Spectris Omega DBCL130 User manual

Type
User manual

This manual is also suitable for

UsersGuide
Shop online at
omega.com
e-mail: info@omega.com
For latest product manuals:
www.omega.com
Above items are included with the DBCL130
Dry Block temperature calibrator
.
Not shown
is the calibration certificate and operator manual
Model DBCL130
Dry Block Temperature Calibrator
P a g e | 2
DBCL130 Dry Block Temperature Calibrator
Introduction
The DBCL130 calibrator provides a safe, dry, constant temperature source for checking and calibrating
a wide range of temperature sensors, systems, indicators and thermometers. It is fast and economical and
can be used either on a bench top or as a portable field unit. The weight of the unit is only 15 pounds/6.8
kilograms. The unit covers the temperature range from -25°C to +130°C in an ambient of 20°C using a
machined aluminum block as the heat transfer medium. The temperature control circuit is built into the
unit.
Features include:
• Maximum temperature of 130°C/266°F
• Minimum temperature of -25°C/-13°F
• Up to eight setpoints can be stored & recalled
• Insert well & ¼” diameter reference well
Even though the unit heats up and cools down rapidly, highly efficient insulation and an internal cooling
fan ensure that the case remains safe enough to handle even at maximum operating temperatures. The
DBCL130 calibrator has been designed to comply with all relevant electromagnetic interference and
electrical safety regulations.
Specification
Figures quoted are at the base of the well at the time of calibration.
Temperature range: -25°C/-13°F to 130°C/266°F at an ambient of 20°C/68°F
NOTE: the typical minimum achievable temperature is 45°C/80°F below the room ambient temperature
Display resolution: 0.1°
Accuracy: ±0.4°C (-20 to 130°C)
±0.7°F (-4 to 266°F)
Stability (10 minutes): ±0.050°C (±0.080°F)
Heat up time -20 to 130°C: 5 minutes
Heat up time 20 to 100°C: 2 minutes
Cool down 20 to -20°C: 3 minutes (ambient of +20°C)
Cool down 130 to 20°C: 2.5 minutes
Immersion Depth: 4" (101mm)
Insert well: ½” diameter x 4” depth
Reference well: ¼” diameter x 4” depth
Fan Cooling: Automatic
Weight: 15 lbs (6.8 Kg)
Dimensions* (H x W x D): 11 x 9 x 8 inches/279 x 229 x 203 mm
*excluding the carrying strap
Electrical supply (this unit will operate on any voltage from 100 to 240 VAC)
Voltage Cycles Power
100 to 230V 50/60Hz 200W
Note: The above specifications are quoted for an ambient temperature range of 10°C/50°F to
30°C/86°F. Outside this range, the quoted figures may deteriorate but the unit will still work
safely. NOTE: The minimum achievable temperature is 45°C/80°F below the room ambient
temperature.
P a g e | 3
Working environment
The calibrator units are designed to work safely under the following conditions:
Ambient temperature range: 5°C/9°F to 40°C/104°F
Humidity: Up to 95% relative humidity, non-condensing
Warning
Warning: HIGH TEMPERATURES ARE DANGEROUS
HIGH TEMPERATURES ARE DANGEROUS: They can cause serious burns to operators and ignite
combustible material. Omega Engineering has taken great care in the design of these units to protect
operators from hazards, but operators should pay attention to the following points:
• USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS
• DO NOT put hot objects on or near combustible objects
• DO NOT operate the unit close to inflammable liquids or gases
• DO NOT place any liquid directly in your unit
• At all times USE COMMON SENSE
Operator Safety
All operators of Omega Engineering equipment must have available the relevant literature needed to
ensure their safety. It is important that only suitably trained personnel operate this equipment in
accordance with the instructions contained in this manual and with general safety standards and
procedures. If the equipment is used in a manner not specified by Omega Engineering, the protection
provided by the equipment to the operator may be impaired. All Omega Engineering units have been
designed to conform to international safety requirements. If a safety problem is encountered, switch off
at the power socket and remove the plug from the supply. Please use caution when removing probes and
inserts as burns to the skin can occur if in contact.
Installation
1. All Omega Engineering units are supplied with a power cable.
2. Before connecting the power supply, check the voltage against the rating plate. Connect the
power cable to a suitable plug according to the table below. Note that the unit must be earth
grounded to ensure proper electrical safety.
Electrical connections:
220V-240V 110V-120V
Live Brown Black
Neutral Blue White
Earth ground Green/yellow Green
The fused plug supplied with the power lead for use in the UK is fitted with the following value
fuse to protect the cable: 230V UK 4 AMP
The fuse in the unit protects the unit and the operator. Note that the model DBCL130 operates
on any mains voltage from 100 to 240VAC. However a properly rated mains cable should be
used based on the mains operating voltage used.
3. Plug the power cable into the socket on the rear of the unit.
4. Place the unit on a suitable bench or flat workspace, or in a fume cupboard if required,
ensuring that the air inlet vents on the underside are free from obstruction.
After use, when you have finished heating samples, remember that parts of the unit may be very
hot. Take the precautions listed earlier.
P a g e | 4
OPERATION
Preparation
1.
The heater design, temperature sensor and control circuit give good temperature control and
uniformity, but make sure that there is a close fit of the probes in the block to allow
efficient heat transfer. Contact us about an insert that more closely fits your probe or device
being calibrated.
2.
Plug the power cable into the socket in the back of the unit. Connect the power cable to the
electrical supply and switch the power on. 1 = power on, 0 = power off.
3.
Clean the heater block cavity out with
shop or canned air to remove any
particulate. Next place the probe insert
into the heater block as shown using the
supplied insert extractor to minimize the
risk of damaging the heater block and/or
probe insert. Never place a hot insert into
a cold heater block or vice versa as the
insert may become jammed which will
damage both parts. Always use the insert
extractor to both install and remove the
probe insert.
4. To prevent damage to the heater block,
insert, peltiers and PRT block sensor DO
NOT use the following in or around the
block;
Oil, Thermal grease, Water
Aluminum oxide sand,
Ceramic fiber insulation or Kaowool
Setting the operating temperature
1. To set the operating temperature required, press and hold either the up or down arrow button
to increment to the value required. Alternatively you can press the («PF) key to move over to
individual digits to set higher values much quicker. After 2 seconds your value will be set &
retained.
2. When you have entered a set temperature the unit will start to heat or cool to that value.
3. Once the process value temperature reaches the set point, allow the block to fully stabilize for
at least 10 minutes before performing a calibration.
4. When calibrating sensors and thermometers start at higher temperatures first and then work
down to lower ones. This will allow devices to be removed safely.
5. Use the main insert well for calibrating sensors and thermometers. The ¼” hole behind the
insert well is for holding your reference thermometer when performing comparison
calibrations. All calibrations should be performed in the probe insert.
Entering up to 8 setpoints for fast recall
1. To input up to 8 setpoints press the first button on the left and then the 2
nd
button from the
left until the top line displays SP-0. Here you can enter up to 8 values for fast future recall. Do
not change any of the settings or values after SP-7. When finished entering values press the
button on the left one time. You’ll need to note which value is in which location for future recall.
2. To select one of the 8 setpoints for use from the main display press the 2
nd
button from the
left so the top line displays M-SP. Next use the up arrow key to select one of the 8 setpoint
values. Press the 1
st
key on the left twice for the value to be accepted.
P a g e | 5
Operation near ambient
For best results, to calibrate at a set point near ambient when cooling down from a higher temperature
set parameter (AL-1) 3 °C/5 °F higher than you room ambient. When heating up from a lower
temperature set (AL-1) 3 °C/5 °F lower than ambient. Access the AL-1 parameter by pressing the mode
button (2
nd
from the left).
Temperature scale conversion
To switch from C to F and vice versa press and hold then select C or F under the d-U parameter.
Next press to display SL-H parameter. Set this value to 266 for degrees F operation and 130 for
degrees C. Press to display SL-L parameter, leave this value at -40 for both degrees C and F. Press
to exit.
Press once and then until the parameter CN5 is displayed. For this value when switching from
F to C divide the value shown by 1.8 and change it. For conversion from C to F multiply by 1.8 and
change. Adjusting calibration parameters is discussed below.
Calibration Adjustment using a reference thermometer
Use the following if you have a traceable reference thermometer and would like to adjust the calibration of the
DBCL130. Calibrations should be performed in a 3/16” or ¼” diameter insert and not the reference well.
Press the first button on the left and then the 2
nd
button from the left until the top line displays CN5. The initial
value shown is the default factory calibration and corresponds to the readings shown on the factory calibration
certificate.
CN5 is a calibration offset, low end adjustment. Example if your reference thermometer reads 0.5 and the
DBC130 display is indicating 0.0 then set the CN5 value to 0.5. Then press the first button on the left. The display
will correct to 0.5 and start to cool down so the block reading and your reference thermometer match. Adjust
parameter CNRt at high end temperatures to correct non-linearity or slope. When done press the button on the left
one time.
Operator maintenance
NOTE THAT THIS EQUIPMENT SHOULD ONLY BE DISMANTLED BY PROPERLY TRAINED
PERSONNEL. REMOVING THE FRONT OR REAR PANELS EXPOSES POTENTIALLY LETHAL
VOLTAGES. THERE ARE NO OPERATOR MAINTAINABLE PARTS WITHIN THE
EQUIPMENT.
In the unlikely event that you experience any problems with your unit which cannot easily be remedied,
you should contact your supplier and return the unit if necessary. Please include any details of the fault
observed and remember to return the unit in its original packing. Omega Engineering will accept no
responsibility for any damage to units that are improperly packed for shipment. If in doubt, contact your
supplier.
1. Cleaning: Before cleaning your unit, ALWAYS disconnect it from the power supply and
allow it to cool to 30° C. Your unit can be cleaned by wiping with a damp soapy cloth. Care
should be exercised to prevent water from running inside the unit. Do not use abrasive
cleaners.
2. Fuse: The unit is protected by a fuse. It should only be changed by suitably qualified
personnel. If the fuse blows persistently, a serious fault is indicated and you may need to
return the unit to your supplier for repair.
.
P a g e | 6
Accessories
The following parts may be obtained from Omega Engineering if replacements or alternatives are
required:
Part Number Description
4163 UK 240 volt power cable with 13amp UK plug (5 amp fuse)
4164 Euro style 240 volt power cable with R/A Schuko plug
4150 US style 120 volt power cable
4168 Unit carrying strap
4285 Insert extractor
DBCL-3052 Soft sided carrying case
DBCL-130-3074 Insert for 1/8” diameter probes
DBCL-130-3075 Insert for 3/16” diameter probes
DBCL-130-3076 Insert for ¼” diameter probes
DBCL-130-3077 Insert for 5/16” diameter probes
DBCL-130-3078 Insert for 3/8” diameter probes
Spare Parts
Part Number Description
4224 Peltier
4174 Temperature controller
4147 PRT
4221 Solid state relay
4223 Power supply
4280 3 PDT power relay
4283 2 amp fuse, 5 x 20mm
P a g e | 7
EU Declaration of Conformity (No. DC18-DBCL)
In accordance with European Parliament and Council Decision No 768/2008/EC Annex III
1. Product model / product:
Product Dry Block Temperature Calibrator
Model/type DBCL400 & DBCL130
Batch/serial no. S/N: 619-2993 & onward
2. Manufacturer
Name Omega Engineering
Address 800 Connecticut Ave, Norwalk, CT 06854
3. This declaration is issued under the sole responsibility of the manufacturer.
4. Object of the declaration:
Product Dry Block Temperature Calibrator
Specification Model DBCL400 operating range ambient +5 to 450°C
Model DBCL130 operating range -25 to 130°C (20°C ambient)
5. The object of the declaration described above is in conformity with the relevant Union
harmonisation legislation:
2014/35/EU The Low Voltage Directive
2014/30/EU The Electromagnetic Compatibility Directive
2011/65/EU The Restriction of Hazardous Substances Directive
6. References to the relevant harmonised standards used or references to the other technical
specifications in relation to which conformity is declared:
Reference & Date
Title
EN 60519-1:2015 Safety in installations for electroheating and electromagnetic
processing. General requirements
EN 61000-6-
2:2005
Electromagnetic compatibility (EMC) - Part 6-2: Generic standards
- Immunity for industrial environments
EN 61000-6-
4:2007 + A1:2011
Electromagnetic compatibility (EMC) - Part 6-4: Generic standards
- Emission standard for industrial environments
EN 50581:2012 Technical documentation for the assessment of electrical and
electronic products with respect to the restriction of hazardous
substances
7. Additional information:
Signed for and on behalf of: Omega Engineering
Place of issue: Hainesport, NJ, USA
Date of issue: July 8, 2019
Name: Darren Sager
Signature:
Darren Sager
REV 5 8/19
omega.com
info@omega.com
Servicing North America:
U.S.A
.
Omega Engineering, Inc.
Headquarters:
Toll-Free: 1-800-826-6342 (USA & Canada only)
Customer Service: 1-800-622-2378 (USA & Canada only)
Engineering Service: 1-800-872-9436 (USA & Canada only)
Tel: (203) 359-1660 Fax: (203) 359-7700
For Other Locations Visit omega.com/worldwide
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves
the right to alter specifications without notice.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords
our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2016 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the
prior written consent of OMEGA ENGINEERING, INC.
FOR WARRANTY RETURNS, please have the
following information available BEFORE contacting
OMEGA:
1. Purchase Order number under which the product
was PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
FOR NON-WARRANTY REPAIRS,
consult
OMEGA for current repair charges. Have
the following information available BEFORE
contacting OMEGA:
1. Purchase Order number to cover the COST
of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems
relative to the product.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED
RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return
package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a
period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month
grace period to the normal one (1) year product warranty to cover handling and shipping time. This
ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no
charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser,
including but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for
any damages that result from the use of its products in accordance with information provided
by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the
company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of
OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the
component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
M0000/0019
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
TEMPERATURE
MU
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
MU
Wire: Thermocouple, RTD & Thermistor
MU
Calibrators & Ice Point References
MU
Recorders, Controllers & Process Monitors
MU
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
MU
Transducers & Strain Gages
MU
Load Cells & Pressure Gages
MU
Displacement Transducers
MU
Instrumentation & Accessories
FLOW/LEVEL
MU
Rotameters, Gas Mass Flowmeters & Flow Computers
MU
Air Velocity Indicators
MU
Turbine/Paddlewheel Systems
MU
Totalizers & Batch Controllers
pH/CONDUCTIVITY
MU
pH Electrodes, Testers & Accessories
MU
Benchtop/Laboratory Meters
MU
Controllers, Calibrators, Simulators & Pumps
MU
Industrial pH & Conductivity Equipment
DATA ACQUISITION
MU
Communications-Based Acquisition Systems
MU
Data Logging Systems
MU
Wireless Sensors, Transmitters, & Receivers
MU
Signal Conditioners
MU
Data Acquisition Software
HEATERS
MU
Heating Cable
MU
Cartridge & Strip Heaters
MU
Immersion & Band Heaters
MU
Flexible Heaters
MU
Laboratory Heaters
ENVIRONMENTAL
MONITORING AND CONTROL
MU
Metering & Control Instrumentation
MU
Refractometers
MU
Pumps & Tubing
MU
Air, Soil & Water Monitors
MU
Industrial Water & Wastewater Treatment
MU
pH, Conductivity & Dissolved Oxygen Instruments
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10

Spectris Omega DBCL130 User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI