Tektronix 8.00E+05 User manual

Category
Music drums
Type
User manual

This manual is also suitable for

User Manual
80E01, 80E02, 80E03, 80E04 & 80E06
Electrical Sampling Modules
071-0434-05
This document applies to firmware version 1.00
and above.
www.tektronix.com
Copyright © Tektronix. All right s reserved. Licensed software products are owned by Tekt ronix or i ts subsidiaries or
suppliers, and are protected by national copyright laws and i nternational treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
TEKT RONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For produc t information, sales, service, and technical support:
H In North America, call 1-800-833-9200.
H Worldwide, visit www. tektronix.com to find contacts in your area.
Warranty 2
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year
from the date of shipment. If any such product proves defecti ve duri ng this warranty period, Tekt ronix, at its option, either
will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the
defective product. Parts, modules and replacement products used by Tektronix for warranty work may be new or
reconditioned to like new performanc e. All replaced parts, modul es and products become the property of Tektronix.
In order t o obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the
warranty peri od and make suitable arrangem ents for the pe rformance of service. Customer shall be responsible for
packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid.
Tektronix shall pay for the return of the product to Customer if the shipment is to a locati on within the count ry in which the
Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any
other charges for products returned to any other locations.
This warranty shall not apply to any defect, fail ure or damage c aused by improper use or im proper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting
from attempts by personnel othe r than Tekt ronix representatives to install, repair or service the product; b) to repair
damage resulting from improper use or connection to incompatible equipment ; c) to repai r any dama ge or malfunction
caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other
products when the effect of such modification or integrati on inc reases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER
WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED
WARRANTIES OF MERCHANT ABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’
RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY
PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL
NOT BE LIABLE FOR ANY INDIRECT , SPECIAL, INCIDENT AL, OR CONSEQUENTIAL DAMAGES
IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY
OF SUCH DAMAGES.
80E00 Electrical Sampling Modules User Manual
i
Table of Contents
General Safety Summary iii...................................
Environmental Considerations v...............................
Preface vii...................................................
Manual Structure vii................................................
Related Documentation vii...........................................
Getting Started 1............................................
Product Description 2..............................................
Options and Accessories 4...........................................
Options 4.....................................................
Standard Accessories 4..........................................
Optional Accessories 4..........................................
Installation 6.....................................................
Electrostatic Discharge 6........................................
Static Controlled Workstation 7...................................
Module Installation 8...........................................
Compensation 9...............................................
Operating Basics 11..........................................
Usage 11.........................................................
Front-Panel Controls 12..............................................
Signal Connector 12.............................................
Channel Selection 12............................................
TEKPROBE Connector 13........................................
TDR On Indic ator 13............................................
System Interaction 13...............................................
Commands From the Main Instrument Front Panel 13......................
Programmer Interface Commands 14...................................
User Adjustments 14................................................
Cleaning 14.......................................................
Reference 17.................................................
Taking TDR Measurement s 17........................................
TDR Measurements Background 27....................................
Finding the Velocity of Propagation and Locating Mismatches 30.........
TDR Measurement Units 30.......................................
Making Accurate TDR Measurements 32............................
Taking Differential and Common-Mode TDR Mea surements 32.............
Connector and Adapter Care Requirements 38...........................
Visual Inspection 38.............................................
Cleaning Connectors 39..........................................
Assembly and Torquing 41........................................
TDR Impedance Measuring 43......................................
Detecting Blown Inputs 43...........................................
EOS (Electrical Overstress) Prevention 44...............................
Specifications 47.............................................
Glossary 53.................................................
Index 55....................................................
Table of Contents
ii
80E00 Electrical Sampling Modules User Manual
List of Figures
Figure 1: Samplin g module block diagram 3.....................
Figure 2: Samplin g module, 80E04 shown 3.....................
Figure 3: Sampling module compartments 6.....................
Figure 4: Installing a sampling module 8........................
Figure 5: Samplin g module, 80E04 shown 12.....................
Figure 6: Vertical Setup dialog box 14...........................
Figure 7: Simplified schematic diagram of step
generator - positive polarity 18..............................
Figure 8: Step generator with a shorted output 19.................
Figure 9: Step generation with a 50 Ω load 19.....................
Figure 10: Step generation with an open circuit 19.................
Figure 11: TDR disp lays for typical loads 21......................
Figure 12: Microstrip discontinuities 27..........................
Figure 13: TDR waveform of microstrip in Figure 12 28............
Figure 14: TDR step and reflection (short) 29.....................
Figure 15: TDR step and reflection
(50 Ω line terminated in 75 Ω)31............................
Figure 16: TDR step of undamaged sampling module 44............
Figure 17: First example of EOS error 45........................
Figure 18: Second example of EOS error showing
cumulative effect 46.......................................
List of Tables
Table 1: Application software version required 1.................
Table 2: Sampling module features 2...........................
Table 3: Standard accessories 4...............................
Table 4: Optional accessories 4................................
Table 5: Torque Wrench Information 42.........................
Table 6: Electrical sampling modules -- Descriptions 47............
Table 7: Electrical sampling modu les -- Sign al acquisition 48.......
Table 8: Electrical sampling modu le (80E04) -- TDR system 50......
Table 9: Electrical sampling modules -- Timebase system 51........
Table 10: Electrical sampling modu les -- Power consumption 51.....
Table 11: Electrical sampling modules -- Mechanical 52............
80E00 Electrical Sampling Modules User Manual
iii
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read
the General Safety Summary in other system manuals for warnings and cautions
related to operating the system.
Ground the Product. This product is indirectly grounded through the grounding
conductor of the mainframe power cord. To avoid electric shock, the grounding
conductor must be connected to earth ground. Before making connections to the
input or output terminals of the product, ensure that the product is properly
grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
Do Not Operat e Without Covers. Do not operate this product with covers or panels
removed.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Wear Eye Prot ection. Wear eye protection if exposure to high-intensity rays or
laser radiation exists.
Do Not Operat e With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operat e in Wet/Damp Conditions.
Do Not Operat e in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
ToAvoidFireor
Personal Injury
General Safety Summary
iv
80E00 Electrical Sampling Modules User Manual
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Terms on t he Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
CAUTION
Refer to Manual
WARNING
High Voltage
Protective Ground
(Earth) Terminal
Symbols and Terms
80E00 Electrical Sampling Modules User Manual
v
Environmental Considerations
This section provides information about the environmental impact of the
product.
Observe the following guidelines when recycling an instrument or component:
Equipment Recycling. Production of this equipment required the extraction and
use of natural resources. The equipment may contain substances that could be
harmful to the environment or human health if improperly handled at the
product’s end of life. In order to avoid release of such substances into the
environment and to reduce the use of natural resources, we encourage you to
recycle this product in an appropriate system that will ensure that most of the
materials are reused or recycled appropriately.
The symbol shown to the
left indicates that this
product complies with the
European Union’s require-
ments according to Direc-
tive 2002/96/EC on waste
electrical and electronic
equipment (WEEE). F or
information about recycl-
ing options, check the
Support/Service section of
the Tektronix Web site
(www.tektronix.com).
This product has been classified as Monitoring and Control equipment, and is
outside the scope of the 2002/95/EC RoHS Directive. This product is known to
contain lead, cadmium, and hexavalent chromium.
Product End-of-Life
Handling
Restriction of Hazardous
Substances
Environmental Considerations
vi
80E00 Electrical Sampling Modules User Manual
80E00 Electrical Sampling Modules User Manual
vii
Preface
This is the user manual for the 80E01, 80E02, 80E03, 80E04, and 80E06
sampling modules. It covers the following information:
H Description of the capabilities of the sampling modules and how to install
them
H Explanation of how to operate the sampling modules: how to control
acquisition, processing, and input/output of information
H List of the specifications of the sampling modules
You may want to visit the Tektronix Website at http://www.tektronix.com for the
latest revision of the user documentation. Select the Manuals link, then enter the
part number or product name to locate the document.
A printed version of this manual is also orderable (see Optional Accessories on
page 4).
Manual Structure
This manual is composed of the following chapters:
H Getting Started shows you how to configure and install your sampling
module.
H Operating Basics describes controlling the sampling module using the front
panel and the main instrument user interface.
H Reference provides additional information including the specifications;
detailed descriptions of all programming commands are found in the Online
Programmer Guide for your main instrument.
Related Documentation
This document covers installation and usage of the sampling module and its
features. For information about the main instrument in which the sampling
module is used, refer to the user documents and online help provided with your
main instrument.
Preface
viii
80E00 Electrical Sampling Modules User Manual
80E00 Electrical Sampling Modules User Manual
1
Getting Started
The Tektronix 80E01, 80E02, 80E03, 80E04, and 80E06 sampling modules are
high-performance sampling modules that can be installed in the following main
instruments:
H DSA8200 Digital Serial Analyzer
H CSA8000, CS A800B, and CSA8200 Communications Signal Analyzers
H TDS8000, TDS8000B, and TDS8200 Digital Sampling Oscilloscopes
Proper operation of the electrical sampling modules requires that the appropriate
application software is installed on the main instrument. In Table 1, each new
release of software supports the additional electrical modules as well as all
electrical modules supported by earlier software versions.
To display the version installed, select About from the Help menu of the main
instrument.
Table 1: Application software ver sion r equired
DSA8200 and TDS/
CSA8000 appli cati on
software version
Additional modules supported
1.0.0 or greater
1
80E01, 80E02, 80E03, 80E04
1.4.0 or greater
1
80E06
1
Product application software version 1.x.x requires the Windows 98 operating
system.
Getting Started
2
80E00 Electrical Sampling Modules User Manual
Product Description
The sampling modules provide the features shown in Table 2.
Table 2: Sampling module feat ures
Feature 80E01 80E02 80E03 80E04 80E06
Number of independent channels 1 2 2 2 1
Rise time 7 ps, typical
1
≤28 ps ≤17.5 ps ≤17.5 ps ≤5.0 ps, typical
1
Bandwidth 50 GHz 12.5 GHz 20 GHz 20 GHz 70 GHz, typical
Displayed noise
3
≤2.3 mV
RMS
≤800 V
RMS
≤1.2 mV
RMS
≤1.2 mV
RMS
≤2.8 mV
RMS
Select channel buttons for quick
trace identificat ion
Yes Yes Yes Yes Yes
Maximum non-destruct ive input
voltage
2 V (DC + peak
AC)
3 V (DC + peak
AC)
3 V (DC + peak
AC)
3 V (DC + peak
AC)
2 V (DC + peak
AC)
Vertical sensitivity, full scale 10mVto1V 10mVto1V 10mVto1V 10mVto1V 10mVto1V
Signal connectors
2
2.4 mm male to
2.92 mm (K)
female
3.5 mm female 3.5 mm female 3.5 mm female 2.4 mm male to
2.92 mm (K)
female
4
Number of TDR channels N.A. N.A. N.A. 2 N.A.
1
The 80E01 module risetime is estimated using the formula risetime = 0.35/bandwidth. The 80E06 module risetime is
estimated using the formula risetime = 0.35/(typical bandwidth).
2
3.5 m m female to 2.4 mm male adapter is provided.
3
Measured at 1 ps/div.
4
Because the 2.4 mm connector of this adapter will mechanical ly i nterface with the 1.85 mm connector of the 80E06, it
serves as a 1.85 mm-to-2.92 mm adapter for the 80E06 module.
As shown in Figure 1, the sampling modules have two independent channels
(80E01 and 80E06 each have one channel), each with its own acquisition
circuitry.
The strobe drive signal from the instrument controls the timing of the strobe
assertion to each acquisition system and guarantees sampling coincidence
between the channels in a sampling module.
CAUTION. To prevent electrostatic damage to the main instrument and sampling
modules, follow the precautions described in this manual and the manuals
accompanying your instrument. (See Electrostatic Discharge on page 6.)
Getting Started
80E00 Electrical Sampling Modules User Manual
3
Sampler
Sampler
Strobe
Generator
Strobe drive
50 Ω
From main instrument
50 Ω
Note: the 80E01and 80E06 are
single channel modules with a
dedicated strobe drive and
generator.
To main instrument
To main instrument
Figure 1: Sam pling module block diagram
TDR on indicator
light (red)
(80E04)
Signal connector
Left channel
Right channel
Hold-down screw
SELECT channel button
TEKPROBE connector
Channel indicator
light (yellow)
Figure 2: Sam pling module, 80E04 shown
Getting Started
4
80E00 Electrical Sampling Modules User Manual
Options and Accessories
This section lists the standard and optional accessories available for the sampling
modules.
The following options can be ordered for the sampling modules:
H Option C3: Three years of calibration services
H Option C5: Five years of calibration services
H Option D3: Test Data for calibration services in Option C3
H Option D5: Test Data for calibration services in Option C5
H Option R3: Repair warranty extended to cover three years
H Option R5: Repair warranty extended to cover five years
The accessories in Table 3 are shipped with the sampling modules.
Table 3: Standard accessories
Item Part number
Certif icate of Traceable Calibration for product at initial shipment Not Orderable
2.4 mm male to 2.92 mm (K) female adapter (80E01 and 80E06 only)
1
015-0703-xx
SMA male 50 Ω termination (one per channel) 015-1022-xx
Transit case, ESD protective 004-5091-xx
Transit case, ESD protective (80E06 only) 004-5165-xx
1
Because the 2.4 mm connector of this adapter will mechanical ly i nterface with the
1.85 m m connector of the 80E06, it serves as a 1.85 m m-to-2.92 mm connector for
the 80E06 module
The accessories in Table 4 are orderable for use with the sampling module at the
time this manual w as originally published. Visit the Tektronix Web site or
consult a current catalog for additions, changes and details.
Table 4: Optional accessories
Item Part number
Sampling module ext ender cable (2 meter)
1
80N01
2X attenuator (SMA male-to-female) 015-1001-xx
5X attenuator (male-to-female) 015-1002-xx
Options
Standard Accessories
Optional A ccessories
Getting Started
80E00 Electrical Sampling Modules User Manual
5
Table 4: Optional accessories (cont.)
Item Part number
Power divider 015-0565-xx
SMA accessory kit 020-1693-xx
Torque wrench, 8 mm (5/16 inch) open end n.a.
3.5maleto3.5femaleSMA 015-0552-xx
Slip-on SM A connector 015-0553-xx
3.5 mm 50 Ω connector (SMA male-to-female) 015-0549-xx
BNC female 75 Ω to 50 Ω type N minimum loss attenuator 131-0112-xx
CSA8000 & TDS8000 Series Service Manual 071-0438-xx
DSA8200 Service Manual 071--2049--xx
Termi nator, ECL 015-0558-xx
Connector saver, 3.5 mm SMA 015-0549-xx
80E00 Electrical Sampling Module User Manual (printed) 071-0434-xx
1
An extender cable extends the reach of a sampling module. You install the extender
between the sampling module and the instrument, allowing you to operate the
sampling m odule out of the module compartment. If you compensate a module in
the main instrument and then move the module to an extender, or visa versa,
re-compensate the module (for more information see Compensation on page 9).
Getting Started
6
80E00 Electrical Sampling Modules User Manual
Installation
The sampling modules fit into the front panel of the main instrument. Figure 3
shows the front panel of a DSA8200 and the locations of the sampling-module
compartments.
Large modules
Small modules
Left-most small module
compartment is not usable if
a large module is installed
in left-most slot
Module ejectors
This small module
compartment is not usable if
a large module is installed
in right-most slot
Figure 3: Sam pling module compar tments
At least one sampling module must be installed in an instrument to sample
signals.
NOTE. Installing a large module in the left-most slot disables the left-most small
module compartment. Installing a large module in the right-most slot disables
the small module compartment that is second from the left.
Each instrument supports four large-compartment channels, two per sampling
module, and eight small-compartment channels, two per sampling module. Eight
of the ten channels are usable at one time.
To prevent electrostatic damage to the main instrument and sampling modules,
follow the precautions described in this manual and the manuals that come with
your instrument.
Circuitry in the sampling module is very susceptible to damage from electrostat-
ic discharge or from overdrive signals. Be sure to only operate the sampling
module in a static-controlled environment. Be sure to discharge to ground any
electrostatic charge that may be present on the center and outer connectors of
cables before attaching the cable to the sampling module.
Electrostatic Discharge
Getting Started
80E00 Electrical Sampling Modules User Manual
7
Know your signal source. If it is capable of delivering overvoltages, it is safer to
not depend on the signal source settings for protection, but instead use an
external attenuator that protects the input from the worst-case conditions. For
example, for a 20 V maximum source connected to a 3 V maximum sampling
module, use a 10X attenuator. Where possible, connect your cables to the signal
source first, and to the sampling module second.
CAUTION. To prevent damage from electrostatic discharge, install 50 Ω
terminations on the sampling-module connectors before removing a sampling
module from an instrument or when it is not in use. Store the sampling module in
a static-fr ee container, such as the shipping container. Whenever you move the
sampling module from one instrument to another, use a static-free container to
transport the sampling module.
To prevent damage to the sampling module, discharge to ground any electrostat-
ic charge that may be present on the center and outer conductors of cables
before attaching the cable to the sampling module.
To prevent damage to the sampling module, do not create an ESD antenna by
leaving cables attached to the sampling-module input with the other end of the
cable open.
To prevent damage to the sampling module or instrument, never install or
remove a sampling module when the instrument is powered-on.
Always use a wrist strap (provided with your instrument) when handling
sampling modules or making signal connections. Wear anti-static clothing and
work in a static-free workstation when using sampling modules.
Use a Tektronix 80A02 EOS/ESD Protection Module if doing TDR work.
To prevent damage to the sampling module or instrument, do not apply a voltage
outside the Maximum Input Voltage (see page 48) for your sampling module.
For information on creating a static-controlled workstation, consult the Electron-
ic Industries Association document EIA-625; Requirements for Handling
Electrostatic-Discharge-Sensitive (ESDS) Devices.
You can use a Tektronix 80A02 EOS/ESD Protection Module to protect the
sampling module from damage due to static discharge from circuit boards and
cables. Use the 80A02 in applications where large static charges can be stored on
the device under test, such as when testing TDR circuit boards or cables.
Refer to the documentation supplied with the 80A02 module for proper
installation and use.
Static Controlled
Workstation
Getting Started
8
80E00 Electrical Sampling Modules User Manual
To install a sampling module, first turn off the instrument using the front-panel
On/Standby switch. Then place the sampling module in a compartment and
slowly push it in with firm pressure. Once the sampling module is seated, turn
the hold-down screw on the sampling module to tighten the sampling module
into place. See Figure 4.
CAUTION. To prevent damage to the sampling module or instrument, never install
or remove a sampling module when the instrument is powered on or when either
input connector is unprotected.
NOTE. When removing your sampling module, first loosen the hold-down screw,
and then use the sampling module ejector on the main instrument to eject the
sampling module.
Small-compartment
ejectors
Electrical sampling module
Hold-down screw
Figure 4: Installing a sampling module
Module Installation
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68

Tektronix 8.00E+05 User manual

Category
Music drums
Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI