Brunson 75, 75-H, 76-RH, 76-RH190, 76-RH190M, 771, 771-H Operation And Field Adjustment Manual

  • Hello! I am an AI chatbot trained to assist you with the Brunson 75 Operation And Field Adjustment Manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Models 771, 75, & 76-RH
Optical Tooling Transits
Instrument
Company
Operation and
Field Adjustment Manual
Helping the World Measure
Since 1927
www.brunson.us
Copyright 2009-2011 Brunson Instrument Company
Kansas City, Missouri
8000 E. 23rd Street
Kansas City, MO 64129
Tel: 816.483.3187
Fax: 816.241.1945
TableofContents
Thankyoufor
purchasing
aBrunsonTransit.
Rememberthatour
customersupportdoes
notstopaftershipmentof
aproduct—wearehereto
helpyouwithany
measurementchallenges
thatyoumayhave.
Chapter1:Introduction
1
WhatisaTransit?
1
WhatDoesaTransitDo?
3
TheThreeDifferentTransits
4
Chapter2:GettingStartedWithYourTransit
7
HandleYourTransitWithCare
7
RemovingFromtheCase
8
UsingtheShiftingCenter
8
RoughLeveling
8
PreciseLeveling
9
TheReticle
11
HowtoProperlyFocus
11
FindingInfinityFocus
11
PointingtheTelescope
12
InstallingLightingAccessories
12
Chapter3:MeasuringTechniques
15
BuckingIn
15
Autoreflection
17
Autocollimation
19
Collimation
23
Collineation
26
ReticleProjection
27
MeasuringWith
Micrometers
28
Chapter4:FieldCalibrationChecks
35
Plate(“Bull’sEye”)Vial
37
MicrometerChecks
38
“DoubleCenter”Check
40
HorizontalAxisRunout
41
VerticalSpindleRunout
42
HorizontalCentering
43
HorizontalCollimation
44
PlumbLineCheck
46
CrossAxisTelescopeCollimation
49
PegTest
51
1 Operation and Adjustment Manual
Inthischapterwewilladdressthefollowingtopics:
 Whatisatransit?
 Whatarethecomponents?
 Whatdoesatransitdo?
 ThedifferencebetweenthethreeBrunsontransits
Themostimportantthingaboutatransitisthatithasatelescopeon
agimballingmechanism,soitcanrotatebackandforthhorizon
tally(azimuthdirection),butitcanalsorotateupanddown
(elevationdirection).
Almosteveryoneisfamiliarwithasurveyor’stransit.However,it
isworthexplainingthegeneraldifferencebetweenourtransitsand
thoseusedbysurveyors.OpticalToolingtransitshaveanumberof
modificationstomakethemdeadlyaccuratewhenshootingover
distanceswhich,toasurveyor,wouldseemshort(lessthanacou
pleofhundredfeet).Inaddition,wedonotneedtotakereadings
intwotelescopepositions(directandreverse)inordertoachieve
thisaccuracy.Thisispossiblebecauseourtransitshaveseveral
specialcharacteristics:
a)Theyhaveanextremelystraightlineofsight.Asyoufocusthe
telescopefromneartofar,thelineofsighttravelsinanextremely
straightline.Thatis,pictureputtinganimaginary,
weightless,tautstringdownthecenterofthetelescope,
stretchingstraightoutacoupleofhundredfeet—the
stringwouldalwaysbeinthecenterofthecrosshairsas
youfocus
fromneartofarandbackagain.
b)Atransit’sthreemajoraxesareallmutuallyorthogonal
(allatrightanglestoeachother).Therearetworota
tionalaxesonyourtransit,andoneaxiswhichdoesnot
rotate(seeFigure1.2).Thefirstrotationalaxisisthever
ticalaxis.Thespindleandbearingswhichcreatethever
ticalaxisarehousedinthetransit’sbase.Theentireup
perportionofthetransit(standardsandtelescope)rotate
horizontallyaroundthisaxis.Thesecondrotationalaxis
isthecrossaxis.Thisisthehorizontalaxisaroundwhich
thetelescoperotates.Thethirdaxisisthelineofsight.
c)Lastly,allthreeofthemajoraxesmeetatoneprecise
point.Justbecausetheaxesareallmutuallyorthogonal
PleaserefertoFigure1.1
forageneralorientationto
componentterminology.
Chapter1‐Introduction
Note:Thismanualappliestothe
followingBrunsonproducts:
 771and771HJigTransits
 75and75HOpticalTransit
Squares
 76RH,76RH190,and76
RH190MTelescopicTransit
Squares
Atthistime,themodels771,771H,
75,and75Harenolongerpro
ducedbythefactory.However,we
haveincludedtheminthismanual
duetotheirpastpopularityandthe
factthatmanyofthese
instruments
remaininthehandsofalignment
technicianswhomayfindthisinfor
mationuseful.
Whatisatransit?
Model 771, 75, and 76-RH190 Transits 2
1. Levelingscrews(4)
2. “Spider”base
3. Platevial
4. Telescope(eyepieceend)
5. Crossaxistelescopeeyepiece
6. Focusknob
7. Telescope(objectiveend)
8. 190xOpticalMicrometer
9. Crossaxistelescopeobjectivelens
10.Verticaltangentclamplock
11.Coincidencelevel
12.Crossaxistelescopelightingadapter
13.Standards
14.Horizontaltangentclamplock
15.Baseplate
16.Horizontaltangentadjustmentscrew
17.Plate
18.Verticaltangentadjustmentscrew
19.Shiftingcenter
Figure1.1
OrientationtoComponents
3 Operation and Adjustment Manual
doesn’tnecessarilymeanthatthelineswouldallintersectatthe
sameexactpointinspace.However,onourtransits,theselinesdo
intersect.Thisisanimportantcharacteristicwhenaccuracyispara
mount,particularlyatshorterdistances.Ofcourse,youcan’tsee
thisintersectionpointitisburiedinthemiddleofthetelescope.
Thesearethethreedistinctcharacteristicsthatourtelescopeshave
whichallowthemtomakeveryprecisemeasurementsoverrela
tivelyshortworkingdistances.However,thereisonemorevery
distinctivecharacteristicwhichseparatesOpticalToolingtransits
fromallotherstheopticalcrossaxis.Our76RH190hasaninfin
ityfocusedtelescope,completewithasecondeyepieceandsecond
setofcrosswires,housedinthecrossaxis.Thiscrossaxistelescope
isusedinanumberofoperationstosetthemaintelescopeperpen
diculartoanothertelescopeoramirror(whichmaybemountedon
arotatingshaftorotherreferenceplane).
Atransitisaveryflexiblemeasuringtool.Itcanestablishvertical
orhorizontalopticalplaneswhichareextremelyflat.Itcanestab
lishopticallineswhichareextremelystraight,extendinginanydi
rection.Oncetheseplanesandlinesareestab
lished,youcanusethemasreferencesfrom
whichtomeasureotherpointsofinterest.You
candeterminewhethermachinecomponents
areparallel,flat,level,square,orstraight.This
meansyoucanevaluatethegeometricrela
tionshipsrelatedtothealignmentofalmost
anypart,structuralcomponent,machine,tool,
surface,substrate,wallorjustaboutany
thingelse.Averyfewexamplesofthiswould
be:
 Alignmentofbearingsandbearing
journals.
 Rollparallelisminpaperandmetals
mills,orprintingpresses
 Machinetoolevaluation:bedlevel
ing,bedflatness,geometricintegrity
ofspindlesand
othercomponents
 Layoutformachining
 Crankcasealignment
 Straightnessofenginebores,turbine
cases,torpedotubes,lineshafts,etc.
 Checkrelationshipsofrightangle
planes
TheOpticalToolinginstrumentsareextremely
flexible.Youcanusethemtomakejustabout
anymeasurementorgeometricevaluationthat
Figure1.2
Sowhatdoesatransitdo?
Model 771, 75, and 76-RH190 Transits 4
youcandesigninyourmind.
Asweindicatedearlier,the76RH190istheonlycurrentlypro
ducedOpticalToolingtransit.Althoughthismanualwaswritten
specificallyforthe76RH190,muchoftheinformationitcontainsis
applicabletotheothertransitsaswell.
Thedifferencesbetweenthe771x,75x,and76xtransitsrelate
solelytothewayinwhichthecrossaxisisconstructedandused.
Alltransitssharethesamebase,standardassembly,bearings,and
telescopeoptics.Onalltransits,thetelescoperotatesinavertical
planearoundahorizontalaxis(the“crossaxis”).Thecrossaxisis
perpendiculartothelineofsight.
Themostbasictransitisthe771or
771H.Thesetransitshavenoop
ticsinthecrossaxis.Theyare
usedas“jig”transitstoestablish
straightlinesandplanes.How
ever,theyhavenocapabilityto
makeanaccuraterightangleturn.
Onestep“up”isthemodel75or
75H(ontheleftinFigure1.3).
Thesetransitshaveacrossaxis
whichishollowandcanbeseen
through.Ononesideofthecross
axisisaclearglasswindow,and
ontheothersideisapartiallyre
flectivemirror.The75(H)does
everythingthatthe771(H)can
do,butcanalsobeusedincon
junctionwithanotherinstrument
toproduceaveryaccurateright
angle.
Themostflexibletransit(andthe
onlyonecurrentlyproduced)
is
themodel76RH190.Thecross
axisofthistransitcontainsatele
scopethatispermanentlyfocusedatinfinityandisequippedwith
reticlelighting.Arightangleeyepieceisprovidedasstandard
equipmentonthecrossaxistelescope.The76RH190maybeused
toestablishstraightlines,horizontalorverticalplanes,andcreate
accuraterightangleswithrespecttoothertransits,alignmenttele
scopes,mirrors,orotherphysicaltargets.Whenthecrosswiresof
thetransit’scrossaxistelescopearesetonthoseofareferencetele
scopeormirror,themaintelescopeofthetransitwillsweepaplane
atpreciserightanglestothecrossaxisreferenceline.
The75Htransit
(above,left)hasbeendis
continuedbutmanyare
stillinuse,andmostof
theinformationcon
tainedinthismanual
stillapplies.
Figure1.3
Thethreedifferenttransits
5 Operation and Adjustment Manual
TheH”inthevariousmodelnumbersreferstoa“Hollow”verti
calaxis.Transitsthathavethe‐Hdesignationallowyoutolook
straightdownthroughahollowverticalspindleandtoseetargets
orcollimatorswhicharedirectlybelowthebaseoftheinstrument.
Whydoweusefour“feet”(levelingscrews)insteadofthree?
Weareoftenaskedthisquestion,sincemosttheodoliteshavethreeleveling
screws.Thereasonisthatwhenadjustingfourlevelingscrews,theinstrument
heightisnotaffected.
Whenthreelevelingscrewsareused,twoof
themmustnecessarilyformapivotaxis,and
thethirdisusedtoraiseorloweronesidetotilt
theinstrument.Itgivestheeffectshownatleft.
Thereddotindicatestherelativeinstrument
heightbetweentwopositionsoftheinstrument
(tiltedandnottilted).
However,whenfourlevelingscrewsareused,
theinstrumentposition(again,reddot),effec
tively,doesnotchange.Thisgivestheeffect
shownatright.Onesidegoesupbythesame
amountastheothersidegoesdown,andthe
instrumentpivotsatitscenter,notononeside
ortheother.
Thisisjustanotherexampleofthewaythatwehavecraftedourinstrumentsto
bemoreaccurateatshorterranges.Overlongerdistances,minorchangesinin
strumentheightarenotimportant.Butwhenmeasuringveryaccuratelyover
shorterdistances,avoidingchangesininstrumentheightcanbecomecrucial.
Model 771, 75, and 76-RH190 Transits 6
7 Operation and Adjustment Manual
Inthischapter,wewilldiscuss
Notethattheseproductsareprecisioninstrumentsandshouldal
waysbehandledwithcare.
 Donotforceanyofthescrews.Afirm(butnottight)ten
siongivesthebestresults.
 Keepyourinstrumentcleananddry.Protectitfromthe
weatherifusedoutdoors.
 Allowyourinstrumentto“soak”foratleast4hoursin
thetemperatureinwhichitwillbeused,sothatshifts
duetothermalchangesareminimized.Ifyouaremov
ingthetransitacrossthermaldifferentialsofmorethan
20°F,allowittosoakfor8hours.Theinstrumentshould
beoutofthecaseduringthistimeforbetteraircircula
tion.
 Whentakingasighting,or
readingoneoftheseinstru
ments,removeyourhands,
andmakesureanyother
pressureisalsoremoved.
Keepyourinstrumentcalibrated.
Make
therecommendedfieldchecks
(seeChapter4)aswellasperforminga
completecalibrationatleastannually.
Notethatthereisawarningprintedon
thesideofyourtransitwhichsays
“Caution—DoNotPlungeTelescope
FromFlange”.Plungingthetelescope
meanstorotateitintheverticalplane,
asyoumightwhenyouwanttolook
furtherupordown.Specifically,
plungingreferstoreversingthedirec
tionofthetelescopebyrotatingit180°
Chapter2‐GettingStartedwithYourTransit
 Howtohandleyourtransit
 Removingitfromthecase
 Usingtheshiftingcenter
 Roughleveling
 Preciseleveling
 TheReticle
 Focusingtheeyepieceand
telescope
 FindingInfinityfocus
 Pointingthetelescope
 Installinglightingaccesso
ries
Figure2.1
Handleyourtransitwithcare
Model 771, 75, and 76-RH190 Transits 8
intheverticalplane.The“flange”isthecomponentuponwhichthe
warningisprinted.Thissimplymeansnottorotatethetelescope
bygrabbingtheflangeandtwistingit.Torotatethetelescopeprop
erly,putyourhandonthetelescopebarrelitself.
Loosenboththehorizontalandverticaltangentclampscrews.
Graspthestandardsfirmlywithonehand,andthebasewiththe
otherhand.Liftgentlyfromcaseandplaceonastandorothersup
port.Whilemaintainingaholdonthestandardswithonehand,
rotatethebasesothatthebottomplateisfirmlythreadedontoany
3½”8externalthreadedmountorstand.Whenreplacingthe
transitintothecase,slightlytightenthehorizontalandverticaltan
gentclamplockscrews—butonlydosogently.
Aspecialcenteringmount,differentthanmostofourinstrument
adapters,maybeemployedtophysicallyplacethehorizontalcen
terofthetransitinaknownlocation.Theshiftingcenterofthe
transitextendsaboveandbelowthebaseplate(visiblebetweenthe
levelingscrewsandalsofrombelowasa2”diameterringcentered
onthebottomofthebaseplate).Thisshiftingcenter(radius=1±
0.0001”)fitsintospecialmountshavinganaccuratelyboredinner
diameter.Ifthetransitistobelocatedononeofthesemounting
rings,thelevelingscrewsshouldbeloosenedslightlysothebase
platecanmovearoundinrelationtotheshiftingcenter.Retighten
thelevelingscrewsonlyafterbaseplateistightenedonthemount.
Oncemountedonastandorotherstableinstrumentmount,thefol
lowingprocesswillroughlevel”thetransit,meaningthatthetran
sitwillbebroughtroughlytoplumb,andbewithin
rangeofthe
precisionlevelvialmountedalongthetelescopeaxis.
Toroughlevel,loosenanytwoadjacentlevelingscrewsandturn
thefourarmʺspiderʺbasesothattwodiagonallyopposedleveling
screwsareinlinewiththeprincipalsightingdirectionthatwillbe
used.Figure2.2illustratesthis,
withthespiderbaselinedup
“square”withthetelescopeandcrossaxis.Retightentheleveling
screws.
Acircular“bull’seye”vialrestsonthehorizontal“plate”abovethe
basebutbelowthetelescopeaxis(Figure2.3).Wecallthisthe
“platevial”.Centerthisbubbleinthefollowingmanner:Puta
Removingfromthecase
Usingtheshiftingcenter
Roughleveling
9 Operation and Adjustment Manual
thumbandforefingeroneachoftwodiagonallyopposedleveling
screws.Tomovethebubble,turnthelevelingscrewsatthesame
time,andinoppositedirections(thumbsmovingtogetheror
thumbsmovingapart).Thebubblewillmoveinthedirectionof
movementofyourleftthumb.Movethebubblesothatitisinthe
middleoftheplatevial,andatleastinlinewiththeinscribedcircle
(itprobablywon’tbeexactlycentered).Nowmoveyourhandsto
theothersetoflevelingscrews.Usingthesameprocess,movethe
bubbledirectlyundertheinscribedcircleinthecenteroftheplate
vial.Youmayhavetorepeatthisentireproc essacoupleoftimesto
getthebubblecentered.Keepalightbutfirmtensiononthescrews
allthetime.Neverputheavytensiononlevelingscrews.
Nowrotatethetransit180°aboutitsverticalspindle.Thebubble
shouldstaycenteredundertheinscribedcircle.Ifitdoesn’t,the
vialisoutofadjustment(refertoChapter4ofthismanualforthe
calibrationproceduretocorrectthis).
Topreciselylevel(sometimescalled“precisionplumb”)thetransit
using the coincidence vial, rotate the standards of
the transit until
thetelescopeisinlinewithonepairofdiagonallyopposedleveling
screws.Plunge(rotate)thetelescopesothatthecoincidencevialis
ontopof the telescope, and level the telescopeby eye (Figure2.4).
To assist you, look at the side of the coincidence vial and position
the bubble roughly in the center of its travel.Gently tighten the
Figure2.2
Figure2.3
Preciseleveling
Model 771, 75, and 76-RH190 Transits 10
verticaltangentlock.
Thebubbleinthecoincidencevialop
erateslikeanyotherlevelvial,butthe
operatorviewsa“foldedimage”of
bothendsofthebubblesimultane
ously.Ifthetelescopeistilted,thetwo
bubbleimagesmoveinoppositedirec
tions.However,whenbothendsof
thebubblearebrought
“intocoinci
dence”(Figure2.5),thevialisdead
level.Youwillseesomethingsimilar
toFigure2.5bylookingintothevial’s
turret.Sincethehumaneyeisex
tremelyskillfulatdetectingeventiny
breaksinotherwisecontinuouslines,
thissystemmakesreadingthelevelextremelysensitiveandaccu
rate.
Rotatetheturretonthecoincidencevialsothatyoucanconven
ientlyviewthesplitimageofthebubbleinside.Areflectorhaving
onewhitesideandonemirroredsideisprovidedtohelpilluminate
theimage.Positionthereflectorforoptimalbubbleillumination.
1. Usingtheverticaltangentadjustmentscrew,bringthe
splitimageofthebubbleintocoincidence(Figure2.5).
2. Rotatethestandards180°sothatthetelescopeisnow
pointingintheoppositedirection.Ifthebubbleisoff
center,removeonehalftheerrorwiththetwoleveling
screwswhicharelocateddirectlyunderthetelescope.
Removetheotherhalfoftheerrorusingtheverticaltan
gentscrew.Rememberthatyoucanseethebubbleby
lookingintothesideofthevialitselfratherthanintothe
turretwindow—thismayhelpyoudeterminewhichway
thebubblemustmove.
3. Nowrotate
thestandards90°sothatthetelescopeislo
cateddirectlyovertheotherpairoflevelingscrews.
Bringthebubbletocenter,usingonlythetwoleveling
screwsunderthetelescope.
4. Fromthisposition,rotatethestandards180°.Again,if
thebubbleisoffcenter,removeonehalftheerrorwith
theverticaltangentscrew,andtheremainingonehalf
errorwiththetwolevelingscrewswhicharelocatedun
derthetelescope.
5. Repeatsteps24,alternatingovereachpairofleveling
screws,untilthebubbleremainsincoincidenceinallfour
“compasspoint”positions.
Figure2.4
Notethatthetransit’sverti
calspindlemaybebrought
intoplumb(exactlyvertical)
evenifthecoincidencevial
itselfisnotcalibratedprop
erlywithrespecttothetele
scopebarrel.
Figure2.5
11 Operation and Adjustment Manual
ThetelescopeofeachOpticalToolinginstrumentcontainsareticle
orcrosshairwhichdefinesthecenterofthelineofsight.When
youlookintoatelescope,youseethereticleʺsuperimposedʺupon
theimageofwhateveryouhavethe
telescopefocusedon.Thereticleis
splitintofourquadrants;twohavea
singleʺwireʺ(filar),theothertwohave
adoubleʺwireʺ(bifilar).Thiscon
figurationwasoriginallyadoptedbe
causethehumaneyeisextremely
goodatdetectingtinydifferencesin
alignmentbetweencloselyspacedob
jects.Therefore,thispatternmakesit
mucheasiertovisuallyalignthereti
clewithotherimages(ex.,opticaltar
gets,scales,andreticleimagesre
flectedfrommirrors).
Focusing sounds easy, but it’s surprising how many people don’t
understand the effectsof something called parallax.Here’s how to
properlyfocusyourinstrument.First,pointthetelescopeatasheet
of white paper or other light colored background.Don’t worry
about focusing the telescope using the focus knob, the image can
remainblurry.Instead,usetheknurledringontheeyepiecetofo
custhereticlesothatitissharpandblack.
Nowpointthetelescopeatatargetofinterestandfocustheimage
usingthefocusknobonthetelescope.Slightlyshiftyourheadleft
andrightwhilelookingatthereticleasitis“superimposed”over
theimageofthetarget.Ifthereticleappearstofloatslightlytothe
leftandrightoverthetarget,
parallaxispresentinthesystem.This
occurswheneitherthetelescope’smainopticalsystemortheocular
(eyepiece)system,orboth,arenotfocusedpreciselyontheplaneof
thereticle.Parallaxcanleadtoerroneousmeasurementsifnotre
moved.Toremovetheparallax,repeatthetwosteps
again,being
carefultoachievecriticalfocusineachstep.
Therearemanyoccasionswhenitwillbeimportanttoputyour
telescopeatinfinityfocus(e.g.,autocollimatingyourtelescopeona
mirror).Infinityisimportantbecause,fromanopticalstandpoint,
lightrayscomingfroman“infinitedistance”areallparalleltoeach
TheReticle(crosshairs)
Howtoproperlyfocus
Thefilar/bifilarreticle
imageasseenwhen
lookingintothetele
scope.
Figure2.6
Findinginfinityfocus
Model 771, 75, and 76-RH190 Transits 12
other,orcollimated.
Thefocusknobonthetelescopehasan
indicatorofthedirectionyoumustturn
ittofindinfinityfocus,butyoucan’t
justturnitallthewayinthatdirection
andexpecttothetelescopetobefo
cusedatinfinity.
Theeasiestwaytofindtrue“infinity
focus”istoholdanopticallyflatmirror
(suchasourmodel185)rightup
againsttheobjectiveendofthetele
scope.Turnonthetransit’sinternalillumination(see“Installing
lightingaccessories”below).Turnthefocusknoballthewayinthe
directionofinfinity(indicatedontopofthefocusknob).Then,turn
theknobslowlyintheoppositedirection,andwatchcloselyforthe
reflectedimageofthereticle.Itcangobyprettyquickly,sogoing
slowlyisimportantwhilelookingthroughthetelescope.Onceyou
havefocusedthetelescopesothatyoucanclearlyseethereflected
reticleimage,youknowthatthetelescopeis,bydefinition,focused
atinfinity.
Releaseboththehorizontalandverticaltangentclamplocks,and
movetheinstrumentbyhand.Precisepointingisaccomplishedby
relockingbothclampswhenyougetclosetothetarget,andthen
turningthetangentadjustmentscrewsforpreciseplacementofthe
reticle.Bothtangentadjustmentscrewsareoutfittedwithwobble
pinswhichallowforsmoothtranslationoftheaxes.Thefinalset
tingshouldbemadebyturningthetangentscrewinaclockwise,or
pushing,direction.Thiswilleliminateanypotentialsubsequent
movementduetofrictionofthespringloadedplungerswhichop
posethetangentadjustmentscrews.
Forcertainproceduresthatwewilldiscussinthismanual,(e.g.,col
limation,autocollimation,andreticleprojection)itisnecessaryto
illuminatethereticleofyourtransitfromtheinside.A
model76x
transithastworeticles(oneinthemaintelescopeandoneinthe
crossaxistelescope)whichmayneedtobeilluminated,depending
uponwhatyouaredoing.
Illuminatingthecrossaxistelescopereticle.Yourtransitcomes
equippedwithalightsourceforthecrossaxistelescope(model76
Thereflectedimageofa
reticle(here,located
aboveandtotherightof
center)willappearonly
atinfinityfocus.
Figure2.7
Installinglightingaccessories
Pointingthetelescope
13 Operation and Adjustment Manual
xonly).ThisLEDlightsourcefitsintothereceptacle(red
arrowinFigure2.8).Thesupplytransformerworkswith
either110Vor220Vsources.Installationorremovalofthe
lightsourceissimpleandquick.Simplyinstallthetwo
wayadapteronthecrossaxislightingreceptacle,andpush
the
LEDlamphousinginuntilitclicksintoplace.Thelight
sourcehasaninlineswitchforeasycontrolofthelight.
Rememberthatthefocusofthecrossaxistelescopeisper
manentlyfixedatinfinity.Thistelescopeismeanttobe
usedasacollimatororautocollimator.
Wewilldiscussthis
more,lateroninthemanual.
Illuminatingthemaintelescopereticle.Whenyouwishto
usethemaintelescopeforcollimation,autocollimation,orreticle
projection,youmustinstallamodel1961combinationeyepiece
unit.Todothis,removetheshortsectionofthemaintelescopebar
rellocatedjustinfrontoftheeyepiece(SeeFigure2.9),andreplace
itwiththe1961eyepieceadapter.Notethatthe1961hasa
knurledlockringwhichallowsyoutotightentheadapterinany
givenorientationonthetelescopebarrel.
Beforeinstallingthiseyepiece
adapteronthetelescope,make
suretheknurledlockring(red
arrowinFigure2.10)issecured
againsttheadapter.Itshouldbe
“bottomedout”onthethreads
ratherthanlooseonthethreads
orneartheopenendofthe
adapter.Thenthreadthe
adapterintothemaintelescope
barrelasfarasyoucan.Next,
backofftheadaptertothede
siredradialorientation(lamp
receiverpointinguporpointing
down)andlockitinplaceby
turningtheknurledlockring
towardthefrontofthetele
scope.
Next,removetheeyepieceassemblythatisstill
con
nectedtotheshortsectionofthetelescopebarrelwhich
wasremoved,andinstallitonthe1961eyepiece
adapter.
Toinstallthelight,removethecapfromthelampre
ceiver(undertherightthumbinFigure2.10).Thenin
stallthetwowayadapterbythreadingitsknurledring
ontothelampreceiver.PlugintheLEDlightsource
untilitclicksintheadapter.
Figure2.8
Figure2.9
Note:Toinstallarightangleview
ingeyepieceonthemaintelescope
(convenientforveryhighangle
shots),youmaywishtoinstalla193
Lratherthanthe1961.Theseac
cessoriesareinstalledinbasically
thesameway,byfirstremovingthe
shorttubesectioninfrontofthe
eyepiece.
Model 771, 75, and 76-RH190 Transits 14
Laterinthismanual,wewill
discusssomeoftheoptical
toolingtechniqueswhichre
quirebacklightingofthereti
cleusingoneoftheseadapt
ers.
Figure2.10
15 Operation and Adjustment Manual
Therearemanyjobsthatyoucandowithyourtransit,whichmay
bepurchasedwitha190typemicrometerinastandardconfigura
tion.Youcanevaluateanumberofalignmentrelatedcharacteris
ticsofmachineryorotherstructures,suchasstraightness,plumb,
level,flatness,andothers.However,toperformcertainOptical
Toolingprocedures,otheraccessoriesarerequired.Wewillmen
tionsomeoftheseaccessoriesaswediscusstherelatedtopics.
Inthischapterwewilldiscussthefollowingtopics:
ThisdiscussionismeanttomakeyouawareofsomeoftheOptical
Toolingproceduresthatyoucanuseinalignmentandmeasure
mentwork.Thismanualisnotagoodsubstituteforourtraining
classes,inwhichweofferhandsontrainingoftheseandmany
otherOpticalToolingtechniques.
We’llexplainwhatthesetermsmean,andhowtoaccomplishthem
usingyourtransit.
In
manyalignmentjobs,itisnecessarytoputthelineofsightof
yourtransitdirectlyonalinecreatedbytwoknownreferencetar
gets.Forexample,inFigure3.1(showingaviewfromoverhead),
sometargetshavebeeninstalledinthefloor.Thesetargetsrepre
sentalinewhich
isparalleltothecenterlineofamachineonwhich
youneedtoaligncomponents.Inpreparationfordoingthisalign
mentjob,itcouldbenecessarytobringtheinstrumentinlinewith
thesetwofloormonuments.
NoteinFigure3.1,thetransitmustnotonlyberotatedso
thatitis
paralleltothelinedefinedbythetargets,butitmustalsobeshifted
sideways.“Buckingin”isarepetitiveprocessinwhichyouadjust
boththeinstrument’sangleaswellasitslateralpositionsothatthe
transitisplaceddirectlyinlinewithtwotargets.
Ortobemoreac
curate,sothatthetransit’sverticalspindleliesintheverticalplane
whichalsorunsthroughthecenterofthefloortargets.
Notethatangularadjustmentsareeasyenoughtomakeusingthe
horizontaltangentadjustmentscrewonthetransititself.However,
Chapter3‐MeasuringTechniques
 “Buckingin”
 Autoreflection
 Autocollimation
 Collimation
 Collineation
 Reticleprojection
 Measuringwithmicrometers
“Buckingin”
Model 771, 75, and 76-RH190 Transits 16
Alwayssetthetelescopeon
thefartargetbymak
ingangularmoves
usingthehorizontal
andverticaltangent
screw(s).
Alwayssetthetelescopeon
theneartargetby
makinglateralmoves
usingalateralslideor
precisionlift.
lateralshiftsoftheentireinstrumentarenotaseasy.Togreatly
simplifythis,werecommendtheuseofalateralslidemountedbe
tweenthetransitandthestandonwhichitismounted.
Stepstobuckthetransitintothetargetline:
1. Roughlypositionthetransit.Setuptheinstrument
standnominallyinlinewiththetwotargets.
2. Roughlevel.Roughlevelthetransitasdescribedin
Chapter2“RoughLeveling”.
3. Preciselevel.Formanyapplications,preciseleveling
yourtransitisrequiredpriortobuckin(coveredin
Chapter2).Technicallyisitpossibletobuckintotargets
withoutpreciseleveling,butthetransitwillnotendupin
theverticalplanedefinedbythosetargetsunlessthetran
sititselfispreciselyplumb.Afulldiscussionof
thisfactisbeyondthescopeofthismanual,butis
coveredinmoredetailinourOpticalTooling
classes.
4. Settofartarget.Focusthetelescopeonthefar
targetandsetthecrosswiresdirectlyoverthattar
getusingthehorizontaland/orverticaltangent
adjustmentscrews.Ifthetransitisbetweenthe
targets(asinFigure3.2),simplyassignoneofthe
targetstobethe“fartarget”,andsetonit.
5. Settoneartarget.Withoutmovingthehori
zontaltangentscreworturningthestandardsof
thetransit,turnthetelescopedownwardandfocus
onthesecondtarget.Notethatifthetransitislo
catedbetweenthetargets(asinFigure3.2),youwill
havetoturnthetelescope“through”the
standards
topointintheoppositedirection.Regardlessof
whetheryouareinasituationlikeFigure3.1or
Figure3.2,thissecondtargetwillinvariablybeoff
toonesideofthefieldofviewunlessyouarein
crediblylucky.Forpurposesofillustration,let’s
saythattheobservedlateraldistancetothenear
targetis“L”(seeFigure3.1or3.2).Nowyoumust
movethetransitlaterallytowardtheneartarget
usingthecrossslide—buthowmuch?The
amountthatyoumovethetransitdependsupon
thegeometryofthesetup.Theequationtocalcu
latethe
properamountofshiftis:
TransitLateralShift=Lx(D2/D1)
where
D1=Distancebetweentargets
Figure3.1
/