor-Based Circuit
± 0.25% of Full Scale Accuracy
Keypad Enable/Disable of Peak High
Or Peak Low Hold
Auto “Tare” (Zero) Feature
Applications
Portable test gauge (± 0.1 % accuracy available).
Process control combined with local indication.
Highly cyclic pressures where mechanical gauges
are failing. Burst tests (with peak hold). Severe envi-
ronments.
Specifications
PERFORMANCE
Standard Pressure Ranges: ....0-20 to 0-10,000 psi Gauge or Absolute
Overpressure: .........................2x full scale without damage
Burst Pressure: ....................... 10x full scale or 20.000 PSIG
whichever Is less
Pressure Cavity Volume: ........0.095 in
Accuracy (linearity, hysteresis,
and repeatability): ................... ±0.25% of full scale (±0.1 % optional)
Compensated Temperature Range:
..................30 °F to 130 °F
Thermal Zero Shift .................... ±0.02% F.S./°F
Thermal Sensitivity Shift: ........... ±0.02% F.S./°F
Resolution (Transducer): ......... Infinite
(Display): ...............1 part in 10,000
Life: ...........................................10 million cycles minimum
ELECTRICAL
Batteries: ................................. 1 ea. 3.6v Lithium (standard)
Life....8,000 HRS
2 ea. 3.6v Lithium (optional)
Life....16.000 HRS
2 ea. 1.5v Alkaline (optional)
Life....3,000 HRS
Display: .................................... 4 digit LCD, 0.75" Digits and
40 segment bar graph
On/O Control: .......................By Light Actuated Switch Circuit
Hold: Peak Hi/Peak Lo ...........Keypad Enable/Disable
Auto “Tare” .............................. Keypad Operated Auto Zero
ENVIRONMENTAL
Storage Temperature: ............-40 °F to +160 °F
Operating Temperature: .........0 °F to +160 °F
Compensated Temperature: ... 30 °F to +130 °F
PHYSICAL
Construction Materials
Case: ........................................ Splash Proof Injection Molded Phenolic,
NEMA 4 Sealing Optional
Wetted Materials: ................... 15-5 or 17-4 P.H. Stainless Steel
316 S.S. available at no extra charge
D) RE-CALIBRATION:
This procedure requires a known pressure source of at least ±0.1% accuracy in order to fully utilize the ac-
curacy potential of the GD2. If not available, gauge can be returned to Simplex for re-calibration.
1. With no pressure applied (port vented), adjust zero as per instruction in A.3.
2. Apply full scale pressure to the pressure port and adjust the span (“S”) pot until display reads the
correct pressure.
3. Recheck zero and re-adjust the zero (“Z”) pot if required.
4. Repeat steps 2 and 3 until no further adjustment is required.
E) BATTERY REPLACEMENT:
1. To access the battery, the unit must be partially disassembled. Remove the thread ring surrounding
the lens by turning counter clockwise. Carefully turn the unit face down with one hand prepared to
catch” the lens and display board as they fall from the housing. Do not allow the display board to fall
more than an inch or two as there are wires connecting it to the main housing.
2. Remove the battery or batteries from the holder(s) and replace with the same type that was re moved,
taking care to observe the correct polarity.
F) ANALOG OUTPUT SIGNAL OPTION:
The analog output signal is available on pins B and C of the circular 6 pin connector located on the
rear of the unit. The signal is short circuit protected and capable of driving low impedance external
loads with fixed gain to a specified output level, generally 0 to 2.0 VDC with 10K Ohm and above
loads. Frequency response is a function of the pressure range, generally 1 to 3 KHZ. Loading of the
analog output will result in a reduction of the battery life expectancy due to the additional current
through the external load.
N86 W12500 Westbrook Crossing
Menomonee Falls, WI 53051
Phone - USA Only: 1 (800) 323-9114
Phone - Inside/Outside USA: 1 (262) 293-1575
Fax: 1 (262) 597-1127
ww w.tksimplex.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4

Simplex GD2 Digital Gauge - 55652_b User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI