2. Preparation
2.1. Before using the scale
1. Please keep the scale in a cool and dry place. Do not store at high temperatures.
2. Do not allow any liquids to come into or contact with the scale. If this has happened, wipe
the scale dry with a cloth.
3. Avoid objects impacting with the scale. Do not drop loads onto the scale or subject the
weigh pan to any strong shock loads. Do not leave loads on the scale for extended periods.
4. The load placed on the platform must not exceed the maximum weighing capacity of the
scale. The LCD display will blank if the scale is overloaded.
5. If the scale is not going to be used for some time, please clean it and store it in a plastic
bag in dry conditions. A desiccant sachet may be included to prevent moisture build up.
2.2. Preparing the scale for use
1. Unpack the scale carefully and place it on a firm surface. Turn the scale over and remove
the single transit locking screw located underneath if it is present.
2. For accurate weight readings locate the scale on a firm level surface free from vibrations
and adjusted to be level with the adjusting feet. The installer can use a builders level or
similar on top of the platform. The bubble should be inside the black marks.
3. Install the indicator (display) in a convenient position by laying it flat or unhinging the stand
recessed into the back of the display for desk mounting. It is possible to disconnect the AC
adapter and operate the scale on the internal battery so that the PT520 becomes truly
portable.
4. Avoid operating the scale in direct sunlight or drafts of any kind.
5. Remove any weight that might be on the platform before the scale is switched on and avoid
leaving weight on the platform for long periods of time.
6. Once the scale has been switched on, it will go through a LCD display test and then re-
zero to be ready for use.
7. For best accuracy switch the scale on for 1~2 minutes before use.
8. Please note that if the battery is flat the indicator will power off and the battery needs to be
changed or the mains adapter used.
9. All goods weighed should be placed in the centre of the platform for accurate weighing.
The overall dimensions of the goods being weighed should not exceed the size of the
weigh pan.
10. The acceleration due to gravity varies around the world. The force on the scale is a product
of the gravity value and the mass placed on the scale to weigh. As the gravity value can
vary with different world locations so can the weight displayed for the same mass. The
Scales have been calibrated for a gravity value (acceleration due to gravity) of g=9.799 m/
s
2.
Discrepancies in readings at the location of installation could be due to a different
gravity value and adjustment of the scale calibration or gravity value may be needed.