Brunson 75 Operation And Field Adjustment Manual

Category
Telescopes
Type
Operation And Field Adjustment Manual

This manual is also suitable for

Models 771, 75, & 76-RH
Optical Tooling Transits
Instrument
Company
Operation and
Field Adjustment Manual
Helping the World Measure
Since 1927
www.brunson.us
Copyright 2009-2011 Brunson Instrument Company
Kansas City, Missouri
8000 E. 23rd Street
Kansas City, MO 64129
Tel: 816.483.3187
Fax: 816.241.1945
TableofContents
Thankyoufor
purchasing
aBrunsonTransit.
Rememberthatour
customersupportdoes
notstopaftershipmentof
aproduct—wearehereto
helpyouwithany
measurementchallenges
thatyoumayhave.
Chapter1:Introduction
1
WhatisaTransit?
1
WhatDoesaTransitDo?
3
TheThreeDifferentTransits
4
Chapter2:GettingStartedWithYourTransit
7
HandleYourTransitWithCare
7
RemovingFromtheCase
8
UsingtheShiftingCenter
8
RoughLeveling
8
PreciseLeveling
9
TheReticle
11
HowtoProperlyFocus
11
FindingInfinityFocus
11
PointingtheTelescope
12
InstallingLightingAccessories
12
Chapter3:MeasuringTechniques
15
BuckingIn
15
Autoreflection
17
Autocollimation
19
Collimation
23
Collineation
26
ReticleProjection
27
MeasuringWith
Micrometers
28
Chapter4:FieldCalibrationChecks
35
Plate(“Bull’sEye”)Vial
37
MicrometerChecks
38
“DoubleCenter”Check
40
HorizontalAxisRunout
41
VerticalSpindleRunout
42
HorizontalCentering
43
HorizontalCollimation
44
PlumbLineCheck
46
CrossAxisTelescopeCollimation
49
PegTest
51
1 Operation and Adjustment Manual
Inthischapterwewilladdressthefollowingtopics:
 Whatisatransit?
 Whatarethecomponents?
 Whatdoesatransitdo?
 ThedifferencebetweenthethreeBrunsontransits
Themostimportantthingaboutatransitisthatithasatelescopeon
agimballingmechanism,soitcanrotatebackandforthhorizon
tally(azimuthdirection),butitcanalsorotateupanddown
(elevationdirection).
Almosteveryoneisfamiliarwithasurveyor’stransit.However,it
isworthexplainingthegeneraldifferencebetweenourtransitsand
thoseusedbysurveyors.OpticalToolingtransitshaveanumberof
modificationstomakethemdeadlyaccuratewhenshootingover
distanceswhich,toasurveyor,wouldseemshort(lessthanacou
pleofhundredfeet).Inaddition,wedonotneedtotakereadings
intwotelescopepositions(directandreverse)inordertoachieve
thisaccuracy.Thisispossiblebecauseourtransitshaveseveral
specialcharacteristics:
a)Theyhaveanextremelystraightlineofsight.Asyoufocusthe
telescopefromneartofar,thelineofsighttravelsinanextremely
straightline.Thatis,pictureputtinganimaginary,
weightless,tautstringdownthecenterofthetelescope,
stretchingstraightoutacoupleofhundredfeet—the
stringwouldalwaysbeinthecenterofthecrosshairsas
youfocus
fromneartofarandbackagain.
b)Atransit’sthreemajoraxesareallmutuallyorthogonal
(allatrightanglestoeachother).Therearetworota
tionalaxesonyourtransit,andoneaxiswhichdoesnot
rotate(seeFigure1.2).Thefirstrotationalaxisisthever
ticalaxis.Thespindleandbearingswhichcreatethever
ticalaxisarehousedinthetransit’sbase.Theentireup
perportionofthetransit(standardsandtelescope)rotate
horizontallyaroundthisaxis.Thesecondrotationalaxis
isthecrossaxis.Thisisthehorizontalaxisaroundwhich
thetelescoperotates.Thethirdaxisisthelineofsight.
c)Lastly,allthreeofthemajoraxesmeetatoneprecise
point.Justbecausetheaxesareallmutuallyorthogonal
PleaserefertoFigure1.1
forageneralorientationto
componentterminology.
Chapter1‐Introduction
Note:Thismanualappliestothe
followingBrunsonproducts:
 771and771HJigTransits
 75and75HOpticalTransit
Squares
 76RH,76RH190,and76
RH190MTelescopicTransit
Squares
Atthistime,themodels771,771H,
75,and75Harenolongerpro
ducedbythefactory.However,we
haveincludedtheminthismanual
duetotheirpastpopularityandthe
factthatmanyofthese
instruments
remaininthehandsofalignment
technicianswhomayfindthisinfor
mationuseful.
Whatisatransit?
Model 771, 75, and 76-RH190 Transits 2
1. Levelingscrews(4)
2. “Spider”base
3. Platevial
4. Telescope(eyepieceend)
5. Crossaxistelescopeeyepiece
6. Focusknob
7. Telescope(objectiveend)
8. 190xOpticalMicrometer
9. Crossaxistelescopeobjectivelens
10.Verticaltangentclamplock
11.Coincidencelevel
12.Crossaxistelescopelightingadapter
13.Standards
14.Horizontaltangentclamplock
15.Baseplate
16.Horizontaltangentadjustmentscrew
17.Plate
18.Verticaltangentadjustmentscrew
19.Shiftingcenter
Figure1.1
OrientationtoComponents
3 Operation and Adjustment Manual
doesn’tnecessarilymeanthatthelineswouldallintersectatthe
sameexactpointinspace.However,onourtransits,theselinesdo
intersect.Thisisanimportantcharacteristicwhenaccuracyispara
mount,particularlyatshorterdistances.Ofcourse,youcan’tsee
thisintersectionpointitisburiedinthemiddleofthetelescope.
Thesearethethreedistinctcharacteristicsthatourtelescopeshave
whichallowthemtomakeveryprecisemeasurementsoverrela
tivelyshortworkingdistances.However,thereisonemorevery
distinctivecharacteristicwhichseparatesOpticalToolingtransits
fromallotherstheopticalcrossaxis.Our76RH190hasaninfin
ityfocusedtelescope,completewithasecondeyepieceandsecond
setofcrosswires,housedinthecrossaxis.Thiscrossaxistelescope
isusedinanumberofoperationstosetthemaintelescopeperpen
diculartoanothertelescopeoramirror(whichmaybemountedon
arotatingshaftorotherreferenceplane).
Atransitisaveryflexiblemeasuringtool.Itcanestablishvertical
orhorizontalopticalplaneswhichareextremelyflat.Itcanestab
lishopticallineswhichareextremelystraight,extendinginanydi
rection.Oncetheseplanesandlinesareestab
lished,youcanusethemasreferencesfrom
whichtomeasureotherpointsofinterest.You
candeterminewhethermachinecomponents
areparallel,flat,level,square,orstraight.This
meansyoucanevaluatethegeometricrela
tionshipsrelatedtothealignmentofalmost
anypart,structuralcomponent,machine,tool,
surface,substrate,wallorjustaboutany
thingelse.Averyfewexamplesofthiswould
be:
 Alignmentofbearingsandbearing
journals.
 Rollparallelisminpaperandmetals
mills,orprintingpresses
 Machinetoolevaluation:bedlevel
ing,bedflatness,geometricintegrity
ofspindlesand
othercomponents
 Layoutformachining
 Crankcasealignment
 Straightnessofenginebores,turbine
cases,torpedotubes,lineshafts,etc.
 Checkrelationshipsofrightangle
planes
TheOpticalToolinginstrumentsareextremely
flexible.Youcanusethemtomakejustabout
anymeasurementorgeometricevaluationthat
Figure1.2
Sowhatdoesatransitdo?
Model 771, 75, and 76-RH190 Transits 4
youcandesigninyourmind.
Asweindicatedearlier,the76RH190istheonlycurrentlypro
ducedOpticalToolingtransit.Althoughthismanualwaswritten
specificallyforthe76RH190,muchoftheinformationitcontainsis
applicabletotheothertransitsaswell.
Thedifferencesbetweenthe771x,75x,and76xtransitsrelate
solelytothewayinwhichthecrossaxisisconstructedandused.
Alltransitssharethesamebase,standardassembly,bearings,and
telescopeoptics.Onalltransits,thetelescoperotatesinavertical
planearoundahorizontalaxis(the“crossaxis”).Thecrossaxisis
perpendiculartothelineofsight.
Themostbasictransitisthe771or
771H.Thesetransitshavenoop
ticsinthecrossaxis.Theyare
usedas“jig”transitstoestablish
straightlinesandplanes.How
ever,theyhavenocapabilityto
makeanaccuraterightangleturn.
Onestep“up”isthemodel75or
75H(ontheleftinFigure1.3).
Thesetransitshaveacrossaxis
whichishollowandcanbeseen
through.Ononesideofthecross
axisisaclearglasswindow,and
ontheothersideisapartiallyre
flectivemirror.The75(H)does
everythingthatthe771(H)can
do,butcanalsobeusedincon
junctionwithanotherinstrument
toproduceaveryaccurateright
angle.
Themostflexibletransit(andthe
onlyonecurrentlyproduced)
is
themodel76RH190.Thecross
axisofthistransitcontainsatele
scopethatispermanentlyfocusedatinfinityandisequippedwith
reticlelighting.Arightangleeyepieceisprovidedasstandard
equipmentonthecrossaxistelescope.The76RH190maybeused
toestablishstraightlines,horizontalorverticalplanes,andcreate
accuraterightangleswithrespecttoothertransits,alignmenttele
scopes,mirrors,orotherphysicaltargets.Whenthecrosswiresof
thetransit’scrossaxistelescopearesetonthoseofareferencetele
scopeormirror,themaintelescopeofthetransitwillsweepaplane
atpreciserightanglestothecrossaxisreferenceline.
The75Htransit
(above,left)hasbeendis
continuedbutmanyare
stillinuse,andmostof
theinformationcon
tainedinthismanual
stillapplies.
Figure1.3
Thethreedifferenttransits
5 Operation and Adjustment Manual
TheH”inthevariousmodelnumbersreferstoa“Hollow”verti
calaxis.Transitsthathavethe‐Hdesignationallowyoutolook
straightdownthroughahollowverticalspindleandtoseetargets
orcollimatorswhicharedirectlybelowthebaseoftheinstrument.
Whydoweusefour“feet”(levelingscrews)insteadofthree?
Weareoftenaskedthisquestion,sincemosttheodoliteshavethreeleveling
screws.Thereasonisthatwhenadjustingfourlevelingscrews,theinstrument
heightisnotaffected.
Whenthreelevelingscrewsareused,twoof
themmustnecessarilyformapivotaxis,and
thethirdisusedtoraiseorloweronesidetotilt
theinstrument.Itgivestheeffectshownatleft.
Thereddotindicatestherelativeinstrument
heightbetweentwopositionsoftheinstrument
(tiltedandnottilted).
However,whenfourlevelingscrewsareused,
theinstrumentposition(again,reddot),effec
tively,doesnotchange.Thisgivestheeffect
shownatright.Onesidegoesupbythesame
amountastheothersidegoesdown,andthe
instrumentpivotsatitscenter,notononeside
ortheother.
Thisisjustanotherexampleofthewaythatwehavecraftedourinstrumentsto
bemoreaccurateatshorterranges.Overlongerdistances,minorchangesinin
strumentheightarenotimportant.Butwhenmeasuringveryaccuratelyover
shorterdistances,avoidingchangesininstrumentheightcanbecomecrucial.
Model 771, 75, and 76-RH190 Transits 6
7 Operation and Adjustment Manual
Inthischapter,wewilldiscuss
Notethattheseproductsareprecisioninstrumentsandshouldal
waysbehandledwithcare.
 Donotforceanyofthescrews.Afirm(butnottight)ten
siongivesthebestresults.
 Keepyourinstrumentcleananddry.Protectitfromthe
weatherifusedoutdoors.
 Allowyourinstrumentto“soak”foratleast4hoursin
thetemperatureinwhichitwillbeused,sothatshifts
duetothermalchangesareminimized.Ifyouaremov
ingthetransitacrossthermaldifferentialsofmorethan
20°F,allowittosoakfor8hours.Theinstrumentshould
beoutofthecaseduringthistimeforbetteraircircula
tion.
 Whentakingasighting,or
readingoneoftheseinstru
ments,removeyourhands,
andmakesureanyother
pressureisalsoremoved.
Keepyourinstrumentcalibrated.
Make
therecommendedfieldchecks
(seeChapter4)aswellasperforminga
completecalibrationatleastannually.
Notethatthereisawarningprintedon
thesideofyourtransitwhichsays
“Caution—DoNotPlungeTelescope
FromFlange”.Plungingthetelescope
meanstorotateitintheverticalplane,
asyoumightwhenyouwanttolook
furtherupordown.Specifically,
plungingreferstoreversingthedirec
tionofthetelescopebyrotatingit180°
Chapter2‐GettingStartedwithYourTransit
 Howtohandleyourtransit
 Removingitfromthecase
 Usingtheshiftingcenter
 Roughleveling
 Preciseleveling
 TheReticle
 Focusingtheeyepieceand
telescope
 FindingInfinityfocus
 Pointingthetelescope
 Installinglightingaccesso
ries
Figure2.1
Handleyourtransitwithcare
Model 771, 75, and 76-RH190 Transits 8
intheverticalplane.The“flange”isthecomponentuponwhichthe
warningisprinted.Thissimplymeansnottorotatethetelescope
bygrabbingtheflangeandtwistingit.Torotatethetelescopeprop
erly,putyourhandonthetelescopebarrelitself.
Loosenboththehorizontalandverticaltangentclampscrews.
Graspthestandardsfirmlywithonehand,andthebasewiththe
otherhand.Liftgentlyfromcaseandplaceonastandorothersup
port.Whilemaintainingaholdonthestandardswithonehand,
rotatethebasesothatthebottomplateisfirmlythreadedontoany
3½”8externalthreadedmountorstand.Whenreplacingthe
transitintothecase,slightlytightenthehorizontalandverticaltan
gentclamplockscrews—butonlydosogently.
Aspecialcenteringmount,differentthanmostofourinstrument
adapters,maybeemployedtophysicallyplacethehorizontalcen
terofthetransitinaknownlocation.Theshiftingcenterofthe
transitextendsaboveandbelowthebaseplate(visiblebetweenthe
levelingscrewsandalsofrombelowasa2”diameterringcentered
onthebottomofthebaseplate).Thisshiftingcenter(radius=1±
0.0001”)fitsintospecialmountshavinganaccuratelyboredinner
diameter.Ifthetransitistobelocatedononeofthesemounting
rings,thelevelingscrewsshouldbeloosenedslightlysothebase
platecanmovearoundinrelationtotheshiftingcenter.Retighten
thelevelingscrewsonlyafterbaseplateistightenedonthemount.
Oncemountedonastandorotherstableinstrumentmount,thefol
lowingprocesswillroughlevel”thetransit,meaningthatthetran
sitwillbebroughtroughlytoplumb,andbewithin
rangeofthe
precisionlevelvialmountedalongthetelescopeaxis.
Toroughlevel,loosenanytwoadjacentlevelingscrewsandturn
thefourarmʺspiderʺbasesothattwodiagonallyopposedleveling
screwsareinlinewiththeprincipalsightingdirectionthatwillbe
used.Figure2.2illustratesthis,
withthespiderbaselinedup
“square”withthetelescopeandcrossaxis.Retightentheleveling
screws.
Acircular“bull’seye”vialrestsonthehorizontal“plate”abovethe
basebutbelowthetelescopeaxis(Figure2.3).Wecallthisthe
“platevial”.Centerthisbubbleinthefollowingmanner:Puta
Removingfromthecase
Usingtheshiftingcenter
Roughleveling
9 Operation and Adjustment Manual
thumbandforefingeroneachoftwodiagonallyopposedleveling
screws.Tomovethebubble,turnthelevelingscrewsatthesame
time,andinoppositedirections(thumbsmovingtogetheror
thumbsmovingapart).Thebubblewillmoveinthedirectionof
movementofyourleftthumb.Movethebubblesothatitisinthe
middleoftheplatevial,andatleastinlinewiththeinscribedcircle
(itprobablywon’tbeexactlycentered).Nowmoveyourhandsto
theothersetoflevelingscrews.Usingthesameprocess,movethe
bubbledirectlyundertheinscribedcircleinthecenteroftheplate
vial.Youmayhavetorepeatthisentireproc essacoupleoftimesto
getthebubblecentered.Keepalightbutfirmtensiononthescrews
allthetime.Neverputheavytensiononlevelingscrews.
Nowrotatethetransit180°aboutitsverticalspindle.Thebubble
shouldstaycenteredundertheinscribedcircle.Ifitdoesn’t,the
vialisoutofadjustment(refertoChapter4ofthismanualforthe
calibrationproceduretocorrectthis).
Topreciselylevel(sometimescalled“precisionplumb”)thetransit
using the coincidence vial, rotate the standards of
the transit until
thetelescopeisinlinewithonepairofdiagonallyopposedleveling
screws.Plunge(rotate)thetelescopesothatthecoincidencevialis
ontopof the telescope, and level the telescopeby eye (Figure2.4).
To assist you, look at the side of the coincidence vial and position
the bubble roughly in the center of its travel.Gently tighten the
Figure2.2
Figure2.3
Preciseleveling
Model 771, 75, and 76-RH190 Transits 10
verticaltangentlock.
Thebubbleinthecoincidencevialop
erateslikeanyotherlevelvial,butthe
operatorviewsa“foldedimage”of
bothendsofthebubblesimultane
ously.Ifthetelescopeistilted,thetwo
bubbleimagesmoveinoppositedirec
tions.However,whenbothendsof
thebubblearebrought
“intocoinci
dence”(Figure2.5),thevialisdead
level.Youwillseesomethingsimilar
toFigure2.5bylookingintothevial’s
turret.Sincethehumaneyeisex
tremelyskillfulatdetectingeventiny
breaksinotherwisecontinuouslines,
thissystemmakesreadingthelevelextremelysensitiveandaccu
rate.
Rotatetheturretonthecoincidencevialsothatyoucanconven
ientlyviewthesplitimageofthebubbleinside.Areflectorhaving
onewhitesideandonemirroredsideisprovidedtohelpilluminate
theimage.Positionthereflectorforoptimalbubbleillumination.
1. Usingtheverticaltangentadjustmentscrew,bringthe
splitimageofthebubbleintocoincidence(Figure2.5).
2. Rotatethestandards180°sothatthetelescopeisnow
pointingintheoppositedirection.Ifthebubbleisoff
center,removeonehalftheerrorwiththetwoleveling
screwswhicharelocateddirectlyunderthetelescope.
Removetheotherhalfoftheerrorusingtheverticaltan
gentscrew.Rememberthatyoucanseethebubbleby
lookingintothesideofthevialitselfratherthanintothe
turretwindow—thismayhelpyoudeterminewhichway
thebubblemustmove.
3. Nowrotate
thestandards90°sothatthetelescopeislo
cateddirectlyovertheotherpairoflevelingscrews.
Bringthebubbletocenter,usingonlythetwoleveling
screwsunderthetelescope.
4. Fromthisposition,rotatethestandards180°.Again,if
thebubbleisoffcenter,removeonehalftheerrorwith
theverticaltangentscrew,andtheremainingonehalf
errorwiththetwolevelingscrewswhicharelocatedun
derthetelescope.
5. Repeatsteps24,alternatingovereachpairofleveling
screws,untilthebubbleremainsincoincidenceinallfour
“compasspoint”positions.
Figure2.4
Notethatthetransit’sverti
calspindlemaybebrought
intoplumb(exactlyvertical)
evenifthecoincidencevial
itselfisnotcalibratedprop
erlywithrespecttothetele
scopebarrel.
Figure2.5
11 Operation and Adjustment Manual
ThetelescopeofeachOpticalToolinginstrumentcontainsareticle
orcrosshairwhichdefinesthecenterofthelineofsight.When
youlookintoatelescope,youseethereticleʺsuperimposedʺupon
theimageofwhateveryouhavethe
telescopefocusedon.Thereticleis
splitintofourquadrants;twohavea
singleʺwireʺ(filar),theothertwohave
adoubleʺwireʺ(bifilar).Thiscon
figurationwasoriginallyadoptedbe
causethehumaneyeisextremely
goodatdetectingtinydifferencesin
alignmentbetweencloselyspacedob
jects.Therefore,thispatternmakesit
mucheasiertovisuallyalignthereti
clewithotherimages(ex.,opticaltar
gets,scales,andreticleimagesre
flectedfrommirrors).
Focusing sounds easy, but it’s surprising how many people don’t
understand the effectsof something called parallax.Here’s how to
properlyfocusyourinstrument.First,pointthetelescopeatasheet
of white paper or other light colored background.Don’t worry
about focusing the telescope using the focus knob, the image can
remainblurry.Instead,usetheknurledringontheeyepiecetofo
custhereticlesothatitissharpandblack.
Nowpointthetelescopeatatargetofinterestandfocustheimage
usingthefocusknobonthetelescope.Slightlyshiftyourheadleft
andrightwhilelookingatthereticleasitis“superimposed”over
theimageofthetarget.Ifthereticleappearstofloatslightlytothe
leftandrightoverthetarget,
parallaxispresentinthesystem.This
occurswheneitherthetelescope’smainopticalsystemortheocular
(eyepiece)system,orboth,arenotfocusedpreciselyontheplaneof
thereticle.Parallaxcanleadtoerroneousmeasurementsifnotre
moved.Toremovetheparallax,repeatthetwosteps
again,being
carefultoachievecriticalfocusineachstep.
Therearemanyoccasionswhenitwillbeimportanttoputyour
telescopeatinfinityfocus(e.g.,autocollimatingyourtelescopeona
mirror).Infinityisimportantbecause,fromanopticalstandpoint,
lightrayscomingfroman“infinitedistance”areallparalleltoeach
TheReticle(crosshairs)
Howtoproperlyfocus
Thefilar/bifilarreticle
imageasseenwhen
lookingintothetele
scope.
Figure2.6
Findinginfinityfocus
Model 771, 75, and 76-RH190 Transits 12
other,orcollimated.
Thefocusknobonthetelescopehasan
indicatorofthedirectionyoumustturn
ittofindinfinityfocus,butyoucan’t
justturnitallthewayinthatdirection
andexpecttothetelescopetobefo
cusedatinfinity.
Theeasiestwaytofindtrue“infinity
focus”istoholdanopticallyflatmirror
(suchasourmodel185)rightup
againsttheobjectiveendofthetele
scope.Turnonthetransit’sinternalillumination(see“Installing
lightingaccessories”below).Turnthefocusknoballthewayinthe
directionofinfinity(indicatedontopofthefocusknob).Then,turn
theknobslowlyintheoppositedirection,andwatchcloselyforthe
reflectedimageofthereticle.Itcangobyprettyquickly,sogoing
slowlyisimportantwhilelookingthroughthetelescope.Onceyou
havefocusedthetelescopesothatyoucanclearlyseethereflected
reticleimage,youknowthatthetelescopeis,bydefinition,focused
atinfinity.
Releaseboththehorizontalandverticaltangentclamplocks,and
movetheinstrumentbyhand.Precisepointingisaccomplishedby
relockingbothclampswhenyougetclosetothetarget,andthen
turningthetangentadjustmentscrewsforpreciseplacementofthe
reticle.Bothtangentadjustmentscrewsareoutfittedwithwobble
pinswhichallowforsmoothtranslationoftheaxes.Thefinalset
tingshouldbemadebyturningthetangentscrewinaclockwise,or
pushing,direction.Thiswilleliminateanypotentialsubsequent
movementduetofrictionofthespringloadedplungerswhichop
posethetangentadjustmentscrews.
Forcertainproceduresthatwewilldiscussinthismanual,(e.g.,col
limation,autocollimation,andreticleprojection)itisnecessaryto
illuminatethereticleofyourtransitfromtheinside.A
model76x
transithastworeticles(oneinthemaintelescopeandoneinthe
crossaxistelescope)whichmayneedtobeilluminated,depending
uponwhatyouaredoing.
Illuminatingthecrossaxistelescopereticle.Yourtransitcomes
equippedwithalightsourceforthecrossaxistelescope(model76
Thereflectedimageofa
reticle(here,located
aboveandtotherightof
center)willappearonly
atinfinityfocus.
Figure2.7
Installinglightingaccessories
Pointingthetelescope
13 Operation and Adjustment Manual
xonly).ThisLEDlightsourcefitsintothereceptacle(red
arrowinFigure2.8).Thesupplytransformerworkswith
either110Vor220Vsources.Installationorremovalofthe
lightsourceissimpleandquick.Simplyinstallthetwo
wayadapteronthecrossaxislightingreceptacle,andpush
the
LEDlamphousinginuntilitclicksintoplace.Thelight
sourcehasaninlineswitchforeasycontrolofthelight.
Rememberthatthefocusofthecrossaxistelescopeisper
manentlyfixedatinfinity.Thistelescopeismeanttobe
usedasacollimatororautocollimator.
Wewilldiscussthis
more,lateroninthemanual.
Illuminatingthemaintelescopereticle.Whenyouwishto
usethemaintelescopeforcollimation,autocollimation,orreticle
projection,youmustinstallamodel1961combinationeyepiece
unit.Todothis,removetheshortsectionofthemaintelescopebar
rellocatedjustinfrontoftheeyepiece(SeeFigure2.9),andreplace
itwiththe1961eyepieceadapter.Notethatthe1961hasa
knurledlockringwhichallowsyoutotightentheadapterinany
givenorientationonthetelescopebarrel.
Beforeinstallingthiseyepiece
adapteronthetelescope,make
suretheknurledlockring(red
arrowinFigure2.10)issecured
againsttheadapter.Itshouldbe
“bottomedout”onthethreads
ratherthanlooseonthethreads
orneartheopenendofthe
adapter.Thenthreadthe
adapterintothemaintelescope
barrelasfarasyoucan.Next,
backofftheadaptertothede
siredradialorientation(lamp
receiverpointinguporpointing
down)andlockitinplaceby
turningtheknurledlockring
towardthefrontofthetele
scope.
Next,removetheeyepieceassemblythatisstill
con
nectedtotheshortsectionofthetelescopebarrelwhich
wasremoved,andinstallitonthe1961eyepiece
adapter.
Toinstallthelight,removethecapfromthelampre
ceiver(undertherightthumbinFigure2.10).Thenin
stallthetwowayadapterbythreadingitsknurledring
ontothelampreceiver.PlugintheLEDlightsource
untilitclicksintheadapter.
Figure2.8
Figure2.9
Note:Toinstallarightangleview
ingeyepieceonthemaintelescope
(convenientforveryhighangle
shots),youmaywishtoinstalla193
Lratherthanthe1961.Theseac
cessoriesareinstalledinbasically
thesameway,byfirstremovingthe
shorttubesectioninfrontofthe
eyepiece.
Model 771, 75, and 76-RH190 Transits 14
Laterinthismanual,wewill
discusssomeoftheoptical
toolingtechniqueswhichre
quirebacklightingofthereti
cleusingoneoftheseadapt
ers.
Figure2.10
15 Operation and Adjustment Manual
Therearemanyjobsthatyoucandowithyourtransit,whichmay
bepurchasedwitha190typemicrometerinastandardconfigura
tion.Youcanevaluateanumberofalignmentrelatedcharacteris
ticsofmachineryorotherstructures,suchasstraightness,plumb,
level,flatness,andothers.However,toperformcertainOptical
Toolingprocedures,otheraccessoriesarerequired.Wewillmen
tionsomeoftheseaccessoriesaswediscusstherelatedtopics.
Inthischapterwewilldiscussthefollowingtopics:
ThisdiscussionismeanttomakeyouawareofsomeoftheOptical
Toolingproceduresthatyoucanuseinalignmentandmeasure
mentwork.Thismanualisnotagoodsubstituteforourtraining
classes,inwhichweofferhandsontrainingoftheseandmany
otherOpticalToolingtechniques.
We’llexplainwhatthesetermsmean,andhowtoaccomplishthem
usingyourtransit.
In
manyalignmentjobs,itisnecessarytoputthelineofsightof
yourtransitdirectlyonalinecreatedbytwoknownreferencetar
gets.Forexample,inFigure3.1(showingaviewfromoverhead),
sometargetshavebeeninstalledinthefloor.Thesetargetsrepre
sentalinewhich
isparalleltothecenterlineofamachineonwhich
youneedtoaligncomponents.Inpreparationfordoingthisalign
mentjob,itcouldbenecessarytobringtheinstrumentinlinewith
thesetwofloormonuments.
NoteinFigure3.1,thetransitmustnotonlyberotatedso
thatitis
paralleltothelinedefinedbythetargets,butitmustalsobeshifted
sideways.“Buckingin”isarepetitiveprocessinwhichyouadjust
boththeinstrument’sangleaswellasitslateralpositionsothatthe
transitisplaceddirectlyinlinewithtwotargets.
Ortobemoreac
curate,sothatthetransit’sverticalspindleliesintheverticalplane
whichalsorunsthroughthecenterofthefloortargets.
Notethatangularadjustmentsareeasyenoughtomakeusingthe
horizontaltangentadjustmentscrewonthetransititself.However,
Chapter3‐MeasuringTechniques
 “Buckingin”
 Autoreflection
 Autocollimation
 Collimation
 Collineation
 Reticleprojection
 Measuringwithmicrometers
“Buckingin”
Model 771, 75, and 76-RH190 Transits 16
Alwayssetthetelescopeon
thefartargetbymak
ingangularmoves
usingthehorizontal
andverticaltangent
screw(s).
Alwayssetthetelescopeon
theneartargetby
makinglateralmoves
usingalateralslideor
precisionlift.
lateralshiftsoftheentireinstrumentarenotaseasy.Togreatly
simplifythis,werecommendtheuseofalateralslidemountedbe
tweenthetransitandthestandonwhichitismounted.
Stepstobuckthetransitintothetargetline:
1. Roughlypositionthetransit.Setuptheinstrument
standnominallyinlinewiththetwotargets.
2. Roughlevel.Roughlevelthetransitasdescribedin
Chapter2“RoughLeveling”.
3. Preciselevel.Formanyapplications,preciseleveling
yourtransitisrequiredpriortobuckin(coveredin
Chapter2).Technicallyisitpossibletobuckintotargets
withoutpreciseleveling,butthetransitwillnotendupin
theverticalplanedefinedbythosetargetsunlessthetran
sititselfispreciselyplumb.Afulldiscussionof
thisfactisbeyondthescopeofthismanual,butis
coveredinmoredetailinourOpticalTooling
classes.
4. Settofartarget.Focusthetelescopeonthefar
targetandsetthecrosswiresdirectlyoverthattar
getusingthehorizontaland/orverticaltangent
adjustmentscrews.Ifthetransitisbetweenthe
targets(asinFigure3.2),simplyassignoneofthe
targetstobethe“fartarget”,andsetonit.
5. Settoneartarget.Withoutmovingthehori
zontaltangentscreworturningthestandardsof
thetransit,turnthetelescopedownwardandfocus
onthesecondtarget.Notethatifthetransitislo
catedbetweenthetargets(asinFigure3.2),youwill
havetoturnthetelescope“through”the
standards
topointintheoppositedirection.Regardlessof
whetheryouareinasituationlikeFigure3.1or
Figure3.2,thissecondtargetwillinvariablybeoff
toonesideofthefieldofviewunlessyouarein
crediblylucky.Forpurposesofillustration,let’s
saythattheobservedlateraldistancetothenear
targetis“L”(seeFigure3.1or3.2).Nowyoumust
movethetransitlaterallytowardtheneartarget
usingthecrossslide—buthowmuch?The
amountthatyoumovethetransitdependsupon
thegeometryofthesetup.Theequationtocalcu
latethe
properamountofshiftis:
TransitLateralShift=Lx(D2/D1)
where
D1=Distancebetweentargets
Figure3.1
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63

Brunson 75 Operation And Field Adjustment Manual

Category
Telescopes
Type
Operation And Field Adjustment Manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI