Sanyo MCO-19M User manual

Type
User manual

This manual is also suitable for

Service Manual
Multi-gas Incubator
SANYO Electric Co., Ltd.
Biomedical Business Division
MCO-19M
MCO-19M(UV)
SM9910245
FILE No.
()
()
Effective models
This service manual is effective for following models.
Name Product code Voltage and Frequency
MCO-19M 823 633 51 115V 60Hz
823 633 54 230/240V 50Hz
MCO-19M(UV) 823 633 71 115V 60Hz
823 633 74 230/240V 50Hz
Contents
㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌 Page
Specifications ------------------------------------------------------- 1
- Structural specifications
- Control specifications
- Performance specifications
Dimensions ----------------------------------------------------------------- 4
Gas circuits ----------------------------------------------------------------- 5
Temperature calibration ----------------------------------------------------------------- 6
CO
2
span calibration ----------------------------------------------------------------- 7
O
2
span calibration ----------------------------------------------------------------- 8
CO
2
/O
2
zero calibration ----------------------------------------------------------------- 9
Components on PCB ----------------------------------------------------------------- 10
Connections on PCB ----------------------------------------------------------------- 11
Electric parts ----------------------------------------------------------------- 12
Wiring diagram ----------------------------------------------------------------- 13
Circuit diagram (MAIN / LCD) ----------------------------------------------------------------- 14
Control specifications ----------------------------------------------------------------- 16
Operational specifications of LCD panel ---------------------------------------------- 23
Prevention of contamination ----------------------------------------------------------------- 71
Specifications of sensors ----------------------------------------------------------------- 72
Mechanism of Double wavelength IR sensor ---------------------------------------------- 74
Water sensor ----------------------------------------------------------------- 75
Parts layout ----------------------------------------------------------------- 76
Setting procedures for optional components -------------------------------------- 77
How to install UV lamp add-on kit MCO-19UVS -------------------------------------- 81
Installation of H
2
O
2
Generator (MCO-HP) ---------------------------------------------- 87
How to install H
2
O
2
decontamination kit (MCO-HL) and Electric lock ------------- 88
How to install Roller Base MCO-18RB ---------------------------------------------- 92
How to install Automatic CO
2
cylinder changeoverMCO-21GC ------------- 93
How to install Automatic CO
2
density calibration add-on kit MCO-SG ---------- 101
Test data ----------------------------------------------------------------- 102
- Temperature pull-up data
- Humidity pull-up data
- Temperature change during power failure
- Temperature recovery data (30sec/60sec door open)
- Humidity recovery data (30sec/60sec door open)
- CO
2
recovery data30sec/60sec door open
- N
2
/O
2
recovery data30sec/60sec door open
- Days before CO
2
tank becomes empty
- Days before N
2
/O
2
tank becomes empty
- Temperature distribution data – 15points measured
Instruction manual ----------------------------------------------------------------- 106
- 1 -
Specifications
ŶStrucutural specifications
Name Multi-gas
Incubator
Model MCO-19M MCO-19M(UV)
External dimensions W620 x D710 x H900 mm
Internal dimensions W490 x D523 x H665 mm
Interior volume 162 L
Exterior Painted steel
Interior Stainless steel containing copper
Outer door Painted steel
Inner door Stainless steel with gasket
Gas tight split door 4 doors made of temperared glass
Tray
3 trays, made of stainless steel containing copper
Dimensions; W450 x D450 x H12 mm, Maximum load; 7 kg/tray
Access port Inner diameter; 30 mm, On the back side
Insulation Rigid-polyurethane foamed-in place
Heating system DHA system
Heater jacket and air jacket
Heater 314 W
Humidifying system Natural evaporation with humidifying pan
Temperature controller PID control system
Temperature display Digital display
Resolution: 0.1
o
C
CO
2
controller PID control system
CO
2
density display Digital display (Resolution: 0.1%
O
2
controller PID control system
O
2
density display Digital display (Resolution: 0.1%
Air circulation Fan assisted
Air filter 0.3μm, Efficiency: 99.97 % or more
UV lamp (Option) 4 W×1 (Ozone-free emission)
Water level sensor Optical type
Alarms
Automatic set temperature alarm, automatic set CO
2
density alarm,
Automatic set O
2
density alarm, Upper limit temperature alarm,
various gas/sensor/heater alarms, Door alarm, UVlamp failure
Remote alarm contacts Allowable contact capacity: DC 30 V, 2 A
CO
2
inlet connection Diameter; 4
6 mm tube
CO
2
inlet pressure 0.03 MPaG
0.3 kg/cm
2
G, 4.3psiG
O
2
inlet connection Diameter; 4
6 mm tube
O
2
inlet pressure 0.05 MPaG
0.5 kg/cm
2
G, 7.1psiG
- 2 -
Name Multi-gas
Incubator
Model MCO-19M MCO-19M(UV)
Accessories
3 trays, 3 sets of tray support, 1 gas tube, 1 humidifying pan
1 gas injection nozzle, 1 tube for gas injection nozzle,
1 pair of stacking plate A and B, 2 protective stickers, 6 tube bands
Weight 94 kg
UV Lamp Add-on kit
MCO-19UVS
MCO-19UVS is standard equipment
Optional accessories
H
2
O
2
generator
MCO-HP
H
2
O
2
Decontamination kit
MCO-HL
H
2
O
2
Decontamination Reagent
MCO-H2O2
Automatic CO
2
density Calibration Add-on Kit
MCO-SG
Gas pressure regulator
MCO-100L
Extra tray
MCO-47ST
including 2 tray supports,
Half tray
MCO-25ST
,
Automatic CO
2
cylinder changeover system
MCO-21GC
Roller base
MCO-18RB
UV lamp replacement kit (MCO-20UV)
Stacking spacer (MCO-18SB, when stacking on the MCO-175)
Stacking spacer (MCO-21SB, when stacking on the MCO-20AIC)
Ethernet (LAN) Interface (MTR-L03), RS232C/RS485 Interface (MTR-480)
SANYO Data acquisition software (MTR-5000)
4-20mA Interface (MCO-420MA, USA only)
- 3 -
ŶControl specifications
Model MCO-19M MCO-19M(UV)
Thermal sensor Thermistor
Temperature display LCD digital display
CO
2
controller Microcomputer; Input by keypad
CO
2
density setting range: 0
䠂䡚
20
䠂䠄
Resolution; 0.1
䠂䠅
Non-volatile memory
CO
2
density display LCD digital display (Resolution; 0.1
)
Water level detection Water level sensor
Optic type
H
2
O
2
decontamination
(Option)
- Microwaved H
2
O
2
decontamination
- Low density H
2
O
2
solution (Oxidor) which can be purchased in the market
- Solution level detection by sensor
Electric lock (Option) Outer door is locked during H
2
O
2
decontamination
UV system Option. UV lamp lighting time can be
extended automatically
UV lamp lighting time can be extended
automatically
Self-diagnosis - Error code and error information are displayed by text.
- Remote alarm contact activates (DC30V, 2A)
with buzzer sounds.
ŶPerformance specifications
Model MCO-19M MCO-19M(UV)
Temperature control range Ambient temp. +5
50
o
C
ambient temp; 5
35
o
C
Temperature distribution ±0.25
o
C
ambient temp; 25
o
C
setting; 37
o
C, CO
2
5%, O
2
5%, no load
Temperature variation ±0.1
o
C
ambient temp; 25
o
C
setting; 37
o
C, CO
2
5%, O
2
5%, no load
CO
2
control range 0
20 %
CO
2
variation ±0.15 %
ambient temp; 25
o
C, setting; 37
o
C, CO
2
5%, O
2
5%, no load
O
2
control range 1~18, 22~80 %
O
2
variation ±0.2 %
ambient temp; 25
o
C, setting; 37
o
C, CO
2
5%, O
2
5%, no load
Chamber humidity 95±5%R.H.
Rated voltage, frequency 110~120V, 60Hz 220V, 60Hz 220~240V, 50Hz
Amperage Max. 3.1A Max. 1.5A Max. 1.5A
Heat emission Max. 1274 kJ/h Max. 1206 kJ/h
Power consumption Max. 354 W Max. 335W
Noise level 30 dB (A scale)
Usable environment condition
Temperature: 5~35
o
C, Humidity: Equal or less than 80%R.H.
Designed performance may not be obtained when ambient temp. is equal
or lower than 15
o
C
Note) Specifications will be subject to change without notice.
- 4 -
Dimensions
Power cord
- 5 -
CO
2
sensor
CO
2
inlet
port
N
2
/O
2
inlet
port
O
2
/N
2
valve A
O
2
/N
2
valve B
Filter
Sample port
(Front)
O
2
sensor
CO
2
valve
Gas circuits
- 6 -
Temperature calibration
Here is the explanation how to calibrate chamber temperature.
Ex) When setting temperature is 37.0 and actual temperature is 36.5
Process Operation MENU
0 Set thermal sensor at the center of the chamber and start the unit running.
Leave it until chamber temperature is stable.
1 Press MENU key in Top screen to call MENU screen. Select Tools. Tools
2 Select Tools (1/2) screen is displayed. Select Temp/CO
2
Calibration.
3 Temp/CO
2
Calibration is displayed and press MENU key to activate MENU key. Select
OK and press ENTER key.
OK
4
Change the value in the column Temp Span to 36.5
by pressing digit shift key and
numerical value shift key.
5 Press MENU key to call MENU screen. Select OK and press ENTER key. OK
6 Automatically reverts to Select Tools (1/2) screen.
*Note (1) It is important to have an accuracy in thermal sensor when measuring center temperature of
the chamber.
(2) Input range:1.0 ~ 1.0. It is unacceptable to input over the range at once.
<Display of Process 1>
<Display of Process 3>
<Display of Process 4>
- 7 -
CO
2
span calibration
(ExWhen setting CO
2
density is 5.0% and actual CO
2
density is 4.5%:
Process Operation MENU
1 Put densitometer thorough the sample port at the front side of the unit to measure
chamber CO
2
density.
2
Select Tools in the Top Screen.
Tools
3
Select Tools (1/2) is displayed. Select temp / CO
2
Calibration and press OK key.
OK
4 By pressing digit shift key and numerical value shift key, input actual CO
2
density
4.5% in the column CO
2
Span. Select OK and press ENTER key.
OK
5
Automatically reverts to Select Tools (1/2) screen.
*Note (1) Input range: +1.0% ~ -1.0% (per input).
(2) Repeat CO2 calibration for several times until displayed CO2 density will match with
actual density.
(3) It is permitted to commence CO2 span calibration only when setting CO2 density is 5.0.
CO
2
span calibration
Display of Process-2
Display of Process-3
- 8 -
O
2
span calibration
(ExWhen setting O
2
density is 5.0% and actual O
2
density is 4.5%:
Process Operation MENU
1 Put densitometer thorough the sample port at the front side of the unit to measure
chamber O
2
density.
2
Select Tools in the Top Screen.
Tools
3
Select Tools (1/2) is displayed. Select temp / CO
2
Calibration and press OK key.
OK
4 By pressing digit shift key and numerical value shift key, input actual O
2
density
4.5% in the column O
2
Span. Select OK and press ENTER key.
OK
5
Automatically reverts to Select Tools (1/2) screen.
*Note (1) Input range: +1.0% ~ -1.0% (per input).
(2) Repeat O2 calibration for several times until displayed CO2 density will match with
actual density.
(3) It is unacceptable to commence O2 span calibration when setting O2 density deviates
rom 18.0 ~ 22.0.
㻌㻌㻌There is a blank in the rignt side of O2 Span column.
Display of Process-2
Display of Process-3
O
2
span calibration
- 9 -
CO
2
/O
2
zero calibration
Prior to commence calibration
zMake sure chamber temperature should be stabilized, CO
2
sensor BOX temperature should be
45±2, setting CO
2
density should be 0.0% and setting O
2
density should be 20.0%.
zMake sure there should not be left CO
2
in the chamber.
* Note) When you re-commence CO
2
calibration, outer door should be opened and make sure
there should not be left CO
2
in the chamber.
Procedure
1. Call Select Tools (1/2) screen to display Temp/CO2/O2 Calibration.
2. Turn the VR3 around CN15 on the main PCB or measure the voltages both ends of R63 to make
CO
2
sensor reference cell voltage
CO2 Zero
to 4.00V±0.03V.
3. Turn the VR2 around CN3 on the main PCB or measure the voltages both ends of R40 to make
CO
2
sensor gas cell voltage
CO2 Span
to 4.00V±0.03V.
4. Input 0.0% by pressing digit shift key and numerical value shift key and “0.1”% besides right to
the “CO2 Zero” being turned to black. (See the picture below)
5. Press MENU to call MENU display. Select OK and press ENTER key.
- 10 -
Components on PCB
CN14
#1 Local power supply (AC)
#2 Baliast (Option for MCO-19M)
CN302
#1 Top heater,Back heater,
Side heater, Frange heater
#3 Bottom side heater
Bottom front heater
Bottom back heater
#5 Front panel heater
Bimetal thermo
Door heater
CN303
#1 CO
2
Heater A/B
#3 Agitating fan motor
CN20
#1-#2 H
2
O
2
relay (Option)
#3-#4 Electric key (Option)
#5-#6 Standard gas valve(Option)
CN21
Control PCB (KEY)
Display PCB (LCD)
CN6
#1-#2 Door Switch
#3-#6 MCO-HP(Option)
#7-#8 Standard gas SW (Option)
CN17
#1-#2 O
2
A valve
#3-#4 O
2
B valve
CN9
#1-#2 CO
2
valve
#3-#4 CO
2
auto changer valve (option)
CN3
#3-#7 CO
2
sensor
CN1
#1-#5 Power transformer
CN2
#1-#2 Switching power supply
CN15
#1-#3
Water sensor
CN16
#1-#4
O
2
sensor
CN10
#1-#2 Remote alarm output
CN11
MTR-480C(Option)
MTR-L03
MTR-420MA
CN4
#1-#2 Overheat protect sensor
#3-#4 Temp.sensor
#5-#6 CO
2
BOX Temp.sensor
R63
For CO
2
zero calibration
R40
For CO
2
span calibration
CN12
#3-#4
Hinge sub heater
CN13
Buzzer PCB
- 11 -
Connections on PCB
Connections on Main PCB
Connector Connects to Usage
Power transformer
CN1
#1 - #2: AC8.5V
#3: GND
#4, #5: AC18V
Power supply to PCB
CN2
Switching power supply
#1 - #2: GND, DC5V
Power supply to PCB
CN3 CO
2
sensor Control of CO
2
density
CN4
#1 - #2:
#3 - #4:
#5 - #6:
Overheat protect sensor
Temp. sensor
CO
2
BOX Temp. sensor
Detection of temperatures in the
chamber
CN6
#1 - #2: Door switch
#3 - #6: MCO-HP Option
#7 - #8: Standard gas switch (Option)
Detection of door opening
CN9
#1 - #2: CO
2
valve DC24V
#3 - #4: CO
2
auto changer valve (Option)
Control of CO
2
valve
CN10
Remote alarm output
#1: COM
#2: N.O.
#3: N.C.
Output for remote alarm contact
Open between #1 and #2 in
normal condition.
CN11 MTR-480 (Option) Communication board
CN12 #3 - #4: Hinge sub heater Control of hinge sub heater
CN13 #1 - #2: Buzzer PCB
CN14
#2: UV lamp, ballast Option for MCO-19M
CN15 #1 - #3: Water sensor
Detection of water level in
humidifying pan
CN16 O
2
sensor Control of O
2
density
CN17
#1 - #2: O
2
A valve
#3 - #4: O
2
B valve
Control of O
2
valve
CN20
#1 - #2: H
2
O
2
relay DC24V䠅䠄Option
#3 - #4: Electric lockDC24V䠅䠄Option
#5 - #6: Standard gas valveDC24V䠅䠄Option
CN21
Control PCBKEY
Display PCBLCD
CN302
#1: Top heater, Back heater, Side heater,
Flange heater
#3: Bottom side heater, Bottom front/back heater
#5: Front panel heater, Bimetal thermostat,
Door heater
Control of heaters
CN303
#1: Temp. fuse 70, CO
2
heater A,
CO
2
heater B
#3: Condensing fan motor
Control of temperature in CO
2
sensor BOX
Control of condensing fan motor
Condensing fan motor
Type FL2-011Y1MP FL2-011Y5MP
Power supply
Type HWS15-5 ZWS15-5/J
Rating 110-115V 220-240V DC5V Rating
IN:85-265VAC 5V, 3A
Temp. fuse 70
Type SF70U SF70U
Power switch
Type BAM210131 BAM210131
Rating 250V, 7A 250V, 7A Rating AC250V, 10A AC250V, 10A
Flange heater
Rating AC115V, 31W AC230V, 31W
Door switch
ᆺᘧ SS160-A15 SS160-A15
Resistance 432.7ȍ 1685.4ȍ ᐃ᱁ 28V, 50MA 28V, 50MA
Door heater
Rating AC115V, 85W AC230V, 85W
Noise filter
Type ZHG2210-11S ZHG2210-11S
Resistance 156.9ȍ 610.3ȍ Rating AC250V, 10A AC250V, 10A
Bottom front heater
Rating AC115V, 11.5W AC230V, 10W
CO2 valve
Type FAB11-X2265 FAB11-X2265
Resistance 1149ȍ 5319ȍ Rating DC24V DC24V
Bottom front/back heater
Rating AC115V, 10W AC230V, 10W
O2 valve A/B
Type FAB216212C FAB216212C
Resistance 1308ȍ 15558.8ȍ Rating
CO2 heater A
Rating AC71V, 19.5W AC141V, 19.5W
Power transformer
Type ATR-R502V ATR-R502V
Resistance 254ȍ 1014ȍ MCO-HL, option) Rating 115V 230V
CO2 heater B
Rating AC45V, 12.5W AC89V, 12.5W
H2O2 relay
Type G2R-1A-T G2R-1A-T
Resistance 160ȍ 639ȍ
(MCO-HL, option)
Rating 24V, 10A, 250V 24V, 10A, 250V
Front panel heater
Rating AC115V, 37W AC230V, 37W
Electric key
Type LE-36 LE-36
Resistance 353.7ȍ 1460ȍ
(MCO-HL, option)
Rating DC24V DC24V
Bottom side heater R, L
Rating AC115V, 14W AC230V, 14W
Standard gas switch
Type SPS-35 SPS-35
Resistance 945ȍ 5290ȍ
(MCO-SG, option)
Rating 5KPA 5KPA
Side heater R, L
Rating AC115V, 30W AC230V, 30W
Standard gas valve
Type FAB11-X2265 FAB11-X2265
Resistance 441ȍ 1763ȍ
(MCO-SG, option)
Rating DC24V DC24V
Top heater
Rating AC115V, 8.0W AC230V, 8W
Resistance 1647ȍ 6590ȍ
Back heater
Rating AC115V, 12W AC230V, 12W
Resistance 1104ȍ 4415ȍ
Hinge sub heater R,L Rating AC36V, 0.3W AC36V, 0.3W
Resistance 4410ȍ 4410ȍ
Bimetal thermostat
Type TH-2U TH-2U
Rating ON:60 OFF:70 ON:60 OFF:70
CO2 sensor
Type IR sensor IR sensor
Rating TPS2534 TPS2534
O2 sensor
Type FCX-MW FCX-MW
Water sensor
Type OLS-10003S OLS-10003S
Rating 5V 5V
Overheat protect sensor
Type 103AT-1 103AT-1
Rating 10Kȍ,25 10Kȍ,25
Temp. sensor
Type 103AT-1 103AT-1
Rating 10Kȍ,25 10Kȍ,25
CO2 BOX temp. sensor
Type 103AT-1 103AT-1
Rating 10Kȍ,25 10Kȍ,25
Ballast
Type EKF-04115B1 EKF-04115B1
(Option for MCO-19M)
Rating 115V, 4W 115V, 4W
UV lamp
Type GL4 GL4
(Option for MCO-19M)
Rating 4W 4W
Glow starter Type FG-7P FG-7P
(Option for MCO-19M)
Power transformer
Type ATR-HN231UT ATR-HN235T
Rating 8.5V, +/-18V 8.5V, +/-18V
AC115V, 60Hz AC230/240V, 50HzAC115V, 60Hz
MCO-19M/19M(UV)
AC230/240V, 50Hz
MCO-19M/19M(UV)
Electric parts
- 12 -
Wiring diagram
- 13 -
Circuit diagram
<Main PCB>
- 14 -
< LCD PCB >
- 15 -
- 16 -
Control specifications
1. Temperature control
(1) Setting range: 0.0Υ~50.0Υ(Default setting: 37.0Υ)
(2) Automatic set temperature alarm
High temperature alarm setting range: Set temperature +1.0~+5.0Υ(Default setting:+1.0Υ)
Low temperature alarm setting range: Set temperature -1.0~-5.0Υ(Default setting:-1.0Υ)
(3) Alarm delay time: 0~15 minutes (Default setting: 15 minutes)
(4) How to set temperature
In TopScreen, press MENU key and select Set to display Stand-by Setting screen.
Move a curser on “Temperature” and set a value by pressing numeric keys.
Press MENU key and select OK and press ENTER key to memorize the value.
2. CO2 control
(1) Setting range: 0.0%~20.0%(Default setting: 0.0%)
(2) Automatic set CO2 density alarm
High CO2 density alarm setting range: Set density +0.5%~+5.0% (Default setting: +1.0%)
Low CO2 density alarm setting range: Set density -0.5%~-5.0% (Default setting: -1.0%)
(3) Alarm delay time: 0~15 minutes (Default setting: 15 minutes)
(4) How to set CO2 density
In TopScreen, press MENU key and select Set to display Stand-by Setting screen.
Move a curser on “CO2 Density” and set a value by pressing numeric keys.
Press MENU key and select OK and press ENTER key to memorize the value.
3. N2/O2 control
(1) Setting range: 1.0%~80.0% (Default setting: 20.0%)
It is uncontrolled O2 density when O2 setting range is exceeding 18.0% and less than 22.0%.
(2) Automatic set O2 density alarm
High O2 density alarm setting range: Set density +0.5%~+5.0% (Default setting: +1.0%)
Low O2 density alarm setting range: Set density -0.5%~-5.0% (Default setting: -1.0%)
(3) Alarm delay time: 0~30 minutes (Default setting: 30 minutes)
(4) How to set O2 density
In TopScreen, press MENU key and select Set to display Stand-by Setting screen.
Move a curser on “O2 Density” and set a value by pressing numeric keys.
Press MENU key and select OK and press ENTER key to memorize the value.
4. Alarms, safety functions and self diagnosis functions
(1) Upper limit temperature alarm
Setting range: Approx.35Υ~51Υ
When a chamber temperature is over the setting upper limit temperature, OVERHEAT indicator
will illuminate, buzzer will sound continuously and remote alarm will be active.
- 17 -
(2) Automatic set temperature alarm
Setting range: Set temperature +/-1.0Υ~+/-5.0Υ (Default setting: +/-1.0Υ)
When a chamber temperature deviates from alarm setting range, the digits displaying current
chamber temperature will start blinking.
The following message will be displayed after an alarm delay time elapses.
“Warning:High Temp” --- For high temperature alarm
“Warning:Low Temp” --- For low temperature alarm
Buzzer will sound intermittently and remote alarm will be active.
(3) Automatic set CO2 density alarm
Setting range: Set CO2 density +/-0.5%~+/-5.0% (Default setting: +/-1.0%)
When a CO2 density deviates from alarm setting range, the digits displaying current CO2 density
will start blinking.
The following message will be displayed after an alarm delay time (0~15minutes) elapses.
“Warning:High CO2 Density” --- For high CO2 density alarm
“Warning:Low CO2 Density” --- For low CO2 density alarm
Buzzer will sound intermittently and remote alarm will be active.
(4) Automatic set O2 density alarm
Setting range: Set O2 density +/-0.5%~+/-5.0% (Default setting: +/-1.0%)
When a O2 density deviates from alarm setting range, the digits displaying current O2 density
will start blinking.
The following message will be displayed after an alarm delay time (0~30minutes) elapses.
“Waring:High O2 Density” --- For high O2 density alarm
“Warning:Low O2 Density” --- For low O2 density alarm
Buzzer will sound intermittently and remote alarm will be active.
(5) Auto Return
If there are no key operations for 90seconds in each setting screen, the setting mode will be
forcibly to the end and Top screen will be appeared.
(6) Key Lock
When Key Lock is active, the following message will be displayed on the stand-by setting screen.
“Key Lock” --- Key Lock is active
It is not permitted to change a stand-by setting value.
(7) Door alarm
Setting range of door alarm delay time: 1~30 minutes (Default setting: 2 minutes)
If an outer door is left opened, the following message will be displayed on the status screen.
“Door:Open” --- Outer door is left opened
Buzzer will sound intermittently after a door delay time elapses.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187

Sanyo MCO-19M User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI