3 Operation
Saline Solution - Priming and Volume Calibration
Thermo Scientific Centra W CW2+ 3-5
Make sure that test materials are used within the manufacturers’ recommended dates.
Erroneous antiglobulin readings can result from any of the following:
•The sample serum/cell ratio is improper.
•The red cells were not sufficiently washed.
•The saline solution was improperly decanted.
•The volume calibration has shifted due to change in the saline level or restriction in the feed
tube.
•The antiglobulin serum is inactive or outdated.
•The antiglobulin serum was not added to a tube(s).
•The final centrifugation (spin) was improper.
•The reading or recording of results is inaccurate.
This Cell Washer provides accurate washing, decanting, and a proper speed for the final spin. However,
the Cell Washer and the procedures outlined in this manual will not eliminate all potential errors. To
minimize errors, proper laboratory procedures require adequate control of all tests performed.
Saline Solution - Priming and Volume Calibration
Long-term exposure to certain preservatives found in azidefree saline solutions may adversely effect
certain plastic components within the Cell Washer. Routine housekeeping and cleaning practices will
help to remove salt deposits in, and prolong the life of, these components.
Priming
Priming is required whenever a new saline supply is connected. To prime the instrument:
1. Connect the saline reservoir to the cube adapter of the pump tubing.
2. Open the cover and place a receptacle under the saline dispenser nozzle in the cover.
3. Press the CALIBRATE key.
WARNING When handling and disposing of blood and blood components follow
standard laboratory procedures. Observe all safety precautions and standard laboratory
practices.
WARNING Some saline solution contains a sodium azide preservative which may react
with the drain plumbing to form dangerous explosive azide salts. Check with the saline
solution supplier before discharging waste solutions directly into normal drains.
Note When priming or calibrating the pump with the cover open always put a container in front of
the nozzle before pressing the calibrate key. (Otherwise, you will get wet!)