Page 13 / 45
REV. STV20110519
5- Secure clamp around actuator and valve body. For 51 (2”) and 76 (3”) secure the 4 bolts that
attach the actuator to the valve body. Tighten the bolts by using a wrench.
6-Actuate the manual actuator to the closed position by turning the wheel in the opposite
direction.
7- For valves with PTFE diaphragms make sure support gasket/o-rings are installed with the
diaphragm. See figure 14. See additional installation procedures for PTFE diaphragms in section
2.2.3.
2.2.3 Additional Procedure for Installation of PTFE Diaphragms
In order to ensure proper sealing of PTFE diaphragms, it is recommended to heat-treat them
before use. Follow these steps for the heat-treatment procedure:
1- Expose the installed diaphragm to steam at around 121° C for approximately 30 minutes. The
valve should be in the closed position, or closed directly after exposure. Alternatively, the
installed diaphragm can be exposed to hot air at around 100° C for approximately one hour.
2- Re-tighten the clamp (1/2”-1.5” sizes) or bolts (2” and 3” sizes), after the valve has cooled
down to ambient temperature.
3- Repeat this process each time the valve insert (actuator and diaphragm) is disassembled and
reassembled.
2.2.4 Position Indicators
An assortment of models and styles of instrumentation are available to sense the actuator
position. Listed below are some of the most common. Please consult Steridose to determine
the best solution to your particular application.
2.2.4.1 Stainless Steel Pneumatic Actuators
For valves 12 (1/2”), 25 (1”), and 38 (1-1/2”)
To accommodate a wide variety of position indicator solutions various adapters may need to be
used on particular assemblies. Please consult with Steridose representatives to determine overall
dimensions and specifications.
The sensors available are:
Open, micro-switch
Open/Close, micro-switch
Open, inductive proximity switch