UVe-LUX METER
Part Number: 017B012
UVe-LUX METER
OPERATING INSTRUCTIONSOPERATING INSTRUCTIONS
Faraday Road, South Dorcan Industrial Estate,
Swindon, Wiltshire, SN3 5HE, UK
T +44 (0)1793 524566
F +44 (0)1793 490459
E sales@magnaux.co.uk
Bahnhofstraße 94-98, 73457 Essingen, Deutschland
T +49 (0)7365 81-0
F +49 (0)7365 81-449
E sales@magnaux.com.de
Lagergrens gata 2, 652 26 Karlstad, Sverige
T +46 (0)54 29 39 50
E info@magnaux.se
7
CONTENTS
1 Introduction 1
2 System components 2
2.1 Sensor housing 2
2.2 Meter housing 2
2.3 Buttons 3
2.4 Batteries 4
3 Display 5
3.1 Measurement display 5
3.2 Main menu 6
3.3 About screen 6
3.4 Brightness 6
3.5 Power Off time 7
3.6 Time/date 7
3.7 Set units 7
3.5 POWER OFF TIME
The POWER OFF TIME menu option allows you to set
an automatic timer to power the meter down, in order
to conserve battery power. The power off timer may be
set between 0 and 15 minutes. If the power off time is
set at 0, the meter will never turn itself off. If the time
is set outside the required range, an ‘ERROR IN TIME’
message will be displayed.
Error message
3.6 TIME/DATE
The SET TIME/DATE menu option allows you to change
the time and date settings.
To set the time:
Select the TIME ONLY option from the menu; ‘ENTER
CURRENT TIME’ will display. Increase or decrease the
time using the up and down arrow buttons. Use the left
and right arrow buttons to move between digits, then
use the accept button to complete.
To set the date:
Select the DATE ONLY option; ‘ENTER CURRENT
DATE’ will display. Increase or decrease the date using
the up and down arrow buttons. Use the left and right
arrow buttons to move between digits, then use the
accept button to complete.
Set time/date
3.7 SET UNITS
White Light Lux Units
Select ‘WL - Lux from the menu to set the unit to Lux
White Light Foot Candles
Select ‘WL-Foot candles to set the unit to ft-c
UV Light Micro Watt
Select ‘UV - µW/cm2 to set the unit to µW
UV Light Milli Watt
Select ‘UV - mW/cm2 to set the unit to mW Set Units
6
1. INTRODUCTION
The UVe-LUX is a compact, lightweight and robust meter specically designed for the
non-destructive testing environment. The meter is a dual white-light (WL) and ultra-violet
(UV) meter designed to give simultaneous readings of both WL and UV.
The meter has the following specication:
Single sensor head measures UV and white light at the same time
320 x 240 pixel 2.8” TFT colour LCD display
Dimmable display to extend battery life
Auto-ranging on both WL and UV scales
Menu driven setup and multi-function buttons
‘Auto Hold’ function to record maximum readings
Adjustable auto-power down to conserve battery life
Meets the requirements for non-destructive testing viewing conditions for
penetrant and magnetic particle inspections (ISO 3059)
Measurement units - WL: lux or fc, UV: μW/cm² 0 m W/cm²
Measuring range - WL: 0.1 to 5000 lux, UVA: 0 to 10,000 μW/cm²
Accuracy - WL: better than 5%, UV: better than 5%
2 x AA Rechargeable batteries
Display of time and date
Save function for logging readings
Detachable light sensor probe
3.2 MAIN MENU
Enter the menu system by pressing the accept button.
Once in menu-mode, use the up and down button to
scroll through menu items and press the Accept button
to select menu items. Throughout the main menu,
helpful hints at the bottom of the display will guide you
to the right button for the desired operation.
Holding the on/off key down for several seconds in
menu mode will power the system down.
Main menu screen
3.3 ABOUT SCREEN
The ABOUT screen (also the splash screen) displays
the software number, meter serial number and the light
probe serial number. This screen also indicates the
period of time left before recalibration is required.
About/splash screen
3.4 BRIGHTNESS
Select this option to adjust the brightness. Use the
arrow buttons to select either BRIGHTER or DARKER,
then press the accept button.
Brightness menu
1
25
2. SYSTEM COMPONENTS
The system consists of a sensor housing containing both the WL and UV sensors, and
the meter housing containing the main system components such as the screen, battery
and user interface. Both the sensor housing and the meter housing are constructed from
reinforced ABS plastic and are rated to IP64 for moisture and dust ingress. The two
housings are linked together with 750mm of exible screened cable for ease of use.
2.1 SENSOR HOUSING
The sensor housing contains the individual WL and UV
detectors in one handy unit. The active area of the sensor
housing is the circular glass panel on the front of the
housing. Ensure that the glass panel remains clean and
is not damaged in any way, as any dirt or damage to the
glass panel will affect the accuracy of the meter.
If the panel requires cleaning, use a damp cloth with
a small amount of detergent to wipe the panel, then
use a dry, lint-free cloth to dry the panel. Do NOT use
abrasive cleaners and solvents as they may damage
the unit.
If the panel gets damaged, return the whole meter to
the manufacturer for repair and re-calibration.
Sensor housing
2.2 METER HOUSING
The meter housing contains the display screen, buttons
and battery for the unit. The meter housing is rated at
IP64 for moisture and dust ingress. If it needs cleaning,
use a damp cloth and mild detergent.
Probe connection Charging socket
3. DISPLAY
In normal operation, when the unit is powered up,
the rst screen to appear is the logo screen, followed
by the splash screen. The splash screen contains
information about the meter, including the serial number
and the date of last calibration for the unit. This screen
can also be found in the main menu under ABOUT.
The LCD display is protected behind a protective
plastic cover, but take care to protect the screen area
from blows, scratches or other damage.
Splash screen
3.1 MEASUREMENT DISPLAY
When the unit is powered up, it will normally enter the measurement display after the
splash screen. The measurement display contains six components:
• Battery level - this gives an indication of the state of the battery life.
• Current time and date.
• WL reading - this is the current rounded reading for the white light sensor.
• WL max - this is the maximum absolute value the meter has observed for white
light. This reading is reset when the cancel button is pressed.
• UV reading - this is the current rounded reading for the UV sensor.
• UV max - this is the maximum absolute value the meter has observed for UV. This
reading is reset when the cancel button is pressed.
If the meter is exposed to light levels greater than its designed range, the meter will
saturate and display ‘9999 to indicate that it has saturated. The meter will return to normal
operation once the light levels have returned to the meter’s designed range.
2.3 BUTTONS
On/Off button:
Press this button (while the unit is off) to switch the UVe-LUX on
Hold down this button (while the unit is on) to switch the UVe-LUX off
Accept button:
Press this button in measurement mode to enter menu mode
Press this button in menu mode to select/approve the highlighted menu
function.
Cancel Button:
Press this button in measurement mode to reset max values
Press this button in menu mode to cancel selection or go back to previous
menu function.
Arrow up button:
Press this button in menu mode to cycle up through the menu options
Use within submenus for incrementing values
Arrow down button:
Press this button in menu mode to cycle down through menu options
Use within submenus for descending values
Arrow right button:
Press this button to move right for selection within submenus
Arrow left button:
Press this button to move left for selection within submenus
4
2.4 BATTERIES
The UVe-LUX is powered by two AA type, rechargeable batteries. You can access the
batteries by unscrewing the screw holding the battery housing cover on the rear of the
meter housing, and removing the cover. It is important to insert the batteries in the correct
orientation as marked in the battery compartment, otherwise damage may occur to the
meter. The battery housing cover should then be replaced and screwed back down.
We recommend that the batteries supplied with the meter should be charged
overnight (the meter must be off whilst the batteries are charging). The charger
supplied with the meter is for the sole purposeof charging the batteries and must
not be used independently (i.e. as an external power adaptor) to run the meter
under normal operating conditions.
Whilst connected to the charger, the top-right of the screen will display ‘CHARGING’.
When the batteries are full, the sreen will display ‘CHARGED’. Once the batteries are
fully charged, the meter will run for approximately eight hours before needing a recharge.
‘BATTERY LOW’ means the batteries need to be recharged or the meter will power itself
down. Ordinary dry cell batteries can be used and no damage will result if the battery
charger supplied is inadvertently used.
Battery terminals
3
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6