Campbell Scientific 4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules Owner's manual

Category
Measuring, testing & control
Type
Owner's manual

This manual is also suitable for

4WFBS120, 4WFBS350, 4WFBS1K
4 Wire Full Bridge Terminal
Input Modules
Revision: 3/12
Copyright © 1996-2010
Campbell Scientific, Inc.
Warranty
“PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are
warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in
materials and workmanship under normal use and service for twelve (12)
months from date of shipment unless otherwise specified in the corresponding
Campbell pricelist or product manual. Products not manufactured, but that are
re-sold by Campbell, are warranted only to the limits extended by the original
manufacturer. Batteries, fine-wire thermocouples, desiccant, and other
consumables have no warranty. Campbell's obligation under this warranty is
limited to repairing or replacing (at Campbell's option) defective products,
which shall be the sole and exclusive remedy under this warranty. The
customer shall assume all costs of removing, reinstalling, and shipping
defective products to Campbell. Campbell will return such products by surface
carrier prepaid within the continental United States of America. To all other
locations, Campbell will return such products best way CIP (Port of Entry)
INCOTERM® 2010, prepaid. This warranty shall not apply to any products
which have been subjected to modification, misuse, neglect, improper service,
accidents of nature, or shipping damage. This warranty is in lieu of all other
warranties, expressed or implied. The warranty for installation services
performed by Campbell such as programming to customer specifications,
electrical connections to products manufactured by Campbell, and product
specific training, is part of Campbell’s product warranty. CAMPBELL
EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. Campbell is not liable for any special, indirect,
incidental, and/or consequential damages.”
Assistance
Products may not be returned without prior authorization. The following
contact information is for US and international customers residing in countries
served by Campbell Scientific, Inc. directly. Affiliate companies handle
repairs for customers within their territories. Please visit
www.campbellsci.com to determine which Campbell Scientific company serves
your country.
To obtain a Returned Materials Authorization (RMA), contact CAMPBELL
SCIENTIFIC, INC., phone (435) 227-9000. After an applications engineer
determines the nature of the problem, an RMA number will be issued. Please
write this number clearly on the outside of the shipping container. Campbell
Scientific's shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____
815 West 1800 North
Logan, Utah 84321-1784
For all returns, the customer must fill out a "Statement of Product Cleanliness
and Decontamination" form and comply with the requirements specified in it.
The form is available from our web site at www.campbellsci.com/repair. A
completed form must be either emailed to repair@campbellsci.com or faxed to
(435) 227-9106. Campbell Scientific is unable to process any returns until we
receive this form. If the form is not received within three days of product
receipt or is incomplete, the product will be returned to the customer at the
customer's expense. Campbell Scientific reserves the right to refuse service on
products that were exposed to contaminants that may cause health or safety
concerns for our employees.
4WFBS120, 4WFBS350, 4WFBS1K
Table of Contents
PDF viewers: These page numbers refer to the printed version of this document. Use the
PDF reader bookmarks tab for links to specific sections.
1. Function........................................................................1
2. Specifications ..............................................................1
3. Measurement Concepts ..............................................2
4. Quarter Bridge Strain ..................................................4
4.1 Quarter Bridge Strain with 3 Wire Strain Element...................................4
4.1.1 Quarter Bridge Strain with 3 Wire Element Wiring .......................5
4.1.1.1 Quarter Bridge Strain with 3 Wire Element Wiring using
a multiplexer.......................................................................5
4.1.2 Quarter Bridge Strain with 3 Wire Element Calculations...............6
4.1.3 Quarter Bridge Strain with 3 Wire Program Examples ..................7
4.1.3.1 CRBasic Programming..........................................................7
4.1.3.2 Edlog ...................................................................................11
4.2 Quarter Bridge Strain with 2 Wire Element ...........................................17
4.2.1 Quarter Bridge Strain with 2 Wire Element Wiring .....................17
4.2.2 Two Wire ¼ Bridge use with Multiplexers and Equations...........18
4.3 Quarter Bridge Strain with Dummy Gage ..............................................18
4.3.1 Quarter Bridge Strain with Dummy Gauge Wiring Setup ............20
4.3.2 Quarter Bridge Strain with Dummy Gauge Calculations..............21
4.3.3 Quarter Bridge Strain with Dummy Gauge Example Programs ...21
4.4 Quarter Bridge Strain Lead Resistance Compensation...........................21
4.4.1 Mathematical Lead Compensation for 3-Wire, ¼ Bridge Strain ..21
4.4.1.1 Mathematical Lead Compensation Circuit and Equations ..22
4.4.1.2 Mathematical Lead Compensation Programs......................23
4.4.2 Shunt Calibration Lead Compensation for 3-Wire, ¼ Bridge
Strain .........................................................................................30
4.4.2.1 Three Wire Gage Circuit with Shunt...................................30
4.4.2.2 Math for Shunt Calibration of 3-Wire, ¼ Bridge Strain
Circuits .............................................................................32
4.4.2.3 Example Programs for Shunt Calibration of 3-Wire,
¼ Bridge Strain Circuits...................................................34
4.4.3 Lead Compensation using Quarter Bridge Strain with 2 Wire
Element .....................................................................................35
4.5 Calculation of Strain for ¼ Bridge Circuits............................................37
i
4WFBS120, 4WFBS350, 4WFBS1K Table of Contents
ii
Figures
1-1. Terminal Input Module with CR1000 .................................................... 1
2-1. Schematic................................................................................................ 2
3-1. Strain definition ...................................................................................... 2
4.1-1. Three wire quarter bridge strain circuit ............................................... 4
4.1-2. 3-wire ¼ bridge strain wiring .............................................................. 5
4.1-3. 3-wire ¼ bridge strain with multiplexer wiring................................... 5
4.2-1. Two wire quarter bridge strain circuit ............................................... 17
4.2-2. Wiring for 2-wire gauges................................................................... 18
4.3-1. Quarter bridge strain circuit with dummy gauge ............................... 19
4.3-2. ¼ bridge strain with remote dummy gauge ....................................... 20
4.3-3. ¼ bridge strain with dummy gauge at datalogger.............................. 20
4.4-1. Three wire ¼ bridge strain circuit...................................................... 22
4.4-2. Shunting remotely across active gauge.............................................. 31
4.4-3. Circuit for shunting across dummy resistor....................................... 31
4.4-4. Wiring for shunt across dummy resistor............................................ 32
4.4-5. Two wire quarter bridge strain circuit ............................................... 35
4.5-1. Strain gage in full bridge ................................................................... 37
Table
4-1. Input Locations Used in CR10(X), 21X, and CR7 Examples .............. 11
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire
Full Bridge Terminal Input Modules (TIM)
1. Function
The 4WFBS120, 4WFBS350, and 4WFBS1K Terminal Input Modules (TIM)
complete a full Wheatstone bridge for a single strain gage or other sensor that
acts as a single variable resistor. The difference between the three models is in
the resistor that matches the nominal resistance of a 120 ohm, 350 ohm, or
1000 ohm quarter bridge strain gage. It can also be used to complete the back
half of a Wheatstone bridge for use in a ¼ bridge strain circuit (1 active
element) using a dummy gage, or in a ½ bridge strain circuit (2 active
elements).
FIGURE 1-1. Terminal Input Module with CR1000
2. Specifications
2:1 Resistive Divider
Resistors:
1 kΩ/1 kΩ
Ratio tolerance @ 25 °C:
±0.01%
Ratio temperature coefficient:
0.5 ppm/°C
(-55°C to 85°C)
Power rating per element:
0.1 W @ 70°C
Completion Resistor: 120, 350, or 1000 Ω
Tolerance @ 25 °C:
±0.01%
Temperature coefficient:
±0.8 ppm °C
-1
(-55°C to 85°C)
Power rating:
0.25 W @ 70°C
1
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
FIGURE 2-1. Schematic
3. Measurement Concepts
Measuring strain is measuring a change in length. Specifically, the unit strain
()
ε
is the change in length divided by the unstrained length
(
)
LL /Δ=
ε
,
and thus is dimensionless.
L +
Δ
L
L
L
T
L
T
+
Δ
L
T
PP
FIGURE 3-1. Strain definition
As the subject is elongated in the longitudinal direction, the material will be
narrowed or thinned down in the transverse direction. The ratio of the
transverse strain to the longitudinal strain is known as the Poisson ratio (
ν).
L
L
L
L
T
T
Δ
Δ
=
ν
3.1
This Poisson ratio is a known property for most materials and is used in some
half bridge strain and full bridge strain circuits.
Strain is typically reported in microstrain
()
με
. Microstrain is strain
expressed in parts per million, i.e.: a change in length divided by one millionth
of the length.
A metal foil strain gage is a resistive element that changes resistance as it is
stretched or compressed. The strain gage is bonded to the object in which
strain is measured. The gage factor,
GF , is the ratio of the relative change in
resistance to the change in strain:
GF R R l l
=
Δ
Δ
//
. For example, a
2
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
gage factor of 2 means that if the length changes by one micrometer per meter
of length
()1
μ
ε
, the resistance will change by two micro-ohms per ohm of
resistance. A more common method of portraying this equation is:
G
G
RGF
R
Δ
=
ε
3.2
Or in terms of micro-strain:
(
)
G
RGF
=
με
V
G
RΔ×
6
101
3.3
Because the actual change in resistance is small, a full Wheatstone bridge
configuration is used to give the maximum resolution. The Wheatstone bridge
can be set up with 1 active gage (Quarter bridge strain circuit), two active
gages (Half bridge strain circuit), or 4 active gages (Full bridge strain circuit).
For each of these Wheatstone bridge circuits there are multiple configurations.
The 4WFBS module provides three resistors that can be used for three of the
arms of the Wheatstone Bridge (Figure 4-1). There are two 1000 ohm
precision resistors for the back plane of the Wheatstone bridge, and a resistor
matching the strain gage's resistance for the bridge arm opposite the gage. The
inputs of the 4WFBS are configured so that this matching resistor can be
bypassed if it is desired to utilize a dummy gauge, or to use two active gauges
(Half Bridge Strain circuit).
For Full Bridge Strain circuits, as all four arms of the Wheatstone bridge are
active gages, there is no need for completion resistors, and thus a 4WFBS
module is not required.
The resistance of an installed gage will differ from the nominal value. In
addition, lead resistance imbalances can result in further unbalancing of the
bridge. A zero measurement can be made with the gage installed. This zero
measurement can be incorporated into the datalogger program such that
subsequent measurements can report strain relative to this zero basis point.
This removes the apparent strain resulting from the initial bridge imbalance.
Strain is calculated in terms of the result of the full bridge measurement. This
result is the measured bridge output voltage divided by the bridge excitation
voltage:
V
ou
ex
/
.
All of the various equations that are used to calculate strain use V
r
, the change
in the bridge measurement from the zero state:
ZeroexoutStrainedexoutr
VVVVV )/()/(
=
3.4
(
)
The result of the zero measurement,
Zero
exout
VV /
, can be stored and used in
the calculation of future strain measurements. Alternatively, the zero reading
value can be left at 0 (zero measurement is neither recorded nor used).
It should be noted the actual result of the full bridge instruction (BrFull) is the
millivolts output per volt of excitation (
1000 VV
out ex
/
). The StrainCalc
3
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
function used in CRBasic uses this raw output as its input to calculate µstrain.
See Section 4.5 Calculation of Strain for ¼ Bridge Circuits for a detailed
derivation of the equations used.
4. Quarter Bridge Strain
A "quarter bridge strain circuit" is so named because an active strain gage is
used as one of the four resistive elements that make up a full Wheatstone
bridge. The other three arms of the bridge are composed of inactive elements.
There are various circuits that use a single active element, including 2-Wire
gauges, 3-Wire gauges, as well as a few circuits that utilize a dummy gauge for
the arm opposite the arm holding the active gage instead of a resistor, R
D
in
Figure 4.1.-1 (See Figures 4.3-1, 4.3-2, and 4.3-3). The 4WFBS TIM modules
can support all types of these ¼ Bridge Strain circuits.
4.1 Quarter Bridge Strain with 3 Wire Strain Element
A 3-wire quarter bridge strain circuit is shown in figure 4.1-1. Strain gages are
available in nominal resistances of 120, 350, and 1000 ohms. The
4WFBSXXX model must match the nominal resistance of the gage when using
the 3-Wire circuit (e.g., the 4WFBS120 is used with a 120 ohm strain gage).
In Figure 4.1-1, R
1
and R
2
are 1000 ohm resistors making up the back plane of
the Wheatstone bridge, as is done in the TIM design. R
D
, the third resistive
element, is the complementary resistor that has a nominal resistance of the un-
strained gage. The 4
th
resistive element is the active strain gage.
Excite V
+
-
R
2
=1 KΩ
R
1
=1 K
Ω
R
D
R
4
= Gauge
L
3
L
2
L
1
FIGURE 4.1-1. Three wire quarter bridge strain circuit
The 3-Wire gage alleviates many of the issues of the 2-Wire gage. As can be
seen in Figure 4.1-1, lead wire L
3
is in the arm of the Wheatstone bridge that
has the completion resistor while lead wire L
1
is in the arm that has the active
gage. L
2
is tied back to the input channel of the datalogger that has an input
resistance greater than 1 Gohm, thus the current flow is negligible, negating
effects of L
2
’s resistance. This circuit nulls temperature induced resistance
changes in the leads as well as reduces the sensitivity effect that the wires have
on the gauge. See Section 4.4 for more on Lead resistance effects and methods
to compensate for them.
4
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
4.1.1 Quarter Bridge Strain with 3 Wire Element Wiring
Figure 4.1-2 illustrates the wiring of the strain gage to the 4WFBS module and
the wiring of the module to the datalogger. It is important that the gage be
wired as shown, and that the leads to the L and G terminals be the same length,
diameter, and wire type. It is preferable to use a twisted pair for these two
wires so that they will undergo the same temperature and electromagnetic field
variations. With this configuration, changes in wire resistance due to
temperature occur equally in both arms of the bridge with negligible effect on
the output from the bridge.
Datalogger
A
ctive Gauge
or G
A
G
L
H
VX or E
X
4WFBSXXX TIM
R
2
=1K R
1
=1K
R
D
Shunt Receptacle
Shunt Receptacle
FIGURE 4.1-2. 3-wire ¼ bridge strain wiring
4.1.1.1 Quarter Bridge Strain with 3 Wire Element Wiring using a multiplexer
When using a mechanical relay multiplexer such as the AM16/32B, the
4WFBS module should normally be placed on the face of the multiplexer
similar as shown in Figure 4.1-3.
H L H LH LHL
H
LHLHLHLHL
H LHL H L
2 345
ODD EVEN
13
4X16
2X32
21
22
23 24
25
N
O
1
HLG
AM16/32B Relay Multiplexer
4
WFBS
23
COM
7
11
12
13
RES
CLK
GND
12V
CR800
CR850
G
12 V
G
C1–C4
C1–C4
CR10X
CR1000
CR3000
G
12 V
G
C1–C8
C1–C8
CR23X
CR5000
12 V
G
C1–C8
C1–C8
21X
+12 V
EXCIT 1–4
C1–C6
CR7
12 V
EXCITATION
725 Card
Control
CR10X
AG
E1–E3
1L
1H
CR23X
EX1–EX4
1L
1H
CR1000
EX1–EX3 or
VX1–VX3
1L
1H
CR3000
CR5000
VX1–VX4
1L
1H
21X
EXCITATION
1–4
1L
1H
CR7
SWITCHED
ANALOG OUT
1L
1H
CR800
CR850
EX1–EX2 or
VX10VX2
1L
1H
Cable Shield
FIGURE 4.1-3. 3-wire ¼ bridge strain with multiplexer wiring
5
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
Although this requires a 4WFBS module for each strain gage, it is important
because placing relays internal a Wheatstone bridge strain system is
discouraged. Any change in resistance of the multiplexer’s relay contacts
would result in a corresponding change in the bridge’s output voltage.
Changes in contact resistance can be induced by temperature fluctuations,
oxidation, environmental conditions, and normal wear of contact surfaces. The
specification for the relays that are used in our multiplexers state that initial
contact resistance will be less than 100 milliohms (AM16/32B). There is not a
specification for change in contact resistance for the relays because there are so
many variables that affect contact resistance. Test reports exist for various test
conditions that show contact resistance changing over time by 10 to 20 milli-
Ohms. These tests were performed using static test temperatures, so it is safe
to assume that real world conditions would result in larger resistance shifts.
When strain gauges are used in the Wheatstone bridge, small changes in
contact resistance result in large apparent strains. To understand the error that
can be introduced from allowing the relay contacts to be internal of the
Wheatstone bridge, let us assume that the two relays carrying the current from
the strain gage vary by 20 milliohms (40 milliohm total variance or ΔR
G
= 40
m ). Inserting this into equation 3.3, using a 120 ohm strain gage with a gage
factor of 2 results in an apparent strain of about 167 με.
(
)
Ω×
=
1202
167
με
V
out ex
/
Ω×× 04.0101
6
4.1.2 Quarter Bridge Strain with 3 Wire Element Calculations
As noted in Section 3, in real life applications the Wheatstone bridge starts out
unbalanced. The strain gauge is never perfectly at its nominal resistance even
prior to installation. The installation process can lead to even more deviation
from this nominal state. In addition, lead resistance can cause an initial
apparent strain reading. To remove this initial offset, a zero measurement can
be made with the gauge installed. This zero measurement can be incorporated
into the datalogger program and subsequent measurements can report strain
relative to this zero basis point.
Strain is calculated in terms of the result of the full bridge measurement. This
result is the measured bridge output voltage divided by the bridge excitation
voltage
V
. (The actual result of the full bridge instruction is the
millivolts output per volt of excitation,
1000
V
out
/ V
ex
) The result of the zero
measurement,
1000
0
VV
out ex
/
can be stored and used to calculate future
strain measurements. The change in the full bridge measurement from the zero
state, V
r
, is used in the calculation of the strain.
VVV V V
r out ex out ex
=
(/)( /)
0
4.1.1
Using V
r
from equation 4.1.1, the strain is calculated using equation 4.1.2.
ε
=
4
12
V
GF V
r
r
()
4.1.2
The calculations are covered in more detail in Section 4.5.
6
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
4.1.3 Quarter Bridge Strain with 3 Wire Program Examples
This section is broken out into CRBasic programs and EDLOG programs.
These programs are only to be used as examples. Besides adding additional
measurement instructions, the programs will need to have the scan and data
storage intervals altered for actual applications. Refer to the datalogger’s
manuals and/or the CRBasic Editor’s help files for detailed information on the
program instructions used as well as additional program examples.
4.1.3.1 CRBasic Programming
Dataloggers that use CRBasic include our CR800, CR850, CR1000, CR3000,
CR5000, and CR9000(X). CRBasic uses the StrainCalc Instruction for
calculating strain from the output of different full bridge configurations:
StrainCalc(Dest,Reps,Source,BrZero,BrConfig,GageFactor,PoissonRatio)
Source is the variable holding the current result from the full bridge
measurement
BrZero is the zero measurement; this parameter uses the results of a previous
full bridge measurement instruction when the gage is at the zero condition
(multiplier=1, offset=0, mV/V) directly.
BRCode for the Bridge Configuration used with the 4WFBS module should be
set to -1 for a quarter bridge strain circuit.
Enter the actual gage factor in the GageFactor parameter.
Enter 0 for the Poisson ratio parameter, which is not used with ¼ Bridge strain
circuits.
Example Program 4.1. CR9000X ¼ bridge Strain with 3 reps
This example program measures the output from the Wheatstone bridge using
the BrFull instruction. The output from this instruction is input into the
StrainCalc instruction in order to calculate the raw µstrain value. This
program does not use a zero offset reading. See Example Program 4.2 for an
example that performs a zero calibration.
' Program name: STRAIN.C9X
Public StrainMvperV(3) : Units StrainMvperV = mV_per_V 'Raw Strain dimensioned source
Public Strain(3) : Units Strain = uStrain uStrain dimensioned source
Public GF(3) 'Dimensioned gauge factor
DataTable(STRAIN,True,-1) 'Trigger, auto size
DataInterval(0,0,0,100) 'Synchronous, 100 lapses, autosize
CardOut(0,-1) 'PC card , size Auto
Sample (3,Strain(),IEEE4) '3 Reps, uStrain, Resolution
Sample (3,StrainMvperV(),IEEE4) ‘3Reps,Stain mVolt/Volt, Resolution
EndTable 'End of table STRAIN
BeginProg 'Program begins here
GF(1) = 2.1 : GF(2) = 2.2 : GF(3) = 2.3 'Initialize gauge factors for Strain( )
7
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
Scan(10,mSec,100,0) 'Scan once every 10 mSecs, non-burst
BrFull(StrainMvperV(),3,mV50,4,1,5,7,1,5000,True,True,70,100,1,0)
StrainCalc(Strain(),3,StrainMvperV(),0,-1,GF(),0) 'Strain calculation
CallTable STRAIN
Next Scan 'Loop up for the next scan
SlowSequence 'Slow sequence Scan to perform temperature
Scan(1,Sec,0,0) ' compensation on DAQ
Calibrate 'Corrects ADC offset and gain
BiasComp 'Corrects ADC bias current
Next Scan
EndProg 'Program ends here
Example Program 4.2. CR9000X ¼ bridge Strain with 3 reps and zero offset
This example program starts out with Example Program 4.1 and adds
instructions (highlighted) to perform a zero calibration. As all strain circuits
have a zero or initial imbalance that is related to the circuit rather than the
member undergoing strain, a zero reading is often used to offset or remove this
apparent strain. Again, see the manual and CRBasic editor’s Help file for
more in-depth discussion on the instructions.
The FieldCalStrain instruction takes care of the underlying math for the
zeroing using equation 4.1.2.
The LoadFieldCal instruction facilitates the reloading of the calibration
factors when the logger is powered up. In addition, the programmer should
create a DataTable (we have called this DataTable Calib in the example) to
store the calibration factors each time a calibration is done.
The NewFieldCal is a Boolean flag variable that is only high during the Scan
that a calibration has been completed. It is used in the DataTable instruction’s
trigger parameter to trigger the table to record a record.
The SampleFieldCal output instruction is used to inform the logger to store all
of the calibration factors that are controlled using the FieldCalStrain
instruction.
'
Program name: STRAIN0.C9X
Public StrainMvperV(3) : Units StrainMvperV = mV_per_V 'Raw Strain dimensioned source
Public Strain(3) : Units Strain = uStrain uStrain dimensioned source
Public GF(3) 'Dimensioned gauge factor
Public ZeromV_V(3), ZeroStrain(3)
Public ZReps, ZIndex, ModeVar
DataTable(STRAIN,True,-1) 'Trigger, auto size
DataInterval(0,0,0,100) 'Synchronous, 100 lapses, autosize
CardOut(0,-1) 'PC card , size Auto
Sample (3,Strain(),IEEE4) '3 Reps, uStrain, Resolution
Sample (3,StrainMvperV(),IEEE4) ‘3Reps,Stain mVolt/Volt, Resolution
EndTable 'End of table STRAIN
DataTable (Calib,NewFieldCal,10) ‘Table for calibration factors from zeroing
SampleFieldCal ‘User should collect these to his computer
EndTable ‘for future reference
8
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
BeginProg 'Program begins here
GF(1) = 2.1 : GF(2) = 2.2 : GF(3) = 2.3 'Initialize gauge factors for Strain( )
ZReps = 3 : ZIndex = 1 ‘initialize cal reps and index pointer
LoadFieldCal(True) ‘Load prior calibration factors
Scan(10,mSec,100,0) 'Scan once every 10 mSecs, non-burst
FieldCalStrain(10,StrainMvperV(),ZReps,0,ZeromV_V(),ModeVar,0,ZIndex,1,0,Strain())
BrFull(StrainMvperV(),3,mV50,4,1,5,7,1,5000,True,True,70,100,1,0)
StrainCalc(Strain(),3,StrainMvperV(),ZeromV_V(),-1,GF(),0) 'Strain calculation
CallTable STRAIN
CallTable Calib
Next Scan 'Loop up for the next scan
SlowSequence 'Slow sequence Scan to perform
Scan(1,Sec,0,0) 'temperature compensation on the DAQ
Calibrate 'Corrects ADC offset and gain
BiasComp 'Corrects ADC bias current
Next Scan
EndProg 'Program ends here
Example Program 4.3. CR1000 ¼ Bridge Strain with 3 reps and zero offset
This example program performs the same tasks as Example Program 4.2, only
it is a CR1000 program instead of a CR9000X program. There are slight
differences such as range codes and the fact that the CR1000 does not have a
Slot parameter for its measurement instructions. This program is more similar
to what a CR800, CR3000, or a CR5000 program would look like than the
CR9000X program.
'
Program name: STRAIN0.CR1
Public StrainMvperV(3) : Units StrainMvperV = mV_per_V 'Raw Strain dimensioned source
Public Strain(3) : Units Strain = uStrain uStrain dimensioned source
Public GF(3) 'Dimensioned gauge factor
Public ZeromV_V(3), ZeroStrain(3)
Public ZReps, ZIndex, ModeVar
DataTable(STRAIN,True,-1) 'Trigger, auto size
DataInterval(0,0,0,100) 'Synchronous, 100 lapses, autosize
CardOut(0,-1) 'PC card , size Auto
Sample (3,Strain(),IEEE4) '3 Reps, uStrain, Resolution
Sample (3,StrainMvperV(),IEEE4) ‘3Reps,Stain mVolt/Volt, Resolution
EndTable 'End of table STRAIN
DataTable (Calib,NewFieldCal,10) ‘Table for calibration factors from zeroing
SampleFieldCal ‘User should collect these to his computer
EndTable ‘for future reference
BeginProg 'Program begins here
GF(1) = 2.1 : GF(2) = 2.2 : GF(3) = 2.3 'Initialize gauge factors for Strain( )
ZReps = 3 : ZIndex = 1 ‘initialize cal reps and index pointer
LoadFieldCal(True) ‘Load prior calibration factors
Scan(100,mSec,100,0) 'Scan once every 10 mSecs, non-burst
FieldCalStrain(10,StrainMvperV(),ZReps,0,ZeromV_V(),ModeVar,0,ZIndex,1,0,Strain())
BrFull(StrainMvperV(),3,mV7_5,1,1,3,2500,True,True,450,500,1,0)
StrainCalc(Strain(),3,StrainMvperV(),ZeromV_V(),-1,GF(),0) 'Strain calculation
CallTable STRAIN
CallTable Calib
Next Scan 'Loop up for the next scan
9
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
Example Program 4.3. CR1000 ¼ Bridge Strain using an AM16/32B
Multiplexer with 16 reps and zero offset
This example program has 16 strain gages multiplexed through an AM16/32
Multiplexer and uses FieldCalStrain for zeroing.
'
Program name: QuarterStrain with Zero and Mux.CR1
' This is only an example program and should be used only for help in creating a usable program
' ----------------------------------------------- WIRING ----------------------------------------------------------
' CR1000 to AM16/32 Multiplexer Control
' C1 (Control Port 1) Res (Reset)
' C2 (Control Port 2) Clk (Clock)
' G GND (Ground)
' 12V 12V
' CR1000 to AM16/32 Common TIMs to AM16/32 Banks
' Diff 1H to Common Even Hi Blk Wire to Bank Odd Lo
' Diff 1L to Common Even Lo TIM H to Bank Even Hi
' EX1 to Common Odd Lo Tim L to Bank Even Lo
' AG to Common Gnd Tim AG to Bank Even AG
'\\\\\\\\\\\\\\\\\\\\\\\DECLARE VARIABLES and CONSTANTS ///////////////////////
Const REPS = 16 'Strain gage sensor count
Public MVpV(REPS) : Units MVpV = mV_V 'mV per Volt output from Bridge Measurement
Public STRAIN(REPS) : Units STRAIN = uStrain 'Variable where uS is stored,
Const BATCH_GF = 2.1 : Public GF(REPS) 'Batch Gage Factor
Public mV_VZero(REPS) : Units mV_VZero = mV_V 'Variable for Zero mV per V reading
Public CalReps, ZeroMode, ZeroStartIdx, ZeroCalAvgs 'Used by wizard for zeroing
Public CalFileLoaded As Boolean
Dim I
'\\\\IF DESIRED (NOT REQUIRED): GIVE STRAIN VARIABLES UNIQUE ALIAS NAMES ////////
Alias STRAIN(1) = Strain1 : Alias STRAIN(2) = Strain2 : Alias STRAIN(3) = Strain3
Alias STRAIN(4) = Strain4 : Alias STRAIN(5) = Strain5 : Alias STRAIN(6) = Strain6
Alias STRAIN(7) = Strain7 : Alias STRAIN(8) = Strain8 : Alias STRAIN(9) = Strain9
Alias STRAIN(10) = Strain10 : Alias STRAIN(11) = Strain11 : Alias STRAIN(12) = Strain12
Alias STRAIN(13) = Strain13 : Alias STRAIN(14) = Strain14 : Alias STRAIN(15) = Strain15
Alias STRAIN(16) = Strain16
'\\\\\\\\\\\\\\\\\\\\\\\\ OUTPUT SECTION ////////////////////////
' Table STRAIN stores uStrain and raw mV per Volt measurements to the PC Card
DataTable(STRAIN,True,-1) 'Trigger, auto size
DataInterval(0,0,0,100) 'Synchronous, 100 lapses
CardOut(0,-1) 'PC card , Autosize
Sample (REPS,STRAIN(),IEEE4) 'Sample uStrain
Sample (Reps,mVpV(),IEEE4) 'Sample raw mV per Volt values
EndTable 'End of table
' Table CalHist uses SampleFieldCal which stores all of the Calibration constants
' When a calibration function is complete, user should always collect this Table as a record
DataTable(CalHist,NewFieldCal,50)
SampleFieldCal
EndTable
'\\\\\\\\\\\\\\\\\\\\\\\\MAIN PROGRAM SECTION ////////////////////////
BeginProg 'Program begins here
For I = 1 To REPS ' For the 16 gages
GF(I) = BATCH_GF 'Assign default gauge factor (2.1) to GF array elements
Next I 'Loop back up until complete
CalFileLoaded = LoadFieldCal(1) 'Load the Cal constants if program signature matches
10
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
Scan(1,Sec,10,0) 'Scan once a Second
PortSet (1 ,1 ) 'Turn on AM16/32 using C1
I = 1
Delay (0,150,mSec) 'required Delay for AM16/32 multiplexer
SubScan (0,0,16)
PulsePort (2,10000) 'Pulse port C2 hi and low to clock the multiplexer
BrFull(MVpV(I),1,mV7_5C,1,VX1,1,2500,True,True,250,500,1,0) 'Full Bridge measurement
StrainCalc(Strain(I),1,MVpV(I),mV_VZero(I),-1,GF(I),0) 'Strain calculation
I = I + 1 'Increment I
NextSubScan
PortSet (1 ,0 ) 'Turn on AM16/32 using C1
FieldCalStrain(10,MVpV(),CalReps,0,mV_VZero(),ZeroMode,0,ZeroStartIdx,ZeroCalAvgs,0,STRAIN())
CallTable CalHist
CallTable STRAIN
Next Scan 'Loop up for the next scan
EndProg 'Program ends here
4.1.3.2 Edlog
The following examples for the CR10(X), 21X, and CR7 all have subroutines
that measures the unstrained "zero" output of the strain gage. The examples
calculate strain using equation 4.1.2 for a strain gage with a GF=2. These are
just examples. Besides adding additional measurement instructions, the
programs will probably need to have the scan and data storage intervals altered
for actual applications. The instructions in the subroutine will also need to be
modified for the actual gage factor.
Dataloggers that use Edlog include CR510, CR10(X), 21X, and CR7. The
Edlog instruction that is used to measure strain gages is Instruction 6 – Full
Bridge.
The Input Locations assignments used in CR10(X), 21X, and CR7 Examples
are listed in Table 4-1.
TABLE 4-1. Input Locations Used in
CR10(X), 21X, and CR7 Examples
Addr Name
1 mVperV
2 mVperV_0
3 Vr
4 uStrain
5 Count
6 GF
7 _4e6
8 Mult
9 1_2Vr
10 Vr_1_2Vr
11
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
Example Program 4.4. CR10X ¼ Bridge Strain with 1 rep and zero offset
;{CR10X}
*Table 1 Program
01: 1 Execution Interval (seconds)
1: If Flag/Port (P91) ;On the first execution (Flag 1 is low)
1: 21 Do if Flag 1 is Low ;or when user sets Flag 1 low
2: 1 Call Subroutine 1 ;call the zeroing subroutine
2: Full Bridge (P6) ;Measure the strain gage
1: 1 Reps
2: 22 ± 7.5 mV 60 Hz Rejection Range
3: 1 DIFF Channel
4: 1 Excite all reps withExchan 1
5: 2500 mV Excitation
6: 1 Loc [ mVperV ]
7: 1 Mult
8: 0 Offset
3: X-Y (P35) ;Subtract zero reading from the
1: 1 X Loc [ mVperV ] ;measurement
2: 2 Y Loc [ mVperV_0 ]
3: 3 Z Loc [ Vr ]
4: X*F (P37) ;Change Vr from mV/V to V/V
1: 3 Loc [ Vr ]
2: 0.001
3: 3 Loc [ Vr ]
;The following instructions calculate microstrain
5: Z=X*F (P37)
1: 3 X Loc [ Vr ]
2: -2 F
3: 9 Z Loc [ 1_2Vr ]
6: Z=Z+1 (P32)
1: 9 Z Loc [ 1_2Vr ]
7: Z=X/Y (P38)
1: 3 X Loc [ Vr ]
2: 9 Y Loc [ 1_2Vr ]
3: 10 Loc [ Vr_1_2Vr ]
8: Z=X*Y (P36)
1: 10 X Loc [ Vr_1_2Vr ]
2: 8 Y Loc [ Mult ]
3: 4 Z Loc [ uStrain ]
; Output Section : This example outputs an average of the 1 second readings ;once per minute.
09: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 1 Interval (same units as above)
3: 10 Set Output Flag High
10: Set Active Storage Area (P80)
1: 1 Final Storage Area 1
2: 1 Array ID ;Set Array ID = 1 for measurement data
11: Real Time (P77)
1: 1110 Year,Day,Hour/Minute
12
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
12: Average (P71)
1: 1 Reps
2: 4 Loc [ uStrain ]
*Table 2 Program
2: 0.0000 Execution Interval (seconds)
*Table 3 Subroutines
1: Beginning of Subroutine (P85) ;Subroutine to measure "zero"
1: 1 Subroutine 1
2: Do (P86) ;This prevents calling subroutine
1: 11 Set Flag 1 High ;until user sets flag 1 low again.
3: Z=F (P30) ;Set counter use for average to 0
1: 0 F
2: 0 Exponent of 10
3: 5 Z Loc [ Count ]
4: Z=F (P30) ;load 4 million (4*uS/S) into input location
1: 4 F
2: 6 Exponent of 10
3: 7 Z Loc [ _4e6 ]
5: Z=F (P30) ;Load Gage Factor into input location
1: 2 F ;Enter the actual Gage Factor here
2: 0 Exponent of 10
3: 6 Z Loc [ GF ]
6: Z=X/Y (P38) ;calculate multiplier to use with strain
1: 7 X Loc [ _4e6 ] ;calculation
2: 6 Y Loc [ GF ]
3: 8 Z Loc [ Mult ]
7: Beginning of Loop (P87) ;Loop through 5 times to obtain average
1: 0 Delay ;zero reading
2: 5 Loop Count
8: Z=Z+1 (P32) ;Increment Counter used to determine
1: 5 Z Loc [ Count ] ;when to output
9: Full Bridge (P6) ;Measure Strain Gage
1: 1 Reps
2: 22 ± 7.5 mV 60 Hz Rejection Range
3: 1 DIFF Channel
4: 1 Excite all reps withExchan 1
5: 2500 mV Excitation
6: 1 Loc [ mVperV ]
7: 1 Mult
8: 0 Offset
10: IF (X<=>F) (P89) ;Check for last pass through loop
1: 5 X Loc [ Count ] ;to set output flag
2: 3 >=
3: 5 F
4: 10 Set Output Flag High
11: Set Active Storage Area (P80) ;Direct averaged "zero" reading
1: 3 Input Storage Area ;to input storage
2: 2 Array ID or Loc [ mVperV_0 ]
13
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM)
12: Average (P71)
1: 1 Reps
2: 1 Loc [ mVperV ]
13: If Flag/Port (P91) ;When average is calculated,
1: 10 Do if Output Flag is High (Flag 0) ;also send it to Final Storage
2: 10 Set Output Flag High
14: Set Active Storage Area (P80) ;Direct Output to Final Storage
1: 1 Final Storage Area 1
2: 11 Array ID ;set Array ID = 11 for zero data
15: Real Time (P77)
1: 110 Day,Hour/Minute
16: Sample (P70)
1: 1 Reps
2: 2 Loc [ mVperV_0 ]
17: End (P95)
18: End (P95)
End Program
Example Program 4.5. 21X ¼ Bridge Strain with 1 rep and zero offset
;{21X}
*Table 1 Program
01: 1 Execution Interval (seconds)
;Other measurements could be inserted here or before the Output section
1: If Flag/Port (P91) ;On the first execution (Flag 1 is low)
1: 21 Do if Flag 1 is Low ;or when user sets Flag 1 low
2: 1 Call Subroutine 1 ;call the zeroing subroutine
2: Full Bridge (P6) ;Measure the strain gage
1: 1 Reps
2: 2 ± 15 mV Slow Range
3: 1 DIFF Channel
4: 1 Excite all reps withExchan 1
5: 5000 mV Excitation
6: 1 Loc [ mVperV ]
7: 1 Mult
8: 0 Offset
3: Z=X-Y (P35) ;Subtract zero reading from the
1: 1 X Loc [ mVperV ] ;measurement
2: 2 Y Loc [ mVperV_0 ]
3: 3 Z Loc [ Vr ]
4: Z=X*F (P37) ;Change Vr from mV/V to V/V
1: 3 X Loc [ Vr ]
2: 0.001 F
3: 3 Z Loc [ Vr ]
14
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46

Campbell Scientific 4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules Owner's manual

Category
Measuring, testing & control
Type
Owner's manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI