SNA-20/-23 Service Manual
2 Safety Instructions
2
Important
Notes
2.1
Safety Instructions
2.1.1
Preventing Electrical Accidents
Safety class
This instrument belongs to safety class I as defined by lEG publication 348 and VDE 0411. The
AC line connector shall not be inserted into AC outlets without a protective ground contact, ex-
cept in
specially certified areas.
Checks before repairs and maintenance
Checking the instrument construction
No changes in instrument construction which reduce safety may be made.
Checking the protective
ground
Check the connection and general condition by visual inspection and measure the resistance
between the protective ground contact and the AC line plug and the instrument enclosure. The
resistance must be
<
0.1
n.
Shake the cable during the measurement. Resistance variations
mean that the cable
is
faulty.
Checking the insulation resistance
Measure the insulation resistance between the AC line terminals which
have
been shorted to-
gether and the protective conductor terminal of the instrument using a
500 V insulation tester.
Ensure that the instrument's
AC
line switch
is
in the ON position. The insulation resistance must
be>
2 MQ.
Opening the instrument
When covers are
removed,
or
parts are extracted with tools, live components may be exposed.
Even contact points
may
be live.
Therefore, before opening the instrument, disconnect all voltage sources.
If it
is
absolutely essential to carry out calibration, maintenance
or
repairs while the instrument
is
opened and connected to a voltage supply, these procedures may only be carried out by a
competent technician
who
is
aware of the associated risks.
Capacitors in the instrument may
still be charged; consult the circuit diagrams.
Fuses
Only specified fuses shall be used.
Repairs, replacing components
Repairs must be carried out by a competent technician. No changes in the instrument construc-
tion, which reduce its safety shall be made.
In
particular, this applies to creepage paths and
component spacing.
Always use original spare parts whenever possible.
Other spare parts may only be used if the
safety specifications of the instrument are not reduced.
2-1