18
If WT RANGES = 2 (Dual Interval), Cont.
LOW RNG= (Allowable values are: 1 through 9,999,999).
HI INT=X
If
LFT=YES
(Allowable values are: 1, 2, 5, 10, 20 or 50).
If
LFT=NO
(Allowable values are: 1 through 99).
HI DPP
(Allowable values are 0, 1, 2 or 3).
0 = XXXXXX 1= XXXXX
.
X 2 = XXXX
.
XX 3 = XXX
.
XXX
INTERVAL
If
LFT=YES
(Allowable values are: 1, 2, 5, 10, 20 or 50).
If LFT=NO (Allowable values are: 1 through 99).
DPP (Allowable values are 0, 1, 2 or 3).
0 = XXXXXX 1= XXXXX.X 2 = XXXX.XX 3 = XXX.XXX
CAP (Allowable values are: 1 through 9,999,999).
Z TRACK (Allowable values are: 0 (disables Zero Tracking), 0.5, or 1 through 9).
Z LIMIT
(YES = 5% scale capacity, NO = Full Capacity, no limit).
PUO
(YES = Auto Re-Zero on Power-Up, NO = No Re-Zero on Power-Up).
TIME
(12 = 12 hour clock (3PM displays 3:00).
(24 = 24 Hour Clock, (3PM displays 15:00).
CLR TARE (YES = Clears Store Tare when Net Weight goes below Zero).
(NO = Stored Tare
is not
cleared when Net Weight goes below Zero).
CLR ID
(YES = Automatically clears ID after print)
(NO = ID
is not
cleared after print).
ANALOG? - Analog to Digital Filtering
If analog to digital filtering setup is desired, press the 1/YES key, then the ENTER key.
Otherwise press the ENTER key to advance to
CALIBRATE?
.
FILTER
(Allowable values are: 0 = disabled, 1 = minimal, 2 = moderate or 3 = custom).
If
FILTER =
0, 1, or 2 advance to the
SAMPLES
.
If
FILTER =
3
F=
(Filter level, allowable values are
0
to
255
).
B= (Break range, allowable values are 1 to 255).
SAMPLES (Allowable values for the sample rate are: 1 through 100).
STABLE (Motion detection, allowable values are 0 through 99 divisions).
STBL CNT
(Consecutive Stable weight readings, allowable values are 1 through 255).
CALIBRATION? - Calibration
If calibration is desired, press the 1/YES key, then the ENTER key. Otherwise press the
ENTER key to advance to
SERIAL?
.
If the indicator was calibrated previously and the four calibration “C” numbers were recorded,
you may enter the “C” numbers instead of using test weights. Please note that if the scale is
used in a commercial application and must be "Legal for Trade", the "C" numbers can
not be
used to re-calibrate.
ENTERING CALIBRATION “C” NUMBERS
1. With
CAL WT 1=
displayed, press the "diamond T"
TARE
key.
2. The display will show C1=XXX.
3. If the
C1=
number displayed is acceptable, press the ENTER key.
4. Otherwise, enter a new
C1=
number, then press the ENTER key.
5. Repeat steps 2 through 4 for
C2=
,
C3=
and
C4=
.