Micro Motion Digital Viscosity Meter - Model 7827 User manual

Category
Measuring, testing & control
Type
User manual

This manual is also suitable for

Installation and Configuration Manual
P/N MMI-20015440, Rev. AA
July 2009
Micro Motion
®
7827 Digital
Viscosity Meter
Short and Long Stem Versions
©2009, Micro Motion, Inc. All rights reserved. Micro Motion is a registered trade name of Micro Motion, Inc., Boulder, Colorado.
The Micro Motion and Emerson logos are trademarks and service marks of Emerson Electric Co. All other trademarks are property
of their respective owners.
Micro Motion pursues a policy of continuous development and product improvement. The specification in this document may
therefore be changed without notice. To the best of our knowledge, the information contained in this document is accurate and
Micro Motion cannot be held responsible for any errors, omissions, or other misinformation contained herein. No part of this
document may be photocopied or reproduced without prior written consent of Micro Motion.
Installation and Configuration Manual i
Contents
Chapter 1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Safety guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 About the meter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.1 What is it? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.2 7827 meter measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.3 What is it used for? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Principle of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Chapter 2 Installation (Short Stem) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Boundary effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Standard installations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3.2 Meter orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.3 Free stream installation - flanged fitting . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3.4 Free stream installation - weldolet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3.5 T-piece installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.3.6 Flow-through chamber installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.4 Installation in the pipeline or system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.5 Typical installations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.5.1 Jacketed pipeline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.5.2 Flow-through chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Chapter 3 Installation (Long Stem) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2 Installation considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2.1 Fluid at the sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2.2 Flow rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2.3 Entrained gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.2.4 Solids contamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.3 Open-tank installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.4 Closed-tank installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6 If the Tank is Pressurized . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Chapter 4 Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.2 EMC and cabling considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.3 Installation and safety in hazardous areas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.4 Installation in non-hazardous areas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.5 Wiring the meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.6 Connecting the 7827 to a 795x series computer. . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.6.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.6.2 Connection diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
ii Micro Motion 7827 Digital Viscosity Meter
Contents
4.7 Checking the installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Chapter 5 Using 7950/7951 Processing Electronics . . . . . . . . . . . . . . . . . . . . 39
5.1 Using the 7950 / 7951 Processing Electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Chapter 6 Calibration Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.2 Factory calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.2.1 Viscosity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.2.2 Density. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.2.3 Primary standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.2.4 Transfer standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.3 In-line calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.3.1 Viscosity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.3.2 Density. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.3.3 Requirements for VOS correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.4 Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.4.1 Viscosity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.4.2 Density. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.5 Calibration certificate example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Chapter 7 General Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.2 General maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.3 Fault analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.3.1 Checking the Signal Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.3.2 Checking the power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.3.3 Checking the installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Appendix A Calculated Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
A.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
A.2 Viscosity equations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
A.2.1 Quality factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
A.2.2 General viscosity equation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
A.3 Density equations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
A.3.1 General density equation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
A.3.2 Temperature correction equation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
A.3.3 Viscosity correction equation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
A.3.4 Pressure correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
A.3.5 Velocity Of Sound correction equation. . . . . . . . . . . . . . . . . . . . . . . . . . . 57
A.3.6 Density scaling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
A.4 Kinematic viscosity equation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Appendix B Safety Certification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
B.1 Safety certification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Installation and Configuration Manual iii
Contents
Appendix C Product Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
C.1 Density / temperature relationship of hydrocarbon products. . . . . . . . . . . . . . . . . . . 61
C.1.1 Crude oil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
C.1.2 Refined products . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
C.1.3 Platinum resistance law . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
C.1.4 Density of ambient air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
C.1.5 Density of water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
C.1.6 Velocity of sound in liquids. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Appendix D Return Policy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
D.1 General guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
D.2 New and unused equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
D.3 Used equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
iv Micro Motion 7827 Digital Viscosity Meter
Contents
Installation and Configuration Manual 1
Installation (Short Stem) Electrical ConnectionsInstallation (Long Stem)Introduction
Chapter 1
Introduction
1.1 Safety guidelines
Handle the 7827 digital viscosity meter with great care.
Do not drop the meter.
Do not use liquids incompatible with materials of construction.
Do not operate the meter above its rated pressure or maximum temperature.
Do not pressure test beyond the specified test pressure.
Ensure all explosion-proof requirements have been applied.
Ensure the meter and associated pipework are pressure tested to 1-1/2 times the maximum
operating pressure after installation.
Always store and transport the meter in its original packaging, including the transit cover
secured by grub screws.
To return a meter, refer to the Return Policy appendix for more information on the Micro
Motion return policy.
Safety messages are provided throughout this manual to protect personnel and equipment. Read each
safety message carefully before proceeding to the next step.
2 Micro Motion 7827 Digital Viscosity Meter
Introduction
1.2 About the meter
1.2.1 What is it?
The 7827 meter is a digital viscosity meter, based on the proven tuning fork technology of Micro
Motion. It is an all-welded sensor designed to be mounted directly into a pipeline or in a tank.
Viscosity and density are determined from the resonance of the tuning fork immersed in the fluid, and
a temperature sensor (PRT) is also fitted within the meter.
The 7827 meter is available in a variety of materials, and the immersed tines can be laminated with
PFA to inhibit the build up of residues such as asphaltenes.
1.2.2 7827 meter measurements
The 7827 meter directly measures the following fluid properties:
Line dynamic viscosity – measured in centiPoise - cP.
Line Density – measured in kg/m
3
.
Temperature – measured in °C or °F.
From these properties, the 7827 meter calculates:
Line and base (referred) kinematic viscosity – measured in centiStokes - cSt.
Line and base (referred) density – API or Matrix.
Referral is made to 15°C, 1.013 bar; or at 60°F, 14.5 psi.
Electronics housing
Spigot
Tines
Stem (available in varying
lengths)
Stem is sealed at both ends
and air filled
Installation and Configuration Manual 3
Introduction
Installation (Short Stem) Electrical ConnectionsInstallation (Long Stem)Introduction
1.2.3 What is it used for?
The 7827 meter is ideally suited to applications where continuous real time measurement of viscosity
is required. The meter is particularly suited where viscosity is an indication of the behavioral
properties of the fluid, for example in applications involving spraying, coating or dipping.
Some uses are in the oil and petrochemical industry for:
•Refining
•Marine
•Power
Heavy fuel oil (HFO) blending and bunkering
1.3 Principle of operation
The 7827 meter operates on the vibrating element principle, the element in this case being a slender
tuning fork structure which is immersed in the liquid being measured.
The tuning fork is excited into oscillation by a piezo-electric crystal internally secured at the root of
one tine, while the frequency of oscillation is detected by a second piezo-electric crystal secured at the
root of the other tine. The sensor is maintained at its first natural resonant frequency, as modified by
the surrounding fluid, by an amplifier circuit located in the electronics housing.
The electronics circuit actually excites the sensor into oscillation alternately at two positions on the
frequency response curve as shown in Figure 1-1. In doing this, the quality factor (Q) of the resonator
may be determined as well as the resonant frequency.
For details of the viscosity and density calculations, see the Calculated Parameters appendix.
Figure 1-1 Frequency response curve showing the quality factor (Q) calculation
0dB level
-3dB level
Frequency
Response
τ
B
τ
A
A
B
4 Micro Motion 7827 Digital Viscosity Meter
Introduction
Installation and Configuration Manual 5
Installation (Short Stem) Electrical ConnectionsInstallation (Long Stem)Introduction
Chapter 2
Installation (Short Stem)
For information on installing a long-stem version of the 7827 digital viscosity meter, see Chapter 3.
2.1 Introduction
There are a variety of external factors that affect the ability of the 7827 digital viscosity meter to
operate successfully. In order to ensure that your system works correctly, the effects of these factors
must be taken into consideration when designing your installation.
There are two main aspects to consider:
The accuracy and repeatability of the measurements
The relevance of the measurements to the overall purpose of the system
Factors which may adversely affect accuracy and repeatability include:
The presence of gas or bubbles within the fluid being measured
Non-uniformity of the fluid
The presence of solids as contaminants
Fouling of the meter
Temperature gradients
Cavitations and swirls
Operating at temperatures below the wax point of crude oils
The correct pipe diameter that corresponds to the calibration of the meter.
In some applications, absolute accuracy is less important than repeatability. For example, in a system
where the control parameters are initially adjusted for optimum performance, and thereafter only
checked periodically.
The term achievable accuracy can be used to describe a measure of the product quality that can be
realistically obtained from a process system. It is a function of measurement accuracy, stability and
system response. High accuracy alone is no guarantee of good product quality if the response time of
the system is measured in tens of minutes, or if the measurement bears little relevance to the operation
of the system. Similarly, systems which require constant calibration and maintenance cannot achieve
good achievable accuracy.
All drawings and dimensions given in this manual are given here for planning purposes only. Before
commencing fabrication, reference should always be made to the current issue of the appropriate
drawings. Contact Micro Motion for details.
For further information on handling and using the meter, see “Safety guidelines” on page 1
6 Micro Motion 7827 Digital Viscosity Meter
Installation (Short Stem)
Factors which may adversely affect the relevance of the measurements could include:
Measurement used for control purposes being made too far away from the point of control, so
that the system cannot respond properly to changes.
Measurements made on fluid which is unrepresentative of the main flow.
2.2 Boundary effects
Any insertion device or meter can only measure the properties of the fluid within the region of fluid to
which it is sensitive.
For practical reasons, it is helpful to consider the sensitive, or effective region, for the viscometer as
an ovoid centered on the tips of the tines with its long axis aligned with the direction in which the
tines vibrate, as shown below. The 7827 meter is insensitive to the properties of the fluid outside this
region and progressively more sensitive to fluid properties the closer the fluid is to the tines. Density
can be considered a “mass centered” effect and viscosity a “surface centered” effect in this
visualization; i.e. the measurement of density is more uniformly sensitive to the density of fluid
throughout the region while viscosity measurement is much more critically sensitive to fluid on the
surface of the tines.
If part of this volume is taken up by the pipework or fittings there is said to be a boundary effect; i.e.,
the intrusion of the pipe walls will alter the calibration. The diagram below illustrates the 7827 meter
installed in a pocket on the side of a 4" (100 mm) horizontal pipe line (viewed from above). The
effective region is completely enclosed within the pipe line and thus is completely fluid.
long axis
short
axis
Installation and Configuration Manual 7
Installation (Short Stem)
Installation (Short Stem) Electrical ConnectionsInstallation (Long Stem)Introduction
This next view shows other pipe outlines superimposed:
The smaller circle represents a 4" (100 mm) vertical pipe, which because the 7827 meter orientation is
constant irrespective of pipe orientation intersects the effective region. The 6" (150 mm) pipe is the
smallest pipe diameter to completely enclose the effective region when the pipe is vertical. Thus
smaller pipe diameters can lead to a variety of different geometries which would each require a
separate calibration.
An alternative condition is shown in the next diagram where the side pocket is extended until it passes
completely through the effective region producing a “core”:
Top or Plan view
4” horizontal pipe
2”Schedule 40
Pocket or “T”
8 Micro Motion 7827 Digital Viscosity Meter
Installation (Short Stem)
From this, it would appear that almost every installation requires a separate in situ calibration – a very
undesirable situation. The problem is resolved by providing standard calibration geometries which
can be used in all pipe work configurations and thereby allow the factory calibration conditions to be
reproduced in the process.
2.3 Standard installations
2.3.1 Overview
To overcome the need for in situ calibration for every installation, three standard installations are
proposed. If an installation conforms to one of these standards, the factory calibration of the 7827
meter is valid, and in-situ calibration unnecessary. Table 2-1summarizes the three installations. For
tank installations, consult Micro Motion.
Table 2-1. Types of standard installations
Installation type Free stream T-piece Flow-through chamber
Description 7827 meter tines are inserted
directly into the main fluid flow.
7827 meter tines are contained
in a side pocket off the main
flow.
7827 meter tines are
contained in a flow-through
chamber in which fluid is
circulated from the main
flow.
Flow rate 0.3 to 0.5 m/s at the 7827 meter. 0.5 to 3 m/s at main pipe wall. 10 to 30 l/min.
Viscosity 0.5 to 12500 cP 0.5 to 100 cP 0.5 to 1000 cP
Temperature
(1)
-50 to 200°C
(-58 to 392°F)
-50 to 200°C
(-58 to 392°F)
-50 to 200°C
(-58 to 392°F)
Installation and Configuration Manual 9
Installation (Short Stem)
Installation (Short Stem) Electrical ConnectionsInstallation (Long Stem)Introduction
2.3.2 Meter orientation
The meter must always be installed horizontally, and orientated to allow flow in the gap between the
tines. This is irrespective of the pipe line orientation, and helps to prevent the trapping of bubbles or
solids on the meter.
Main flow pipe size 100 mm (4") horizontal
150 mm (6") vertical, or larger.
100 mm (4") horizontal or
larger.
Any.
Advantages Simple installation in large bore
pipes.
Ideal for clean fluids and
non-waxing oils.
Suitable for line viscosity
measurement and simple
referrals.
Simple installation in large
bore pipes.
Ideal for clean fluids and
non-waxing oils.
Suitable for line viscosity
measurement and simple
referrals.
Adaptable installation to
any diameter main pipe
and for tank applications.
Ideal for flow and
temperature conditioning.
Suitable for complex
referrals and for use with
heat exchangers.
Suitable for step changes
in viscosity.
Fast response.
Ideal for analyser
cubicles.
Not recommended
for
Dirty fluids.
Low or unstable flow rates.
Where step changes in
viscosity can occur.
For small bore pipes.
Dirty fluids
Low or unstable flow rates.
Where step changes in
viscosity can occur.
for small bore pipes.
Where temperature effects
are significant.
Uncontrolled flow rates.
Careful system design
required to ensure
representative
measurement.
Frequently requires the
use of a pump.
(1)
Approval for use in hazardous areas is limited to –40 to +200°C (–40 to +392°F)
Table 2-1. Types of standard installations continued
Installation type Free stream T-piece Flow-through chamber
10 Micro Motion 7827 Digital Viscosity Meter
Installation (Short Stem)
Figure 2-1 Meter orientation
Note: All drawings and dimensions given in the following sections are derived from detailed
dimensional drawings. They are given here for planning purposes only. Before commencing
fabrication, reference should always be made to the current issue of the appropriate drawings -
contact Micro Motion for details.
2.3.3 Free stream installation - flanged fitting
Conditions:
Flow: 0.3 to 0.5 m/s (at the meter)
Viscosity: 0.5 to 12,500 cP
Temperature: -50 °C to 200 °C (–58 °F to 392 °F)
[-40 °C to 200 °C (-40 °F to 392 °F) in hazardous areas]
Note: The thermal mass of the flanges may affect the response time of the meter to temperature
changes.
The view shown below is schematic to show the dimensions of the side pocket, which is fabricated by
the end user.
Bubbles rise!
Solids sink!
the slot
must be
vertical
.
For ALL pipe and flow directions
.
the meter
must be
horizontal
Installation and Configuration Manual 11
Installation (Short Stem)
Installation (Short Stem) Electrical ConnectionsInstallation (Long Stem)Introduction
The pocket geometry must be consistent with 2" schedule 40 tube in both internal diameter and
minimum wall thickness, such as:
Internal diameter: 2" (52.5 mm)
Wall thickness: minimum 0.15" (3.912 mm)
Weld neck or slip-on flanges may be used, according to the flange rating selected. However, for
higher rated flanges, only slip-on flanges may give the necessary clearances.
2.3.4 Free stream installation - weldolet
This is the preferred option where temperature variations are a critical factor. The reduced thermal
mass of the weldolet's taper-lock fitting renders it more able to track rapid changes in temperature.
Conditions:
Flow: 0.3 to 0.5 m/s (at the meter)
Viscosity: 0.5 to 12,500 cP
Temperature: -50 °C to 200 °C (–58 °F to 392 °F)
[-40 °C to 200 °C (-40 °F to 392 °F) in hazardous areas]
The weldolet has a 1.5" taper lock fitting, and is supplied to be welded on 4", 6", 8" or 10" pipelines.
Use of the weldolet ensures that the tines of the 7827 meter are orientated correctly and are fully
inserted into the fluid stream.
Before fitting the weldolet, the pipeline must be bored through at 2.1" (52.5 mm) diameter to accept
the viscometer. The weldolet must be welded to the pipeline concentrically with the pre-bored hole.
The view shown below is a schematic to show the relevant dimensions.
circlip
4” or larger;
horizontal
6” or larger;
vertical
2.75”
(70 mm ±2 mm)
7.75”
(197 mm)
Free Stream; flanged
2” (52.3 mm)
wall thickness
at least
0.15” (3.912 mm)
2” Schedule 40
4.37”
(111 mm)
0.47” (12 mm)
PTFE
ring
circlip
PTFE
ring
circlip
PFA
ring
circlip
12 Micro Motion 7827 Digital Viscosity Meter
Installation (Short Stem)
Figure 2-2 Free stream 1.5" Swagelock fitting
The installation will conform generally to Schedule 40 pressure ratings. The weldolet fabrication is
rated to 100 Bar at ambient temperature.
Note: Correct installation and pressure testing of the fitting is the responsibility of the user.
2.3.5 T-piece installation
Conditions:
Flow: 0.5 to 3.0 m/s (at the pipe wall)
Viscosity: 0.5 to 100 cP
Temperature: -50 °C to 200 °C (–58 °F to 392 °F)
[-40 °C to 200 °C (-40 °F to 392 °F) in hazardous areas]
The thermal mass of the flanges may affect the response time of the meter to temperature changes.
Flow velocity at the pipe wall and fluid viscosity must be within the limits shown to ensure that the
fluid within the pocket is refreshed in a timely manner. This installation will not respond as rapidly as
the free-stream installation to step changes in viscosity.
The view shown is a schematic to show the dimensions of the side pocket, which is fabricated by the
end user.
10” (254 mm)
Horizontal: 4” or larger
Vertical: 6” or larger
Weld
4.4” (111 mm)
Free stream weldolet
to suit pipe diameter
(4, 6, 8 or 10” N.B.)
2.1” (52.5 mm)
min
Installation and Configuration Manual 13
Installation (Short Stem)
Installation (Short Stem) Electrical ConnectionsInstallation (Long Stem)Introduction
The pocket geometry must be consistent with 2" schedule 40 tube in both internal diameter and
minimum wall thickness, i.e.:
Internal diameter: 2.1" (52.5 mm)
Wall thickness : minimum 0.15" (3.912 mm)
Weld neck or slip-on flanges may be used, according to the flange rating selected. However, for
higher rated flanges, only slip-on flanges may give the necessary clearances.
2.3.6 Flow-through chamber installation
Flow-through chambers are fabricated by Micro Motion, and are available with either weld prepared
ends or with flange or compression fittings for connection into the process pipe lines. They are
available with 1" NB, 2" NB, or 3" NB inlet and outlet pipes.
Note: The length of the inlet and outlet pipes must not be altered, otherwise the temperature response
and stability of the fitting may be adversely affected.
Conditions:
Flow: constant, between 10 and 30 l/min for 2" sch 40 calibration bore section, 5–300 l/min
for 3" sch 80 calibration bore.
Viscosity: 0.5 to 1000 cP
Temperature: -50 °C to 200 °C (–58 °F to 392 °F)
[-40 °C to 200 °C (-40 °F to 392 °F) in hazardous areas]
Pressure: 70 bar @ 204 °C, subject to process connections.
The PT100 is a direct insertion type, without a thermowell, and uses a ¾" Swagelok connection.
The diagram below shows an example of this type of standard installation.
“T” piece Flanged
D
4” or larger;
horizontal
or vertical
6.9”
(175 mm ±2
mm)
7.75”
(197 mm)
2” (52.3 mm)
wall thickness
at least
0.15” (3.912 mm)
2” Schedule 40
PFA
ring
circlip
0.47” (12 mm)
4.37” (111 mm)
14 Micro Motion 7827 Digital Viscosity Meter
Installation (Short Stem)
The three compression fittings on the flow pockets (½" drain, ¾" temp probe, and 1-½" mounting nut
for the meter) are rated to above the working pressure of the flow pocket. The fittings may be
Swagelok or Parker; both are used in manufacture.
The fittings are certified to the following standards:
Swagelok: SO9001 / 9002, ASME,TUV,CSA,DNV
Parker: ISO 9001 / 9002, TUV, DNV, LLOYDS
2.4 Installation in the pipeline or system
Viscosity is a highly sensitive indicator of change in a fluid – a key reason why viscosity
measurement is increasingly being chosen as a process measurement.
This sensitivity means that the measurement can be very sensitive to extraneous effects and therefore
great care must be taken to consider all the factors which affect measurement when assessing the
installation requirements.
Like many other meters, the optimum performance of the viscometer depends upon certain conditions
of the fluid and configuration of the process pipe-work. By introducing appropriate flow conditioning,
the optimum performance of the 7827 meter can be achieved at any chosen location in the process
system.
Dimensions shown in inches (mm)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76