Taylor Technologies K-1100 Owner's manual

Type
Owner's manual

Taylor Technologies K-1100 is a versatile device designed to accurately measure orthophosphate levels in water samples, ranging from 5 to 100 ppm. It employs a colorimetric method to determine the concentration of orthophosphate, providing reliable results within a minute. The kit encompasses essential components like test tubes, a slide comparator, and reagents, enabling convenient and precise testing. The device's user-friendly design and straightforward procedure make it suitable for various applications, including water quality monitoring, environmental analysis, and educational purposes.

Taylor Technologies K-1100 is a versatile device designed to accurately measure orthophosphate levels in water samples, ranging from 5 to 100 ppm. It employs a colorimetric method to determine the concentration of orthophosphate, providing reliable results within a minute. The kit encompasses essential components like test tubes, a slide comparator, and reagents, enabling convenient and precise testing. The device's user-friendly design and straightforward procedure make it suitable for various applications, including water quality monitoring, environmental analysis, and educational purposes.

COLOR COMPARISON TEST
ORTHOPHOSPHATE (5-100 ppm)
COMPONENTS:
1 x 2269 Cap, Dispenser, 18 mm, natural (or #2280, 28 mm, white)
5 x 4023 Test Tube, Calibrated (5 mL), glass
1 x 4027 Funnel, 58 mm, plastic
1 x 5160 Instruction
1 x 6003 Brush, Test Tube
1 x 6009 Filter Paper, 9.0 cm, #610, 100/box
2 x 9021 Test Tube, Mixing, Calibrated (5, 10, 14, 15, 17.5 mL), glass w/ stopper
1 x 9110 Slide Comparator, Orthophosphate, Stannous Chloride, 5-100 ppm
1 x 9190 Base, Slide Comparator (for test tubes)
1 x R-0601-C Molybdate Reagent*, 2 oz (or R-0601-E, 16 oz)
1 x R-0602P-I Stannous Chloride Powder, 10 g
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
Orthophosphate Test
1. Filter water to be tested to clarify.
2.Rinseandllonemixingtesttube(#9021)to5mLmarkwithlteredsample.
3.Diluteto15mLmarkwithdistilledordeionizedwater.Stopperandmix.Removestopper.
Transfertotwo5mLtesttubes(#4023)to5mLmark.
4.Wipedryandplaceintwoholesadjacenttocenterholeofcomparatorbase(#9190).
5.Rinseandllsecondmixingtesttube(#9021)to5mLmarkwithlteredsample.
6.Cap R-0601 Molybdate Reagent with dispenser cap (#2269 or #2280). Add R-0601
MolybdateReagentto15mLmark.Stopperandmix.Removestopper.
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548) • 410-472-4340 11/16
7.Add2leveldippersR-0602PStannousChloridePowder.Stopperandmix.
8.Wipedryandplaceincenterholeofcomparatorbase(#9190).WAIT1MINUTE.Readbefore
5minutes.
9.PlaceSlideoncomparatorbase.Matchcolorincentertesttubewithacolorstandard.The
SlideisinproperalignmentforacolormatchwhenawhitelineontheSlideisdirectlyabove
thewhitelineonthecomparatorbase.Recordaspartspermillion(ppm)orthophosphate.
NOTE: Ifsamplecolormatchesorisdeeperthanhighestcolorstandard:Dilutesamplewith
distilled,deionized,orphosphate-freewater.Repeattest.Forsampledilutedwith1
partwater,multiplyreadingby2;for2parts,multiplyby3,etc.
*WARNING: MolybdateReagent(R-0601)contains5–10%(w/w)sulfuricacid,acorrosive
acid.
Instr. #5160
SLIDE & MIDGET COMPARATOR TESTS
ORTHOPHOSPHATES & POLYPHOSPHATES IN GENERAL
This method is based on the formation of phosphomolybdic acid, which is subsequently
reduced with stannous chloride to produce a blue color whose intensity is proportional to
the phosphate (PO
4
) present. Interferences are reduced to a minimum. The presence of
10,000 ppm of silica (SiO
2
), 100 ppm of ferrous or ferric iron, or 1000 ppm of sulte causes
no error.
The water to be tested must be carefully ltered with a good grade of lter paper to free it
from sludge and scale. Mere traces of sludge will cause serious errors.
If repeated ltration does not clarify the water, add 0.5 g potassium nitrate crystals to 25 mL
of the sample water, heat to boiling, cool, dilute to 25 mL with distilled water, and lter.
If the water is very highly colored by tannin or other highly colored organic matter, so that a
phosphate determination is dicult or impossible, it can be eectively decolorized with
activated charcoal. Add half a teaspoon of activated charcoal to a pint container of the
colored sample water. Close the container and shake sample well. Decolorization should
take place in 5 to 15 minutes. The treated sample is then ltered as instructed above. From
this point on follow the procedure outlined in the Orthophosphate or Polyphosphate tests.
31 Loveton Circle, Sparks, MD 21152 U.S.A.
800-TEST KIT (837-8548) 410-472-4340 8/16
Instr. #5159
  • Page 1 1
  • Page 2 2

Taylor Technologies K-1100 Owner's manual

Type
Owner's manual

Taylor Technologies K-1100 is a versatile device designed to accurately measure orthophosphate levels in water samples, ranging from 5 to 100 ppm. It employs a colorimetric method to determine the concentration of orthophosphate, providing reliable results within a minute. The kit encompasses essential components like test tubes, a slide comparator, and reagents, enabling convenient and precise testing. The device's user-friendly design and straightforward procedure make it suitable for various applications, including water quality monitoring, environmental analysis, and educational purposes.