Instruction Manual
748183-K
April 2002
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755
NOTE
For NEMA 3R service, all conduit must be
connected through approved fittings.
The analyzer is supplied, as ordered, for
operation on 120 VAC or 240 VAC, 50/60 Hz.
Make sure that the power source conforms to
the requirements of the individual instrument,
as noted on the name-rating plate.
2-4 ELECTRICAL CONNECTIONS
a. Line Power Connections
Electrical power is supplied to the
analyzer via a customer-supplied three-
conductor cable, type SJT, minimum wire
size 18 AWG. Route power cable through
conduit and into appropriate opening in
the instrument case. Refer to Installation
Drawing (632349 or 643127). Connect
power leads to HOT, NEUT, and GND
terminals on TB1, Figure 2-1. Connect
analyzer to power source via an external
fuse or breaker, in accordance with local
codes.
NOTE
Do not draw power for associated
equipment from the analyzer power
cable.
b. Recorder Output Selection and Cable
Connections
If a recorder, controller, or other output
device is used, connect it to the analyzer
via a 22 or 24 AWG two-conductor
shielded cable. Route the cable through
conduit to the analyzer, and into the case
through the appropriate opening shown in
Installation Drawing (632349 or 643127).
Connect the shield only at the recorder
end.
NOTE
Route recorder cable through a
separate conduit, not with power cable
or alarm output cable.
Cable connections and output selection
for potentiometric and current-actuated
devices are explained below.
NOTE
Do not allow internal cable service
loop to touch the shock-mounted
detector assembly or associated
sample inlet and outlet tubing. This
precaution ensures against possible
transmission of mechanical vibration
through the cable to the detector,
which could cause noisy readout.
Potentiometric Output
1. Insert RECORDER OUTPUT Selector
Plug (Figure 2-2, page 2-4) in position
appropriate to the desired output: 10
mV, 100 mV, 1 V, or 5 V.
2. On TB2 (Figure 2-1, page 2-3)
connect leads of shielded recorder
cable to MV+ and COM terminals.
3. Connect free end of output cable to
appropriate terminals of recorder or
other potentiometric device:
a. For device with a span of 0 to
10mV, 0 to 100mV, 0 to 1V, or 0 to
5V, connect cable directly to input
terminals of the device, making
sure polarity is correct.
b. For device with intermediate span,
i.e., between the specified values,
connect cable to device via a
suitable external voltage divider, as
shown in Figure 2-3, page 2-5.
Isolated Current Output (Option)
The isolated current output board (Figure
2-2, page 2-4) is optional, and can be
adjusted for either 0 to 20 mA or 4 to 20
mA. The adjustments made on this board
are for zero and span. To set output:
1. With analyzer meter at zero, adjust
R1 for desired zero level (typically 0
for 0 to 20 mA, 4 for 4 to 20 mA).