M1 Series

Itm M1 Series User guide

  • Hello! I've reviewed the m1 Quick Reference guide and am ready to help you with any questions you may have about this gauge. I can provide information on its various settings, such as pressure units, averaging modes, power management options, and troubleshooting tips found in the document.
  • How do I access the setup options?
    How do I return to operating mode?
    What does Err 1 mean?
    What does Err 2 mean?
Power On
Power O
Zero
Backlight
20 PSI, 1.4 bar,
140 kPa, 1.4kg/cm2
1
sec >
setup
setup
2
sec
1
sec
m1 Quick Reference
www. .com information@itm.com1.800.561.8187
All units are not available with every pressure range.
*H20 icons and temperature settings will only display on H20 enabled units.
Gauge Setup Basics
Gauge Setup Map
1 Access Setup
1
x
2 View Setup options 3 Select an option
Maximum, Minimum, or Averaging
(ashing HI, LO, and AVG icons)
Hi On ( )
Lo On ( )
A On ( )
=+1
=
PSI
mmHg
"Hg
mmH2O*
"H2O*
kg/cm2
mbar
bar
MPa
kPa
2
x
Pressure Units
=+1
=
t: 60F
t: 68F
next
( )
next
( )
next
( )
next
( )
next
( )
enter
( )
(Averaging window: 2 to 20 samples)
Note: Pressing the enter button always returns the gauge to operating mode. If you want to set multiple Display Functions,
press the setup button to navigate from one Display Function to the next, then press the enter button when you are done.
m1Quick Reference
www. .com information@itm.com1.800.561.8187
Note: Whenever one of the rd settings is selected,
the battery icon will change from to
indicating that the M1 is in Ultra Low Power Mode.
Gauge Setup Map continued
Troubleshooting
Err 1 displayed:
The M1 checks the internal calibration every time it’s turned on.
Contact factory for further instructions.
Err 2 displayed:
The M1 has tried to display a number too large
for the display due to electrical malfunction or numerical error.
Contact factory for further instructions.
Err 5 or Err 6 displayed:
The M1 pressure sensor is exhibiting behavior out of normal
operating condition. Contact factory for sensor replacement.
Power Management *
+
=+1
=
next
( )
Lt: 15 (light stays on for 15 seconds)
Lt: 60 (light stays on for 60 seconds)
Lt: On (light does not turn o automatically)
3
x
Backlight Timer
To Clear Peaks or Line Pressure
next
( )
setup
Note: If there is no gauge activity for 10 seconds, the gauge
will return to operating mode. Any Setup options that you have
selected will be enabled.
Ultra Low Power Mode
crystalengineering.net
*This menu can be accessed only when the gauge is o.
*AtPower On,” this setting is displayed as No Auto o
*AtPower On,” this setting is displayed as Auto o 1H
*
O - -*. . . . . (4/1 second . . . . . 600 hours)
O 1H**. . . (4/1 second . . . . . 600 hours)
rd 4.0. . . . . . . (1/15 seconds . . . 3,000 hours)
rd 2.0. . . . . . . (1/ 30 seconds. . . 6,000 hours)
rd 1.0. . . . . . . (1/1 minute . . . . . 10,000 hours)
rd 0.2. . . . . . . (1/5 minutes . . . . 25,000 hours)
rd 0.1. . . . . . . (1/10 minutes . . . 30,000 hours)