  • Hello! I've reviewed the Quick Guide for the Invacare Softform Active 2 RX pump mattress system. This document includes information about the control unit indicators, how to install the system, and instructions for performing CPR. I'm ready to answer your questions about the device and its operation described in the manual.
  • What does the Mute button do?
    What do the red LED indicators mean?
    How do I perform CPR?
Invacare UKOpera ons Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ, UK
60132743-A 2022-01-10 1 / 3
Control unit menu display
KJ
G
H
I
C D E F
A
B
QuickGuide
Invacare® Softform Active® 2 with RX pump
Pos.DescriptionFunction
AMutebutton
The relevant audible / visible indicator turns on when low pressure,
power failure or alternate failure is recognised. To mute the audible
signal, press the Mute button. The visible indicator will ash u ntil the
problem is solved. If the problem indicated has not been solved within
10 minutes, the indicator will sound again. Press the Mute button to
mute the audible indication, the indicator will not sound again. Refer to
Identifying and repairing faults on page 3 of this guide.
BSystemincycle/Setupindicator
(OperativeLED)
GreenLEDindicatorasheswhilstsystemissettingupSolidgreenLED
indicatesnormaloperational.
CAlternatingsystem/Timingfailure
indicator
RedLEDindicatorashesandaudiblebuzzersoundsifthesystem
recognisesacycletimeproblem.
DPowerfailureindicatorRedLEDindicatorashesandaudiblebuzzersoundsifthesystemhas
nopowerconnected.Thiscouldbeintheeventofapowerfailure,for
examplecausedbypowercordunpluggedorpoweroffwhencontrol
unitisworking.Whenpowerisrestored,controlunitwillautomatically
startworking.
EPressurefailure/Lowpressure
indicator
RedLEDindicatorashesandaudiblebuzzersoundsifthesystem
recognisesalowpressurefailure.
Operation
1. The power indicator will flash and control unit will enter static mode initially until pressure reaches 15 mmHg. Then control
unit will enter alternate mode and power indicator will remain on. The cycle time is set at 10 min, and the pressure is
set at 60 ± 3 mmHg.
Therstinationofnewmattress:theindicatorfunction(visibleandaudibleindicators)willactivateifthemattress
doesnotinatecompletelywithin15minutes.
Operationfailures
• During control unit operation, low pressure indicator will activate within 1 minute if air hoses are disconnected or
air cells are broken.
• In the case of abnormal alternate or no alternate, the audible signal will be activated and alternate failure indicator
will flash.
InvacarePortugal,Lda
RuaEstradaVelha949
4465-784LeçadoBalio
Portugal
Invacare Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ
UK
Invacare UKOpera ons Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ, UK
60132743-A 2022-01-10 2 / 3
Installing the system
KJ
G
H
I
C D E F
A
B
QuickGuide
Invacare® Softform Active® 2 with RX pump
1.Hangthecontrolunitbymeansofthe2builtinhangers
Aattheendofthebedorplaceitonahorizontal
surface.(Placingthecontrolunitontheoorwillnot
affecttheperformance,butmayexposeittoaccidental
damage.)
2.ConnecttheCPRhoseEtoairoutletsonthemattress
andcontrolunitB,ensuringthatthehoseisnotkinked
ortwisted.Allowspaceforbedrailstodropfreely.
3.Connectthemainspowercabletothepowersupply
socketCofthecontrolunitandasuitableoutlet.
4.Switchonmainspoweroftheoutletifpresent.
5.MovethemechanicalswitchDontheleftsideofthe
ControlunittotheOnposition.Afaintsinglebeep
soundsandthesystemincyclelightasheswhilethe
systempowersup.
CPR procedure
The Invacare® Softform Active® 2 Rx mattress has been fully
tested to comply with the current CPR standard of 5 - 6 cm
compression depth. This was achieved at all stages of
inflation/deflation.
1.HolddownRedCPRbuttonA.
2.Pullhoseconnectorrmlya wayfromthecontrolunitB.
3.Switchoffthecontrolunit.
Mattresswillstarttodeate.Thedeationtimeis
20seconds.
4.WhenCPRiscompletereactivatethesystemfollowing
section Installing the system.
InvacarePortugal,Lda
RuaEstradaVelha949
4465-784LeçadoBalio
Portugal
Invacare Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ
UK
Invacare UKOpera ons Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ, UK
60132743-A 2022-01-10 3 / 3
QuickGuide
Identifying and repairing faults
There are audio and visual alarms present on the control unit.
Problem/AlarmCauseSolution
MattressCPRhosedisconnected.ConnectCPRhoseconnector,lockitinplace.
Powercableandfusehasbeenchecked,control
unitdoesnotoperate.
SendcontrolunitbacktoInvacareforrepair.
Mattressnotinating
(notalternating
properly).
Alternatingsystem
/Timingfailure
indication
Majorleakinaircellorcompleteairinsert.Replaceleakingaircell.
CPRhoseortubeconnectorskinkedorsplit.UnkinkorreplacesplitCPRhoseortube
connectors.
Notalternating,rotorfailure.SendcontrolunitbacktoInvacareforrepair.
Noair(controlunitfailure).SendcontrolunitbacktoInvacareforrepair.
Controlunitoff.Checkpowersource,turnuniton.
Nopower/
Powerfailure
indication
Powercorddisconnected.Connectpowercordandensurethepower
sourceison.
Nopowerinthepoweroutlet.Switchonthepoweroutlet(ifpoleswitch
available).
Havethepoweroutletrepairedbyanelectrician.
Poweroutage.Waituntilthepowersourcehaspower.
Fuseblown.Changefuseonpowerinletconnectorwithspare
fuseoridenticalreplacementonly(consulta
trainedengineerifyouareunsurehowtochange
afuse).
DisconnectionofCPR(connectionhose).Connecthoseproperly.
Disconnectionofconnectortubestoaircellsin
airinsert.
Checkindividualaircellsininsertarecorrectly
connectedtoconnectortubes.
Pressurefailure/Low
pressureindication
KinkedconnectionhoseMakesurethattherearenokinksorbendsin
thehose.