Mammotome MEP10 User guide

Type
User guide

Mammotome MEP10, a versatile ultrasound probe, empowers you with precise tissue sampling capabilities. Designed for compatibility with the Mammotome Elite Holster, this 10G probe excels in capturing high-quality samples. Its 136mm needle length (without coaxial cannula) and 19.1mm aperture length provide ample reach and accuracy. Experience the ease of interrupting biopsy cycles and priming the probe with saline for optimal performance. Trust Mammotome MEP10 for reliable and efficient tissue sampling procedures.

Mammotome MEP10, a versatile ultrasound probe, empowers you with precise tissue sampling capabilities. Designed for compatibility with the Mammotome Elite Holster, this 10G probe excels in capturing high-quality samples. Its 136mm needle length (without coaxial cannula) and 19.1mm aperture length provide ample reach and accuracy. Experience the ease of interrupting biopsy cycles and priming the probe with saline for optimal performance. Trust Mammotome MEP10 for reliable and efficient tissue sampling procedures.

Quick Reference Guide
After cleaning and disinfecting, place the holster into the charging base. The
holster will emit a two-tone beep to indicate correct placement into the base.
When charging, set the holster on top
of this guide for quick, easy reference.
Holster On Charger
SOLID = Battery is fully charged.
FLASH = Battery is charging.
FLASH = Battery is charging. Holster must remain on charging base.
Holster Not On Charger
SOLID = Sufcient charge available to complete one or more procedure(s).
FLASH = Sufcient charge available to complete one procedure.
SOLID = Insufcient charge. Return holster to charging base.
Ensure the probe is outside the patient.
Disassemble the probe from the holster and replace it with a
new probe.
If the indicator light does not turn off, contact Mammotome
Technical Support.
Check the battery indicator to ensure there is sufcient charge.
Ensure the alignment and locking tabs are securely in place.
Ensure it is the initial use of the probe.
Replace the probe with a new probe.
If the indicator light does not turn on, contact Mammotome
Technical Support.
If the battery indicator light is not illuminated after probe loading:
If the fault indicator light is illuminated:
Troubleshooting
Charging the Holster
Needle Rotation (optional)
Pull the needle rotation knob out (B). Then rotate to desired clock position
and release (C).
Step 1
Grasp the probe body and align the alignment tab with the holster
notch (A). Push the probe base into the holster while maintaining the
alignment (B).
Step 2
Push the probe body firmly into the holster until a click is heard,
indicating the locking tabs have engaged the holster (C).
Note: Automatic initialization will begin.
Interrupting the Biopsy Cycle
Interrupt a biopsy cycle (as described below) leaving the aperture
exposed (A). Use a sterile syringe to insert 0.5cc of saline into the
needle aperture (approximately 7-10 drops). Press the active biopsy
button to advance the cutter.
Push the active biopsy button once to initiate the biopsy cycle. To
interrupt the biopsy cycle, push the active biopsy button again when
the cutter is retracting from position 1 to 3 (B) or when the cutter is
advancing from position 3 to 2. The cutter will retract to position 3,
and the biopsy indicator light will ash orange. Complete the cycle by
pushing the active biopsy button.
Introduce the probe through the incision and position the probe tip in
the target area with the tip past the lesion. Imaging guidance should be
used as appropriate. Push the active biopsy button on the holster once to
acquire a tissue sample. The biopsy cycle proceeds: a vacuum is created,
the inner cutter of the probe is retracted, tissue is drawn into the aperture,
and the cutter rotates and translates forward, cutting tissue (A).
When the cutter reaches the distal end of the aperture, the vacuum will
transport the sample to the specimen collection cup.
Ensure the probe is attached to the holster. Press and hold the
two inactive biopsy buttons simultaneously for three seconds.
There will be an audible indicator and all three indicator lights will
ash green (A).
Press and release the desired active biopsy button. The active
biopsy button is stored in the device for future use (B).
Please read the full Instructions For Use before using this product.
If the holster is left undisturbed for 10 minutes after loading a probe,
the holster will enter a sleep mode to conserve battery charge. During
sleep mode, the biopsy and battery indicator lights will turn off and the
buttons will be inactive. Once the holster is moved, the device will wake
from sleep mode. Any time the holster is motionless for one hour with
or without a probe attached, the holster will go into a deep sleep mode.
Place the holster into the charging base without the probe and remove
from the charging base to turn the holster back on.
Disconnect the integrated coaxial cannula and remove the probe by
pressing the coaxial release button, leaving the cannula in place.
Removing the Probe from the Coaxial Cannula
Post Sampling Instructions
Biopsy Button Selection
Loading the Probe
Priming the Probe
Tissue Sampling Procedure
Step 1
Rotate the specimen collection cap counterclockwise 1/4 turn (A).
Step 2
Remove the cap and tissue basket together (B).
Step 3
Press the tissue basket release button to remove the tissue basket from
the specimen collection cap for ease of transport to pathology (C).
Note: Sterile forceps may be used to aid in removing the samples from the
tissue basket, if needed.
1
2
3
5
4
6
While pressing down on the probes locking tabs, pull the probe out of the
holster cavity (A). Dispose of the probe in accordance with the institutions
policy for infectious waste disposal (B).
Removing the Probe from the Holster
A. A.
B. C.
A. B.
C.
A.
B.
B.
A.
A.
B.
B.
C.
Attribute MEP10 MEP13
Needle Length (w/o coaxial cannula) 136mm 136mm
Needle Length (w/ coaxial cannula) 122mm 122mm
Needle Diameter 3.5mm 2.6mm
Aperture Length 19.1mm 18.4mm
Dead Space 10.0mm 10.4mm
Tip to Center 19.6mm 19.6mm
Code Description Qty
MEH1 Mammotome® Elite Holster
(includes charger base and power cord) 1
MEP10 10G Mammotome® Elite Ultrasound Probe 5
MEP13 13G Mammotome® Elite Ultrasound Probe 5
MEI10 10G Mammotome® Elite Introducer Stylet 5
MEI13 13G Mammotome® Elite Introducer Stylet 5
For complete product details, see Instructions for Use.
Product may not be approved or available in your region. Please check with your local Mammotome representative.
© 2022 Devicor Medical Products, Inc. All rights reserved. Devicor, Mammotome and Mammotome Elite are registered trademarks of Devicor Medical Products, Inc., in the USA and optionally in
other countries. Other logos, product and/or company names might be trademarks of their respective owners. MDM# 12-0041 Rev 9/22
mammotome.com
Product Codes
Technical Support
Order Placement
Technical Specications
Contact your Customer Service Representative at 1-877-926-2666 between the hours of 8:00am and 5:00pm EST. Shipping
details and an order conrmation number will be provided to you upon completion of your order.
Send your order via fax to your Customer Service Representative at 1-888-260-6362.
Send your order via email to your Customer Service Representative at us.customerservice@mammotome.com.
For technical support contact us via phone at 1-877-926-2666 (option 3) or by email at TechnicalSupport@mammotome.com.
  • Page 1 1
  • Page 2 2
  • Page 3 3

Mammotome MEP10 User guide

Type
User guide

Mammotome MEP10, a versatile ultrasound probe, empowers you with precise tissue sampling capabilities. Designed for compatibility with the Mammotome Elite Holster, this 10G probe excels in capturing high-quality samples. Its 136mm needle length (without coaxial cannula) and 19.1mm aperture length provide ample reach and accuracy. Experience the ease of interrupting biopsy cycles and priming the probe with saline for optimal performance. Trust Mammotome MEP10 for reliable and efficient tissue sampling procedures.