Tettex MIDAS 2881 Operating Instructions Manual

Category
Measuring, testing & control
Type
Operating Instructions Manual
Operating Instructions
HAEFELY TEST AG
MIDAS 2881
Mobile Insulation
Diagnosis &
Analysing System
Version 2.0
(LAN version only)
4843930
Art.Nr. 4843930 Operating Instructions MIDAS 2881
Date Version Responsible Changes/Reasons
04.16 2.0 SG 2881 LAN Version,
replace the firewire model
WARNING
Before operating the instrument, be sure to read and fully
understand the operating instructions. This instrument
produces hazardous voltages. It is the responsibility of the user
to ensure that the system is operated in a safe manner.
This equipment contains exposed terminals carrying hazardous
voltages. There are no user serviceable components in the unit.
All repairs and upgrades that require the unit to be opened must
be referred to HAEFELY TEST AG or one of their nominated
agents.
This Manual is only for Midas 2881 with Ethernet (LAN)
connection to computer, for those equipped with FireWire
connection type refers to the corresponding manual.
HAEFELY TEST AG and its sales partners refuse to accept any responsibility for consequential or
direct damage to persons and/or goods due to none observance of instructions contained herein or
due to incorrect use of the equipment.
Further be aware that safety is the responsibility of the user!
Any correspondence regarding this instrument should include the exact type number, instrument
serial number and firmware version number. With the exception of the firmware version number, this
information can be found on the registration plate on the right panel of the instrument. The firmware
version specified in the miscellaneous folder under the setup menu.
Unauthorized opening of the unit may damage the EMI protection of the system and will reduce its
resistance to interference and transients. It may also cause the individual unit to be no longer
compliant with the relevant EMC emission and susceptibility requirements. If the unit has been
opened, the calibration will be rendered invalid.
Note
HAEFELY TEST AG has a policy of continuing improvement on all their products. The design of this
instrument will be subject to review and modification over its life. There may be small discrepancies
between the manual and the operation of the instrument, particularly where software has been
upgraded in the field.
HAEFELY TEST AG retains the right to modify the functionality, specification or operation of the
instrument without prior notice.
All rights reserved. No section of this manual may be reproduced in any form, mechanical or
electronic without the prior written permission of HAEFELY TEST AG.
2016, HAEFELY TEST AG, Switzerland
Manual Conventions
In the manual, the following conventions are used:
Indicates a matter of note.
If it refers to a sequence of operations, failure to follow the instructions could result in errors in
measurement.
Indicates hazards.
There is a risk of equipment damage or personal injury or death. Carefully read and follow the
instructions. Be sure to follow any safety instructions given in addition to those for the site at which
tests are being performed.
Foreword
Welcome as a new user of the Insulation Diagnosis System MIDAS 2881. Thank you for placing your
confidence in our product.
With the purchase of this measuring instrument you have opted for all the advantages that have built
a world-wide reputation for a Tettex Instrument: Robustness, performance and quality assured. As a
result this instrument provides a solution which achieves the optimal combination of traditional know-
how and leading edge technology.
This operating manual is designed for completeness and easy location of the required information.
Customers who already have experience with this kind of equipment will find this document to be of
assistance as an extended help. A keyword index at the end of the operating manual greatly eases
use.
If you find a mistake or inconsistency in the operating manual then please feel free to inform our
Customer Support department with your corrections so that other users may benefit.
Abbreviations, definitions
Wherever possible the corresponding IEC definitions are used. The following abbreviations and
definitions are used in this manual:
CN Standard capacitor (Measurement reference, built-in the instrument)
CX Test object capacitance (e.g. power transformer, generator, motor etc.)
HV High voltage
cos Power Factor
PF Power Factor
tan Dissipation factor
DF Dissipation factor
DUT Device Under Test (Test Object)
ppm Parts per million
Introduction I
Contents
1Introduction 1
1.1Receiving Instructions ..................................................................................... 1
1.2General ........................................................................................................... 1
1.3Hardware ........................................................................................................ 1
1.4Software ......................................................................................................... 2
1.5Scope of Supply .............................................................................................. 2
1.6Optional Accessories ...................................................................................... 3
2Technical Data 4
3Safety 6
3.1General ........................................................................................................... 6
3.2Personnel Safety ............................................................................................ 6
3.3Safety Features .............................................................................................. 7
3.4Safety Precautions .......................................................................................... 7
3.5Summary ........................................................................................................ 8
4Theory 9
4.1Why is Insulation Tested? ............................................................................... 9
4.2What is Loss Factor? ...................................................................................... 9
4.3What is Dissipation Factor tan ? ................................................................. 10
4.4The Difference between Power Factor and Dissipation Factor ..................... 11
4.5Apparent Power, Real Power, Reactive Power ............................................. 11
4.6Test Instruments ........................................................................................... 12
4.7Evaluation of Test Results ............................................................................ 12
4.8Supplementary Test Methods ....................................................................... 16
4.9Standard Capacitor, Measuring Current & Limits .......................................... 16
4.10Parallel & Series Equivalent Circuits ............................................................. 17
5Functional Description 18
5.1System Overview .......................................................................................... 18
5.2V-potential point and Guarding ..................................................................... 19
5.3Test Modes ................................................................................................... 21
5.4Interference Suppression .............................................................................. 23
6Operation Elements 24
6.1Individual Parts ............................................................................................. 24
6.2Laptop as System Controller......................................................................... 25
6.2.1Interfaces ...................................................................................... 25
6.3Computer connection .................................................................................... 25
6.4Instrument Side Panel .................................................................................. 26
6.4.1Measuring Inputs .......................................................................... 26
6.4.2High Voltage and Power Outputs .................................................. 27
6.5HV GND Connection Surveillance ................................................................ 29
II Introduction
6.6Safety Ground (Earthing) .............................................................................. 29
6.7Emergency Stop ........................................................................................... 30
6.8Warning Lamp Bar ........................................................................................ 30
7Software 31
7.1General ......................................................................................................... 31
7.1.1Start-up ......................................................................................... 31
7.1.2Main Window ................................................................................ 32
7.1.3Title bar ......................................................................................... 33
7.1.4Alarm Messages ........................................................................... 34
7.1.5Function Keys ............................................................................... 35
7.2File Manager ................................................................................................. 35
7.2.1File Selector Dialog ....................................................................... 36
7.2.2Report ........................................................................................... 37
7.2.3Data Format Files ......................................................................... 39
7.3Display of Measurement Values ................................................................... 39
7.4Tab sheet SETUP ......................................................................................... 43
7.4.1Menu DUT Info .............................................................................. 44
7.4.2Menu Conditions (Temperature correction)................................... 45
7.4.3Menu Settings ............................................................................... 47
7.4.4Menu Options ................................................................................ 50
7.4.5Menu Auxiliary .............................................................................. 53
7.5Tab sheet MANUAL ...................................................................................... 55
7.5.1Tools Window ............................................................................... 57
7.5.2Definition of Columns for Measuring Spreadsheet ........................ 58
7.5.3Formulas in Measuring Grids ........................................................ 59
7.5.4Menu Signal Analysis .................................................................... 61
7.6Tab sheet SEQUENCE ................................................................................. 65
7.6.1Definition of Spreadsheet Sequence ............................................. 65
7.6.2Sequence Measurement ............................................................... 68
7.6.3Edit Sequence Limiters ................................................................. 70
7.6.4Starting Sequence ........................................................................ 72
7.6.5Sequence with External AC Power Source ................................... 74
7.7Tab sheet ANALYSIS ................................................................................... 75
7.7.1Spreadsheet Measurement ........................................................... 75
7.7.2Graphic Analysis ........................................................................... 76
7.7.3More Analysis ............................................................................... 77
7.8Remote Operation ........................................................................................ 78
7.8.1Characteristics of the interface ...................................................... 79
7.8.2General Commands ...................................................................... 82
7.8.3System control commands ............................................................ 84
7.8.4Measurement commands ............................................................. 86
7.8.5Alarms ........................................................................................... 88
8Accessories and Options 90
8.1Office Software ............................................................................................. 90
8.2Safety Strobe Light ....................................................................................... 90
8.3Oil Test Cell .................................................................................................. 90
8.4Resonating Inductor ...................................................................................... 91
8.4.110kV Resonating Inductor 5288A ................................................. 92
8.4.215kV Resonating Inductor 5289 .................................................... 95
8.5Current Booster .......................................................................................... 102
8.6Care and Maintenance ............................................................................... 104
Introduction III
9Instrument Storage 105
10Packing and Transport 106
11Recycling 107
12Trouble Shooting 108
12.1Software Updates ....................................................................................... 108
13Conformity 109
Appendix 110
14Applications Guide 111
14.1Bushings ..................................................................................................... 111
14.1.1Spare Bushings ........................................................................... 113
14.1.2Installed Bushings ....................................................................... 114
14.1.3Measuring Data Interpretation ..................................................... 118
14.2Transformers .............................................................................................. 119
14.2.1Power and Distribution Transformers .......................................... 119
14.2.2Shunt Reactors ........................................................................... 125
14.2.3Current Transformers .................................................................. 126
14.2.4Voltage Transformers ................................................................. 127
14.2.5Short Circuit Impedance ............................................................. 128
14.2.6Excitation Current Measurement ................................................. 130
14.3Rotating Machines ...................................................................................... 132
14.3.1Test Procedure ........................................................................... 132
14.3.2Measuring Data Interpretation ..................................................... 133
14.3.3Measuring high cap values using the Resonating Inductor ......... 135
14.3.4Operation of Resonating Inductor with Midas ............................. 136
14.4Liquid Insulation .......................................................................................... 137
14.4.1Test Procedure ........................................................................... 137
14.4.2Measuring Data Interpretation ..................................................... 138
14.5Cables ........................................................................................................ 139
14.5.1Test procedures on different cables ............................................ 139
14.5.2Test Procedure Example ............................................................ 140
14.5.3Measuring Data Interpretation ..................................................... 141
14.6Capacitors .................................................................................................. 141
14.6.1Measuring high cap values using the Resonating Inductor ......... 142
14.7Circuit Breakers .......................................................................................... 143
14.7.1Dead Tank Breaker ..................................................................... 145
14.7.2Live Tank Breaker ....................................................................... 146
14.7.3Measuring Data Interpretation ..................................................... 147
14.8Surge (Lightning) Arresters ......................................................................... 147
14.8.1Test Levels .................................................................................. 147
14.8.2Test Procedures .......................................................................... 148
14.8.3Measuring Data Interpretation ..................................................... 150
Index 151
Introduction 1
1 Introduction
1.1 Receiving Instructions
When taking delivery, any possible transport damage should be noted. A written record should be
made of any such damage. A suitable remark should be recorded on the delivery documents.
A claim for damage must be reported immediately to the transport company and to the Customer
Support Department of HAEFELY TEST AG or the local agent. It is essential to retain the damaged
packing material until the claim has been settled.
Check the contents of the shipment for completeness immediately after receipt (See chapter "Scope
of Supply“). If the shipment is incomplete or damaged then this must be reported immediately to the
transport company and the Customer Support Department of HAEFELY TEST AG or the local agent.
Repair or replacement of the instrument can then be organised immediately.
1.2 General
The MIDAS provides determination of the capacitance and dielectric loss of liquid and solid insulation.
Measurements can be carried out on solid insulation such as cables, capacitors, power transformers,
generators, motors, bushings etc. and on insulating oil with an optional oil test cell.
Operation is achieved via a sunlight readable touch screen and offers optimal user friendliness.
Operation is simple thanks to the user dialogue system and on-line help.
The system is ideal for high and low voltage measurements over a wide frequency range. The test
setup has been specially developed for efficient use in maintenance measurements, production and
quality control. Thanks to it’s high precision, the system is also ideal for laboratory and development
use.
Once a measurement on a specific device is done I can be recalled and repeated in the same way
and results can be compared graphically. So the system shows fast and easy a trending analysis of
your equipment under test.
1.3 Hardware
The measuring instrument is fully automatically balanced and the measurement values are calculated
and displayed. The measuring instrument is provided with two measurement inputs, HV Supply, HV
Ground and Safety Ground. Over 20 various parameters can be measured respectively calculated.
The instrument , as a double vector meter, recognises the type of test object ( inductive / capacitive)
and determines and displays its values automatically.
Advanced noise reduction is provided for field measurements where the measurement results might
otherwise be falsified due to interferences.
The combination of the measuring instrument with built-in high voltage supply, standard capacitor CN
and all suitable connection cables provides the user with a complete measurement system.
2 Introduction 2
1.4 Software
With the powerful operating software manual measurements and automated test sequences can be
defined and executed. The user simply enters the desired high voltage values and frequencies. The
measurement values are then automatically read and displayed on the computer screen in curve and
tabular format.
With the occasion to integrate bitmaps, comments and hints, a guided step by step measurement
macro can be done. So it is possible to grant a reproducible defined measurement procedure.
1.5 Scope of Supply
The standard scope of supply includes the following items:
Qty Description
1 MIDAS Instrument trolley with measuring hardware and HV supply *
* MIDAS standard type supplies 12kV,
* MIDAS G type supplies 15kV
Control Computer
1 Fully installed Laptop in shell case (MIDAS 2881)
Cables in rugged trolley including
Shielded measuring cable, unipolar, LEMO plug, 20m, with clamp, indicator blue
Shielded measuring cable, unipolar, LEMO plug, 20m, with clamp, indicator white
HVGND measuring cable, unipolar, LEMO plug, 20m, with clamp, indicator yellow
HV supply cable, unipolar, double shielded, 20m, yellow
Safety Ground cable, 20m, with locking pliers, yellow/green
Country-specific mains cord, 2P & E, 10A , 2m
HV connection hook, used with HV supply cable
HV connection clamp, used with HV supply cable
Handheld Safety Switch with cable, LEMO plug, 10m
Small set of accessories
1 Operating Instruction (this manual) and Test Certificate
1 USB Memory Stick with software and manual in electronic format
Introduction 3
1.6 Optional Accessories
This chapter describes the accessories that can be used in conjunction with the MIDAS. Contact us
directly if you have a special application, as the following list is only a part of the broad range of
accessories available.
Type / Series Description
288x TEMP Laser infrared, contact-less Thermo/Hygrometer. For determination of: tank(oil)
temperature, air temperature and air humidity.
288X CASE Rugged field case for transportation of MIDAS
288X RACK Mechanical kit for 19” rack mounting of MIDAS
288x SAFE Safety Strobe Light with magnetic base for mounting e.g. on a transformer tank,
providing additional visual warning of high voltage presence.
MIDAS OFFICE MIDAS software for customer personal office PC. Used for data visualisation,
staff education, test planning and preparation, reporting.
5287 Current Booster to increase test current (while voltage decreases), especially for
short circuit impedance testing of power transformers to diagnose transformer
winding deformation.
5288A 10kV Resonating inductor (creates LC parallel resonant circuit) to increase test
current up to max. 4.4A. Used for testing of high capacitance values up to 1uF
5289 15kV Resonating inductor (creates LC parallel resonant circuit) to increase test
current up to max. 6A. Used for testing of high capacitance values up to 1.56 uF
6835 Mobile Test cell for on-site maintenance measurements on liquid insulations
(10kV max.)
4 Technical Data 4
2 Technical Data
HV Power Supply
MIDAS 2881 MIDAS 2881G
Output Voltage Up to 12 kV Up to 15 kV
Output Voltage Regulation 1 % rdg 1 V 1 % rdg 1 V
Output Frequency @ Nom Volt 45Hz …. 70 Hz 45Hz …. 70 Hz
Output Frequency @ 5kV 15 Hz … 1000 Hz 15 Hz … 1000 Hz
Output Current
1
@ Nom Volt 115 mA continuous , > 275mA max 1 min. 115 mA continuous , > 275mA max 1 min.
Output Current @ 10kV 165 mA continuous , > 400mA max 1min. 165 mA
continuous , > 400mA max 1 min.
Output Power
1
max > 4000 VA > 4000 VA
Output PD level
2
max 500 pC 500 pC
Duty cycle 1650VA 1650 .. 2000 VA 2001 .. 3000 VA >3001
continuous 30 min. ON / 1h 5 min. ON / 1h 1 min. ON / 1h
1 Can be expanded with optional Resonating Inductor
2 Can be reduced with additional filters on coupling capacitor
Measuring Unit (MIDAS 2881 & MIDAS2881G)
Range Resolution Accuracy
Test Voltage < 1MV 1 V 0.2 % rdg 1 V
Dissipation Factor (tan ) 0 … 10’000.% 0.001% 0.5 % rdg 0.01.%
Power Factor (cos ) 0 ... 100.% 0.001% 0.5 % rdg 0.01.%
Quality Factor 0.01 ... 10000 0.0001 0.5% rdg 0.0001
Capacitance 0.01pF
0.2 % rdg 0.2pF
Inductance 0.1mH
0.5 % rdg 0.5 mH
Test Current Input A, B & HVGND 20.uA … 15 A 0.1 uA 0.1 % rdg 1 uA
Ref Current Input Cn ext 20.uA … 300 mA 0,1 uA 0.1 % rdg 1 uA
Test Frequency 15 Hz … 1000 Hz 0.01 Hz 0.1 % rdg 0.1 Hz
Apparent Power S 1 MVA 0.1 mVA 0.8 % rdg 1 mVA
Real Power P 1 MW 0.1 mW 0.8 % rdg 1 mW
Reactive Power Q 1 Mvar 0.1 mvar 0.8 % rdg 1 mvar
Accuracy @ frequency 15Hz…100Hz ; In & Ix > 20A ; Ix/In: 0.01 .. 10’000 ; With Internal Cn
Internal reference capacitor
Capacitance 100 pF
tan < 0.00002
Capacitance constancy < 0.01.% / year
Temperature coefficient < 0.01.% / K
Mains Power supply
Voltage / Frequency 90 VAC … 264 VAC / 50 … 60 Hz
Input Power 1 kW
Environmental
Operating Temperature -10 .. 50 °C
Storage Temperature -20 .. 70 °C
Humidity 5 .. 95 % r.h.
Protection classes IP22, IEC 61010, CE mark, general IEC 61326-1,
IEC 61000-4-X, 61000-3-X, EN 55011, ANSI/IEEE C37.90
Safety Specification VDE 0411/part 1a , IEC/EN 61010-1:2002
Technical Data 5
Mechanical
Weight & Dimensions Instrument 58 kg (128 lbs) 34 x 47 x 104 cm (13.5” x 18.5” x 41”)
Trolley 11 kg (25 lbs) 33 x 68 x 112 cm (13” x 26.8” x 44”)
General
Control unit Lap top Computer Windows 7 based.
Measuring unit With Internal and external Power supply and nominal capacitor
Interfaces Ethernet connection to external laptop
Recorded Values DF(tan ), DF(tan )@20°C , DF%(tan ), DF%(tan )@20°C , PF(cos ), PF(cos
)@20°C , PF%(cos ), PF%(cos )@20°C, QF (quality factor), QF (quality factor)
@20°C CP (ZX= CP RP), RP (ZX= CP RP), CS (ZX= CS + RS), RS (ZX= CS + RS) ,
LS (ZX= LS + RS), RS (ZX= LS + RS), LP (ZX= LP RP), RP (ZX= LP RP),
Standard capacitor Cn, URMS, URMS 3, ITest eff, IRef eff, Im, IFe, Impedance Zx,
Phase-angle (Zx), Admittance Yx, FrequencyTest, FrequencyLine , App. Power
S, Real Power P, Reactive Power Q, Real Power @2.5 kV, Real Power @10
kV, TemperatureAmbient 5, TemperatureInsulation 5 , Rel. Humidity 5,
Temp.Corr.Factor K, Connection mode, Settings, all Notes and Comments,
Time, Date
Software Windows 7
Data format XML, CSV
5 measured by external temperature/humidity probe
6 Safety 6
3 Safety
This warning sign is visible on the MIDAS. Meaning: This equipment should only be operated after
carefully reading the user manual which is an integral part of the instrument.
Haefely Test AG and its sales partners refuse to accept any responsibility for consequential or direct
damage to persons and/or goods due to none observance of instructions contained herein or due to
incorrect use of the MIDAS.
Further be aware that Safety is the responsibility of the user !
Remember - Hazardous voltage can shock, burn or
cause death !
3.1 General
Safety is the most important aspect when working on or around high voltage electrical equipment.
Personnel whose working responsibilities involve testing and maintenance of the various types of high
voltage equipment must have understood the safety rules written in this document and the associated
safety practices specified by their company and government. Local and state safety procedures
should also be consulted. Company and government regulations take precedence over Tettex
recommendations.
The MIDAS generates high voltage and is capable of causing serious even lethal electrical shock. If
the instrument is damaged or it is possible that damage has occurred, for example during
transportation, do not apply any voltage.
The instrument may only be used under dry operating conditions. The use of MIDAS is prohibited in
rain or snow.
Do not open the MIDAS, it contains no user replaceable parts.
Do not switch on or operate a MIDAS instrument if an explosion hazard exists.
3.2 Personnel Safety
The MIDAS should not be operated by a crew smaller than two people. Their function can be
described as follow:
Test Operator The person who is making the test connections and operates the MIDAS. He must be
able to have a clear view of the device under test and the area where the test is performed.
Safety Observer The person who is responsible for observing the performance of the test, seeing
any safety hazard, and giving warning to crew members.
Both persons should perform no other work while the MIDAS is energized.
While making the various types of connections involved in the different tests, it may be necessary for
personnel to climb up on the equipment, but no one should remain on the equipment during the test
itself.
Safety 7
Non-test related persons who are working in proximity to the area where testing is performed must be
informed. Consistent visual and verbal signals should be agreed and followed.
Perform only one job at a time on any equipment. The situation in which two crews are doing different
tasks with the same equipment at the same time is an open invitation for confusion, trouble, and
danger to the personnel.
People with heart pacemakers should not be in the vicinity of this system during operation.
3.3 Safety Features
Beside an Emergency Stop switch the MIDAS is equipped with an external Safety Switch (spring-
release type or a 'dead man' type). The Safety Switch should be controlled by the second test crew
member (safety observer). Without the Safety Switch the instrument can not be activated.
Prior to making the first measurements, the Safety Switch operator should verify the correct
operations of the switch.
It is recommended that the Safety Switch be the last switch closed. It must remain open until all
personnel are safely in the clear. If unauthorized personnel should enter the area, or if some other
undesirable situation should develop, the Safety Switch operator should release the switch
immediately, and then notify the MIDAS operator.
The Safety Switch should be used at all times. Never short circuit it and do not use fixed mechanical
locking devices for depressing the switch button. The switch button must be manually operated at all
times.
For visual warning of high voltage presence a warning lamp bar is located on the top rear side of the
instrument. Optional a strobe light is delivered which can be mounted on the device under test.
The MIDAS is equipped with a HV GND connection surveillance. The high voltage can only be
switched on when the earth circuit is properly connected. The instrument indicate the status
“Grounded” or “Open” by LED and by software.
A separate green/yellow earth cable is provided for the purpose of safety grounding the instrument.
The earth cable should be connected to the Earthing Screw on the back of the MIDAS at one end and
to the station grounding system at the other end.
The green / yellow safety ground cable should be the FIRST lead to be connected to the set.
3.4 Safety Precautions
All tests must be performed with the device under test completely de-energized and isolated from its
power systems. The equipment, its tank or housing must be disconnected from all buses and properly
earthed, so that all induced voltages or trapped charges are neutralized. Only when the measurement
procedure is actually being performed the grounds should be temporarily removed.
The MIDAS must be solidly earthed with the same ground as the device under test. When the MIDAS
is permanently housed in a vehicle, the MIDAS ground should be bounded to the vehicle chassis,
which in turn is grounded.
Exposed terminals of equipment should not normally be allowed to 'float'. They should be grounded
directly or through the low voltage leads (INPUT V) of the MIDAS, unless otherwise specified.
Testing of high voltage equipment involves energizing the equipment through the MIDAS. This can
produce dangerous levels of voltage and current. Care must be taken to avoid contact with the
equipment being tested, its associated bushings and conductors, and with the MIDAS cables.
Especially the high voltage test cable should not be held during energization of the MIDAS. Flashover
of the test specimen or the MIDAS can generate transient voltages of sufficient magnitude to puncture
the insulating jacket of the high voltage test cable.
8 Safety 8
It is strongly recommended that the test crew make a visual check to ensure that the equipment
terminals are isolated from the power system. If there is real possibility that the device under test fails
precautions such as barriers or entrance restrictions must be taken against harm in the event of
violent failure.
Proper clearance between the test equipment and the device under test must be ensured during the
presence of high voltage. Barriers and safety tapes can be established around the test area to prevent
unintentional entry into the live area. It must also be guaranteed that extraneous objects like ladders,
buckets, etc. can not enter the test area.
After the MIDAS is properly grounded, the remaining test leads and the High Voltage Test Cable are
plugged into their receptacles. Do not connect test leads to the equipment terminals until after
the leads are connected to the MIDAS.
The proper procedures for connecting the MIDAS leads to the device under test is described in detail
in chapter "Accessories and Options" The safety observer should supervise this procedure at all
times.
The MIDAS operates from a single-phase power source. It has a three wire power cord and requires
a two-pole, three terminal, live, neutral and ground type connector. Do not bypass the grounding
connection. Any interruption of the grounding connection can create electric shock hazard. The
power input connection should be the last step in setting up the instrument.
After the tests are completed, all test leads should be disconnected first from the device under
test and earthed before they are disconnected from the instrument.
The green / yellow safety ground cable should be the LAST lead to be disconnected from the
set.
Do not disconnect the voltage cables from unless the MIDAS Voltage is set to HV OFF, and the
Safety Switch is released. Attempts to disconnect leads while the MIDAS is energized may
result in a serious and possibly lethal electrical shock.
3.5 Summary
Note: Many accidents that happen around high voltage equipment involve personnel who are familiar,
and perhaps too familiar, with high voltage equipment. Staying alert and ever watchful requires
constant training and awareness of the inherent hazards. The greatest hazard is the possibility of
getting on a live circuit. To avoid this requires constant vigilance - for oneself and for one's fellow
workers.
In addition to the obvious dangers, personnel should be alert to recognize subtle dangers as well. For
example, during transformer excitation-current tests, the floating terminals may have significant
voltages induced in them by simple transformer action. Therefore, all terminals of a device under
test, unless grounded, should be considered to be live while the test is in progress.
When potential transformers or any transformers are interconnected, voltage can be back-fed through
the secondary windings to produce high voltage on the primary although the primary is seemingly
isolated from the power system. This entail a second important rule - all terminals of a device under
test should be completely isolated.
Finally it should be noted that the MIDAS is relatively heavy. We recommend that at least two people
are used to slide the MIDAS and three to lift it. Special care must be taken in lifting or sliding the
Instrument into or from a vehicle so as not to incur bodily injury.
Remember - Safety, FIRST, LAST, ALWAYS !
Theory 9
4 Theory
4.1 Why is Insulation Tested?
All transformers, high voltage switchgear, motors and electrical equipment accessories have a high
voltage lifespan. From the first day of use the equipment is subject to thermal and mechanical
stresses, foreign particle ingress and variations in temperature and humidity. All of these influences
raise the working temperature of the equipment when switched on. This heating accelerates chemical
reactions in the electrical insulation, which result in a degradation of the dielectric characteristics. This
process has an avalanche character i.e. the changing electrical characteristics of the insulation
increase the loss factor and produce heating which further degrades the insulation. If the loss factor
of the insulation is periodically monitored and recorded, it is possible to predict and / or avoid
catastrophic failure of the electrical equipment.
At the beginning of the public electricity supply industry, methods and processes were sought to avoid
unexpected losses caused by equipment defects. One method that provided repeatable data and
offered simple on-site measurement was the measurement of capacitance and loss factor (power
factor) of the equipment insulation.
In cases where loss factor tests were regularly carried out, and the relevant test results compared with
earlier results, the deterioration of the insulation was noted and necessary preventative measures
were carried out. Based on this groundwork, a series of test procedures were developed and
described in various IEEE, ANSI and IEC documents and standards to specify the insulation quality
for various types of electrical equipment.
In order to define acceptable loss factor values, a „data record service“ was developed based upon
statistical data that related to specific equipment types and models. Standard measurements of
capacitance and loss factor of the electrical insulating medium were carried out to ensure that the data
was comparable. The loss factor was calculated and the results were corrected by energy
comparison to a value for a test voltage of 10kV. Some test results were further multiplied by a
temperature correction factor to produce 20°C compatible values. Any results that are now acquired
at different Test Levels and temperatures are recalculated for 10kV and / or 20°C and then recorded
and compared. In this way the degradation of the insulation characteristics over a specified period of
time can be determined. With the test result history an experienced engineer is able to take the
necessary maintenance actions based upon changes in the value of loss factor.
4.2 What is Loss Factor?
Loss factor is the total energy that will be used by the equipment during normal service. In particular,
the insulation loss factor is any energy that is taken by the flow of current through the resistive
component of the insulation. The earth path varies according to the type of electrical equipment. For
example, switchgear will probably develop tracking to earth at right angles to the floor connections. In
transformers paths can develop in the insulation resistance between the windings or between the
windings and housing (tank). In all cases the result is a loss factor in the form of heating.
Note: In this text loss factor (losses, watts) is referred to, in contrast with total loss factor. Total loss
factor is normally used to describe the total losses of the transformer under load and should not be
confused with the energy that is lost due to degradation of the insulation.
10 Theory 10
4.3 What is Dissipation Factor tan ?
To specify the insulation loss factor, the test object must be considered in the test arrangement as a
capacitor. Consider all test objects e.g. transformers, bushings, busbars, generators, motors, high
voltage switchgear etc. are constructed from metal and insulation, and therefore possess associated
capacitive properties. Every test object consists of various capacitances together with the insulation
and the internal capacitance to earth. The figure shows the components that comprise a capacitance
and the diagram for a simple disc capacitor.
Figure 1- Disc Capacitor
d
A
C
where:
A electrode face
d distance between the electrodes
C capacitance
0 dielectric constant of air (0=8,8542•10-12 F/m)
r relative dielectric constant dependent upon
material
= 0r, dielectric constant
In an ideal capacitor the resistance of the insulation material (dielectric) is infinitely large. That means
that, when an AC voltage is applied, the current leads the voltage by exactly 90° as it flows as pure
current.
After further consideration it must be realized that every insulation material contains single free
electrons that show little loss under DC conditions with P= U2/R. Under AC a behaviour called
dielectric hysteresis loss occurs which is analogous to hysteresis loss in iron.
As losses therefore occur in every insulation material, an equivalent diagram of a real capacitance can
be constructed as follows:
Figure 2 - Parallel equivalent diagram of a lossy
capacitance with vector diagram
Loss factor (Dissipation Factor)
RCR
X
I
I
Q
PC
C
R
C
R
1
tan
Power Factor
PF I
I
P
S
RR
C

cos tan
tan
12
UTest applied test voltage
IC current through capacitance
IR current through resistance (insulating material)
C ideal capacitance
R ideal resistance
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164

Tettex MIDAS 2881 Operating Instructions Manual

Category
Measuring, testing & control
Type
Operating Instructions Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI