4WFB120, 4WFB350, 4WFB1K 4 Wire Full Bridge Terminal Input Modules
5.2.1 CR9000(X)
This example program is slightly different in operation than the examples for
the other dataloggers. Data are only output to data table STRAINS when the
user sets Flag(1). Every measurement is output (rather than averages like in
the other examples) while Flag(1) is high.
' Program name: STRAIN.DLD
Public Count, ZStrain, StMeas, Strain, Flag(8) 'Declare all variables as public
'Data Table STRAINS samples every measurement when user Sets Flag(1) High
DataTable(STRAINS,Flag(1),-1)
DataInterval(0,0,0,100) 'Interval = Scan, 100 lapses
Sample (1,Strain,Ieee4)
EndTable
'DataTable ZERO_1 stores the "zero" measurements
DataTable(ZERO_1,Count>99,100) 'Trigger on Count 100
Average(1,ZStrain,IEEE4,0)
EndTable
'Subroutine to measure Zero, Called when user sets Flag(2)low
Sub Zero
Count = 0 'Reset Count
Scan(10,mSec,0,100) 'Scan 100 times
BrFull(ZStrain,1,mV50,5,1,6,7,1,5000,1,0,0,100,1,0)
Count = Count + 1 'Increment Counter used By DataTable
CallTable ZERO_1 'Zero_1 outputs on last scan (Count=100)
Next Scan
ZStrain = ZERO_1.ZStrain_Avg(1,1) 'Set ZStrain = averaged value
Flag(1) = True
End Sub
BeginProg
Scan(10,mSec,0,0) 'Scan 10(mSecs)
If Not Flag(2) Then Zero
BrFull(StMeas,1,mV50,5,1,6,7,1,5000,1,0,0,100,1,0)
StrainCalc(Strain,1,StMeas,ZStrain,-1,2,0)
CallTable STRAINS 'Strains outputs only when Flag(1)=True
Next Scan
EndProg