Operating Instructions—
7S1
1
Source resistance for
the voltage
at the OFFSET OUT
monitor
jack is 10
kTi;
therefore, meter loading
may be
a
factor if other than
an infinite-impedance
meter is used.
The accuracy of the
DC offset voltage
measurement
depends on the accuracy
and
the
loading effect of
the meas-
uring
device. The following measuring
devices are
recom-
mended,
in order of preference,
for monitoring voltage
at
the OFFSET
jacks.
(1)
Differential,
non-loading DC voltmeter
accurate within
0.2% or better. This
type
of
device provides absolute
offset
voltage measurements
accurate within 2%. Meas-
urements of small changes
in offset voltage can be
made more accurately than
2%.
(2)
Vacuum-tube voltmeter with an input impedance of
at
least
10 megohms. Accuracy of the VTVM should
be as
high as practical.
(3)
Zero-center
±1
mA milliammeter with
as
high an accur-
acy as
practical. The milliammeter should
be
connected
directly between
the OFFSET
OUT monitor jack and
ground. When using
a
milliammeter,
1
milliampere is
equivalent to 1 volt
of the actual
offset
voltage. The
loading effect is within 2%
+
the
error
percentage
deter-
mined by: 100
(Rmeter/10000).
Slideback Measurement Procedure.
To measure the
volt-
age difference between two points on
a
waveform (such as
peak or peak-to-peak volts), proceed
as
follows:
1.
Set
the
DC
OFFSET
control
to about midrange.
2.
Apply the signal to be measured
to
the sampling head
input. Adjust for a stable display with about 7 divisions of
vertical deflection
between
the
two points of the signal to
be measured.
3. With the DC
OFFSET
controls, move one of the
points to be measured to
the
center
line
of
the graticule
and
measure
the voltage
at
the OFFSET OUT monitor
jack. Use
one of
the
measuring devices mentioned previously.
4. With the DC OFFSET control, move the display so
the other point to be measured is aligned with the center-
line of the graticule and again measure the voltage
at
the
OFFSET
OUT monitor jack.
5.
Find the difference between the voltage measured in
step 3 and
the voltage measured in
step
4,
and divide by
10.
The result
is the voltage
difference,
in volts,
between
the
two points on the
waveform. This
voltage
tolerance
is
±2%
of
the input signal plus
the
tolerance and loading
effect
of
the
measuring device.
Use of
Smoothing
Random
process sampling
does
not generally
permit the
use of
smoothing. This process
requires
unity loop
gain to
construct
a
coherent display from
the
samples taken
at ran-
dom. When
using random
process sampling,
operate the
7S11 with the NORMAL
switch
pushed in, and
adjust the
DOT RESPONSE
control
to unity loop
gain.
Normal process
sampling
does permit the
use of smooth-
ing. Time and amplitude noise
may sometimes
be objection-
able when
operating at minimum
deflection factors
or max-
imum sweep rates. For
Smoothing
operate the
7S11 with
the SMOOTH switch pushed
in.
This will reduce the random
noise
by about one half
by
decreasing the gain of the sampling
feedback loop. Fig.
2-7
shows the advantage of using
smoothing
when observing a
low-amplitude
signal.
Dot Density.
Normally the
risetime of the display will
not
be
significantly affected
in smoothed
operation if the
dot density is sufficient. If,
however, the display waveform
shape is affected when the
SMOOTH switch is
pushed in, a
compromise must
be made between smoothing
and dot
density. Fig.
2-7C
illustrates
the effect produced when the
dot density
is
low
with smoothed operation. This effect
can
be
compared
to the
high
dot density of the
same input
signal as shown in Fig. 2-7B.
"False"
Displays
Because sampling displays only
a
fraction of any
one
cycle of
a
repetitive signal,
it is sometimes possible
to
obtain
a
false
equivalent-time
display of the signal. The
incorrect
equivalent-time display
occurs when the sampling
rate is
an exact multiple of
the signal rate. Each sample
vertical
position represents the
correct signal amplitude,
but
an incorrect
number of samples
is taken,
causing the false
equivalent-time
display.
False equivalent-time
displays can
be
detected
by
chang-
ing the sampling sweep unit
Scan control; i.e., changing the
dot
density.
Occasionally
such
a
false equivalent-time dis-
play
occurs when the
sampling
sweep unit triggering circuit
is
adjusted to free run, and at the same time the triggering
signal
amplitude synchronizes it falsely.
Any time the dis-
play equivalent time changes
when changing the dot den-
sity, alter the Time/Div switch
to obtain a display with an
2-10