Thermo Fisher Scientific MagMAX™ Express-96 Magnetic Particle Processor Owner's manual

Type
Owner's manual

This manual is also suitable for

MagMAX™ Express 96
User Manual
Rev. 1.5
Contents
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 2
© Copyright 2008, 2010 Applied Biosystems. All rights reserved.
Information in this document is subject to change without notice. Applied Biosystems assumes no responsibility for any errors that
may appear in this document.
APPLIED BIOSYSTEMS DISCLAIMS ALL WARRANTIES WITH RESPECT TO THIS DOCUMENT, EXPRESSED OR
IMPLIED, INCLUDING BUT NOT LIMITED TO THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. IN NO EVENT SHALL APPLIED BIOSYSTEMS BE LIABLE, WHETHER IN CONTRACT, TORT, WARRANTY,
OR UNDER ANY STATUTE OR ON ANY OTHER BASIS FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE,
MULTIPLE OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT,
INCLUDING, BUT NOT LIMITED TO, THE USE THEREOF.
For Research Use Only. Not for use in diagnostic procedures.
NOTICE TO PURCHASER: License Disclaimer
Purchase of this product alone does not imply any license under any process, instrument or other apparatus, system, composition, reagent or
kit rights under patent claims owned or otherwise controlled by Applied Bisosytems, either expressly or by estoppel.
TRADEMARKS:
Applied Biosystems, and AB (Design) are registered trademarks, and MagMAX is a trademark of Applied Biosystems or its
subsidiaries in the U.S. and/or certain other countries.
All other trademarks are the sole property of their respective owners.
Part Number N07849 Rev. B
January 2011
Contents
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 3
Contents
1Safety Symbols and Markings . ....................................................................................... 7
2About the User Manual. ................................................................................................. 10
3Introduction to the MagMAX™ Express 96. ................................................................. 13
4Functional Description. .................................................................................................. 14
5Installation ...................................................................................................................... 19
6Routine Operation. ......................................................................................................... 36
7Maintenance . .................................................................................................................. 46
8Technical Specifications . ............................................................................................... 50
9Troubleshooting Guide. ................................................................................................. 53
10 Warranty Information . ................................................................................................... 58
11 Ordering Information . .................................................................................................... 62
12 Glossary and Abbreviations . ......................................................................................... 63
13 Index................................................................................................................................ 66
14 Appendix A. MagMAX™ Express 96 Brief User’s Guide . ............................................ 69
15 Appendix B. MagMAX™ Express 96 Feedback Form . ................................................. 70
16 Appendix C. Certificate of Instrument Decontamination . .......................................... 71
17 Appendix D. Addresses. ................................................................................................. 73
Table of Contents
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 4
Table of Contents
1 Safety Symbols and Markings . ....................................................................................... 7
1.1 Instrument safety and guidelines for use . ....................................................................... 8
1.1.1 Chemical Safety Guidelines. .......................................................................................... 8
1.1.2 Biohazard Warning. ........................................................................................................ 9
2 About the User Manual. ................................................................................................. 10
2.1 Preface............................................................................................................................. 10
2.1.1 Purpose of This Guide. ................................................................................................. 10
2.1.2 Audience........................................................................................................................10
2.1.3 Assumptions. ................................................................................................................ 10
2.1.4 Text Conventions . ........................................................................................................ 11
2.1.5 User Attention Words . ................................................................................................. 11
2.1.6 Safety Alert Words . ...................................................................................................... 11
2.2 How to Obtain More Information. .................................................................................. 11
2.2.1 Related Documentation . .............................................................................................. 11
2.3 How to Obtain Support . ................................................................................................. 12
3Introduction to the MagMAX™ Express 96. ................................................................. 13
3.1 Intended use.................................................................................................................... 13
3.2 Principle of operation . .................................................................................................... 13
3.3 Advantages of processing magnetic particles with MagMAX™ Express 96 . ................... 13
4Functional Description. .................................................................................................. 14
4.1 Instrument layout. ........................................................................................................... 14
4.1.1 Front view......................................................................................................................14
4.1.2 Back / Side view . .......................................................................................................... 15
4.2 MagMAX™ Express 96 magnetic particle processor. ................................................... 16
4.3 Principle of magnetic particle processing . .................................................................... 17
4.4 Working with a magnetic rod. ........................................................................................ 17
4.4.1 Collecting magnetic particles. ...................................................................................... 17
4.4.2 Releasing magnetic particles. ...................................................................................... 17
4.4.3 Washing magnetic particles. ........................................................................................ 18
4.4.4 Incubation ......................................................................................................................18
4.4.5 Changing the volume during the magnetic particle processing. ............................... 18
5 Installation ...................................................................................................................... 19
5.1 What to do upon delivery. .............................................................................................. 19
5.1.1 How to unpack. ............................................................................................................. 19
5.1.2 Checking delivery for completeness . .......................................................................... 19
5.1.3 Checking for damage during transport . ...................................................................... 19
5.1.4 Environmental requirements . ...................................................................................... 19
5.1.5 Things to avoid. ............................................................................................................ 20
5.1.6 Technical prerequisites. ............................................................................................... 20
5.2 Setups before you put the instrument into operation . ................................................. 20
5.2.1 How to release the transport locks. ............................................................................. 20
5.2.2 How to ensure startup. ................................................................................................. 24
5.2.3 How to fit the subassemblies of the instrument into place. ....................................... 25
Table of Contents
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 5
5.3 Operational check . .......................................................................................................... 31
5.4 How to pack the instrument for transportation. ............................................................ 31
5.4.1 How to refit the transport lock of the heating block. .................................................. 31
5.4.2 How to refit the transport lock of the tip comb holder. .............................................. 34
6 Routine Operation. ......................................................................................................... 36
6.1 Do’s and Don’ts of the MagMAX™ Express 96 . ............................................................ 36
6.1.1 Do’s ............................................................................................................................... 36
6.1.2 Don’ts............................................................................................................................ 36
6.2 Switching on ....................................................................................................................36
6.3 How to use the keypad and display. .............................................................................. 37
6.4 How to handle tip combs . .............................................................................................. 38
6.5 How to start..................................................................................................................... 39
6.6 Changing the heating block. ........................................................................................... 43
6.7 Shutdown........................................................................................................................ 45
6.8 Emergency situations . .................................................................................................... 45
7Maintenance . .................................................................................................................. 46
7.1 Regular and preventive maintenance . ........................................................................... 46
7.1.1 How to clean the turntable. .......................................................................................... 46
7.1.2 How to clean the magnetic rods. ................................................................................. 47
7.1.3 How to clean the shield plate . ..................................................................................... 47
7.2 Disposal of materials . ..................................................................................................... 47
7.3 Decontamination procedure . ......................................................................................... 47
7.4 How to pack for service. ................................................................................................. 48
7.5 Disposal of the instrument . ............................................................................................ 49
8Technical Specifications . ............................................................................................... 50
8.1 General specifications . ................................................................................................... 50
8.2 Safety specifications. ...................................................................................................... 51
8.2.1 In conformity with the requirements. .......................................................................... 51
9Troubleshooting Guide. ................................................................................................. 53
9.1 Error messages and warnings. ....................................................................................... 53
9.2 Troubleshooting guide . .................................................................................................. 54
9.3 Frequently asked questions (FAQ) about MagMAX™ Express 96 processing. ........... 55
9.4 Service request protocol . ............................................................................................... 57
10 Warranty Information . ................................................................................................... 58
11 Ordering Information . .................................................................................................... 62
11.1 List of accessories. .......................................................................................................... 62
11.2 List of spare parts ........................................................................................................... 62
12 Glossary and Abbreviations . ......................................................................................... 63
12.1 Keywords for web pages. ............................................................................................... 64
12.2 Literature ......................................................................................................................... 64
13 Index................................................................................................................................ 66
14 Appendix A. MagMAX™ Express 96 Brief User’s Guide . ............................................ 69
15 Appendix B. MagMAX™ Express 96 Feedback Form . ................................................. 70
16 Appendix C. Certificate of Instrument Decontamination . .......................................... 71
17 Appendix D. Addresses. ................................................................................................. 73
List of Figures and Tables
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 6
List of Figures and Tables
Fig. 3.1 MagMAX™ Express 96 magnetic particle processor........................................... 13
Fig. 4.1 MagMAX™ Express 96 front view without see-through lid ................................. 14
Fig. 4.2 MagMAX™ Express 96 front view with see-through lid and plates..................... 14
Fig. 4.3 MagMAX™ Express 96 back view. ...................................................................... 15
Fig. 4.4 MagMAX™ Express 96 side view. ....................................................................... 15
Fig. 4.5 MagMAX™ Express 96 magnetic particle processor........................................... 16
Fig. 4.6 Inverse magnetic particle processing. ................................................................. 17
Fig. 4.7 A concentration step during magnetic particle processing ................................. 18
Fig. 5.1 MagMAX™ Express 96 transport locks fitted . ...................................................... 21
Fig. 5.2 Removing the transport lock of the tip comb holder (A). ..................................... 21
Fig. 5.3 Removing the transport lock of the tip comb holder (B)...................................... 22
Fig. 5.4 Tip comb holder transport lock removed. ............................................................ 22
Fig. 5.5 Removing the transport lock of the heating block (A).......................................... 23
Fig. 5.6 Removing the transport lock of the heating block (B). ......................................... 23
Fig. 5.7 Fastening the transport lock of the heating block to its storage site .................... 24
Fig. 5.8 Mains supply cable and serial cable connected. ................................................. 25
Fig. 5.9 Fastening/changing the heating block. ................................................................ 25
Fig. 5.10 Two different heating blocks available . ............................................................... 26
Fig. 5.11 Shield plate prior to fitting . .................................................................................. 26
Fig. 5.12 Inserting the shield plate . .................................................................................... 27
Fig. 5.13 Fastening/loosening the shield plate. .................................................................. 27
Fig. 5.14 Shield plate fitted . ............................................................................................... 28
Fig. 5.15 Interchangeable MagMAX™ Express 96 heads and tip combs available............ 28
Fig. 5.16 MagMAX™ Express 96 heads in their storage boxes . ......................................... 29
Fig. 5.17 Inserting the MagMAX™ Express 96 head . ......................................................... 30
Fig. 5.18 MagMAX™ Express 96 head for deep well plate fitted....................................... 30
Fig. 5.19 See-through lid fitted. .......................................................................................... 31
Fig. 5.20 Unfastening the transport lock screw from its storage site.................................. 32
Fig. 5.21 Refitting the transport lock of the heating block (A)............................................. 32
Fig. 5.22 Refitting the transport lock of the heating block (B). ............................................ 33
Fig. 5.23 Fitting the transport lock of the tip comb holder (A). ........................................... 34
Fig. 5.24 Fitting the transport lock of the tip comb holder (B) . ........................................... 34
Fig. 5.25 Fitting the transport lock of the tip comb holder (C). ........................................... 35
Fig. 5.26 Both transport locks refitted. ................................................................................ 35
Fig. 6.1 MagMAX™ Express 96 keypad. ........................................................................... 37
Fig. 6.2 Tip comb package with two tip combs . ............................................................... 38
Fig. 6.3 Separating the two tip combs . ............................................................................. 38
Fig. 6.4 Storing the other tip comb on a MagMAX™ Express 96 plate for further use ...... 39
Fig. 6.5 Combining the tip comb and MagMAX™ Express 96 plate (A) ........................... 39
Fig. 6.6 Combining the tip comb and MagMAX™ Express 96 plate (B) ........................... 40
Fig. 6.7 Loading the tip-plate. ........................................................................................... 40
Fig. 6.8 Loading the plates ................................................................................................ 41
Fig. 6.9 Plates loaded ........................................................................................................ 41
Fig. 6.10 Tip comb in the tip comb holder . ........................................................................ 42
Fig. 6.11 Shield plate in function. ....................................................................................... 42
Fig. 6.12 MagMAX™ Express 96 plate. .............................................................................. 43
Fig. 6.13 Fastening/changing the heating block . ............................................................... 44
Fig. 6.14 Two different heating blocks available . ............................................................... 44
Fig. 11.1 KingFisher Software CD . ..................................................................................... 62
Fig. 14.1 MagMAX™ Express 96 keypad . .......................................................................... 69
Table 6.1 Processing volumes vs. plate types. .................................................................... 43
Chapter 1: Safety Symbols and Markings
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 7
1 Safety Symbols and Markings
These symbols are intended to draw your attention to particularly important
information and alert you to the presence of hazards as indicated.
Safety symbols and markings used on the MagMAX™ Express 96
Power ON
Power OFF
Warning: Hot surface, risk of burns.
Warning: Risk of body parts, hair, jewelry or clothing getting caught in a moving
part.
Serial number
Catalog number
Date of manufacture
Consult instructions for use
WEEE symbol This product is required to comply with the European Union’s Waste
Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC.
A black label with the following text (Fig. 4.3, item 6):
WARNING: DISCONNECT SUPPLY BEFORE SERVICING
AVERTISSEMENT: COUPER L'ALIMENTATION AVANT L'ENTRETIEN ET LE
DEPANNAGE
Warning markings used in the documentation
Warning: Risk of electric shock.
Chapter 1: Safety Symbols and Markings
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 8
Warning: Biohazard risk.
Warning: Risk of injury to the user(s).
Other markings used in the documentation
Note: Marks a hint, important information that is useful in the optimum
operation of the system, or an item of interest.
1.1 Instrument safety and guidelines for use
1. Always follow basic safety precautions when using the MagMAX™ Express
96 to reduce the risk of injury, biohazardous contamination, fire, or electrical
shock.
2. Read the user manual in its entirety prior to operating the instrument.
Failure to read, understand, and follow the instructions in the manual may
result in damage to the instrument, injury to laboratory and operating
personnel or poor instrument performance.
3. Observe all Warning and Note statements as well as safety symbols and
markings on the instrument and in the documentation.
4. The device shall be operated only with software specifically designed for
the device.
5. Never open any other covers of the MagMAX™ Express 96 than the see-
through lid (Fig. 4.2, item 1) or the sliding door (Fig. 4.2, item 2) while the
instrument is plugged into a power source.
6. Only when the instrument is switched off, can you push in the heads or turn
the turntable manually.
7. Never force a microplate into the instrument.
8. The MagMAX™ Express 96 is intended for laboratory research use only.
Observe proper laboratory safety precautions, such as wearing protective
clothing and following approved laboratory safety procedures.
9. Preventative maintenance instructions should be followed closely to keep
the instrument in the best condition for maximum reliability. A poorly
maintained instrument will not give the best results.
1.1.1 Chemical Safety Guidelines
To minimize the hazards of chemicals:
Read and understand the Material Safety Data Sheets (MSDSs) provided by
the chemical manufacturer before you store, handle, or work with any
chemicals or hazardous materials.
Minimize contact with chemicals. Wear appropriate personal protective
equipment when handling chemicals (for example, safety glasses, gloves,
or protective clothing). For additional safety guidelines, consult the MSDS.
Chapter 1: Safety Symbols and Markings
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 9
Minimize the inhalation of chemicals. Do not leave chemical containers
open. Use only with adequate ventilation (for example, fume hood). For
additional safety guidelines, consult the MSDS.
Check regularly for chemical leaks or spills. If a leak or spill occurs, follow
the manufacturer’s cleanup procedures as recommended in the MSDS.
Comply with all local, state/provincial, or national laws and regulations
related to chemical storage, handling, and disposal.
1.1.2 Biohazard Warning
Biological samples such as tissues, body fluids, infectious agents, and blood of
humans and other animals have the potential to transmit infectious diseases.
Follow all applicable local, state/provincial, and/or national regulations. Wear
appropriate protective equipment, which includes but is not limited to:
protective eyewear, face shield, clothing/lab coat, and gloves. All work should
be conducted in properly equipped facilities using the appropriate safety
equipment (for example, physical containment devices). Individuals should be
trained according to applicable regulatory and company/institution
requirements before working with potentially infectious materials. Read and
follow the applicable guidelines and/or regulatory requirements in the
following:
U.S. Department of Health and Human Services guidelines published in
Biosafety in Microbiological and Biomedical Laboratories (stock no. 017-
040-00547-4; http://bmbl.od.nih.gov).
Occupational Safety and Health Standards, Bloodborne Pathogens (29
CFR§1910.1030; http://www.access.gpo.gov/nara/cfr/waisidx_01/
29cfr1910a_01.html).
Your company’s/institution’s Biosafety Program protocols for working
with/handling potentially infectious materials. Additional information about
biohazard guidelines is available at: http://www.cdc.gov.
Chapter 2: About the User Manual
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 10
2 About the User Manual
The user manual has been written for the actual user (for example, laboratory
technician) and provides information on the MagMAX™ Express 96 magnetic
particle processor, including the installation and operating instructions.
Read the manual in its entirety prior to operating the instrument.
The user manual has been designed to give you the information you need to:
Review safety precautions
Install the MagMAX™ Express 96
Use the MagMAX™ Express 96 in routine jobs – the processing step
Perform basic cleaning and maintenance procedures
Troubleshoot the instrument performance
The user manual also describes all the features and specifications of the
MagMAX™ Express 96 hardware and on-board software.
Sections 4.2 through 4.4 explain the processing principles and procedures.
In Chapter 9 Troubleshooting Guide you will find explanations of all error
messages and a problem-solving guide. The user should be familiar with the
contents of Chapter 7 on maintenance.
For warranty and ordering information, refer to Chapters 10 Warranty
Information and 11 Ordering Information.
For further software-related issues, refer to the KingFisher Software User
Manual (Cat. no. 1508540).
In an effort to produce useful and appropriate documentation, we appreciate
your comments on this user manual.
2.1 Preface
2.1.1 Purpose of This Guide
The Applied Biosystems MagMAX™ Express 96 User Guide provides reference
material related to the Applied Biosystems MagMAX™ Express 96 Magnetic
Particle Processor. It describes the system, theory of operation, and how to
prepare, maintain, and troubleshoot the system.
2.1.2 Audience
This guide is intended for novice and experienced MagMAX™ Express 96 users
who will:
Install the Applied Biosystems MagMAX™ Express 96
Perform routine runs using the Applied Biosystems MagMAX™ Express 96
Perform planned maintenance procedures
2.1.3 Assumptions
This guide assumes that your Applied Biosystems MagMAX™ Express 96 has
been installed by an Applied Biosystems technical representative. This guide
also assumes that you have a working knowledge of the Windows operating
system.
Chapter 2: About the User Manual
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 11
2.1.4 Text Conventions
This guide uses the following conventions:
Bold text indicates user action. For example:
Type 0, then press Enter for each of the remaining fields.
Italic text indicates new or important words and is also used for emphasis.
For example:
Before analyzing, always prepare fresh matrix.
A right arrow symbol () separates successive commands you select from
a drop-down or shortcut menu. For example:
Select File Open Spot Set.
Right-click the sample row, then select View Filter View All Runs.
2.1.5 User Attention Words
Two user attention words appear in Applied Biosystems user documentation.
Each word implies a particular level of observation or action as described
below:
Note: – Provides information that may be of interest or help but is not critical to
the use of the product.
IMPORTANT! – Provides information that is necessary for proper instrument
operation, accurate chemistry kit use, or safe use of a chemical.
Examples of the user attention words appear below:
Note: The Calibrate function is also available in the Control Console.
IMPORTANT! To verify your client connection to the database, you need a valid
user ID and password.
2.1.6 Safety Alert Words
Safety alert words also appear in user documentation. For more information,
see “Safety Symbols and Markings” on page 7.
2.2 How to Obtain More Information
2.2.1 Related Documentation
The following related documents are shipped with the system:
Applied Biosystems MagMAX™ Express 96 User Guide – Describes the
Applied Biosystems MagMAX™ Express 96 hardware and software and
provides information on preparing, maintaining, and troubleshooting the
system.
Applied Biosystems MagMAX™ Express 96 Service Guide – Provides
explanatory procedures for servicing the Magnetic Particle Processors.
Chapter 2: About the User Manual
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 12
2.3 How to Obtain Support
For the latest services and support information for all locations, go to
http://www.appliedbiosystems.com, then click the link for Support.
At the Support page, you can:
Search through frequently asked questions (FAQs)
Submit a question directly to Technical Support
Order Applied Biosystems user documents, MSDSs, certificates of analysis,
and other related documents
Download PDF documents
Obtain information about customer training
Download software updates and patches
In addition, the Support page provides access to worldwide telephone and fax
numbers to contact Applied Biosystems Technical Support and Sales facilities.
Chapter 3: Introduction to the MagMAX™ Express 96
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 13
3 Introduction to the MagMAX™ Express 96
3.1 Intended use
The MagMAX™ Express 96 magnetic particle processor (Fig. 3.1) is intended for
professional research by trained personnel. The instrument is intended for
automated transfer and processing of magnetic particles in a microplate scale.
Use for self-testing is excluded. It is recommended that Good Laboratory
Practices (GLP) are followed to guarantee reliable analyses.
Refer to Chapter 8 Technical Specifications.
3.2 Principle of operation
The MagMAX™ Express 96 magnetic particle processor (Fig. 3.1) is designed
for automated transfer and processing of magnetic particles in microplate
format.
The technology of the MagMAX™ Express 96 system is based on the use of
magnetic rods covered with a disposable, specially designed tip comb and
plates. The instrument functions without any dispensing or aspiration parts or
devices.
Samples and reagents including magnetic particles are dispensed into the plates
according to the corresponding instructions. The protocol that is selected by
the user via the keypad and display have already been preloaded into the on-
board software. KingFisher PC software can be used to create and run proto-
cols.
Fig. 3.1 MagMAX™ Express 96 magnetic particle processor
3.3 Advantages of processing magnetic particles with MagMAX™ Express 96
The MagMAX™ Express 96 magnetic particle processor provides several
advantages relating mainly to the principle of operation in that it:
Enables automation of complicated manual steps
Enables simultaneous processing and purification
Incubation is enhanced due to the heating option
Enables concentration of the sample during processing
Facilitates a good collection of bead-bound sample due to the efficiency of
the magnet
Enables a quicker reaction and a more efficient wash due to the
simultaneous shaking during incubation and washing
No carryover
No cross contamination
Facilitates the whole processing with the aid of an internal program/
KingFisher Software.
Chapter 4: Functional Description
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 14
4 Functional Description
4.1 Instrument layout
4.1.1 Front view
1. Display
2. Keypad
3. Loading station
4. Heating block
5. Interchangeable MagMAX™ Express 96 head
6. Shield plate
7. Turntable
8. Plate station (1 – 8)
Fig. 4.1 MagMAX™ Express 96 front view without see-through lid
1. See-through lid
2. Sliding door
Fig. 4.2 MagMAX™ Express 96 front view with see-through lid and plates
2
6
5
1 4
2
1
3
8
7
Chapter 4: Functional Description
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 15
4.1.2 Back / Side view
1. Cooling-air inlet
2. Type label
3. USB port (not in use)
4. Ethernet (not in use)
5. Serial port RS-232C
6. Warning marking
7. Mains power supply connector
8. ON/OFF switch
9. Cooling-air outlet
Fig. 4.3 MagMAX™ Express 96 back view
Fig. 4.4 MagMAX™ Express 96 side view
1 5 7 8 9
3 4 6
2
Chapter 4: Functional Description
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 16
4.2 MagMAX™ Express 96 magnetic particle processor
The MagMAX™ Express 96 (Fig. 4.5) has room for eight plates. The tip combs
are compatible with the plates. During the individual steps, the plates are kept
stationary, and the only moving assembly is the processing head with tip comb
and magnetic rods. The head consists of two vertically moving platforms. One
is needed for the magnetic rods (96 pieces) and the other one for the plastic tip
comb (Fig. 5.15).
Up to eight plates can be simultaneously on the turntable. However, during one
sample processing, the protocol enables the use of more than eight plates in
total. One tip comb contains 96 tips used for processing 96 samples at a time.
Before starting the magnetic particle processing via the keypad and display, the
samples and reagents are dispensed into the plates and the tip comb is placed
onto a MagMAX™ Express 96 plate, from which it is automatically loaded. The
plates are placed onto the turntable into the corresponding plate stations
according to the protocol instructions. During the operation, the sliding door
can be closed or open (Fig. 4.2) or the whole see-through lid can be absent. The
closed door protects the processing against environmental contamination.
The operating principle employed is MPP (inverse magnetic particle processing)
technology (Fig. 4.6). Rather than moving the liquids, the magnetic particles are
moved from plate to plate containing specific reagents, in contrast to the
external magnet method. Magnetic particles are transferred with the aid of
magnetic rods covered with a disposable, specially designed plastic tip comb.
Fig. 4.5 MagMAX™ Express 96 magnetic particle processor
Chapter 4: Functional Description
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 17
4.3 Principle of magnetic particle processing
Fig. 4.6 Inverse magnetic particle processing
4.4 Working with a magnetic rod
Working with magnetic particles can be divided into five separate processes:
collecting magnetic particles
releasing magnetic particles
washing magnetic particles
incubation
concentration
4.4.1 Collecting magnetic particles
During the collection of the magnetic particles, the magnetic rod is fully inside
the tip. The magnetic rods together with the tip comb move slowly up and
down in the plate and the magnetic particles are collected onto the edge of the
tips. The magnetic rods together with the tip comb, having collected the
magnetic particles, can be lifted out of the plate and transferred into the next
plate.
4.4.2 Releasing magnetic particles
After collection of the magnetic particles, the magnetic rods together with the
tip comb are lifted from the plate, the magnetic rods are lifted off and the tip
comb is lowered into the next plate containing a reagent.
Magnetic particles are released by moving the tip comb up and down several
times at considerably high speed until all the particles have been mixed with the
substance in the next reaction.
Chapter 4: Functional Description
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 18
4.4.3 Washing magnetic particles
Washing the magnetic particles is a frequent and an important processing
phase. Washing is a combination of the release and collection processes in a
plate filled with washing solution.
To maximize washing efficiency, the magnetic rods together with the tip comb
are designed to have minimized liquid-carrying properties.
4.4.4 Incubation
To keep the magnetic particle suspension evenly mixed in long-running
reactions, the tip comb can be moved up and down in the solution.
4.4.5 Changing the volume during the magnetic particle processing
The volume of the first plate can be larger than the volume of the next plate,
and this is used for concentration purposes (see Fig. 4.7 below).
Fig. 4.7 A concentration step during magnetic particle processing
Chapter 5: Installation
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 19
5 Installation
5.1 What to do upon delivery
5.1.1 How to unpack
Move the packed instrument to its site of operation. Unpack the MagMAX™
Express 96 instrument and accessories carefully with the arrows on the
transport package pointing upwards. The following notes and instructions are
sent with the instrument and are immediately available when you open the
package:
the packing instructions/packing list
the MagMAX™ Express 96 Feedback Form
the Warranty Certificate card
the MagMAX™ Express 96 User Manual/Quick Reference Guide
Note: Do not touch or loosen any screws or parts other than those specifically
designated in the instructions. Doing so might cause misalignment and
will void the instrument warranty.
Warning: The MagMAX™ Express 96 weighs approximately 28 kg [62 lbs.] without
the transport package and should be lifted with care. It is recommended
that two persons lift the instrument together, taking the proper
precautions to avoid injury.
To lift the instrument, put your fingers under the bottom on either sides and lift
it with your back straight, while another person is helping offset the weight.
Retain the original packaging for future transportation. The packaging is
designed to assure safe transport and minimize transit damage. Use of
alternative packaging materials may invalidate the warranty. Also retain all
instrument-related documentation provided by the manufacturer for future use.
5.1.2 Checking delivery for completeness
Check the enclosed packing list against order. If any parts are missing, contact
your local Applied Biosystems representative or Applied Biosystems.
5.1.3 Checking for damage during transport
Visually inspect the transport package, the instrument and the accessories for
any possible transport damage.
If the carton has been damaged in transit, it is particularly important that you
retain it for inspection by the carrier in case there has also been damage to the
instrument.
Visually check all interconnections in the basic instrument. Check that there are
no loose parts inside the instrument.
If any parts are damaged, contact your local Applied Biosystems representative
or Applied Biosystems.
5.1.4 Environmental requirements
When you set up your MagMAX™ Express 96, avoid sites of operation with
excess dust, vibrations, strong magnetic fields, direct sunlight, draft, excessive
moisture or large temperature fluctuations.
Chapter 5: Installation
Applied Biosystems MagMAX™ Express 96 User Manual; Rev. 1.5; January 2011 20
Make sure the working area is flat, dry, clean and vibration-proof and leave
additional room for cables, lids, etc.
Make sure the ambient air is clean and free of corrosive vapors, smoke and
dust.
Make sure the ambient temperature range is between +10°C (50°F) and
+40°C (104°F).
Make sure relative humidity is between 10% and 80% (non-condensing).
Leave sufficient space (at least 10 cm) at both sides and at the back/side of the
unit to allow adequate air circulation.
The MagMAX™ Express 96 does not produce operating noise at a level that
would be harmful. No sound level measurements are required after installation.
Note: Do not operate the instrument in an environment where potentially
damaging liquids or gases are present.
5.1.5 Things to avoid
Do not smoke, eat or drink while using the MagMAX™ Express 96. Wash your
hands thoroughly after handling test fluids. Observe normal laboratory
procedures for handling potentially dangerous samples. Use proper protective
clothing. Use disposable gloves. Be sure the working area is well-ventilated.
Never spill fluids in or on the equipment.
Note: The MagMAX™ Express 96 should not be kept in close proximity to
magnetic tapes, computer discs or other magnetic storage systems,
such as credit cards, as these can be damaged by the strong magnetic
field of the MagMAX™ Express 96 heads.
Do not hold the MagMAX™ Express 96 heads close to a PC display,
since this may cause damage to the display.
Do not use metal tools when handling MagMAX™ Express 96 heads.
Warning: This product contains very strong permanent magnets.
People wearing a pacemaker or metallic prostheses should not use this
product. A pacemaker or prostheses may be affected or damaged if it
comes in close contact with a strong magnetic field.
5.1.6 Technical prerequisites
Place the instrument on a normal laboratory bench. The net weight of the unit is
approx. 28 kg [62 lbs.].
The instrument operates at voltages of 100 – 240 Vac and the frequency range
50/60 Hz.
5.2 Setups before you put the instrument into operation
5.2.1 How to release the transport locks
Note: THERE ARE TWO TRANSPORT LOCKS PRESENT (Fig. 5.1): the tip comb
holder transport lock, and the heating block transport lock. Remove both
the transport locks. Ensure that the transport locks have been released
before you put the instrument into operation.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73

Thermo Fisher Scientific MagMAX™ Express-96 Magnetic Particle Processor Owner's manual

Type
Owner's manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI