Roche cobas u 411, a reliable and user-friendly analyzer, empowers you with accurate results for various clinical chemistry parameters. Its compact design fits seamlessly into small labs or point-of-care settings, enabling quick and convenient testing. The intuitive interface simplifies operation, while the comprehensive menu of assays caters to a wide range of routine and specialized testing needs. Experience fast and precise results with Roche cobas u 411, enhancing patient care with its efficient and dependable performance.

Roche cobas u 411, a reliable and user-friendly analyzer, empowers you with accurate results for various clinical chemistry parameters. Its compact design fits seamlessly into small labs or point-of-care settings, enabling quick and convenient testing. The intuitive interface simplifies operation, while the comprehensive menu of assays caters to a wide range of routine and specialized testing needs. Experience fast and precise results with Roche cobas u 411, enhancing patient care with its efficient and dependable performance.

cobas u 411 System
Short Guide version 1.0
COBAS, COBAS U and
LIFE NEEDS ANSWERS
are trademarks of Roche.
©2007 Roche Diagnostics
Roche Diagnostics GmbH
D-68298 Mannheim
Germany
www.roche-diagnostics.com
05064724018 - Ch. 00101181 - V.1.0
Processing samples from a work list
(barcoded ID entry)
1. Touch < Workplace >.
2. Touch < Work List >.
3. Scan the first sample barcode.
4. Touch
to enter color and clarity
(if necessary).
5. Repeat steps 3-4 for remaining samples.
6. Touch
to print the work list.
Note: Samples must be processed the order listed.
7. Load the first test strip.
Remove a test strip from the vial.
Dip the strip into the sample and wipe off
excessive sample on the edge of the tube.
Place sample test strip onto loading area.
8. Repeat step 7 for the remaining samples on the
work list.
Power-on / Daily checks
1. Power on the analyzer.
2. Login.
3. Empty waste area (if necessary).
4. Check paper.
Processing individual sample
(manual ID entry)
1. Touch < Workplace >.
2. Touch < Sample Entry >.
3. Enter the sample information:
Sample ID
Color and clarity (if necessary)
4. Touch
.
5. Load test strip:
Remove a test strip from the vial.
Dip the strip into the sample and wipe off
excessive sample on the edge of the tube.
Place sample test strip onto loading area.
6. Repeat steps 3-5 for the next sample.
Processing individual sample
(barcoded ID entry)
1. Touch < Workplace >.
2. Touch < Work List >.
3. Scan the sample barcode.
4. Touch
to enter color and clarity
(if necessary).
5. Load test strip:
Remove a test strip from the vial.
Dip the strip into the sample and wipe off
excessive sample on the edge of the tube.
Place sample test strip onto loading area.
6. Repeat steps 3-5 for the next sample.
Processing controls
1. Touch < Workplace >.
2. Touch < Run Controls >.
3. Select the control level that will be analyzed.
4. Load control test strip:
Remove a test strip from the vial.
Dip the strip into the appropriate control
material and wipe off excessive control on
the edge of the control tube.
Touch .
Place control test strip onto loading area.
5. Repeat steps 2-4 for remaining control levels.
Reviewing results
Display Results
1. Touch < Workplace >.
2. Touch
< Control List > to view control results
or
< Sample List > to view sample results.
3. Highlight the desired sample.
4. Touch
.
Print / Store / Delete Control Results
1. Touch < Workplace >.
2. Touch < Control List >.
3. Touch
.
4. Type the date range of the controls that
will be affected.
5. Touch
,
or
.
Print / Upload / Store / Delete Sample Results
1. Touch < Workplace >.
2. Touch < Sample list >.
3. Touch
.
4. Type the first and last sample sequence numbers
in the < To > and < From > fields.
5. Touch
,
,
or
.
Maintenance
Daily
1. Empty waste area.
2. Clean the test strip transfer system.
Monthly
1. Calibrate.
2. Clean interior of instrument.
3. Clean exterior of instrument.
Note: Verify the analyzer is powered OFF.
As Required
1. Replace main fuse.
2. Replace internal calibration strip.
Calibration
3. Touch < Utilities >.
4. Touch < Tools >.
5. Touch < Calibration >.
6. Touch
.
7. Touch
.
8. Place the calibration strip onto loading area.
9. Verify the calibration results.
cobas u 411 System
Short Guide version 1.0
  • Page 1 1
  • Page 2 2

Roche cobas u 411 Short Guide

Type
Short Guide

Roche cobas u 411, a reliable and user-friendly analyzer, empowers you with accurate results for various clinical chemistry parameters. Its compact design fits seamlessly into small labs or point-of-care settings, enabling quick and convenient testing. The intuitive interface simplifies operation, while the comprehensive menu of assays caters to a wide range of routine and specialized testing needs. Experience fast and precise results with Roche cobas u 411, enhancing patient care with its efficient and dependable performance.